Proactive Protekt PMNEBRDK-BF User manual

Protekt® Pediatric Nebulizer
w/Disposable and Reusable Neb Kit,
w/Pediatric Mask - Blue Fish
MODEL NO: PMNEBRDK-BF
Instruction Manual
The Proactive Compressor Nebulizer is an AC-powered air
nebulizer compressor system intended to provide a source of
compressed air for medical purposes for use in home healthcare.
It is to be used with a nebulizer kit to produce medicated aerosol
particles for respiratory therapy for both children and adults.
Important:
Read these instructions carefully before using this device.
Caution
Caution: Federal law restricts this device to sale by or on
the order of aphysician
Use the device only for its intended purpose as described in
these instructions for use as a compressor nebulizer system for
aerosol therapy. Follow the instructions of your doctor. Any other
form of use constitutes improper use and may be dangerous. The
manufacturer cannot be held liable for any damage caused by
improper, incorrect, or unreasonable use, or if the unit is connected
to electrical installations not complying with safe regulations.
Keep this manual for future reference.
Do not operate the unit in the presence of any anesthetic,
flammable mixtures, or oxygen.
Correct device operation can be affected by electromagnetic
interference. If this device interferes with another electrical apparatus,
remove it and plug it into a different electrical outlet.
In case of failure and/or malfunction, do not open the
compressor housing, as this is not a serviceable product.
The proper function of the device can be compromised if original
replacement parts (e.g., filter) are not used.
When using any electrical appliance, certain important safety
measures must always be observed, including the following:
• Never submerge the unit in water, as the device is not
protected against water penetration
• Never touch the unit if hands are wet or moist
• Do not leave the unit outdoors
• Place on stable and horizontal surface when operating
• Make sure the air vent openings area is unobstructed
• Do not allow unsupervised children or physically challenged
individuals to use the unit
• Turn the unit off with the power switch
,
do not pull plug out from
electrical outlet while the power is on
Make sure that the electrical rating shown on the rating plate on
the bottom of the unit corresponds to your main voltage and
frequency before plugging in the device.
Do not use electrical adapters including extension cords.
Do not leave the unit plugged in when not in use. Unplug the
device from the electrical outlet when not in operation.
Follow the manufacturer’s instructions for installing the device. The
manufacturer cannot be held responsible for any damage caused
by incorrect installation.
The power supply cord cannot be replaced by the user.
The power supply cord should always be fully unwound in order
to prevent dangerous overheating.
Before performing any cleaning operation or filter replacement,
disconnect the device from the main supply by unplugging it or
by turning the main supply switch OFF.
Some parts of the unit are small enough to be swallowed by children.
Do not let unattended children have access to the unit.
To avoid strangulation and entanglement, keep power cord and air
tubes out of reach of young children.
If you decide not to use the device, it is recommended to dispose
of it according to local regulations.
Remember to:
• Use this device only with medications prescribed by your doctor
• Use the accessories recommended by your doctor for therapy
Verify that the medications prescribed by your doctor can be used
with common compressor nebulizer systems for aerosol therapy.
Proactive Medical offers a limited 5-year warranty on the
compressor unit to be free from defects in material and
workmanship under normal use and operation for a period of 5
years from date of original retail purchase. This warranty does not
extend to failures resulting from accidental damage, misuse,
negligence, abuse, alteration, or improper operation. This warranty
does not extend to non-durable components which are subject to
normal wear and need periodic replacement. All items returned
must be properly packaged and shipped in a manner to avoid
shipping damage.
Instructions for Use
Warranty
Examine the unit before the first use in order to check
for possible shipping damage. When taking a treatment sit
upright and relaxed.
1. After unpacking the device, check the compressor and
accessories for visible damage or defects such as cracks in the
plastic housing, which could expose electrical components.
Contact your retailer to report any damaged product.
2. Before using the device, follow the cleaning instructions as
detailed in this user manual.
3. Attach the supply tube to the compressor and to the bottom of
the nebulizer cup.
4. Open the nebulizer cup by pushing down on the cap
and twisting counter-clockwise.
5. Add prescribed medication into nebulizer cup then turn top
on until secure.
6. Attach mouthpiece or face mask to the nebulizer bottle.
7. Connect the tube to the nebulizer bottle and the nebulizer machine.
8. Ensure connections are secure before turning on
the nebulizer machine.
9.
Plug the compressor into the electrical outlet, making sure the
electrical rating shown on the rating plate on the bottom of the unit
corresponds to your main voltage.
10.To start treatment, switch the compressor on by pressing
the ON/OFF switch onto the “I” position.
ATTENTION:This device rated for intermittent use
30 minutes ON /30 minutes OFF.
11. When mist is visible, begin inhaling with the mouthpiece.
12. When the treatment has been completed, switch the unit
OFF by setting the switch position to “O” and disconnect the
plug from the electrical outlet.
13. Wash the nebulizer and accessories as
described in the cleaning section.
If the device does not turn on:
1. Make sure the plug is firmly inserted into the electrical outlet.
2. Make sure the device has been operating within the limits
indicated in this instructional manual (30 minutes ON/30 minutes
OFF).
If the device does not nebulize, or the nebulization
is weak:
1. Make sure the ends of the air tube are fitted tightly onto the main
unit and onto the nebulizer (see connected kit and unit diagram
below)
2. Check whether the nebulizer mask or mouthpiece kit cup chamber
is empty or overfilled with medication.
3. Check whether the nebulizer nozzle is obstructed
(see medicine cup diagram below)
4. Check whether the baffle is inserted properly. The baffle fits over
the medicine cup’s interior cone (see medicine cup diagram below).
Connect Mouthpiece or Mask
Fit tube ends
tightly
Medicine cup
Exploded view
Baffle
Instructions for Use Troubleshooting
Closed view
Nozzle
Cleaning of the compressor
Use a clean dry cloth and mild detergents to clean the outside of
the unit.
ATTENTION:
During cleaning, make sure the internal parts of the unit
are not in contact with liquids and the power cord is disconnected.
Cleaning and disinfection of the accessories
After each use, disassemble the kit into 3 sections. Wash in warm
soapy water and rinse well. Wipe with a paper towel and allow to air dry.
Do not boil or autoclave angled mouthpieces, air tubes, or masks.
It is important that the cleaning instructions are carefully
followed to ensure optimal operation.
Disposable Nebulizer Kit
with Green Nozzle
The disposable nebulizer kit should be replaced after a long
period of inactivity, or if it is obstructed by dry medication, dust, etc.
With normal use, the disposable nebulizer kit should last 1 month.
Reusable Nebulizer Kit
with Beige Nozzle
The reusable nebulizer kit should be replaced after a long period
of inactivity, or if it is obstructed by dry medication, dust, etc. With
normal use, the reusable nebulizer kit should last 6 months.
Replacement of the filter
The filter is to be replaced after approximately 30 hours of
use, or when it appears dirty. Open the filter holder to replace
the filter with a new one.
Do not use the device without the filter.
Never open the device. There are no serviceable parts within the unit.
The compressor does not require lubrication or maintenance.
Proactive Medical Products
270 Washington Street
Mount Vernon, NY 10553
Tel: 855-237-7622 / Fax: 845-205-6005
www.proactivemedical.com
Cleaning
Replacement of Nebulizer Kit
Maintenance and Repair

Other Proactive Medical Equipment manuals

Proactive PROTEKT 600 User manual

Proactive

Proactive PROTEKT 600 User manual

Proactive PROTEKT 33350 User manual

Proactive

Proactive PROTEKT 33350 User manual

Proactive PROTEKT DASH 32500 User manual

Proactive

Proactive PROTEKT DASH 32500 User manual

Proactive PROTEKT 600 LIFT User manual

Proactive

Proactive PROTEKT 600 LIFT User manual

Proactive PROTEKT 600 STAND User manual

Proactive

Proactive PROTEKT 600 STAND User manual

Proactive PROTEKT 500 LIFT User manual

Proactive

Proactive PROTEKT 500 LIFT User manual

Proactive alevia 715-425 User manual

Proactive

Proactive alevia 715-425 User manual

Proactive alevia User manual

Proactive

Proactive alevia User manual

Proactive Elle 715-432 User manual

Proactive

Proactive Elle 715-432 User manual

Proactive Protekt Akra PMSEBHR User manual

Proactive

Proactive Protekt Akra PMSEBHR User manual

Proactive PROTEKT STS COMPACT 500 User manual

Proactive

Proactive PROTEKT STS COMPACT 500 User manual

Proactive 20110 User manual

Proactive

Proactive 20110 User manual

Proactive Libera User manual

Proactive

Proactive Libera User manual

Proactive Protekt Aire 94001 User manual

Proactive

Proactive Protekt Aire 94001 User manual

Proactive Protekt Aire 4000DX User manual

Proactive

Proactive Protekt Aire 4000DX User manual

Proactive TheraBelt User manual

Proactive

Proactive TheraBelt User manual

Proactive Protekt Aire 2000 User manual

Proactive

Proactive Protekt Aire 2000 User manual

Proactive thera3+ User manual

Proactive

Proactive thera3+ User manual

Proactive TheraHeat 715-440 User manual

Proactive

Proactive TheraHeat 715-440 User manual

Proactive Protekt Akra PMSEBHR User manual

Proactive

Proactive Protekt Akra PMSEBHR User manual

Popular Medical Equipment manuals by other brands

Dr. Langer Medical TWISTER Instructions for use

Dr. Langer Medical

Dr. Langer Medical TWISTER Instructions for use

Inspiration Tecotherm Neo Instructions for use

Inspiration

Inspiration Tecotherm Neo Instructions for use

Great Bear Healthcare Night Bag GB5 Instructions for use

Great Bear Healthcare

Great Bear Healthcare Night Bag GB5 Instructions for use

Enraf Nonius Curapuls 670 operating instructions

Enraf Nonius

Enraf Nonius Curapuls 670 operating instructions

OPED VACOCAST DIABETIC Leaflet

OPED

OPED VACOCAST DIABETIC Leaflet

Atlas Ernie and Bert user guide

Atlas

Atlas Ernie and Bert user guide

TekTone NC403TS user guide

TekTone

TekTone NC403TS user guide

Boscarol OB1000 Series user manual

Boscarol

Boscarol OB1000 Series user manual

Neurosoft Neuron-Spectrum-AM Technical manual

Neurosoft

Neurosoft Neuron-Spectrum-AM Technical manual

Yamakin Luna-Wing instruction manual

Yamakin

Yamakin Luna-Wing instruction manual

Invacare R100 user manual

Invacare

Invacare R100 user manual

Otto Bock 17LK3 Series Instructions for use

Otto Bock

Otto Bock 17LK3 Series Instructions for use

PFM Medical EWF 600 operating manual

PFM Medical

PFM Medical EWF 600 operating manual

TNI 20s manual

TNI

TNI 20s manual

Braun Omnitest 3 Guide

Braun

Braun Omnitest 3 Guide

North-Vision Tech Wrist-100 user manual

North-Vision Tech

North-Vision Tech Wrist-100 user manual

Covidien DAR Instructions for use

Covidien

Covidien DAR Instructions for use

Bovie Icon Gi user guide

Bovie

Bovie Icon Gi user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.