Legend Pro™System User Manual Version 7
Proprietary Information of Pollogen a company of Lumenis 8
Only move the system by using its handle. Pushing or pulling the system without using its handle can
turn the system over and may result in damage to the system.
Switch off the system when moving the unit.
Before connecting the power supply cord, make sure that the ON/OFF switch is in the OFF position (“0”).
Check all connections before plugging in the System’s power supply cord.
Pay attention to the screen quality, to ensure that it is acceptable during the first few seconds of the
system check (touch screen becomes active). At this time, the Applicator's LED flashes (first green then
orange) as part of the System’s self-check. In case of display failure, stop system operation.
In case of uncertainty regarding potential side effects, have the patients consult their physician and
bring consent for treatment.
Restart the system in the event that a wrong or incomplete program upload occurred while uploading
the program version using the USB memory stick.
Make sure that the counter on the display is stopped when the foot switch is released.
Patients should remove all jewelry prior to treatment.
Do not use Legend Pro™ System in close proximity to a short-wave or microwave therapy equipment.
Before every treatment, verify that the applicator is intact.
Make sure to replace the applicator if it falls before, during or after a treatment.
Always use tips which were approved and supplied by Lumenis or its legal representatives.
When treating with TriPollar® programs, patients should remove all jewelry prior to treatment
2.3 Warnings
Before attempting to operate the Legend Pro™ System, be sure to read this manual to become familiar
with all safety requirements and operating procedures.
No modification of this equipment is allowed!
When treating facial area, never treat the area below the eyebrow & never treat over the eye lids &
eyes themselves. Never treat over the lips. Do not treat genitals, the chest, breasts or underarms.
Do not connect a patient to High Frequency surgical equipment during treatment.
Treatment should not be applied over swollen, infected, inflamed areas or skin eruptions.
Verify that fan is functioning when operating the System. In case of fan failure do not operate the
device, call your Lumenis representative for service.
Do not operate the unit in the presence of electromagnetic interference. Electromagnetic interference
may be caused by electro surgery, diathermy, magnetic resonance imaging or other type of equipment.
Portable and mobile are RF communication equipment & may affect the operation of this device.
Using power that is higher than needed can cause over-treatment of the skin.
When treating with TriPollar® programs, do not treat using mineral oils.
When treating with TriPollar® programs, do not heat the skin over 43ᵒC (109.4ᵒF).
Treatment should not be applied over, or in proximity to, cancerous lesions.
When treating with TriPollar® programs, to avoid the risk of potential burns, applicator must always be
in constant movement while in contact with the skin. Never stop applicator movement while in contact
with the patient.
There may be risks associated with bipolar RF energy delivery to the skin that are yet unknown. Any
procedure perforating the skin can cause discomfort, pain, bleeding, infection, swelling, edema, scar
formation, permanent marking, and pigment alteration. Therefore, potential risks which could result
from RF and electrode insertion include discomfort during and after the procedure, pain, infection,
scarring, swelling or edema, and permanent discoloration.
Do not allow patients to contact metal surfaces.
The Applicator should be used at least 1 cm away from a cochlear implant.
Check device for any damage prior to use.
Do not activate the applicator when it is not in contact with the patient’s skin.
Discontinue use immediately in case of excessive heat or burning sensation.