PremiumLift AirCare Max-Pro ACMaxPP User manual

AirCare Max-Pro
Digital Pump
User Manual

2www.premiumlift.com.au

E
sales@harvesthealthcare.co.uk www.harvesthealthcare.co.uk
CONTENTS
WARNINGS & CAUTIONS
GENERAL INFORMATION
1
DEFINITION OF THE GROUPS MENTIONED
2
NON-COMPLIANT USE
3
SAFETY INSTRUCTIONS
3.1
GENERAL SAFETY INSTRUCTIONS
3.2
SAFETY INFORMATION FOR THE OPERATOR
3.3
SAFETY INFORMATION FOR THE USER
3.4
SYMBOLS USED
3.5
CLEANING & DISINFECTION
3.6
SERVICING & MAINTENANCE
3.7
SERVICE LIFE & DISPOSAL
TECHNICAL SPECIFICATION
OVERVIEW
OPERATION
CLEANING & CARE
ROUTINE MAINTENANCE & SERVICING
TROUBLESHOOTING
PARTS LIST
GUARANTEES & WARRANTIES
NOTE
4-5
6-10
7
7
7
7
8
8
9
10
10
10
11
12 -13
14
15
16-17
18-21
22
23
24

4
www.premiumlift.com.au
READ THIS INSTRUCTION MANUAL AND OBSERVE SAFETY
INSTRUCTIONS.
WARNING
•
This pump must be properly installed and operated as directed by this user manual.
•
The pump should bechecked regularly toensurecorrect operation. Loss of function
will remove all pressure relieving properties.
•
Thispump,whenconnectedto a pressure relieving system,is intended foruseas part
of a pressure ulcer prevention program; do not rely solely onthis device to achieve
theresult.Themedicalprofessionalisresponsibleforapplyingbestmedicaljudgment
when using this system.
•
Select the correct setting for the occupant’s weight andtherapy required. Care should
be taken not to accidently change pressures once set as the effectiveness of the
therapy may be reduced.
•
All hoses must befree of kinks, twists andmust be properly connected and positioned
so as not to cause any obstruction.
•
Do not positionthe pump in a way that prevents access to the disconnection device
(mains power plug).
•
Ensure the mains lead or pump cannot become trapped orcrushed, e.g. by raising or
lowering of bed or bed rails or any other movingobject.
•
Check the mains lead is damagefree and positionedso as not to cause an
obstruction or injury e.g. strangulation or triphazard.
•
Ensure that the electricity supply is of the type stated on the pumpunit.
•
Do not place pump on or near a heat source or cover withbedding.
•
PremiumLiftadvise againstsmokingwhilstthesystemis inuseto preventthe
accidental ignition of associated items which may be flammable, such as bed linen.
Weadvise a risk assessmentiscarriedout before using any medical equipment to
protect the user and service user.
WARNINGS &
CAUTIONS

www.premiumlift.com.au
5
•
Do not expose the pump to liquids.
•
Wireless equipment such as mobile phonesshould bekept at least 10ft/3m away from
the system.
•
The mattress/cushion and pump should be cleaned between patientuses.
•
Do not use bleach, chlorine releasing agents in concentrations over 1000 ppm,
solventsoralcohol-basedcleansers,e.g.Phenicol,Hibiscrub,Clearsol,Stericoland
Hycolineasthesewilldestroythemattressmaterial.Fullcleaninginstructionscanbe
found on page 15.
•
Suitableforcontinuoususe.
•
Do not modify the mattress/cushion or pump unit in anyway.
•
Do not connect to any other medical device orequipment.
•
Not for use in an oxygen enriched environment.
•
Not for use in an outdoor environment.
•
Store the system in a clean and dry environment, out of directsunlight.
Electrical equipment can be hazardous. Only authorised technical
personnel should remove the rear pump case for maintenance.
Removal of the case by unqualified personnel will invalidate the
warranty.
Before cleaning the unit ensure that the electrical supply to the pump
has been disconnected by removing the plug from the power supply.
WARNINGS &
CAUTIONS

6
www.premiumlift.com.au
BEFORE USING THIS SYSTEM FOR THE FIRST TIME:
-
Read through this instruction manual conscientiously from start to finish.
-
Note the safety instructions which must be observed.
PremiumLiftproductsareISO Certified and TGA Registered andmeetallsafetyand
functionality requirements.
These safety requirements can only be met if the user is satisfied with the proper condition of
the product (including accessories) before use.
GENERAL INFORMATION
AirCare Max-Pro is a digital pump designed to provide controlled compressed air to a
pressure care mattress. The inflated cells within the supplied mattress
providean
alternatingpressuresurfacedesignedtoreducesurfacepressureontheskinand therefore
reduce risk of the development of pressure ulcers in susceptable patients.
This pump has audibleandvisualalarms forlow pressure, powerfailure andcyclefault,static
mode and a lock out facility. Extra features include maxi firm mode and four
choices of cycle
time - 6, 9, 12 and 25 minutes.
Harvest Healthcare supply a pump and mattress but it is the user’s responsibility to ensure
each system meets the guidelines set out in the pump and mattress guide (shown below).
Failureto do this could put the service user at risk. This could prolong the recovery process or
increase the risk of further tissue damage.
Forinformationonourmattressrangepleaserefertothemattressuser manualsavailableon
our website www.premiumlift.com.au
GENERAL
INFORMATION

www.premiumlift.com.au
7
1
DEFINITION OF THE GROUPS MENTIONED
OPERATOR
Anoperatoris any natural orlegal person who uses theequipment oron whoseinstruction
itisused(e.g.nursinghomes,specialisedretailers,healthinsurancecompanies,medical
suppliers).
USER/CARE PERSONNEL
Users are persons who as a result of their vocational training, experience or briefing are
authorised to operate the equipment.
Furthermore,theuser/carepersonnelcanrecogniseandavoidpotentialdangersandassess
the clinical condition of the service user.
PATIENT/OCCUPANT/SERVICE USER
The person in need of pressure care support, due to being identified as at risk of developing
of pressure ulcers by a suitably qualified carer or other.
QUALIFIED PERSONNEL
Qualified personnel areemployees of the operator who as a result of their vocational
training or briefing are entitled to deliver, assemble, disassemble and transport the product.
2
NON-COMPLIANT USE
Allusesdeviatingfromtheintendedpurpose,whichmay alsobe hazardous as aresult.
This includes for example:
•
Incorrect installation.
•
Operation by persons who have not been instructed in itsuse.
•
Using the system with non-approved parts/accessories.
•
Using the system if any of the components are damaged orfaulty.
3
SAFETY INSTRUCTIONS
3.1
GENERAL SAFETY INSTRUCTIONS
During the briefing, specific attention must be drawn to any potential dangers
which can occur despite correct operation. Before putting the product into service
for the first time, the Instruction Manual must be read conscientiously and in detail
by the user/care personnel.
Programmingof thesystemshouldbe carriedout by competenttrainedpersonnel.
Use only spares and accessories supplied or approvedby PremiumLift.
Only suitably trained personnel are allowed to operate the system.
GENERAL
INFORMATION

8
www.premiumlift.com.au
The mains cable must be free and not be allowed to be caught up in the bed’s
moving mechanisms. The mainscable may be damagedor itmay bepulled out of
its socket and electric leads exposed as a result.
Ifthe mains cable orthe mains plug are damaged, the relevant part must be
replaced. This work shouldbe carried outby the manufacturer orauthorised
service agents.
When connecting the mains plug do not use multiple sockets since liquids may
penetrate into these (fire hazard and electric shock).
3.2
SAFETY INFORMATION FOR THE OPERATOR
With the help of this instruction manual, instruct each user in the safe operation of
this system before it is put into service for the first time.
Advise the user of any hazards which may occur if not handled correctly.
Onlypersonswhohavebeenproperlyinstructedmayoperatethissystem.This
also applies for persons who only operate the system on a temporary basis.
3.3
SAFETY INFORMATION FOR THE USER
Ensure that the operator instructs you in the safe operation of this system.
In addition, pay particular attention to the ‘Warnings and Cautions’ (page 4-5) and the
‘General Safety Information’ as described in 3.1.
Ifthere is a suspected fault or damage, unplug the mains plug from the socket and follow
the power down procedure (page 16). Clearly mark ‘Out of Order’, take out of service
immediately and inform the person in charge withoutdelay.
GENERAL
INFORMATION

www.premiumlift.com.au
9
3.4
SYMBOLS USED
Thissymbolindicatesgeneralhazards.Thereisadangertolifeandhealth.
ConformitymarkinaccordancewiththeMedicalDeviceDirective(93/42EEC).
The symbol for Protection Class II device, doubleinsulated.
The symbol for Medical Device according to EN 60601
Handle with care.
This way up.
Keep dry.
Recycling symbol. Refers to packaging that can be recycled(cardboard).
Fragile, handle with care.
This product must be disposed of in a designated refuse bin for waste
electronicdevices(WEEE)intheEuropeanUnion.Donotdisposeofasnormal
domestic waste.
No smoking. No naked flames.
Read instructions / consult manufacturers guide.
This symbol indicates electrical hazards. There is danger to life and health.
GENERAL
INFORMATION

10
www.premiumlift.com.au
3.5
CLEANING & DISINFECTION
Do not immerse electrical components in water but wipe with a damp cloth only.
The electrical components must not be cleaned with a high-pressure cleaner or
water jet. Disinfection by wiping only is allowed.
Full cleaning and disinfection instructions can be found on page 15.
3.6
SERVICING & MAINTENANCE
Servicing must only be carried out by qualified personnel.
A technical check and/or safety inspection must beconducted at least once a year
or after a lengthy break in use.
Any defects, damage or signs of wear must be rectified without delay. Only original
spare parts from PremiumLift may be used, otherwise all guarantees or
warranties will be invalidated.
See pages 16-17 for ‘Routine Maintenance’ and ‘Routine Servicing’.
3.7
SERVICE LIFE & DISPOSAL
The systemmust not bedisposedof as normal domestic waste after its service life,
but must be disposed of in a designated refuse bin for waste electronic devices
(WEEE)in the European Union. Do not dispose of as normal domestic waste.
Ourfullterms& conditionsincludingproductwarrantiesareavailableby requestorcan be
found on our website www.premiumlift.com.au
PARTS AND DATA MAY UNDERGO FURTHER DEVELOPMENT AND
THEREFORE DEVIATE FROM THE DETAILS GIVEN.
GENERAL
INFORMATION

www.premiumlift.com.au
11
AirCare Max-Pro TECHNICAL SPECIFICATION
Pump Model No.
Operating Cycle
Dimensions
Weight
Air Flow Output
Pressure Setting
ACMaxPP
6 / 9 / 12 / 25 minutes
345 x 115 x 185mm
3.5
kg
14 lpm
Manual
PUMP POWER REQUIREMENTS
Power Rating
Voltage
Fuse
Medical Classification
Safety Standards
13W
AC 240V / 50Hz
T 1AH 250V
Medical Device Class 1
EN 60601-1. EN 60601-1-2. ARTG No.398711
SPECIFICATION

12
www.premiumlift.com.au
OVERVIEW
PUMP OVERVIEW
A Feed Tube Connection
B Main Power Switch
C Cycle Times 6/9/12/25min
D PressureRange
E Alarm Mute
F Alternating Mode (4 Cycle Times)
G Static Mode (4 Lights ON)
H Maxi-Inflate Mode
I Comfort Level (kg)
J Lock/ Unlock Button
A
Feed Tube Connector
Pipes from the mattress connect here. The pump must be able to be disconnected in
an emergency for CPR to be administered.
B
Power Switch
The power switch is on the left hand side of the pump. When pressedto the ON
position the switch will illuminate to indicate that there is a power supply to the
pump. With the switch in the ON position press the power button on the control panel
to activate the pump. Toturn the pump off press the power switch on the control
panel.
C
G/F
D
I
K B J H E
A

13
www.premiumlift.com.au
C
Cycle/Alternating Time
The cycle time can be selected by pressing the button on the control panel. The
cycle time options are 6, 9, 12or 25 minutes. Consult a medical professional for
cycle times above ten minutes. The pump always start with 6min cycle.
D
Pressure Range
Pressure range digital display 30-80 mmHg
E
Alarm Mute
Thepump isfitted with an audibleand visual alarm. Press the ‘Mute’button to
reset the alarm. Initiallythe alarm will sound after 40mins if themattress hasnot
fully inflated while inthe start-up mode. Once safely inflated the alarm will sound if
it detects a cycle fault or low pressure fault on thesystem.
F
Alternating Mode
When pump is switched on for the first time it will inflate all the cells.
The alternating mode LED will illuminate to indicate that the pump is in low-pressure
mode.
Once all the cells have been inflated the alternating mode will activate (normal
pressure mode). The cells in the mattress will then inflate and deflated to prove
pressure relief for the service user.
G
Static Mode
The ‘Static Mode’ can be operated by the user to inflate all the cells for patient care
andtransfers.Press the‘Cycle Time” buttonuntil all 4 LEDwillilluminategreen for
“Static Mode”
H
Maxi-Inflate Mode
Press the ‘Maxi-Inflate’button to automatically inflate the mattress to its maximum
pressure.Thiswilltakeapproximately20mins.Duringnormalpressureoperationsthe
pump will monitor the pressures.
I
Comfort Level (kg)
The pump has a scale printed on the front panel in kilograms. Press the comfort
control button (-) or (+) until the nearest desired weight setting is illuminated on
the display panel. When the mattress is in use the user can adjust the comfort
setting as required for the service user’s needs. A slight increase of the comfort
setting may be required if the service user needs more support when the bed is
profiled. When the bed platform is laid flat readjust the setting as required for the
service user.
J
Lock Button
Press “Lock” to lock pump panel.
Press “Unlock” to unlock pump panel.
K
Bed Hooks (Back of Pump)
Attach the pump to the foot end of the bed. Adaptor brackets are available to fit
different sizes of beds. If the pump cannot be attached to the foot end of the bed it is
the user’s responsibility to ensure that the pump is placedin such a position that it will
not cause harm or injury to the service user. The pump unit must be placed so
that it can be clearly seen and monitored at all times.
OVERVIEW

14
www.premiumlift.com.au
INFLATING THE MATTRESS/CUSHION
1
Plug in at the mains supply, press the ON button on the side of the pump and then
on the control panel. The control panel will light up to indicate that there is a power
supply.
2
The pump is in start up mode. Once the mattress is fully inflated it will be ready to use.
While in start up mode do not select static mode. This will cause the pump to
alarm. If the pump alarms, switch off the pump and switch it back on to reinflate the
mattress.
The mattress is now ready to be used.
Set the pump to run at the correct pressure to suit the weight of the
service user. Refer to the Pump Control Panel guide on the front of
the pump.
To turn off the system press the power button to OFF and unplug from
the mains supply.
TRANSPORT MODE
If the pump needs to be disconnected for any reason attach the transport cap to the feed
tubes connector. This will prevent deflation retaining the remaining air within the mattress.
The alternating action stops in this mode.
SETTING UP PROCEDURE
It is important to follow the correct setting up procedure to ensure the patient receives
sufficient support whilst achieving maximum pressure relief and comfort.
Failure to follow this procedure could result in the patient being put at risk.
1
When the mattress is fully inflated set the pump to the approximate weight of the
serviceuser.Positiontheserviceuserinthelyingposition.Ifrequiredadjusttheweight
setting to meet the patient’s comfort needs.
2
Ensurethatthecomfortsettingis alteredcorrectly whenthe patient moves from alying
to a sitting position andviceversa.
NOTE: The static mode is only to be used for patient care. Switch off after use.
OPERATION

15
www.premiumlift.com.au
PUMPS
For general cleaning wipe with a soft cloth dampened with a mild detergent and water
solution. This may befollowedby either wiping with asodium hypochlorite solutiontoa
dilution of 1000ppm (parts per million) or by using alcoholwipes.
DO NOT USE HYPERCARBONATE, PHENOL-BASED CLEANING
SOLUTIONS, ABRASIVE COMPOUNDS OR CLEANING PADS.
NOTES FOR CARE OF ASSOCIATED PRODUCTS
•
Followingtheuseof adetergentand/ordisinfectantsolutionrinsethe mattresscover
with clean water using a clean cloth and allow to dry.
•
Frequentor prolongedexposuretohighconcentrationsof disinfectantsolutionswill
reduce the useful life of the mattresscover.
•
Where high concentration disinfectants e.g. > 10,000ppm chlorine releasing
agent(e.g.Haztaborbleach)orcombinedcleaning/chlorinereleasingagent(e.g.
Chlorcleam, Actichlor) and detergent solutions are used to remove blood or other
body fluids, mattresses should be thoroughly rinsed with clean water to remove any
residues.Thiswillhelp preventany longtermcompatibilityissuesassociatedwith
disinfectant residues.
•
Alternatively,disinfectionmaybeachievedbylaunderingattemperaturesnot
exceeding 80°C for 10 minutes which may include a chlorine rinse.
TRANSPORT & STORAGE
Storage conditions as follows:
–15 °C without relative humidity control and +40 °C at a relative humidity up
to 93%, non-condensing. Atmospheric pressure range of 700 hPa to 1060 hPa.
Suitable for all standard modes of transport when in the correct packaging.
Operation conditions:
A temperature range of +5 °C to +35 °C and relative humidity range of 15% to 93%, non-
condensing.Operationalatmosphericpressure700hPato1060hPa.Suitableforpollution
degree 2. Operational altitude 2000 m.
Transportation of the mattress system:
Themattressshouldbe looselyrolledlengthwisewiththecover innermost,takingcare not
to strain the feed pipes. It can then be stored / transported in the carry bag with the pump,
mains cable and this booklet. Do not stack bagged mattresses more than two high.
CLEANING & CARE

16
www.premiumlift.com.au
These checks should be carried out at each decontamination process, i.e.
between patients or patient occupancy and weekly for longer term patients.
PUMP
1
Check the pump casing for cracks or other damage that could bedangerous.
2
Check the power cord (ensure there are no barewires).
Ifanyfaultsaredetectedremovetheequipmentfromuseand contacttheHarvestHealthcare
Service Department or your distributor.
OTHER COMPONENTS
•
Check air cells and mattress interior for signs of damage or contamination, e.g.
staining or fluid ingress at each decontamination process, i.e. between patients or
patient occupancy (or weekly for longer termpatients).
•
Theindividualcellscanbewipedcleanwithamildantisepticsolution.
•
AllcellsarereplaceableandcanbesourcedfromPremiumLift.
POWER UNIT
Disconnectthepowerunitfromtheelectricitysupplybeforecarryingoutmaintenance,repairs
or cleaning.
Check all electrical connections and power lead for signs of wear and damage.
The power unit can be wiped down with a detergent or disinfectant solution. Do not use
solvents. Unsuitable for sterilisation.
Theuseofinappropriatecleaninganddetergentmaterialsonmedicalequipmentcould damage
surfacesandmay compromisetheability todecontaminate medical devices adequately ormay
interfere with device function.
When the pump or mattress is no longer required dispose of the equipment in accordance
withthelocalregulations.WEEErequirementsdoapplytothepumpunit andanyelectrical
components including cables which are used for or with thisproduct.
MAINTENANCE

17
www.premiumlift.com.au
SERVICING YOUR SYSTEM
The AirCare Max-Pro pump should beserviced every 12 months by PremiumLift approved
personnel using genuine PremiumLift spareparts.
Failure to follow the service schedule may invalidate future
warranty claims (‘Guarantees & Warranties’ can be found on page
23).
SERVICING

1
8
T
+4
4
(0)
17
09
37
71
72
F
+4
4
(0)
17
09
37
71
73
TROUBLESHOOTING
FAULT
CHECK THAT
STAGE 2 CHECK
IF PROBLEM PERSISTS
Pump shows no
indication that it is
powered up.
1.
Mains plug is plugged inand
power switched on.
2.
The power switch onthe
pump is switched on.
3.
Thefusein themains plugis
not blown.
4.
The wall socket that the
pump is connected to is
working correctly.
1.
Connect the pump to the
nearest(working)mainsoutlet.
2.
Replace theplugfuse withthe
correct fuse.
3.
Trya differentdeviceinthe
mains outlet.
Contact PremiumLift technical
support.
Before calling:
Pleaseensureyouhavetheserial
numberandmodelof equipment.
Please record details of the results of
the recommended tests. (Notes pages
are provided at the back of this user
manual).
18

FAULT
CHECK THAT
STAGE 2 CHECK
IF PROBLEM PERSISTS
The pump appears
to be running but the
mattress is not inflating
correctly and or the
low pressure light is
illuminated.
PLEASE NOTE
Inflation can take up to 30
minutes.
1. The hoses are routed
correctly (not kinked) and
connected to the pump
correctly.
1.
Disconnect and then re-
connect the hoses to the outlet
on the side of the pump.
2.
OpenthenreclosetheCPR
valve,makesurethevalve
is not trapped in the bed
mechanism.
Contact PremiumLift technical
support.
Before calling:
Pleaseensureyouhavetheserial
numberandmodelof equipment.
2. The CPR valve is not trapped
and is in the closed position.
Please record details of the results of
the recommended tests. (Notes pages
are provided at the back of this user
manual).
1.
Therearenoleaksinthe
mattress.
2.
Thetubesinthemattress are
not disconnected orkinked.
1. Replace any damaged or
leaking mattress parts with
the correct genuine Harvest
Healthcare spare parts.
2. Straighten out any kinked
pipes and reconnect any
disconnected joints.
19
www.premiumlift.com.au
19
19
20

TROUBLESHOOTING
FAULT
CHECK THAT
STAGE 2 CHECK
IF PROBLEM PERSISTS
Some of the cells
appear to be deflated.
This is normal for alternating pressure therapy.
Themattressis madeup of individual aircells. Thealternating sectionis split up into2 sections consistingof odd
cells e.g. 1,3,5 etc and even cells e.g. 2,4,6 etc. These two sections will alternate through a 10 minutecycle in
which time both sets of alternating air cells will have inflated and deflated sequentially.
The system does
not appear to be
alternating.
1.
Check that the static
mode is not turned on.
2. Carefully mark one ofthe
inflated cells with a pen.
3. Ensure that there are no
kinks in the pipeworkdown
the side of themattress.
4. Check the cycle time on the
controlpanel.Recommended
setting 10 minutesunless
a health professional has
recommended otherwise.
1.
Monitor the cell for 7minutes
to see if it deflates.
2.
Straightenoutanykinked
pipes.
Contact Harvest Healthcare technical
support.
Before calling:
Pleaseensureyouhavetheserialnumber
and model of equipment.
Please record details of the results of the
recommended tests. (Notes pages are
provided at the back of this user manual).
20
20
www.premiumlift.com.au
Table of contents
Other PremiumLift Medical Equipment manuals
Popular Medical Equipment manuals by other brands

vygon
vygon Plexygon user manual

Romidan
Romidan RomiApex A-15 user manual

LKC TECHNOLOGIES
LKC TECHNOLOGIES RETeval Quick reference guide

Chattanooga
Chattanooga Intelect TranSport 2783 user manual

Mettler Electronics
Mettler Electronics Sonicator Plus 992 Maintenance manual

Planmeca
Planmeca ProX user manual

Ferno
Ferno 26-S Use and maintenance manual

swissflex
swissflex uni 20_75RF operating instructions

Weinmann
Weinmann MEDUMAT Standard2 Step-By-Step Instructions

Invacare
Invacare 1085140 Assembly, installation and operating instructions

Chummie
Chummie TCE25 Users manual & quick start guide

Blatchford
Blatchford Silcare Walk SWTTCP Series Instructions for use