push buon switch has two posions: depressing the buon latches it into
the lower Sample posion. Pressing it again releases it into its upper posion. The
knob may be turned to the le or right to change the zero seng of the display. The unit must
be operated while grounded.
Push the slide switch (s1) to its upper posion (towards the display) to select
. The display will come ON, a decimal point will be displayed, and the meter will mea-
sure up to 19.99kV at one inch in 10 volt increments. Placing the slide switch in the lower posi-
on will select . There will be no decimal point in the display and the meter will measure
up to 1999 volts at one inch, in one volt increments. Moving the switch to its center posion
will turn the meter OFF.
Ground the case by wearing a funconal wrist strap that is properly grounded. Turn the meter
on by selecng the desired range. Press the buon down so that it latches in
the lower or posion, and is not indicated in the meter display. Adjust the dis-
play to zero (0) volts by rotang the control.
The PFM-711A meter case is conducve. It provides the ground reference for
its measuring circuit. For accurate measurements, it is necessary that the person holding the
meter be properly grounded, or the meter has a ground connecon made to the metal snap on
the case.
Turn the meter to the desired range. Hold the front of the meter approximately one inch
from the surface to be measured. Adjust the distance from the test surface unl the RED rang-
ing lights form a Bull’s Eye. Allow the instrument to stabilize. Record the Measurement.
With the meter in the measurement posion described above, press the but-
ton so that it latches in the upper or hold posion. This will freeze the reading on the display,
allowing the meter to be moved to a posion where it may be more easily read and document-
ed. When the meter is in the posion, the word “” will appear In the display.
A. The PFM-711A is factory calibrated and other than baery replacement, external cleaning, and oc-
casional adjustment of the front panel knob range posion, general user maintenance is not
required. The case has been sealed and . If for
any reason you believe the meter is not working correctly, contact Prostat Corporaon for assis-
tance.
Other than baery replacement & front panel zero range adjustment, there are no
user serviceable parts. Any unauthorized service will void the warranty and result in
addional repair charges.
8Rev. C / September 21, 2017
PFM-711 Electrostac Field Meter