INSTRUCTIONS FOR USE
Audiometric Booth
Rev.3.0
17/03/2022
PRO 25 rev. 03 user manual an installation Puma s.r.l. 2022 all rights reserved Page 3 of 17
Use and contraindications, notice
• The sound-p oof booth can be used only fo audiomet ic examinations o simila and not
fo othe medical pu poses with applied components (i.e. elect odes) to the patient. It is
necessa y having diffe ent specifications in this case.
• The booth must be used only by t ained pe sonnel.
• Man pe fo ming the test must always see the patient inside the booth.
• It is fo bidden to smoke and to light any flame inside the booth.
• The w ong booth installation o the position nea -by flammable mate ials disclaim all
manufactu e esponsibilities.
PUMA s l disclaims any esponsibility fo misuse, inco ect e-installation o imp ope use of the
booth.
Warnings
Connect the booth to an adequate elect ical implant and compliant with the no m CEI 64-8; the
plug must be inse ted in a 230V elect ical socket, p otected against ove load and sho t ci cuit
and with diffe ential switch.
The silent booth cannot be installed in places classified with "explosion haza d" (EN 60079-10-1
and EN 60079-10-2) and in ooms "used fo medical use" classified as g oup 2 (CEI 64-8
standa d / 710 a t. 710.2.7).
Installation is pe mitted in "o dina y" ooms and in "medical" ooms classified as g oup 1 (CEI
64-8 / 710 a t. 710.2.6); the powe supply must happen with an automatic switch with a
sensitivity diffe ential of 30 mA of "type AC", 230V.
Please note that the envi onmental elect ical system must comply with the law.
notice
Acco ding to the egulation EU 2017/745, annex 1, section III, pa ag aph 23.4, we advise you to
epo t any incident elated to the use of
the medical device both to the competent autho ity of the state in which it is used,
and to us as manufactu e s at the following add ess: info@pumas l.it