Promeba PA-137 User manual

USER GUIDE
VACUUM SPLINT PA-137 AND PA-186
Review 2021/09


01 INTRODUCTION 05
01.1 Using this manual 05
01.2 Legend of symbols 05
01.3 Servicing request 05
01.4 Demolition 06
01.5 Labelling 06
01.6 Contraindications and adverse effects 06
01.7 Physical requirements of the operators 06
01.8 Intended purpose 06
01.9 General warnings 06
01.10 Specific warnings 09
01.11 Residual risk 09
01.12 Reference standards 10
01.13 Life span 10
02 PRODUCT DESCPRIPTION 10
02.1 Product Features 10
02.2 Components 10
02.3 Use Method 11
03 OPERATION 11
03.1 Transport 11
03.2 Storage 12
03.3 Maintenance 12
04 MAINTENANCE REGISTER 13
05 LEGAL NOTICES 15
06 PRODUCT WARRANTY 16


USER GUIDE // VACUUM SPLINT PA-137 AND PA-186
Review 2021/09
01 INTRODUCTION
5
01 INTRODUCTION
01.1 Using of manual
The manual provides using and maintenance instructions of the product as well as technical
aspects, functioning, spare parts and safety.
It is recommended before the operation of the product to read carefully this manual in order to
avoid damages caused by a misuse.
Do not lose this document. It should be accessible to any doubt that could appear by medical
personnel. Remember that a good use and maintenance are necessary for the proper operation of
the product.
01.2 Legend of Symbols
SYMBOL EXPLANATION / DESCRIPTION
MANUFACTURER symbol. This symbol is accompanied by the name and address of the
manufacturer, adjacent to the symbol (PRODUCTOS METÁLICOS DEL BAGÉS S.L., Ctra.
C-16 Km 59.5, 08650 - Sallent (Barcelona) - Spain).
Indicates the manufacturer’s reference number to identify the medical device.
PROMEBA, S.L. uses this symbol to set each internal reference for each configuration and
business variant.
Indicates the manufacturer’s serial number to identify a specific medical device.
Indicates the manufacturing date. The symbol must be accompanied by a manufacturing
Date (yyyy-mm), adjacent to the symbol.
It is placed to inform that the product is a“Medical Device”.
CE symbol without the intervention of a Notified Organism, as a Medical Device classified as
Class I according to the EU Regulation 2017/745 on Medical Devices.
Symbol for the Unique Device Identifier.
Symbol“See instructions for use or operating instructions”.
Symbol “caution”. This symbol is placed to warn of the need for the user to refer to important
precautionary information in the operating instructions, such as warnings and cautions not
otherwise found on the label.
Symbol“Caution”. For a general warning.
Warning, crushing of hands
-20ºC
+40ºC
Indicates the temperature limits. The upper and lower temperature limits should be indicated
adja- cent to the horizontal lines.
01.3 Servicing request
For information of the correct interpretation of the instruction manual, the use, maintenance,
installation and restoration of the product, please contact Promeba customer service: T. 93 837 12
Sallent (Barcelona) · SPAIN.

USER GUIDE // VACUUM SPLINT PA-137 AND PA-186
Review 2021/09
6
01.4 Demolition
When the devices are no longer suitable for use, if they have not been contaminated by any
particular agent, they can be disposed of as normal solid waste, otherwise, follow the current
demolition regulations.
01.5Labelling
Each product incorporates an identification label, placed on the device itself and/or on the box. It
must never be removed or covered. This label includes the serial number and the product code.
Please keep these numbers so you can inform the dealer if necessary.
1.6 Contraindications and adverse effects
The use of this device, if used as described in this manual, does not present any contraindications
or collateral effects.
01.7Physical requirements of the operators
Promeba Vacuum splint is destined to professional use only.
The operators must be trained in efficient, effective and safe patient transport
01.8Intended purpose
It is mainly used for hospitals, sports, rescue activities, ambulance and battle fields, for the member’s
immobilization both adult and infantile.
01.9General warnings
1. The product must be used by trained personnel only, having attended specific
training for this device and not for similar products.
2. At least every 6 months, is important to verify if updated User Manuals are
available or there is any change involving your product. This information is freely
available on the website http://www.promeba.com
3. Before carrying out any kind of operation on the appliance (training,
installation, use), the operator must carefully read the enclosed instructions,
paying particular attention to the correct safety precautions and to the procedures
to be followed for installation and for correct use.

USER GUIDE // VACUUM SPLINT PA-137 AND PA-186
Review 2021/09
01 INTRODUCTION
7
4. In the case of any doubts as to the correct interpretation of the instructions,
please contact Promeba, S.L. for any necessary clarifications.
5. Do not allow untrained persons to help during the use of the device, because
they could cause damage to the patient or to themselves.
6. Periodically check the device, carry out the prescribed maintenance and
respect the life span indicated by the manufacturer in this user manual.
7. Before each use of device the perfect operating state of the device must be
checked as specified in the instruction manual. If any damage or abnormalities
which could in any way influence the correct functioning and the safety of the
device, of the patient and/or of the user are detected, the device must be
immediately removed from service and the manufacturer must be contacted.
8. If any failure or incorrect functioning of the device is detected, it must be
immediately substituted with a similar item so that the rescue procedures are
guaranteed without any interruption.
9. Use of the device in anyway other than described in this manual is forbidden.
10. Do not alter or modify in any way the device; any such interference could
cause malfunctions and injury to the patient and/or rescuer.
11. The device must not in any way be tampered with (modification, adjustment,
addition, replacement). In such cases all responsibility will be denied for any
malfunctions or injuries caused by the appliance itself; moreover CE certification
and product warranty will be considered void.
12. Handle with care.
13. Ensure that all the necessary precautions are taken in order to avoid the
hazards that can arise as the result of contact with blood or body fluids.
14. Register and store with these instructions: lot number, place and date of
purchase, first date of use, date of checks, name of users, any comments.
15. When the device is being used, the assistance of qualified staff must be
guaranteed.
16. Do not store the device underneath any heavy objects which could cause
structural damage.
17. Store in a cool, dry, dark place and do not expose to direct sun.
18. Store and transport device in its original packaging.
19. The device must not be exposed or come into contact with any source of
combustion or inflammable agents.
20. Position and adjust the device taking care not to cause any obstruction to
rescuers and or any other rescue equipment.

USER GUIDE // VACUUM SPLINT PA-137 AND PA-186
Review 2021/09
8
21. Attention: laboratory testing, post production tests and instruction manuals
cannot always consider every possible scenario for use. This means that in some
cases the performance of the product could be notable different from results to
date obtained. Instructions are continually being updated and are under tight
surveillance of fully qualified staffs with adequate technical formation.
22. Both public and private operators are obliged to report any accident involving
any medical device to the Ministry of Health and the manufacturer as specified and
within the time given by European regulations.
23. Both public and private operators are obliged to inform the manufacturer of the
measures to be taken to guarantee the safety and health of patients and users of
any medical device.
24. As a distributor or end user of the products manufactured and/or distributed by
Promeba, S.L., you are strictly obliged to know the legal provisions in force in the
country of destination of the goods, applicable to the devices to be supplied
(including the regulations relating to technical specifications and/or safety
requirements) and, therefore, understand the necessary requirements to ensure
compliance of the products themselves with all the legal requirements of the
territory.
25. Promptly notify PROMEBA regarding any revisions to be made by
manufacturer in order to guarantee the conformity of the product to the territory’s
legal specifications (including those resulting from rules and/or norms of other
nature).
26. Act with all due care and diligence, and contribute to ensure conformity to
general safety requirements of all devices marketed in the territory, by providing
final users with all necessary information for carrying out periodical checks on
their devices, as specified in the relevant user manual.
27. Actively contribute to product safety checks on products sold, by
communicating any relevant risk analysis information both to the manufacturer and
to any competent authorities so that the necessary actions can be promptly taken.
28. Be aware that in the event of any failure to conform to the above mentioned
requirements you will be deemed fully responsible for all damages that might
occur. Therefore we expressly disclaim any responsibility and/or liability for your
non-compliance with the present“Regulatory provisions”.

9
USER GUIDE // VACUUM SPLINT PA-137 AND PA-186
Review 2021/09
01.10 Specific warnings
1. Establish a maintenance program and periodic testing, identifying a reference employee.
The person to whom the ordinary maintenance of the device is entrusted must ensure the
basic requirements foreseen by the manufacturer in the user’s manual.
2. Use only accessories/spare parts that are original or approved by Promeba, S.L. , when
carrying out any operation, to avoid causing any alteration or modification to the device,
otherwise we assume no responsibility for the proper functioning or damage resulting from
device to the patient or the operator and warranty will be considered void.
3. Never leave the patient unassisted on the device, because he may be injured.
4. Do not use bleach to disinfect the product. Use a hydroalcoholic based disinfectant and
wash with water.
5. The device and all its components, after washing, should be allowed to dry completely
before storing.
6. Keep clean as routine (incl. disinfect).
7. Follow procedures approved by Emergency Medical Services for immobilization and
transportation of the patient.
8. Follow procedures approved by Emergency Medical Services.
9. Avoid contact with sharp objects.
10. Use the vacuum splint only as described in this user’s manual.
01.11 Residual risks
The residual risks listed below have been identified only with reference to the intended use
of the device:
1. Use by untrained personnel may result in injury to patient, operator, or third parties.
2. Inappropriate disinfection procedures can create a risk of cross infection.
3. Failure to comply with the warnings for operators can create risks.
4. Failure to read and understand the product instructions can result in injury to the patient
and operators.

10
USER GUIDE // VACUUM SPLINT PA-137 AND PA-186
Review 2021/09
01.12 Reference standards
REFERENCE
TITLE OF DOCUMENT
(UE) 2017/745
Reglamento UE del 05 de Abril del 2017 sobre productos sanitarios
As a distributor or end user of the products manufactured and/or distributed by
Promeba, S.L., you are strictly obliged to know the legal provisions in force in the
country of destination of the goods, applicable to the devices to be supplied
(including the regulations relating to technical specifications and/or safety
requirements) and, therefore, understand the necessary requirements to ensure
compliance of the products themselves with all the legal requirements of the territory.
01.13 Life span
If used as described in the following instructions, this device has a useful life of 10 years
from the date of purchase.
This useful life can be extended with annual reviews carried out by the manufacturer, which
uses specialized and authorized internal and external technicians.
In case these annual checks are not carried out, the device must be disposed of according
to the information in paragraph 01.4 and the manufacturer must be notified.
Only the manufacturer or an authorized center can extend the life of the device, if it meets
the safety requirements.
Promeba, S.L. will not accept any responsibility for malfunction or damage caused by the
use of devices that have not been checked by the manufacturer or authorized center, or that
have exceeded the maximum allowed useful life.ODUCT DESCRIPTION
02 PRODUCT DESCRIPTION
02.1 Product Features
The PA-137 and PA-186 models have been designed for the members' immobilization both adult and infantile,
by means of vacuum.
The pieces are made of Poliamida by covering of PVC, and you refill of pearl pre-expanded type P4. Each one is
provided with an air valve with system ON/OFF.
For its capacity of adjustment, they are easy and comfortable of applying in any type of situation, allowing at all time
the managing breathing techniques, or if it was necessary, the accomplishment of resuscitation’s actions.
02.2 Components
PA-137
Adult
1 unit Arm sling

11
USER GUIDE // VACUUM SPLINT PA-137 AND PA-186
Review 2021/09
PA-137
Adult
1 unit
Leg sling
PA-186
Child
3 units
Sling
02.3. Use Method
1. The kit of vacuum splint is only for professional use.
2. To place it to follow the protocols of its use.
3. The immobilization must include the joints proximal and distal to the fracture.
4. In case of wounds or opened fractures, these must be covered with sterile dressings before placing the rod.
03. OPERATION
03.1. Transport
1. Be transferred by common transporting tools: plane, ship or road.
2. Vacuum splint can be manually handling
3. A whole carton packaging
4. Attachments: certificate, user’s manual.

12
USER GUIDE // VACUUM SPLINT PA-137 AND PA-186
Review 2021/09
03.2. Storage
1. Store this vacuum splint in the place of damp proof and non-corrosion environment.
2. Storage, transportation, as shown by the box stacking.
03.3. Maintenance
1. Inappropriate use and a lack of maintenance can cause damage to people and objects.
2. Though a specific maintenance is not detailed, if one advises to verify once to the month (in case of normal use),
that the rod does not lose the vacuum. It will be verified removing all the air to the rod, closing the valve and after 3
hours, to verify that it has not recovered air.
In case some piece was losing vacuum, it must be replacement very soon, by the new one.

13
USER GUIDE // VACUUM SPLINT PA-137 AND PA-186
Review 2021/09
04. MAINTENANCE REGISTER
The product must be used by trained personnel only, having attended specific training for this
device and not for similar products.
Keep this document at least 10 years from the end of life of the device.
PLACE AND DATE
NAME OF
THE
OPERATOR
NAME OF TRAINER
TYPE OF TRAINING

USER GUIDE // VACUUM SPLINT PA-137 AND PA-186
Review 2021/09
14
Perform the required maintenance as indicated by the manufacturer in this user’s
manual. Keep this document at least 10 years from the end of life of the device.
DATE
TYPE OF SERVICE
(Maintenance /
verification /
extension of useful
life)
OPERATIONS
OF MAINTENANCE
REALIZED
PERSON IN CHARGE
OF SERVICE
(Operator / Authorized /
Center / Manufacturer)

USER GUIDE // VACUUM SPLINT PA-137 AND PA-186
Review 2021/09
15
45
15
21
05. LEGAL NOTICE
This document may contain technical inaccuracies
or typographical errors.
Changes are periodically added to the information
herein; these changes will be incorporated in new
editions of the publication.
Promeba, S.L. reserves the right to make any
modification or improvement in the products
described in this publication if it is appropriate.
Promeba, S.L. may have patents or patent
applications which address themes described in this
document. The possession of this document does
not entitle no license to these patents.
The information contained in this document does
not affect or change the specifications or product
specifications or warranties of Promeba, S.L.
No part of this document shall operate as express
or implied license or compensation under the
intellectual property rights of Promeba, S.L. or third
party.
All the information contained in this document has
been obtained in specific environments and it is
presented as an illustration. The results obtained in
other operating environments may vary.
Promeba, S.L. can use or distributes the
information that the client provides in whatever way
they see fit, without incurring any obligation with the
client.

USER GUIDE // VACUUM SPLINT PA-137 AND PA-186
Review 2021/09
16
06. PRODUCT WARRANTY
Promeba, S.L. guarantees that its products have
satisfactorily passed all established quality controls,
both functional and material. The duration of the
guarantee is 2 years from the date of purchase of
the product.
This guarantee will be valid only when it is
presented with the original invoice or proof of
purchase (indicating the date of purchase, model
and the name of the distributor) together with the
defective product during the period covered by the
guarantee. Promeba, S.L. reserves the right not to
offer the free warranty service if the indicated
documents are not presented or if the information
they contain is incomplete or illegible.
1. This warranty will not apply if the model name
or serial number of the product has been altered,
erased, disappeared or becomes illegible.
2. This guarantee does not cover transportation
costs or risks derived from the transportation of
your pro- duct to and from Promeba, S.L.
3. This warranty does not cover any of the
following:
a) Periodic maintenance and repair or replacement
of parts derived from normal wear and tear.
b) Expendable material (components that are
expected to need periodic replacement during the
life of the product, such as non-rechargeable
batteries, light bulbs, etc.).
c) Damages or defects derived from the improper
use, operation or treatment of the product and not
due to normal use of the product.
d) Damages derived from:
i. Misuse, including:
- Treatment that results in damages or physical,
superfi- cial or appearance changes of the product.
- Installation, use or storage of the product in a way
that does not respect the instructions described by
Promeba, S.L.
- Maintenance of the product in a way that does not
respect the instructions of Promeba, S.L. for proper
maintenance.
- Installation or use of the product in a way that
does not respect the technical or safety regulations
of the country where it is used or installed.
iii. Use of components not provided with the
product or incorrect installation of accessory parts
not previously tested.
iii. States or defects of the system in which the
product is used or incorporated with the exception of
other pro- ducts Promeba, S.L. designed for use
with the product.
iv. Use of the product with accessories, peripheral
units and other products of a type, condition or
standards not established by Promeba, S.L.
v. The manufacturer or distributor will be solely
responsible for deciding whether to send the parts
for repair, or to replace the product in its entirety. In
no case will operators be sent to repair or
replacement of the product.
Except in the cases mentioned above, Promeba,
S.L. will make no warranties in relation to product,
performance, accuracy, reliability, or fitness for logic
purpose of the equipment or other type. If this
exception is not legal or contemplated by current
law, Promeba, S.L. will limit or exclude your
warranties only to the extent permitted by
applicable law.
The only obligation on the part of Promeba, S.L. in
connection with this warranty is to repair or replace
parts subject to the terms and conditions of this
warranty.
Promeba, S.L. is not responsible for loss or damage
to products, this warranty or others, including
economic loss or non-assessable damage; the
price paid for the product; loss of profits, income,
information, usufruct or use of the product or
associated products or indirect, accidental or critical
loss or damage.
This clause refers to whether the loss or damage is
due to deterioration or inoperability of the
associated product due to defects or unavailability
of Promeba, S.L., which has caused downtime, loss
of user time or business interruption.
In cases where the law prohibits or limits these
liability exclusions, Promeba, S.L. will exclude or
limit his liability only to the extent permitted by
applicable law. For example, there are countries
that prohibit the exclusion or limitation of damages
caused by negligence, reckless negligence, willful
misconduct, fraud and similar actions. The
responsibility of Promeba, S.L. in this guarantee
will not exceed, in any case, the price paid for the
product, but if the current law allows only limitations
of greater responsibilities, these will apply.

USER GUIDE // VACUUM SPLINT PA-137 AND PA-186
Review 2021/09
Todos los derechos reservados. Reservado el derecho a modificaciones sin previo aviso.
Promeba, S.L. no se considera responsable de los daños causados por la falta o la
inexactitud de la información aquí mencionada.
All rights reserved. Variations can be done without notice.