Prius Healthcare USA Duet User manual

Content
1. The Purpose of this Manual....................................................................................................1
2. Product Description...................................................................................................................1
3. Technical Data..............................................................................................................................4
4. Operation Instruction ...............................................................................................................5
5. Cleaning...........................................................................................................................................8
6. Storage and Care .........................................................................................................................9
7. Maintenance and Troubleshooting .................................................................................. 10
8. EMC Related Notification...................................................................................................... 11
9. Expected Service Life ............................................................................................................. 14
10. Warranty................................................................................................................................... 14

Warning
❖Connect the Master Control unit to a proper power source.
❖Don’t use the system in the presence of any flammable gases.
( such as Anesthetic Agents)
❖Keep the pump and mattress away from open flame.
❖Keep sharp objects away from the mattress.
❖The device is not AP/APG protected.
❖Do not place a heating device on or close to the mattress system.
Caution
❖*
The Alternating System should always be used in accordance with your
Institutions pressure care guidelines.
❖*
Re-positioning of the patient is always recommended when using an
alternating pressure air mattress (APAM).
❖*
The Control unit can only be repaired by an authorized technician.
❖*
Do not drop the control unit.
❖
Do not store the system in direct sunlight or extreme cold conditions
❖*
Operation Temp: 15-35°C ( 59-95℉) R.H. : 30-75 %

1
1. The Purpose of this Manual
This operation manual is mainly focused on the set up, cleaning and routine
maintenance of the Duet LAL Alternating Anti-Decubitus System. We
recommend you keeping this manual handy to answer most of the question
related to the system and please read the whole manual before setting up.
2. Product Description
The Duet LAL Alternating Anti-Decubitus System is a unique and innovative
specialty mattress replacement unit. The system utilizes true low air loss
technology with a high flow rate that provides pressure management for the
treatment of pressure ulcers. It features continuous lateral rotation therapy in
40 degree, which gently turns the patient from side to side to significantly
lower the risk of infection, pneumonia and other pulmonary complications –
illnesses that significantly ad to patient care costs and length of stay.
The Duet Features
⚫User-friendly controls
⚫Large LCD display on each function status
⚫Rapid CPR deflation
⚫Patient Care mode provides quick maximum inflation within seconds to
help transfers and nursing procedures
⚫Incorporate sensor technology with Auto mode to constantly monitoring
the mattress pressure based on inputting patient’s height and weight
⚫Lock out function avoids tampering with settings
⚫Fowler mode gives added support to help prevent bottoming out while
patient is in sitting position
⚫Highly vapor permeable and oversized pliable quilted nylon top cover
providing low shear, friction and moisture protection
⚫Low Air Loss, Pulsation, Alternation, Static and Lateral Rotation for
outstanding pressure redistribution outcomes
⚫2“ convoluted foam base for addition safety support Recommended
maximum safe working load as following:
➢36“: 350lbs
➢42“: 500lbs
➢48“: 600lbs
Intended Users
•Healthcare professionals or caregivers who are at least fifteen years in age,
with the ability to read and understand English and Westernized Arabic
Numerals.
•This device should not be operated by patients.

2
Therapy Mode
Selection for
Alternation,
Pulsation and
Static
LCD Display
Rotation Therapy
Selection
Maximum
Inflation
Cycle Time
Preset Key
Power Switch
Audio Alarm
ON/OFF
Auto
Panel Lock
Power Failure
Sensor Disconnect
Bottom Out
Patient’s
BMI Setting
Comfort
Setting
Caution
Alternating pressure therapy is not recommended to patient who has
serious pain or pain-sensitive symptom. In this case, we recommend the
application of foam mattress which can be found in MOXI
ENTERPRISES, LLC product range.

3
LCD Display introduction
LOCK
STATIC
FULL TURN
ALTERNATION
CYCLE TIME
MUTE
PULSATION
LEFT TURN
TURNING
CYCLE TIME
FOWLER
ALTERNATION
RIGHT TURN
HEIGHT
AUTO
PATIENT CARE
NO TURNING
FUNCTION
WEIGHT
BOTTOM
OUT

4
3. Technical Data
Master Control Unit
Model No.
FC-MOX0027
Model Name
Duet
Size (inch) LxWxH
12.2“x 6.7“x 13.2“
Weight
14.3lbs
Cycle Time (min)
3 - 95
Min Operating Pressure
8 +/- 5mmHg
Max Operating Pressure
33 +/- 5mmHg
Rated Voltage
AC 110-120V
Max Current
5 Amp
Fuse Rating
T5AH 250V
Rated Frequency
60 Hz
Classification
Class I, Type BF
Not AP/APG type
Mode of Operation
Continuous
Environment
(Temperature)
Operation: 15°C to 35°C (59°F to 95°F)
Storage:5°C to 60°C (41°F to 140°F)
Environment (Humidity)
Operation: 30% to 75% non-condensing
Storage: 30% to 90% non-condensing
Standard
IEC 60601-1,
CAN/CSA C22.2 No. 60601-1,
IEC 60601-1-2
Mattress Replacement
Model No
FM-MOX0009(36“)
FM-MOX0044(42“)
FM-MOX0045(48“)
Size (inch)
36”(W)x80”(L)x10”(H)
42“ (W)x80”(L)x10”(H)
48“ (W)x80”(L)x10”(H)
Weight (lbs)
39.6
Cells Number
18 cells
Cells Material
Nylon coated with PU
Cover Material
Nylon woven fabric w/ PU coating finish
Base Material
Woven Polyester fabric w/ PVC backing

5
Symbol Definition
4. Operation Instruction
1. Power On/Stand-By
Plug the power cord to the socket and switch on the power at control
panel, the orange LED will illuminate, it means Stand-By. Press the
button Power On/Stand-By button, the LED light will turn to green,
and the control unit will start to operate.
2. Mode Selection
Under normal operating, press mode button switches to select therapies. Each
mode needs manually switch. Please must have healthcare professionals or
caregivers to check the setting every two hours.
2.1 Static Mode
The system will only provide low air loss therapy.
2.2 Alternate Mode
The system will provide 3-1 altternation. Press "CYCLE" button and
adjust the alternation cycle time by pressing the +/- button. The
cycle can be set from 3 to 95 minutes.
Refer to Accompanying Documents
Waste Disposal
Type BF Applied Part
Alternating Current
Caution
Keep Dry
Manufacturer

6
2.3 Pulsation Mode
The system will provide pulsation pressure-relieving therapy.
3. Patient Care
3.1 Patient Care provides quick maximum inflation within seconds to help
transfers and nursing procedures.
3.2 Press the Patient Care button for fast
inflation.
3.3 When the mattress is fully inflated, the caregiver can transfer the patient
onto the mattress.
3.4 Press Patient Care button again to return previous setting.
3.5 When Patient Care function is activated for over 30 minutes, the system
will default back to previous setting automatically.
4. Turning Function
4.1 Press the TURN button to select from LEFT/RIGHT/FULL TURN to enable
rotation pressure-relieving therapy.
4.2 Press CYCLE button to set for rotation cycle
time from 3 to 95 minutes and hold postion
“Hd“.
4.3 Hold function is only engaged in Left and Right Turn
5. Auto Function
5.1 When Auto Function is activated, control unit is automatically optimizing
patient’s comfort setting base on patient’s BMI input. The BMI setting is the
guideline and the proper adjustment of the pressure level will be applied
according to individual patient.
5.2 Press SET button, when both the height and weight indication flash, press
+/- to adjust the metric or imperial unit.

7
5.3 Press SET button, when the height indication flash, press +/- to adjust the
height of the patient.
5.4 Press SET button, when the weight indication flash, press +/- to adjust
the weight of the patient.
5.5 Press SET to exit menu when finish.
6. Alarm On/Off
Press Alarm on/off to switch on/off the audio alarm
7. Lock Button
When th pump has been activated, the pump will automatically lock
on after 2 minutes no operation, and press the lock On/Off button for
3 seconds to unlock.
8. CPR
The air hose connectors can be disconnected from the controller
to quick release the air when in an emergency situation where
CPR is to be performed.
9. Fowler mode
While the patient is in sitting position over 30 degrees, the fowler
mode will be activated automatically.
10.Comfort Level Setting
Press the +/- to adjust the comfort level. The following steps are the suggested
pressure level settings when the mattress in alternation mode only.
⚫Make sure the mattress already fully inflated and connected to the control
unit.
⚫With the patient on the bed, healthcare professionals or caregivers press
the SET button to set auto detect function finding the best matching level
based on the user weight and height.
⚫Suggested comfort level can be adjusted according to
patients‘ requirements with no less or more than 2 levels based on the auto
detected level setting.
⚫Suggested weight for 36“ mattress is no more than 350 lbs; 42“ mattress is
no more than 500 lbs and 48“ mattress is no more than 600 lbs.
⚫Suggeseted cycle time is no more than 15 minutes
Please be aware that once patients have any uncomfortable
feelings or symptoms, patients must notify healthcare professionals
or caregivers to change product settings suit for patients’ condition.

8
5. Cleaning
The Mattress
The mattress should be cleaned on the bed weekly using a damp soft cloth and
mild detergent. If top cover or base cover becomes grossly soiled, put on clean
gloves, plastic gown and eye protection before removing top and base covers and
disposing according to standard hospital procedures for contaminated waste and
replace with clean covers. The mattress should check and clean each time before
use or once a week.
Covers can be washed and thermally disinfected in a washing machine by
following below procedure: (Never use phenol based cleaning solutions).
Industrial
Break washes
Cold
10 minutes
Main washes
60°C (140°F)
6 minutes
Main washes
70°C (158°F)
10 minutes
Extraction
2 minutes
Cold Rinses
Extraction
5 minutes
Domestic
Pre-wash
Cold
Main Wash
70°C (158°F)
10 minutes
Extraction
2 minutes
Cold Rinses
Extraction
5 minutes
Tumble Drying or Tunnel Drying is not recommended.
Mattress Cells can be wiped over with a solution of sodium hypochlorite1000ppm
or any other non-phenolic germicidal solution.
The Master Control Unit
The pump unit should also be cleaned weekly using a damp soft cloth and mild
detergent.
The pump casing is manufactured from ABS plastic and if the case is soiled the
pump can be wiped down with a sodium hypochlorite solution to dilution of
1000ppm or any EPA- approved hospital grade disinfectant. (Do not use phenol
based cleaning solution).
The air filter should also be cleaned and checked as often as possible at a
minimum of every six months. Air Filter can be removed by pinching center of
the filter and pulling outward from the back of the control unit.
Caution
SWITCH OFF THE ELECTRICAL SUPPLY TO THE PUMP AND DISCONNECT THE
POWER CORD FROM THE MAIN SUPPLY BEFORE CLEANING AND INSPECTION

9
Replace Air Filter
1.Remove air filter and replace with a new one.
2.Use a soft bristle to remove dust and difficult dried-on soil.
NOTE:
1. Do not use phenol based cleaning solutions.
2. Switch off the electrical supply to the pump and disconnect the power cord
from the main supply before cleaning and inspection)
Waste Disposal
This Product has been supplied by an environmentally
conscious manufacturer that complies with the WEEE.
This product may contain substances that could be harmful to
the environment if disposed of in places that are not approved
by your state, local or federal laws. Please be environmentally
responsible and recycle this product through your recycling
facility at its end of life.
6. Storage and Care
Master Control Unit:
⚫Check the power cord and plug for abrasions or excessive wear.
⚫Plug in the unit and verify air flows from the units hose connection ports.
⚫Place in plastic bag for storage.
Mattress:
⚫Check the air manifold for kinks or breaks. Replace if necessary.
⚫Twist the CPR plug at the head of the mattress and disconnect the air feed
tubes. All the air will now be expelled. Starting at the head end, the
mattress can now be rolled. Use the base mounted straps for
containment.
⚫Place in plastic bag of storage.
It is recommended the following guidelines are used whenever this system is
being stored or transported another location:
Temperature limitations: 5°C ~ 60°C
Relative Humidity: 30% to 90%
Filter

10
7. Maintenance & Troubleshooting
No daily maintenance is required. It is intended this equipment should only be
serviced by properly qualified, authorized technical personnel. In case of minor
trouble please refer as following Troubleshooting.
Symptom
Inspection Procedures
Possible Solution
Air is pumping out from
the control unit but
mattress is not inflating.
1. Is the power source
correct? Improper voltage
may cause the pump to
function abnormally and
damage the control unit.
2. Is there any kinking tube?
3. Is there any air leakage
from air cells?
4. Is there any air leakage
from air tube between
mattress and control unit?
5. Has the air tube connector
been connected properly?
1. Use power regulator
2. Adjust the air tubes to
enable smooth air flow.
3. Replace with new air cells.
4. Replace with new air tubes.
5. Re-connect the air tubes.
6. Refer to service if problem
persist.
The control unit is not
functioning
1. Check the power cord and
the power voltage
2. Check the fuse.
1. Use a power regulator.
2. Replace with a new fuse
3. Refer to service if problem
persist.
Some of the air cells are
not properly inflated.
1. Is the connection between
air cells and the manifold
kinked?
2. Is there any air leakage
from the air cells?
1. Check for any kinking
between air cells and
manifold.
2. Replace new air cell if faulty.
3. Refer to service if problem
persist.
Sensor Disconnect
1. Check the sensor pad to
see if the sensor pad
connect properly?
2. Check the sensor pad to
see if there is any damage
or broken on the sensor
pad.
1. Connect the sensor pad to
the properly
2. Replace a new sensor pad
3. Refer to service if problem
persist.
Bottom Out
1. Check the tubing to see if
the mattress connect with
the pump propoerly?
2. Check the hardness of the
mattress to see if the
mattress is too soft?
3. Check the sensor pad to
see if there is any damage
or broken on the sensor
pad.
1. Connect the tubing to the
pump properly.
2. Adjust the pressure setting
or replace a new mattress.
3. Replace a new sensor pad.
4. Refer to service if problem
persist.

11
8. EMC Related Notifications
Guidance and manufacturer’s declaration – electromagnetic emissions
The Duet control unit is intended for use in the electromagnetic environment specified below. The
customer or the user of the Duet control unit should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment –guidance
RF emissions
CISPR 11
Group 1
The Duet control unit uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Duet control unit is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Recommended separation distances between portable and mobile RF
communications equipment and the Duet control unit
The Duet control unit is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Duet control unit can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Duet control unit as recommended below,
according to the maximum output power of the communications equipment.
Rated
maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80
MHz
d = 1,2
80 MHz to 800 MHz
d = 1,2
800 MHz to 2,5 GHz
d = 2,3
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.

12
Guidance and manufacturer’s declaration – electromagnetic immunity
The Duet control unit is intended for use in the electromagnetic environment specified below. The
customer or the user of the Duet control unit should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output
lines
± 2 kV for power
supply lines
± 1 kV for
input/output
lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Mains power quality should be that of
a typical commercial or hospital
environment.
interruptions
and
voltage
variations
on power
supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the Duet
control unit requires continued
operation during power
mains interruptions, it is
recommended that the Duet control
unit be powered from an
uninterruptible power supply or a
battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.

13
Guidance and manufacturer’s declaration – electromagnetic immunity
The Duet control unit is intended for use in the electromagnetic environment specified below. The
customer or the user of the Duet control unit is responsible for making sure that it is used in such an
environment.
Immunity
test
IEC 60601 test
level
Compliance
level
Electromagnetic environment –guidance
Conducted
RF
IEC 61000-
4-6
Radiated RF
IEC 61000-
4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2,5
GHz
3 Vrms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any part of
the Duet control unit, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1,2
d = 1,2 80 MHz to 800 MHz
d = 2,3 800 MHz to 2,5 GHz
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,a
should be less than the compliance level in each
frequency range.b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the Duet control unit is used exceeds the applicable RF
compliance level above, the Duet control unit should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the Duet control unit.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

14
9. Expected Service Life
The Duet control unit has an expected service life of five years. To maintain the
condition of the pump, have the pump serviced regularly according to the
schedule recommended by Prius Healthcare USA. Do NOT use unapproved
accessories or attempt to modify, disassemble or otherwise misuse the Duet LAL
Alternating Anti-Decubitus System.
10. Warranty
⚫Prius Healthcare USA guarantees that this equipment is free from defects in
material and workmanship. Our obligation under this warranty is limited to
the repair of equipment returned to the service address given below,
transportation charges prepaid, within 12 months after delivery to the
original purchaser for all equipment.
⚫We agree to service and/or adjust any equipment returned for that purpose
and to replace or repair any part, which is proven to be defective at no charge.
⚫This warranty excludes equipment damage through shipping, tampering,
improper maintenance, careless, accident, negligence or misuse, or products
which have been altered, repaired or dismantled other than with the
manufacture’s written authorization and by its approved procedures and by
properly qualified technicians.
⚫In no event shall Prius Healthcare USA Products be liable for any direct,
indirect of consequential damages or losses resulting from the use of
equipment.

Prius Healthcare USA
4027 Tampa Road, #3200
Oldsmar, FL 34677, USA
TEL: (813)854-5464
FAX: (813)854-5442
AL300214 V1.01
This manual suits for next models
1
Table of contents
Other Prius Healthcare USA Medical Equipment manuals
Popular Medical Equipment manuals by other brands

TopTens
TopTens DT6030 Quick reference guide

Sunrise Medical
Sunrise Medical Jay J2 2300 SERIES User instruction manual & warranty

Raz
Raz Jaz-AP User and Dealer Manual

BIOPAC Systems
BIOPAC Systems NIBP250 user guide

Microline
Microline PowerPack Dual Control Footswitch Instructions for use

RC ART
RC ART DENS-GLASSES operating manual