manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Ashcroft
  6. •
  7. Measuring Instrument
  8. •
  9. Ashcroft 1132 User manual

Ashcroft 1132 User manual

Other Ashcroft Measuring Instrument manuals

Ashcroft Digital Test Gauge User manual

Ashcroft

Ashcroft Digital Test Gauge User manual

Ashcroft N5500 Instruction Manual

Ashcroft

Ashcroft N5500 Instruction Manual

Ashcroft P6500 Instruction Manual

Ashcroft

Ashcroft P6500 Instruction Manual

Ashcroft TC User manual

Ashcroft

Ashcroft TC User manual

Ashcroft DURALIFE 1009 Installation and operating instructions

Ashcroft

Ashcroft DURALIFE 1009 Installation and operating instructions

Ashcroft a4a Manual

Ashcroft

Ashcroft a4a Manual

Ashcroft F5510 User manual

Ashcroft

Ashcroft F5510 User manual

Ashcroft 1131 User manual

Ashcroft

Ashcroft 1131 User manual

Ashcroft F5509 Instruction Manual

Ashcroft

Ashcroft F5509 Instruction Manual

Ashcroft 8008 User manual

Ashcroft

Ashcroft 8008 User manual

Ashcroft A5400 Instruction Manual

Ashcroft

Ashcroft A5400 Instruction Manual

Ashcroft F5503 User manual

Ashcroft

Ashcroft F5503 User manual

Ashcroft 2074 User manual

Ashcroft

Ashcroft 2074 User manual

Ashcroft 2074 User manual

Ashcroft

Ashcroft 2074 User manual

Ashcroft 28HPX User manual

Ashcroft

Ashcroft 28HPX User manual

Ashcroft 2032 User manual

Ashcroft

Ashcroft 2032 User manual

Ashcroft A Instruction Manual

Ashcroft

Ashcroft A Instruction Manual

Ashcroft t5500e Instruction Manual

Ashcroft

Ashcroft t5500e Instruction Manual

Ashcroft T5500 User manual

Ashcroft

Ashcroft T5500 User manual

Ashcroft P5500 Instruction Manual

Ashcroft

Ashcroft P5500 Instruction Manual

Ashcroft 50 HPS User manual

Ashcroft

Ashcroft 50 HPS User manual

Ashcroft F5504 User manual

Ashcroft

Ashcroft F5504 User manual

Ashcroft F5509 User manual

Ashcroft

Ashcroft F5509 User manual

Ashcroft DG25 User manual

Ashcroft

Ashcroft DG25 User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Installation Instructions for
ASHCROFT®Model 1132
Differential Pressure Gau e/Switch
© Ashcroft nc., 250 East Main Street, Stratford, CT 06614-5145, USA, Tel: 203-378-8281, Fax: 203-385-0357, www.ashcroft.com
All sales subject to standard terms and conditions of sale. &M008-10087-11/01 AMR 2C09/09
For efficient working of your gauge, please read all instructions
carefully before attempting to install.
CAUTION: Do not exceed maximum operating pressure given
on the gauge label. Check fluid compatibility with wetted parts
before use.
OPERATING PRINCIPLE
High and Low pressures are separated by a sensor assembly
consisting of a magnet, diaphragm and a range spring. The
difference in pressure causes the sensor assembly to move in
proportion to the change against a range spring.
A rotary magnet, located in a separate body compartment and
isolated from the acting pressures is rotated by a magnetic
coupling following the linear movement of the sensor assembly.
A pointer attached to the rotary magnet indicates differential
pressure on the dial.
SWITCH: Reed switches (optional) are located adjacent to the
pressure chamber and are activated by the magnetic field of
the sensor assembly.
Note: The gauges are calibrated to give ±2% full scale accura-
cy on ascending readings.
INSTALLATION
For better performance the gauge should be mounted horizon-
tally with the dial vertical. Depressurize the system and con-
nect the high and low pressure lines of your system to the
“High” and “Low”ports of the gauge, respectively.
t is recommended to use “O” rings with male connectors to
avoid excessive tightening and to prevent leakage (for parallel
threads). The gauge is now ready for operation. Apply “High”
and “Low”pressures simultaneously, to avoid damage to the
internal parts.
Model 1132 consists of a small convoluted diaphragm to
sense the pressure difference. t can withstand maximum
operating pressure up to 1500 psi for all ranges. f pressure
exceeds the rated maximum pressure, “O” rings and
diaphragm inside the pressure chamber, will be damaged. f
maximum operating pressure is within the allowable limit of
1500 psi, but the differential pressure exceeds gauge range,
there will be no damage to the gauge. The pointer will only go
to the extreme right end of the scale.
HIGH
POINTER
ROTARYMAGNET
MAGNET
CONVOLUTEDDIAPHRAGM
LOW
REEDSWITCH(OPTIONAL)
20
10
0
30
40
50
RANGESPRING
Model 1132
Small Convoluted Diaphra m
PRECAUTIONS
Do not connect “High” and “Low” ports to the wrong pipe ends.
Do not subject the gauge to excessive vibration. The gauge is
never to be used in an area where a magnetic field is present.
t may show the wrong readings.
As the gauge works on a magnetic coupling principle, use only
non-magnetic fittings, parts, etc. in areas closer than 2˝ on all
sides, otherwise calibration will be affected. Panel mounted
gauges should be installed in non-ferrous panel material.
However, gauges with 2˝ and 2.5˝ dials, mounted in steel
panels, may require pointer adjustment. This is done by the
manufacturer, upon request.
Do not try to open any part of the gauge for any reason, be-
cause if not reassembled properly calibration will be affected.
SWITCH SETTING
Please follow these instructions when your differential pressure
gauges are supplied with a switch. The switches are normally
factory set to save the customer time. However, they are field
adjustable.
CAUTION: Supply voltage should not exceed switch rating. For
higher supply, use of relay circuit is recommended.
SWITCH ADJUSTMENT
The following procedure must be done by putting the gauge on
test bench or while in actual service. The switch adjustment
screw is located on the plastic cover. Rotate the screw counter
clockwise to increase the setpoint and clockwise to decrease
the setpoint. One or two trials may be necessary to attain the
exact setpoint.
ncrease setpoint
counter clockwise
View from high pressure side
Load
L
(+)
N
(-)
N
(-)
R
RB
B
Switch #1Switch #2
Load
L
(+)
N
(-)
RB
Switch #1
6
6
5
5
4
4
3
3
2
2
1
1
6
6
5
5
4
4
3
3
2
2
1
1
Lor(+) Lor(+)
Load
Nor(-)Nor(-)
Nor(-)
1
3
2
E
B
R
E
1
3
2SPST
Switch
Load for
Load for
Switch #1
3
2
E
1
3
1
E
R
B
2
B
Switch #2
Switch #1Switch #2
View of plug after removing the socket View of plug after removing the socket
Installation Instructions for
ASHCROFT®Model 1132
Differential Pressure Gau e/Switch
© Ashcroft nc., 250 East Main Street, Stratford, CT 06614-5145, USA, Tel: 203-378-8281, Fax: 203-385-0357, www.ashcroft.com
All sales subject to standard terms and conditions of sale. &M008-10087-11/01 AMR 2C09/09
Load
NC NO
1
3
2
E
N
(-)
L
(+)
N
(-)
Load
Load
NC
NO
B
RY
6
6
5
5
4
4
3
3
2
2
1
1
N
(-)
L
(+)
N
(-)
N
(-)
Load
Load
Load
Load
N
(-)
NC
NC
NO
NO
B
RYY
B
6
6
5
5
4
4
3
3
2
2
1
1
Lor(+)
Nor(-)
Nor(-)
Y
R
3
12
EB
SPDT
Switch
Switch #1Switch #1Switch #2
SPST SWITCH
Specifications
Contact Rating : 10 VA ac (rms) or dc (max)
Switching Current : 0.5 Amp ac (rms) or dc (max)
Switch Voltage : 100 Vac/Vdc (max)
SPDT SWITCH
Specifications
Contact Rating : 3 VA ac (rms) or dc (max)
Switching Current : 0.3 Amp ac (rms) or dc (max)
Switch Voltage : 30 Vac/Vdc (max)
ONE SPST SWITCH
Reed Switches and
Terminal Strip Connection
TWO SPST SWITCHES
Reed Switches and
Terminal Strip Connection
Reed Switches and
DIN Plu Connection
View of socket for supply connections
Reed Switches and
DIN Plu Connection
View of socket for supply connections
Reed Switches and
DIN Plu Connection
View of socket for supply connections
ONE SPDT SWITCH
Reed Switches and
Terminal Strip Connection
TWO SPDT SWITCHES
Reed Switches and
Terminal Strip Connection
View of plug after removing the socket
R = Red; B = Black; Y = Yellow; L = Live or + Supply; N = Neutral or – Supply. *Body to be suitably grounded while using gauge + switch and only switch.