5
IV. Specifications (23±5°C)
Range of Wind Velocity:
Units Range Resolution Threshold Accuracy
m/s 0.0 - 45.0 0.1 0.5 ±3% ± 0.3
ft/min 0 - 8800 10 60 ±3% ± 40
knots 0.0 - 88.0 0.1 0.6 ±3% ± 0.4
Km/hr 0.0 - 140.0 0.1 1.0 ±3% ± 0.8
mph
(optional) 0.0 - 100.0 0.1 0.7 ±3% ± 0.4
m/s :meter per second ft/min : feet per minute
knots :nautical miles per hour Km/hr : kilometers per hour
mph : miles per hour
Unit Conversion table:
m/s ft/min knots Km/hr mph
1 m/s 1 196.87 1.944 3.60 2.24
1 ft/min 0.00508 1 0.00987 0.01829 0.01138
1 knot 0.5144 101.27 1 1.8519 1.1523
1 Km/hr 0.2778 54.69 0.54 1 0.6222
1 mph 0.4464 87.89 0.8679 1.6071 1
Range of Temperature:
Range Resolution Accuracy
°C 0 - 60.0 0.1 ±0.8
°F 32.0 - 140.0 0.1 ±1.5
Bearing : Sapphire jewel bearing
Temperature sensor : K-type thermocouple
Mounting Nut : 1/4" x 20
Operating Temperature : Meter 0 °C ~ 50°C (32 °F ~ 122°F)
Vane
0°C ~ 60°C (32 °F ~ 140°F)
Operating Humidity : Less than 80% RH
Operating Pressure : 500 mB ~ 2 Bar