Praxisdienst TA60V User manual

Content
Explanation ..................................................................................................................................... - 1 -
Intellectual Property .......................................................................................................................- 1 -
Statement ....................................................................................................................................... - 1 -
Warranty and Repair Service ..........................................................................................................- 2 -
After-Sales Service ..........................................................................................................................- 2 -
Foreword .........................................................................................................................................- 3 -
Explanation ..................................................................................................................................... - 3 -
User .................................................................................................................................................- 3 -
Images .............................................................................................................................................- 3 -
Conventions .................................................................................................................................... - 3 -
1.0 Safety ........................................................................................................................................ - 4 -
1.1 Safety Information ............................................................................................................ - 4 -
1.1.1 Warning ................................................................................................................. - 4 -
1.1.2 Caution .................................................................................................................. - 5 -
1.1.3 Caution .................................................................................................................. - 6 -
1.2 Device Symbol .................................................................................................................. - 7 -
2.0 Overview ...................................................................................................................................- 9 -
2.1 System Overview ........................................................................................................... - 9 -
2.1.1 Intended Use ......................................................................................................... - 9 -
2.1.2 Contraindications .................................................................................................. - 9 -
2.1.3 Product Description ...............................................................................................- 9 -
2.2 Device View .................................................................................................................... - 10 -
2.2.1 Front side view of the anaesthesia machine ...................................................... - 10 -
2.2.2 Breathing circuit view ..........................................................................................- 12 -
2.2.3 Anaesthesia machine rear view(with stand).................................................- 13 -
2.2.4 Anaesthesia machine rear view(Wall)...........................................................- 14 -
3.0 Installation Guide ................................................................................................................... - 15 -
3.1 Stand Installation Guide ................................................................................................. - 15 -
3.2 Anesthesia machine mounted on stand ........................................................................ - 15 -
3.3 Anesthesia vaporizer mounted to anesthesia machine .................................................- 16 -
3.4 CO2 absorber Canister installation and removal ........................................................... - 16 -
3.4.1 CO2 absorber Canister installation ..................................................................... - 17 -
3.4.2 CO2 absorber canister disassembly .................................................................... - 17 -
3.5 Installation and disassembly of waste gas canister and waste gas recycling pipe ........ - 17 -
3.5.1 Installation of waste gas canister and waste gas recycling pipe .........................- 18 -
3.5.2 Dismantling of waste gas canister and waste gas recycling pipe ....................... - 19 -
3.6 Wall-mounted anesthesia machine ............................................................................... - 19 -
4.0 Test before operation ............................................................................................................. - 21 -
4.1 Pre-Operation Testing Requirements .............................................................................- 21 -
4.2 Pre-Operation Checklist ..................................................................................................- 21 -
4.3 Check the system ............................................................................................................- 22 -

4.4 Manual vent leak test .....................................................................................................- 23 -
4.5 Air Supply Pipeline Test .................................................................................................. - 23 -
4.6 Respiratory System Test ..................................................................................................- 23 -
4.6.1 APL Valve Test ...................................................................................................... - 24 -
4.7 Preparation before Surgery ............................................................................................- 24 -
5.0 Operation ................................................................................................................................- 25 -
5.1 Oxygen Flow Setting ....................................................................................................... - 25 -
5.2 Set Up the Vaporizer .......................................................................................................- 25 -
5.2.1 Add Anesthetic .................................................................................................... - 25 -
5.2.2 Discharge anesthetic ........................................................................................... - 27 -
5.3 Manual Ventilation Operation ........................................................................................- 28 -
6.0 Maintenance ...........................................................................................................................- 29 -
6.1 Maintenance Timetable ................................................................................................. - 29 -
6.2 Disinfection and cleaning methods ................................................................................- 29 -
6.2.1 Wipe .................................................................................................................... - 30 -
6.2.2 UV exposure ........................................................................................................ - 30 -
7.0 Appendix .................................................................................................................................- 31 -
7.1 Attachment List ...............................................................................................................- 31 -
A.0 Product Specification ............................................................................................................. - 33 -
A.1 Safety Specification ........................................................................................................ - 33 -
A.2 Environment Specification ............................................................................................. - 33 -
A.3 Physical Specifications ....................................................................................................- 33 -
A.4 Air System Specifications ............................................................................................... - 34 -
A.5 Respiratory System Specifications ................................................................................. - 34 -
A.6 Vaporizer ........................................................................................................................ - 34 -
A.7 No Power AGSS .............................................................................................................. - 35 -

- 1 -
Explanation
Thank you for purchasing the Anesthesia Machine.
Before using the product, please carefully read the contents of this instruction manual in order to
use the product correctly.
After reading, please keep this instruction manual properly so that you can refer to it at any time
when you need it.
Intellectual Property
The intellectual property rights of this instruction manual and its corresponding products belong
to this company. Without the written consent of this company, no individual or organization can
copy, modify or translate any part of this instruction manual.
Statement
This company has the final right to interpret this instruction manual.
This company is responsible for the safety, reliability and performance of products only if all of
the following requirements are met, namely:
Assembly operations, expansions, readjustments, improvements, and repairs are performed by
this company approved professionals;
All repairs involve the replacement of parts and accessories and consumables used in conjunction
with them are the original (original) of this company or approved by this company;
The relevant electrical equipment complies with the national standards and the requirements of
this instruction manual;
Product operation is carried out in accordance with this instruction manual.
Product Name :
Anesthesia Machine
Model:
TA60V
Structure and
Composition:
This product consists of an anesthesia machine, a vaporizer, a stand
a breathing circuit and accessories.
Scope of Application:
This anesthesia machine is used to provide inhalation anesthesia
and respiratory support during animal anesthesia.
Manufacturer's name:
TOOTOO MEDITECH CO., LTD.
Address:
3A, Phase II, Smart Home, No. 76, Baohe Avenue, Baolong Communi
ty, Baolong Street, Longgang District, Shenzhen, Guangdong, China
Production Date:
See the nameplate on the back cover of the product
The Edit Date of of the
Manual:
2022-01

- 2 -
Warranty and Repair Service
The warranty period of the purchased product is subject to the sales contract.
Consumables: Refers to disposable consumables that need to be replaced after each use or
fragile materials that are replaced regularly. Consumables have no warranty.
The warranty period starts from the "Installation date" filled in the "Equipment Warranty Card"
attached to the product. The "Equipment Warranty Card" is the only proof for calculating the
warranty period. In order to protect your rights, please fill in the warranty card after the
installation of the equipment, and hand over the second copy of the warranty ("this company's
retention") to the installer or mail it back to this company User Service Department.
Please note that the following situations will not be covered under warranty:
1. The customer fails to fill in and return the warranty card within 30 days after the completion of
the installation and acceptance;
2. The serial number of the equipment provided by the customer is incorrect (our company
confirms the warranty with the serial number of the equipment).
During the warranty period, the products can enjoy free after-sales service; but please note that
even during the warranty period, if the product needs to be repaired due to the following
reasons, this company will implement a charged maintenance service, and you need to pay the
maintenance fee and accessories fee :
Man-made damage;
Improper use;
Irresistible natural disasters;
Replace or use parts, accessories not approved by this company or repaired by persons not
authorized by this company;
Other failures not caused by the product itself.
The failure of the instrument caused by the use of consumables not approved by this company is
not within the scope of maintenance services provided by this company.
After the warranty period expires, this company can continue to provide paid maintenance
services.
If you do not pay or delay in paying the fee for the maintenance service, this company will
temporarily suspend the maintenance service until you pay.
We strongly recommend that you use this product in a manner consistent with this manual and
under the conditions and environment specified in this manual.
After-Sales Service
Company name:TOOTOO MEDITECH CO., LTD.
Address:3A, Phase II, Smart Home, No. 76, Baohe Avenue, Baolong Community, Baolong Street,
Longgang District, Shenzhen, Guangdong, China
Postcode:518029
Company website:www.newtootoo.com
Tel:+86-755-28306385

- 3 -
Foreword
Explanation
This manual describes the use, function and operation of the product in detail. Before using this
product, please read and understand the contents of this manual carefully to ensure the correct
use of this product and the safety of animals and operators.
This manual introduces this product according to the most complete configuration, so some
contents may not be applicable to the purchased product. If you have any questions, please
contact this company.
Please place this manual near the product so that it can be easily and promptly obtained when
needed.
User
This instruction manual is only for use by professionally trained anesthesiologists
Images
All illustrations provided in this manual are for reference only, and the settings or data in the
illustrations may not be exactly the same as the actual display seen on the product.
Conventions
Bold italics: Indicates quoted chapters.
→:This symbol is used to indicate the steps in the operation

- 4 -
1.0 Safety
1.1 Safety Information
WARNING —Indicates a potentially hazardous or unsafe operation which, if not avoided, could
result in death or serious personal injury or property damage.
CAUTION —Indicates a potentially hazardous or unsafe operation, which, if not avoided, could
result in minor personal injury, product failure, damage, or property damage.
NOTE - Emphasizes important notices, provides instructions or explanations for better use of this
product.
1.1.1 Warning
Warning : do not operate the anesthesia machine before reading this
manual。
Warning: Before operation, ensure that the machine, connecting line and
accessories can work normally and safely。
Warning: Do not use this machine in an environment where flammable
or explosive materials are placed to prevent fire or explosion。
Warning: Repairs or upgrades to the machine may only be performed by
service personnel trained and authorized by the company. The
fresh air must not be turned off until the anesthesia vaporizer is
turned off. The anesthesia vaporizer cannot be turned on without
fresh air. Otherwise, a high concentration of anesthetic vapor can
enter the machine pipeline and the surrounding air and cause
harm to people and things.
Warning:Before moving the anesthesia machine, remove items from the
top plate and stand to avoid a tipping hazard of the anesthesia machine.

- 5 -
Warning: Do not use the anesthesia machine when there is a leak in the
respiratory system
Warning: Check the passive AGSS processing system specifications as
well as the specifications of the anesthesia machine to ensure compatibility
and prevent mismatched processing systems.
Warning: Using the incorrect connector can be dangerous. Make sure all
components are using the correct connectors.
Warning : Disposable items such as single-use breathing hoses, soda
lime may be considered potentially biohazardous and should not be
reused. When handling these items, follow hospital and local regulations
on contaminants and biohazards.
Warning:Do not use the top plate to push and lift the machine. Please use
the handle
Warning:If the anesthesia machine is not disinfected in time after use,
there is a risk of cross infection. Do not perform maintenance on
the machine while the animal is using it
1.1.2 Caution
Caution: For animal safety, please use the accessories specified in this
manual。
Caution: When the machine and its accessories are about to exceed their
lifespan, they must be disposed of in accordance with relevant
local regulations or hospital regulations。

- 6 -
Caution: Please install or transfer the machine properly to prevent the
machine from falling, bumping, being damaged by strong
vibration or other mechanical external forces
Caution: When the trolley is configured, the anesthesia machine can be
tilted 10° to maintain stability. Please do not hang objects on both sides of
the anesthesia machine in order to avoid the danger of dumping.
Caution : Please effectively fix the machine placed on the top cover to
avoid accidental sliding.
Caution : The use and storage of the air source hose assembly in an
environment such as ultraviolet rays, oxidants, high temperature and
humidity should be prevented or avoided, so as to prevent the aging of the
hose and the pressure release from hose assembly to could cause harm to
people, animals and objects.
Caution: This machine is not intended for use in a magnetic resonance
(MRI) environment
Caution : Unlocked casters may move unexpectedly when a stand is
configured. Casters should be locked while using this machine.
1.1.3 Caution
Caution: Install the machine in a location that is easy to observe, operate
and maintain. Please observe the relevant information of the
machine at a position in front of the anesthesia machine within 4m
from the front panel of the anesthesia machine.

- 7 -
Caution:Please place this manual near the machine so that it can be easily
and promptly obtained when needed.
Caution : This manual introduces this product according to the most
complete configuration, and the purchased product may not have
some configurations or functions.
1.2 Device Symbol
Symbol
Description
Symbol
Description
Caution! Read
product manual
Defibrillation Proof
Type BF Applied Parts
Oxygen Port
Flowmeter
Flowmeter knob, turn
counterclockwise to
increase flow
O2 Flush
ACGO Switch
APL Pressure
adjustment knob,
turn clockwise to
increase the set
pressure
Inspiratory Valve Port
Expiratory Valve Port
Breathing Bag Port
ACGO Port
MIN
Minimum
MAX
Maximum
Maximum load
Soda lime filling
maximum
CO2 Absorber
Canister locking and
unlocking
AGSS Exhaust Port
IPX1
Protection level
against water, dust,
etc.
Serial Number

- 8 -
Manufacturing date
avoid rain
face up
Atmospheric
pressure limit
Humidity Limit
temperature limit
Fragile handle
Stacking layer limit
Not for use in a
magnetic resonance
(MRI) environment
This product contains some harmful substances, which can be used
with confidence within the environmental protection use period. After
the environmental protection use period is exceeded, it should enter
the recycling system. The environmental protection use period of this
product is 20 years.

- 9 -
2.0 Overview
2.1 System Overview
Please refer to the "3.0 Installation Guide" of this manual for the installation of the anesthesia
machine。
2.1.1 Intended Use
This anesthesia machine is used to provide inhalation anesthesia and respiratory support during
animal anesthesia.
Warning:
The user of this anesthesia machine should be a full-time
anesthesiologist and have received training in the use of the anesthesia
machine。
2.1.2 Contraindications
Unknown
2.1.3 Product Description
This product consists of an anesthesia machine , a vaporizer, a stand, a breathing circuit
and accessories
This anesthesia machine provides the following ventilation modes:
Manual ventilation mode
Acyclic ventilation mode

- 10 -
2.2 Device View
2.2.1 Front side view of the anaesthesia machine
2-1 Front side view of main unit

- 11 -
Serial
Part
Description
A1
Vaporizer
A device that can effectively output the concentration of
anesthetics. Each vaporizer is suitable for a specific
anesthetic.
A2
Vaporizer base
It is used to install the vaporizer. After the vaporizer is
removed, it can also ensure the normal ventilation of the air
circuit.
A3
Oxygen Port
Oxygen connection port
A4
Storage basket
Used to store various medical items.
A5
Casters
Machine can be moved through casters, all four casters with
brakes
A6
Air pressure
gauge
Used to display the input air pressure
A7
ACGOS Switch
Used to turn on/off the ACGO function. For fresh gas output
via ACGO interface.
A8
O2 Flush
Provides a fixed flow of oxygen to the inspiratory branch of
the respiratory system
Caution:In the case of testing pressing the rapid
Oxygenation button, please ensure that the
connection to the animal has been disconnected
A9
APL vale
Rotary pressure regulator for setting the limiting pressure of
the breathing system during manual ventilation. The scale
above it represents the approximate pressure. Turning the
APL valve clockwise increases the pressure limit and
counterclockwise decreases the pressure limit. After pressing
the APL valve, the limiting pressure will increase by about
30cmH2O on the original basis, and it will return to the
original setting pressure after releasing it.。
A10
Airway
Pressure Gauge
Indicates airway pressure
A11
loop
Integrated recirculating breathing circuit
A12
flowmeter
Displays the flow of oxygen
Table 2-1 Main unit front side view parts list

- 12 -
2.2.2 Breathing circuit view
Image 2-2 Breathing circuit view
Serial
Part
Description
B1
Inspiratory check valve
observation window
The status of the inspiratory check valve can be
observed from outside the device.
B2
Expiratory check valve
viewing window
The status of the expiratory check valve can be observed
from outside the device
B3
Inspiratory port
Inspiratory branch connector
B4
Test plug
When testing for leaks, it is used to connect the outlet
end of the breathing circuit
B5
Expiratory port
Expiratory branch connector
B6
ACGO Port
For connecting to an open breathing circuit
B7
Breathing bag Port
For connecting the storage bag
B8
CO2 absorber canister
Containers for filling soda lime, absorbing CO2
Table 2-2 Breathing Circuit View Parts List

- 13 -
2.2.3 Anaesthesia machine rear view(with stand)
Picture 2-3 Anaesthesia machine rear view
Serial
Parts
Description
C1
Hook
For hanging breathing circuit or other accessories
C2
AGSS Port
for exhaust
C3
CO2 absorber canister
lock and unlock handle
Used to control the installation of CO2 absorber canister.
C4
Exhaust gas recovery line
Connection line for connecting the AGSS port to the
exhaust gas tank
C5
Waste gas tank
Used to absorb anesthetic gas from AGSS exhaust gas
Table 2-3 Anaesthesia machine rear side view parts list

- 14 -
2.2.4 Anaesthesia machine rear view(Wall)
Picture 2-4 Anaesthesia machine rear view(Wall)
Serial
Part
Description
D1
Wall hook
When the main unit is wall mounted on the wall, hang on
the wall bracket
D2
Wall support pad
When the host is wall mounted on the wall, support the
host so that the host is perpendicular to the wall
Table 2-3 Anaesthesia machine rear view (wall mounted) parts list

- 15 -
3.0 Installation Guide
Warning:
Disposable items such as single-use breathing hoses, soda lime may be
considered potentially biohazardous and should not be reused. Dispose
of these items in compliance with hospital and local regulations for
contaminants and biohazards
Caution:
This manual introduces this product according to the most complete
configuration. The purchased product may not have some configurations
or functions。
Note: The part numbers in all the figures in this chapter, if they appear in the previous chapters,
have the same numbers and the same names as the previous chapters.
3.1 Stand Installation Guide
If you choose a stand, please install the stand by referring to the "accompanying documents" in
the stand box.
3.2 Anesthesia machine mounted on stand
Serial
Item
F1
Anesthesia Machine
F2
Stand support plane
F3
Locking screw
(1) Place the anesthesia machine on
the support plane of the stand,
aligning the mounting screw holes
(2) Screw the locking screw upwards
from the lower part of the support
surface of the stand into the
anesthesia machine, and tighten
the locking screw.
(3) Refer to the picture after
installation2-1。
Image 3-2 Anesthesia machine
installed on the stand

- 16 -
3.3 Anesthesia vaporizer mounted to anesthesia machine
3.4 CO2 absorber Canister installation and removal
The views of the CO2 absorber canister before and after installation are as follows:
Serial
Item
A1
Vaporizer base
G1
Connector
G2
Locking knob
A2
Vaporizer
(1) Align the mounting hole of the
evaporator with the connector on the
base of the evaporator, and then install
the evaporator on the bottom of the
evaporator.
(2) Rotate the locking knob on the top of the
evaporator clockwise, and the rotation
angle is 90 degrees. Lock the evaporator.
(3) Refer to the picture after installation 2-1。
Image 3-3Vaporizer mounted to
anesthesia machine diagram
Image 3-4-1 Schematic diagram before
installation
Image 3-4-2 Schematic diagram after
installation

- 17 -
Serial
Part
Description
B8
CO2 Absorber Canister
Vessel for filling soda lime, absorbing CO2
C3
CO2 Absorber Canister
Lock and Unlock Handle
Used to control the installation of CO2 absorber canister.
H1
CO2 Absorber Canister
Support
When rotating the canister disassembly knob, the support
will rise or fall
3.4.1 CO2 absorber Canister installation
(1) Make sure that the CO2 absorber canister is filled with qualified soda lime, and the
filling amount does not exceed the MAX marking line。
(2) Make sure the lock and unlock handle of the CO2 absorber canister is in the
vertical position as shown in Figure 3-4-1。
(3) Place the CO2 absorber canister on the canister support, making sure to place it
on the center of the positioning。
(4) Rotate the lock and unlock handle of the CO2 absorber canister clockwise to the
horizontal position shown in Figure 3-4-2。
(5) CO2 absorber canister is installed, as shown in the figure after installation 3-4-2。
3.4.2 CO2 absorber canister disassembly
The dismantling of the CO2 absorber canister also refers to the replacement of soda
lime. The operation steps are as follow:
(1) Before disassembly of the CO2 absorber canister as shown in the figure 3-4-2
(2) Lock and unlock the canister and turn it counterclockwise to the vertical position,
image 3-4-1
(3) Remove the CO2 absorber canister
(4) Pour out the used soda lime from the CO2 absorber canister (dispose of soda lime
in accordance with hospital and local regulations on contaminants and biohazards)
(5) After the disassembly of the CO2 absorber canister is completed, if you want to
continue to use the equipment, or before using the equipment next time, please
operate according to 3.4.1 and install the CO2 absorber canister.
3.5 Installation and disassembly of waste gas canister
and waste gas recycling pipe
The schematic diagram of the waste gas canister and the waste gas recycling pipe before
installation is as follows, please refer to the diagram after installation 2-3。
Table of contents
Other Praxisdienst Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Top shelf Orthopedics
Top shelf Orthopedics Economy Hinge Knee Instructions for use

SAM MEDICAL
SAM MEDICAL BLEEDING CONTROL KIT manual

Hitachi
Hitachi S3ESEL instruction manual

Top shelf Orthopedics
Top shelf Orthopedics RecoveryWrap System Instructions for use

Thuli Tables
Thuli Tables The Junior Portable owner's manual

AAT
AAT c-max U1 C1414 Translation of the original user manual