Practicon® Acu-Tip™
Cauterizing Instrument
Instructions
Medical Purpose/Indication
• Acu-Tip is a thermocautery device used for localized soft
tissue trimming and hemostasis during oral surgery.
Operating Principle
• This handheld, battery-powered device delivers thermal
energy through a heating element to the patient by pressing
an activation switch.
Patient Population
• No restriction.
Site of Use and Site Conditions
• Site of Use: No restrictions.
• Site Condition: Aseptic.
Intended User Profile
• Qualifications: Licensed medical professional.
• Knowledge: Read and understand these supplied instructions.
• Experience: Training on soft tissue surgical techniques or training
under surveillance/supervision.
HANDLE
Read instructions
before use NON
STERILE
Non-sterile
For professional
use only
CAUTERY
%
STORAGE
10 to 75% non-condensing
OPERATE
30 to 75% non-condensing
TRANSPORT
10 to 100% including condensation
-40°C
10°C
10°C +70°C TRANSPORT
40°C OPERATE
30°C STORAGE
NON-STERILE / SINGLE-USE TIPS
Do not use if packaging
is damaged or opened
Read instructions
before use
For professional
use only 2Do not reuse. This is a
single use device.
CAUTIONS
• DO NOT autoclave the Acu-Tip handle. Damage may occur.
• DO NOT use this device if the packaging is damaged or opened.
• DO NOT apply excessive force on the cautery during use. This
may cause the tip to bend or break.
• The recommended duty cycle is 2 seconds ON and 6 seconds
OFF for no longer than 15 minutes continuous time. Excessive
activation may cause handle to overheat.
WARNINGS
• Tip (heating element) cannot be eectively cleaned or sterilized
by the user and therefore cannot be reused safely. It is intend-
ed for single use only. Any attempt by the user to clean and
resterilize the tip may result in device failure, bio-incompatibility,
infection, or other risks to the patient.
• Always remove the tip when the device is in storage or not in
use. Failure to do so could result in fire.
• DO NOT use in the presence of flammable gases/materials or
oxygen rich environments. Fire could result.
• Acu-Tip produces heat in excess of 1,000°F, which can cause
burns or fire from misuse.
• Prolonged tissue exposure to cautery may lead to damage of
adjacent tissue.
• Cautery tip will remain hot immediately after deactivation and
may cause inadvertent burns to user or patient.
• When ready for disposal, deactivate the cautery before dis-
carding the device in a biohazard or “sharps” container. First,
remove the cautery tip, then remove batteries by turning the
rear cap. Failure to follow this instruction could result in unin-
tended activation of the cautery and may cause a fire in the
disposal container.
1-800-959-9505 • Fax 1-800-278-0885
practicon.com • Greenville, NC 27834
Made in China • IFU 7051711_4
MC-17078 Rev. 6
2018-10-02
Manufactured for
Replacing Batteries
1. Remove cautery tip before beginning battery replacement procedure.
2. To remove depleted batteries, unscrew the rear cap (E) by turning counter-clockwise.
3. Remove the cap and tilt the handle so the batteries slide out. A gentle downward shake may be required.
4. Insert new AA alkaline batteries positive end first (never use carbon zinc batteries in this device).
5. Replace rear cap and turn clockwise until tightly closed.
Cleaning the Handle
1. Use only disinfectant wipes to clean the Acu-Tip handle.
2. DO NOT put handle in autoclave.
Applying Barrier Sheath
1. Remove cautery tip from handle before applying barrier sheath.
2. Holding the barrier sheath (A) with the open end up, insert the handle into the sheath.
3. Slide the handle forward, until the nose of the handle is at the end of the sheath.
IMPORTANT: Visually inspect barrier sheath for damage after application to the handle and prior to placing in sterile field.
Replacing Tips
1. Handle the tip with care, avoiding the fragile wire. Always hold the tip by the plastic grip (B) to avoid damage.
2. Insert the cautery tip into the handle by pushing the two electrodes posts (C) through the barrier sheath.
3. In order to make proper contact with the batteries, insert the tip all the way into the handle until the plastic grip is against the front of
the handle. DO NOT FORCE. It should slide easily into place.
4. Activate the unit by depressing the power button (D).
5. After use, remove tip and barrier sheath and dispose of properly.
Replacement parts
Additional Acu-Tip Replacement Tips and UIKsheath™ Barriers are available from Practicon:
70-51712 Acu-Tip Short Replacement Tip
70-51713 Acu-Tip Long Replacement Tip
70-51714 Acu-Tip 45˚ Replacement Tip
70-19721 QUIKsheath Barriers Low-Speed Straight are approximately 8" long. Non-sterile. 500 per box.
A
B C D
E
1-800-959-9505 • Fax 1-800-278-0885
practicon.com • Greenville, NC 27834
Made in China • IFU 7051711_4
MC-17078 Rev. 6
2018-10-02
Manufactured for

Popular Medical Equipment manuals by other brands

Spencer Rock Pin Use and maintenance manual

Spencer

Spencer Rock Pin Use and maintenance manual

Orliman COL-240CSB manual

Orliman

Orliman COL-240CSB manual

bk ultrasound bk3000 user guide

bk ultrasound

bk ultrasound bk3000 user guide

Braun Aesculap Acculan 4 Instructions for use/Technical description

Braun

Braun Aesculap Acculan 4 Instructions for use/Technical description

Gi Paragon 203 quick start guide

Gi

Gi Paragon 203 quick start guide

Lloyd 402 Flexion Elevation Operations & parts manual

Lloyd

Lloyd 402 Flexion Elevation Operations & parts manual

Sirona orthophos xg 3 installation instructions

Sirona

Sirona orthophos xg 3 installation instructions

ZOLL LifeVest 4000 Patient manual

ZOLL

ZOLL LifeVest 4000 Patient manual

Jula 951-106 User instructions

Jula

Jula 951-106 User instructions

videomed Truelink 4 Instructions for use

videomed

videomed Truelink 4 Instructions for use

Olympus MU-1 Maintenance manual

Olympus

Olympus MU-1 Maintenance manual

EyeQue Insight product manual

EyeQue

EyeQue Insight product manual

Ossur LIVINGSKIN Instructions for use

Ossur

Ossur LIVINGSKIN Instructions for use

medi 3 C Instructions for use

medi

medi 3 C Instructions for use

Packard Bell PBO-1 user guide

Packard Bell

Packard Bell PBO-1 user guide

Sanyo mpr-214f - Commercial Solutions Refrigerator brochure

Sanyo

Sanyo mpr-214f - Commercial Solutions Refrigerator brochure

Invacare Matrx PS user manual

Invacare

Invacare Matrx PS user manual

HeartSine samaritan PAD SAM 300P user manual

HeartSine

HeartSine samaritan PAD SAM 300P user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.