PROVA AFLEX 3030 Manual

Prova
INSTRUCTION MANUAL
Flexible AC Current Probe
Model Prova AFLEX 3030
Prova In trument Inc.

Table of Content
1. Warning Page 2
2. Panel and part de cription Page 3
3. Operating in truction Page 4
4. Electrical pecification Page 10
5. General pecification Page 11
6. Battery replacement Page 12
7. Maintenance and cleaning Page 13

EN 61010-2-032
CAT III 600V
Pollution Degree 2
Definition of Symbol :
Caution : Refer to accompanying document
Caution : Ri k of electric hock
In trument i protected by double in ulation
Do not apply around or remove from HAZARDOUS LIVE conductor .
Over-voltage Category I (CAT I)
Equipment for connection to circuits in which measures are taken to limit the transient
over-voltages to an appropriate low level.
Over-voltage Category II (CAT II)
Energy-consuming equipment to be supplied from the fixed installation.
Over-voltage Category III (CAT III)
Equipment in fixed installations.
WARNING : If the flexible current probe is used in a manner not specified by the
manufacturer, the protection provided by the flexible current probe may be impaired.
1

1. WARNING
Do not u e the flexible current probe before you read the following in truction .
1 Do not install the flexible current probe around bare conductors carrying a voltage
from 30V to 600V unless you are wearing protective clothing and gloves suitable for
high-voltage work.
2 Always inspect and check any damage of the current probe assembly before usage.
Do not use the flexible current probe if any damage is found.
3 Do not use the flexible current probe on circuit rated higher than 600V in installation
Category III.
2

2. PANEL AND PARTS DESCRIPTION
1. Coupling assembly
2. Flexible loop
3. Banana plugs for output
4. Output range select switch
5. Power on LED
6. Low battery voltage LED
7. Current range select switch
8. Battery cover
3

3. OPERATING INSTRUCTION
4

1. Connect the banana plugs to the measuring instrument. Select the corresponding
sensitivity on your instrument. Select AC range if a multimeter is used.
2. On the flexible current probe, select the desired range (30A, 300A or 3000A and
output (300mV or 1V .
3. Connect the flexible probe around the conductor.
4. Make sure the current flowing direction is in consistent with the arrow marked on the
probe coupling. If the flexible current probe is connected in the right orientation, the
correct phase will be displayed on the oscilloscope.
5. Keep the probe coupling more than 25mm away from the conductor.
WARNING : Alway wear appropriate glove in operation.
WARNING : Do not select AC + DC range for measurement. Select AC range.
If a RMS multimeter is used, the AC current value can be calculated from the AC voltage
reading.
5

Example 1 : Current range 300A and output 300mV.
The AC voltage reading is 100mV
The actual AC current= 300A / 300mV * 100mV = 100A
6

Example 2 : Current range 3000A and output 300mV.
The AC voltage reading is100mV
The actual AC current = 3000A / 300mV * 100.0mV = 1000A
7

Example 3 : Current range 300A and output 1V.
The AC voltage reading is 333.3mV.
The AC actual AC current = 300A/V * 333.3mV = 100A
8

Example 4 : Current range 3000A and output 1V.
The AC voltage reading is 333.3mV
The actual AC current = 3000A/1V * 333.3mV = 1000A
9

4. ELECTRICAL SPECIFICATIONS
(23° ± 3°, 50/60Hz, conductor located at the center of the flexible loop
DC offset : < 20mVDC (300A AC range
<100mVDC (30A Ac range
< 20mVDC (3000A CA
Noise: <2mACV
RMS
Bandwith (-3dB : 10 à 10KHz
Load impedance: 1KΩ
Batteries: 2 "AA" 1.5V Alkaline (200mA hrs
Battery life : 300 hours
10

5. GENERAL SPECIFICATIONS
Indoor u e
Altitude : 2000m
Probe length 3000-24: 24 po / 610 mm
Minimum bending diameter: 35mm
Connector diameter : 23mm
Cable diameter : 14mm
Cable length from probe to case : 1700mm
Cable length from case to output : 1700mm
Batteries : Two 1.5V SUM-3
Dimensions (case : 130mm (L x 80mm (W x 43 mm (H
5.1" (L x 3.1" (W x 1.7" (H
Operating temperature : -10 to 85°C
Operating humidity : 15% to 85% RH
Storage temperature : -20 to 85°C
Storage humidity: 15% to 85% RH
Weight : 430g
Accessories : 1 instruction manual, 2 1.5V battery
11

6. BATTERY REMPLACEMENT
1. Remove the flexible current probe from any circuit to be measured.
2. Disconnect the flexible current probe from any measuring instrument.
3. Turn the power off.
4. Remove screw and then the battery cover.
5. Replace the old batteries with new batteries.
6. Replace the battery cover and fasten the screw.
12

7. MAINTENANCE AND CLEANING
1. Always inspect the flexible current probe for any damage. If user finds any
damage, do not use the flexible current probe. Return the probe to a qualified
person for repair or replacement
2. Clean the probe assembly with a damp cloth and a mild detergent.
3. Remove the battery if the flexible current probe is not to be used for a long period.
MTP Warranty
MTP In trument warrants this instrument to be free of defects in parts and workmanship for
one (1 year from date of shipment. This warranty does not apply to defects resulting from
action of the user such as misuse, improper wiring, operation outside of specification, improper
maintenance or repair, or unauthorized modification. MTP In trument specifically disclaims
any implied warranties or merchantability or fitness for a specific purpose and will not be held
liable for any direct, indirect, incidental or consequential damages. MTP In trument total
liability is limited to repair or replacement of the product.
13

Le In trument
MTP
In trument
Head Office
4409, Charleroi treet
Montréal-North, Quebec
H1H 1T6
Telephone : (514) 326-7167 Fax: (514) 326-7835
Toll Free Number: 1-888-326-7167
Web Site : www.mtpinc.com E-mail : inf[email protected]