QRS QRS-101 User manual

QRS®-101
QRS®-101 P
OPERATING INSTRUCTIONS
GB
The 456 Magnovit International Produktions- und Handels AG has taken care in preparation of this manual, but makes no
expressed or implied warranty of any kind and assumes no responsibility for errors or omissions.
All rights reserved. No part of this manual may be reproduced, in any form or by any means (electronic, mechanical, or otherwise)
without the prior written permission of the 456Magnovit International Produktions- und Handels AG.
456,QWHUQDWLRQDO$*
Sales4560DJQRYLW$*
,QGXVWULHVWUDVVH · )/5XJJHOO
ZZZTUVPDJQRYLWFRP · RIILFH#TUVPDJQRYLWFRP
Manufacturer:
gbo Medizintechnik AG
Kleiststraße 6 · D-64668 Rimbach
Date of issue:
3
QRS®-101
QRS®-101 P

GB
NOTES IN ACCORDANCE WITH EC-DIRECTIVE
AND MEDICAL DEVICE DIRECTIVE (MDD)
QRS®-101 and QRS®-101 P are mains-powered magnetic field therapy devices of protection class I.
The devices are in accordance with the EC directive for medical devices (93/42/EWG) and therefore carry the
CE sign with the registration number of the notified body for medical devices. The according graphical
symbol is placed on the type plate.
According to the MDD, QRS®-101 and QRS®-101 P are class IIa devices.
The manufacturer is only responsible for the safety, operational reliability and functionality of the device if:
• the device is used in accordance with the operating instructions;
• the electrical installation of the location where the device will be used meets the respective current
requirements of electrical safety;
• the device is not used in hazardous environments and humid locations;
• mountings, enhancements, re-adjustments, modifications or repair works are carried out only by
personnel authorized by the manufacturer;
• the operator regulation of this EC directive is observed within the scope of MDD.
Technical support may be obtained by the manufacturer, dealers or service authorized by the manufacturer.
The product’s duration of life as scheduled by the manufacturer is 10 years.
QRS®-101 and QRS®-101 P are electronic devices. For their disposal the according regulations for electronic
devices have to be observed. Incidentals have to be disposed with residual waste.
On request, the manufacturer will provide you with further technical descriptions for all repairable parts of the
device, such as circuit diagrams, spare parts lists, and adjustment instructions as far as these are necessary for
the qualified technical staff of the operator.
Comments on electromagnetic compatibility (EMC)
Medical, electrical devices are subject to special precautions concerning the EMC. They must be installed and
operated according to the EMC-advice given in the accompanying documents. In particular medical, electrical
devices may be influenced by portable and mobile RF-communication devices.
The manufacturer guarantees the conformity of the unit with the EMC-requirements only when using
accessories which are listed in the EC declaration of conformity. The usage of other accessories my cause an
increased emission of electromagnetic disturbances or may lead to a reduced electromagnetic immunity.
The unit must not be arranged physically close to other devices or stacked with other devices. If such an order
is necessary nevertheless, the unit must be observed in order to check it for the intentional operation.
You find more EMC-comments in the chapter “Warnings and Safety Precautions” of this manual as well as in
the Technical Information on the next two pages.
4

GB
In accordance with the EMC-regulations for medical products we are obliged by law to provide the
following information.
Guidance and manufacturer’s declaration – electromagnetic emissions
The equipment is intended for use in the electromagnetic environment specified below. The customer or the user
of the equipment should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions, CISPR 11
Group 1
The equipment uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions, CISPR 11 Class B
The equipment is suitable for use in all
establishments, including domestic establish-
ments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions,
IEC 61000-3-2 (*) Class A
Voltage fluctuation/flicker emissions,
IEC 61000-3-3 (*) Complies
(*) Note: For devices with a power consumption between 75 W and 1000 W only.
Guidance and manufacturer’s declaration – electromagnetic immunity
The equipment is intended for use in the electromagnetic environment specified below. The customer or the user
of the equipment should assure that it is used in such an environment.
Immunity test IEC 60601 - test level Compliance level Electromagnetic environment –
guidance
Electrostatic
discharge (ESD),
IEC61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative h
umidity should be at least 30%.
Electrical fast
transient/burst,
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for input/output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge,
IEC 61000-4-5
±1 kV phase-to-phase
conductors
±2 kV phase-to-earth
±1 kV phase-to-phase
conductors
±2 kV phase-to-earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines,
IEC 61000-4-11
<5% Uo!for ½cycle
(>95% dip)
40% Uo!for 5 cycles
(60% dip)
70% Uo!for 25 cycles
(30% dip)
<95% Uofor 5 s
(>5% dip)
<5% Uo!for ½cycle
(>95% dip)
40% Uo!for 5 cycles
(60% dip)
70% Uo!for 25 cycles
(30% dip)
<95% Uofor 5 s
(>5% dip)
Mains power quality should be that
of a typical commercial or hospital
environment.
If the user of the equipment requires
continued operation during power
mains interruptions, it is recommended
that the equipment be powered from
an uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz) magnetic field,
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
Note: Uois the a.c. mains voltage prior to application of the test level.
NOTES IN ACCORDANCE WITH EC-DIRECTIVE
AND MEDICAL DEVICE DIRECTIVE (MDD)
5

GB
NOTES IN ACCORDANCE WITH EC-DIRECTIVE
AND MEDICAL DEVICE DIRECTIVE (MDD)
Guidance and manufacturer’s declaration – electromagnetic immunity
The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment
should assure that it is used in such an environment.
Immunity test IEC 60601- test level Compliance level Electromagnetic environment –
guidance
Portable and mobile RF com-
munications equipment should
be used no closer to any part
of the equipment, including
cables, than the recommended
separation distance calculated
from the equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
Conducted RF,
IEC 61000-4-6
3 Veff
150 kHz to 80 MHz
3 Veff d=1.23P
Radiated RF,
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m d=1.23P
for 80 MHz to 800 MHz
d=2.33P
for 800 MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter in
watts according to the transmitter
manufacturer and d is the recom-
mended separation distance in
meters (m).
Interference may occur in the
vicinity of equipment marked with
the following symbol:
Recommended separation distances to portable and mobile RF communication equipment
The equipment is intended to be operated in an electromagnetic environment, where radiated RF interference is controlled. The user
can help in avoiding interferences by means of meeting minimum separation distances between portable and mobile RF communication
equipment (transmitters) according to the maximum output power of the communication equipment.
Rated power of the
transmitter (W)
Separation distance according to the tranmission frequency (m)
150 kHz to 80 MHz
d=1.23
3
P
80 MHz to 800 MHz
d=1.2
3
P
800 MHz to 2,5 GHz
d=2.3
3
P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
6

GB
CONTENTS
NOTES IN ACCORDANCE WITH EC-DIRECTIVE AND MEDICAL DEVICE DIRECTIVE 4
OVERVIEW 8
1. INTRODUCTION 9
1.1 Intended use 9
1.2 View of QRS®-101 and QRS®-101 P 10
1.3 Description of the Keys 11
1.4 Description of the Display 11
2. START OF OPERATION 12
2.1 Transport and Assembly 12
2.2 Power Supply 12
2.2.1 Mains Fuses 13
2.3 Connecting the Applicator 13
2.4 Turning on the Device 14
3. FUNCTIONAL CHARACTERISTICS 15
3.1 Basic Functions 15
3.1.1 Selecting the Applicator Jack 15
3.1.2 Turning the Display Light On/Off 15
3.1.3 Turning the Speaker On/Off 16
3.1.4 Setting the Time 16
3.1.5 Setting the National Language 17
3.1.6 Restore Basic Settings 18
3.2 Home-Therapy 18
3.2.1 Fine-tuning of Intensity and Therapy Duration 19
3.3 Chip Card Mode 20
3.4 Timer Mode 21
4. APPLICATORS 23
4.1 The Coil Pillow K1 24
4.1.1 Dosage Guidelines for the Pillow Applicator 25
4.2 The Coil Mat M1 25
4.2.1 Application of Coil Mat M1 25
4.3 The Magnetic Field Bar MFS1 (Pen Applicator) 26
5. THE QRS® MAGNETIC FIELD THERAPY 27
5.1 Three Reliable Programs: Basis-, Vital- and Relax-Programs 28
6. TROUBLESHOOTING 29
6.1 Error Messages on the Display 29
6.1.1 Error Codes During Timer Therapy 30
6.2 Further Error Situations 31
7. MAINTENANCE 32
7.1 Safety Controls 32
7.1.1 Statutory Requirements and Regulations 32
7.1.2 Performance of Safety Checks 32
7.2 Cleaning, Disinfection and Care 33
8. CONTRAINDICATIONS AND SAFETY PRECAUTIONS 34
8.1 Operating Staff Security 34
8.2 Application for Baby and Child 35
8.3 Further Additional Reactions in the Application 35
8.4 Explanation of the Signs used 35
9. TECHNICAL DATA 36
10. ACCESSORIES 37
11. INDEX 38
7

GB
OVERVIEW
Chapter 1 describes the device’s basic characteristics and offers a brief
introduction into its operation.
Chapter 2 explains the device‘s setup and initiation.
Essential settings are described.
Chapter 3 describes all of the device’s functions and their operation.
Chapter 4 states indications about the use of applicators and accessories.
Chapter 5 explains how to carry out the therapies.
Chapter 6 lists possible failures, their indication as well as their possible causes.
Chapter 7 states indications of safety-related checks according to the MDD as
well as routine maintenance.
Chapter 8 lists possible contraindications of QRS®-Therapy, in addition to advices
for avoiding hazards.
Chapter 9 states all of the device‘s relevant technical data.
Chapter 10 depicts the device’s scope of supply and further accessories including
the corresponding part numbers.
Chapter 11 Index
8

GB
1.1 INTENDED USE
The Quantron Resonance System is the result of more than 20 years of fundamental research by leading
international scientists. A qualitative breakthrough has been achieved with this system in the domain
of magnetic field therapy, placing this therapy on a sound scientific basis, which will ensure its place in
medicine.
The Quantron Resonance System, in short QRS®, serves for pain therapy and has proven successful in many
indications even without further pharmaceutical therapies.
It supports both cell revitalization and cell metabolism by targeted ion transport. The whole metabolism is
activated, the immune system stabilised and cell regeneration improved, thereby strengthening weakened
body functions in a natural way.
According to intense international investigation and users‘ journals no harmful side effects could be located
to date (since 1993 about 200,000 devices of QRS patent have been field-tested).
The method is internationally patented (patent no. EP 0594 655).
Through the applicators the control unit QRS®-101 and QRS®-101 P create a low frequency, variable, vibrant
magnetic field of the body’s frequency pattern with a precisely defined wave shape. The intensity/frequency
is adapted according to the vital parameters and specific to the biological frames of the cell structures to be
treated as indicated. Since the individual electromagnetic hypersensitivity varies depending on the patient’s
health and his blood’s acidity, the magnetic field intensity may be altered to achieve the ideal effect. The
magnetic field intensity must be explicitly chosen via operation.
QRS®-101 and QRS®-101 P work with field strengths up to 40 micro Tesla and therefore well below the WHO
limit or so-called geomagnetic earth radiation. For comparison:
Highest QRS®-101 / QRS®-101 P intensity 40 µT
Terrestrial magnetic field 50 µT
WHO limit 100 µT
On this scale the therapy‘s intensity may be adapted to the current patient’s state of health manually.
The treatment time can be perfectly adapted to each person’s individual needs. After an individually set time
(1-60 minutes) the control unit automatically switches the magnetic field off.
The positive physiological effects which are started through the use of the QRS®magnetic field usually take
place within 8 hours (healthy organism). The regular recommendation is a treatment in the morning, in the
midday and in the evening.
The QRS®-101 is a „Home Device“. With this device it is possible to carry out therapies with the Basis, Vital
and Relaxed programmes without a chip card. Moreover, a chip card created at the doctor‘s surgery can be
used as an electronic prescription (section 3.3).
The QRS®-101 P is the P device. With this device only chip card operation is possible (section 3.3). It can be
used at the surgery as an additional therapy device.
1. INTRODUCTION
9

QRS®-101
QRS®-101 P
1.2 VIEW OF QRS®-1 01
ANDQRS
®-1 0 1 P
10
GB

GB
17:34 THERAPY 2
1.3 DESCRIPTION OF THE KEYS
QRS®-1 01 A N D Q R S ®-1 0 1 P
1.4 DESCRIPTION OF THE DISPLAY
QRS®-101 and QRS®-101 P have a 16-digit display. Their contents depend on the current device function. For
detailed information please refer to chapter 3, Functional Characteristics.
Example:
17:34 current time
THERAPY 2 selected therapy program
(e.g. the 2nd program on a chip card)
I/O
I/O
Start
Stop
2
8
min
16
min
24
min
1
On/Off101
A/B
OK
sensitive
medium
intensive
relax
basis
vital
timer 1-key
key Í
program-
key
display-
backlight
On/Off
Start-/
Stop-Taste
On/Off switch
key Î
applicator
selection- and
confirmation key
speaker
On/Off
intensity
and fine tuning
therapy duration
and fine tuning
I/O
I/O
Start
Stop
16
min
On/Off
101 PROFESSIONAL
A/B
OK
vital
speaker
On/Off
key Í
timer 2-
und Taste Î
display-
backlight
On/Off
11

GB
2.1 TRANSPORT AND ASSEMBLY
The image above shows supply and fuse on the device‘s upper quoin.
2. START OF OPERATION
12
power inlet fuse carrier applicator B applicator A
QRS®-101
and
QRS®-101
P
are mobile, mains-powered appliances that shall not be moved during intended
operation. Any location on a level surface is suitable. The devices should not be placed in front of a heater
or radiator (leave approx. 1 m clearance). There are no further requirements concerning wall distance or
ventilation.
The device meets the requirements DIN/VDE 0750, EN 60601-1, EN 60601-1-2 and belongs to the protection
class I. It is part of class IIa within the scope of the Medical Device Directive.
!ATTENTION
The unit is not designed to be operated in places with the inherent risk of explosions. If it is
used in dangerous areas of anaesthesia departments, the possibility of an explosion cannot
be excluded.
!ATTENTION
In order to prevent the risk of electric shock, the unit must be connected only to a supply
network with protective conductor.
If the patient and/or an applicator(-cable) is directly exposed to a radiator of a medical device for high
frequency heat therapy, the damage of the device or danger to the patient cannot be excluded. As a rule, a
distance of 3 m is sufficient.
2.2 POWER SUPPLY
QRS®-101
and
QRS®-101
P
are intended to be connected to mains voltages of 115 V or 230 V and a mains
frequency of 50 or 60 Hz. Within this range there is no need for further change or alteration on the devices.
Connect the device’s power jack to a grounded socket by means of the attached power cord.

GB
!NOTE
QRS®-101 and QRS®-101 P allow time-controlled operation (Timer Mode). In this case the
device starts automatically at a predefined time, selected either via key programming (see
3.4, Timer Mode) or via chip card (see 3.3, Chip Card Mode). This functionality is available
since the control unit permanently surveys its time-controlled routines in the so-called stand-
by mode, even if it has been turned off via the on/off switch. It is thus mandatory that the
control unit is continuously connected to the power supply through its mains cord.
Please do not remove the mains cord from the device and never connect the device to the
power supply via a switched supply (power strip).
2.2.1 MAINS FUSES
Mains fuses serve to protect the device from further damage in case of a serious error. Mains fuses are not
subject to aging or wear. Thus, a defect mains fuse always indicates an internal device error.
If the device does not show any function after being switched on (display stays dark), first make sure that the
socket and mains cord conduct line voltage. Should this be futile, please check the mains fuse and replace it
by a new fuse with the same values if it is defect.
The mains fuse is located in the fuse carrier as shown in the image above. The fuse is accessible by gently
pressing the fuse carrier toward the device and turning it about 1/8th turn counter-clockwise.
!ATTENTION
Please contact an authorized service partner in case of doubt after a defect fuse or after
multiple defects.
2.3 CONNECTING THE APPLICATOR
Please connect the applicator(s) (pillow, mat, pen) to the 15-pole jack(s) (submin-D).
!ATTENTION
The applicator’s 15-pole jacks are provided with 2 thumb screws for attachment. Always
hand-screw these into the stud bolts of the device’s jacks. Never apply tools to tighten
these screws.
13

GB
2.4 TURNING ON THE DEVICE
• To switch on the device please press the - key.
• The illuminated display shows the following text
(V1.1 is the current software version’s label and may change during the device’s development cycle.)
• The device now performs an automatic function test. The display shows
• In case of grave errors the display shows an error code (see chapter 6, Troubleshooting).
• If everything works properly. The display will show
• If required please adapt application-specific settings (see chapter 3.1).
• The device is now ready for operation, it displays the last operating state and waits for your input.
SYSTEMCHECK . . .
SYSTEM OK . . .
!NOTE
QRS®-101 and QRS®-101 P come with a time-out, turning the device off 8 minutes after not
having been used (8 minutes after the end of therapy or 8 minutes after the last key-click).
After this shutdown the device has to be restarted as described above.
On/Off
QRS 101 – V1.1
QRS 101 P – V1.1
14

GB
3. FUNCTIONAL CHARACTERISTICS
3.1 BASIC FUNCTIONS
3.1.1 SELECTING THE APPLICATOR JACK
QRS®-101 and QRS®-101 P are equipped with jacks for 2 applicators (see image in chapter 2, Start of
Operation). The physical identification for those 2 jacks are „A“ (more to the side of the case) and „B“
(almost quoin center). Some QRS®-System‘s applicators may be attached to these two jacks. If 2 applicators
are connected – which have not already been predefined by an electronic prescription (see 3.3, Chip Card
Mode), you may chose the applicator to be applied for the next therapy on the device.
The display of the selected applicator ensues by pressing the
A/B
OK
-key. All of the QRS®-System‘s applicators are
coded, thus QRS®-101 and QRS®-101 P may identify the connected applicator‘s type. Therefore the display will
not show the applicator’s physical identification („A“ or „B“) but the descriptive, logical identification („mat“,
„pillow“ or „pen“). You may for example see the following display:
This display will be shown for 1.5 seconds before returning to the display of the operating mode.
You may switch to the respective other applicator by pressing the
A/B
OK
-key a second time while the above
stated display is still visible.
PILLOW –>
!NOTE
You may only switch to another applicator if 2 applicators are actually connected to the
device. If only one applicator has been connected, it may be displayed but the cursor symbol
(–>) will not appear on the display.
3.1.2 TURNING THE DISPLAY LIGHT ON/OFF
The display’s backlight may be turned on or off. If it is turned on it stays active during the whole operation.
If it is turned off the display will light up for every operation (keys) but shuts down independently after one
minute.
The backlight status may be checked by simply pressing the
I/O
-key. Depending on its status one of the
following displays will appear
BACKLIGHT ON
BACKLIGHT OFF
or
15

GB
These are always visible for 1.5 seconds before returning to the previous status. If the
I/O
-key is being
pressed while the display shown above is visible, the display light’s status will be changed, i.e. the switched-
on display light will be turned off and vice versa.
3.1.3 TURNING THE SPEAKER ON/OFF
QRS®-101 and QRS®-101 P contain a speaker, called ‚beeper‘ in the following. This beeper signalizes a key-click,
an error or the end of therapy with different sounds, each.
The key-click’s acoustical echo, as well as the error signal may not be turned off, contrary to the acoustic signal at
the end of therapy, since the latter disturbed some patients.
The end of therapy’s signal status may be checked by simply pressing the
I/O
-key. Depending on its
status one of the following displays will appear
These are always visible for 1.5 seconds before returning to the previous status. If the
I/O
-key is being
pressed while the display shown above is visible, the end of therapy signal’s status will be changed, i.e. the
switched-on signal will be turned off and vice versa.
TH.–END BEEP ON
TH.–END BEEP OFF
or
3.1.4 SETTING THE TIME
QRS®-101 and QRS®-101 P possess a clock (battery backed RTC). To set the time the device’s user interface offers
you convenient menu guidance. You may reach the corresponding menu by
• turning the device off,
• pressing the
!
!"#$
!%&'%
-key and keeping this key pressed, while the device stays turned off and
• turning the device on via the
On/Off
You are now positioned in the menu to set the time and see the display (exemplary time)
TIME: 12.34
16

GB
The blinking cursor is positioned on the time’s first digit, „1“, as shown above. You may now change the time
by increasing the digit positioned below the cursor by pressing the Î-key, or to lessen it via the
Í-key. The
A/B
OK
-key confirms the current value and goes to the next position.
After confirming the last time digit you automatically reach the menu to set the date. The corresponding
display may look as follows:
Please proceed accordingly to set the date. By confirming the last date digit the menu is closed and the
device will be started with its standard functionality.
DATE: 22.10.07
!NOTE
• You may leave the menus to set time and date at any time by pressing the
Start
Stop
-key.
The currently made (incorrect) settings will then not be effective.
• Only the German format for time and date are available, exclusively.
3.1.5 SETTING THE NATIONAL LANGUAGE
You may operate QRS®-101 and QRS®-101 P in different languages. To set the standard language the device’s
user interface offers you convenient menu guidance. You may reach the corresponding menu by
• turning the device off,
• pressing the
A/B
OK
-key and keeping this key pressed, while the device stays turned off and
• turning the device on via the
On/Off
-key.
You are now positioned in the menu to set the national language and see the display
You may now move the cursor
1
or
2
through the menu to chose among various languages. The
abbreviation of the currently selected language is marked between brackets „[“ and „]“.
The following languages are available:
• DE German
• EN English
• FR French
• IT Italian
• ES Spanish
• SE Swedish
<– . . . ES [DE] EN . . . –>
17

GB
Pressing the
A/B
OK
-key confirms the current selection, leaves this menu and starts the device and its standard
functionality.
!NOTE
You may leave the menu to set the language at any time by pressing the der
Start
Stop
-key.
The currently made (incorrect) setting will then not be effective.
3.1.6 RESTORE BASIC SETTINGS
Through a certain hot key you may restore the factory-made basic device settings.
You may restore these basic settings by
• turning the device off,
• pressing both
I/O
and
On/Off
-keys simultaneously, keeping them pressed, while the device
stays turned off and now
• turning the device on via the
On/Off
-key.
QRS®-101 and QRS®-101 P will now reset to the following basic settings:
• display light OFF
• speaker ON
• language DE (German)
Basic settings only for QRS®-101
• therapy program relax
• therapy intensity medium
• therapy duration 8 min
• timer 1 OFF and 06:00 o‘clock, „vital“, „medium“, 8 min.
• timer 2 OFF and 22:00 o‘clock, „relax“, „medium“, 8 min.
3.2 HOME-THERAPY (ONLY QRS®-101)
The home therapy describes a mode of operation where a QRS®-101 therapy is performed via simple key
programming.
The operation is quite simple. For each therapy 3 parameters have to be defined:
• the therapy program (relax, basis or vital),
• the therapy intensity (sensitive, medium or intensive) and
• the therapy duration (8, 16 or 24 minutes).
18

GB
You may select the stated parameters via control keys on the keyboard, arbitrarily. The therapy will be started
after setting these parameters by pressing the
Start
Stop
-key.
Before starting the therapy you may typically see the following display:
Here, the current time is shown on the display’s left side, the right side will alternately display the selected
parameters for the therapy program, the therapy intensity or the therapy duration (e.g. therapy intensity, as
depicted above).
After starting the therapy the display will change to the following typical view:
On the display’s left side the current therapy program and selected therapy intensity are shown, alternately,
the right side displays the remaining time until the end of therapy. The time (of day) and the pre-defined
therapy duration are not displayed during the therapy.
16:41 INTENSIVE
INTENSIVE 06:21
!NOTES
• The home therapy may not be performed if a chip card has been inserted into the card
reader. Please remove the chip card, first.
• The therapy intensity „intensive“ is not available for the therapy program „relax“. The
intensity „medium“ will always be chosen, whenever the intensity „intensive” is being
selected.
3.2.1 FINE-TUNING OF INTENSITY AND THERAPY DURATION
When necessary the intensity and duration of the home-therapy can be fine tuned.
The therapy duration can be adjusted starting from any time key (8 min, 16 min, 24 min). Press the key for
2 sec. until the time in the display flashes and shows e.g. „16 MIN.“. By pressing the key the time decreases
in 1 minute steps and by pressing the key the time increases in 1 minute steps.
The intensity can be adjusted starting from any of the intensity keys (sensitive, medium, intensive). Press the
key for 2 sec. until the intensity in the display changes and shows e.g. „INT=05“. By pressing the key the
intensity decreases and by pressing the key the intensity increases. The intensity range starts from „SE“
(=sensitive) up to 1 -10. The therapy program „relax“ allows only values up to 5.
After having reached the desired setting press the confirmation key („OK“) and continue with normal
operation.
19

GB
3.3 CHIP CARD MODE (QRS®-101 AND QRS®-101 P)
A chip card contains an electronic prescription and will be given to you by your physician or therapist. The chip
card may include up to 4 different therapies. All required parameters have been saved on the corresponding
chip card.
To perform a therapy through chip card, please insert it into the card reader. At this point it is irrelevant whether or
not QRS®-101 resp. QRS®-101 P have already been turned on.
After inserting a valid chip card you will alternately see one of the following 3 typical displays:
Here, the display’s left side shows the current time, its right side displays the chip card‘s therapy index (therapy
1, 2, 3 or 4). If several therapies (max. 4) have been prescribed on the chip card, you may page within those with
the keys
1
and
2
.
Following the therapy number the plain text therapy description will be displayed, for example:
This plain text has been chosen by your physician or therapist and serves solely as mnemonical description to
identify the therapy. If this text is longer than 16 characters they will be displayed as a sort of ticker, as running
text.
Following the therapy description the respective applicator jack will be displayed, finally. This is
either the or .
Afterwards the display routine will repeat by presenting the time and therapy number..
The therapy is started via the
Start
Stop
-key.
!NOTES
• The chip card depicts the direction in which it has to be inserted into the card reader.
All other possible directions (3 out of 4) will cause the following error:
• If a chip card is inserted in the control unit all pressing of control keys to perform a home
therapy will be ignored.
CARD NOT VALID
16.41 THERAPY 2
MORNING THERAPY
APPLICATOR A
APPLICATOR B
20

GB
After starting the therapy the display will show the following typical contents:
On the display’s left side the current therapy number and therapy description (plain text) will be shown,
alternately, on its right side you can see the remaining time until the end of therapy. The time of day, the
corresponding applicator jack, as well as the preselected therapy duration are not displayed during the
therapy.
THERAPY 3 06:21
!NOTES
• In making a chip card the maximum number of therapies to be performed by the patient
will be prescribed. This credit of therapy units will be deduced by one after performing
each therapy. As soon as this credit has been exhausted the corresponding therapy may no
longer be selected from this chip card. Please contact your physician or therapist as soon as
possible.
• If a QRS®-101 or QRS®-101 P control unit, containing a chip card with automatical, time-
controlled start of therapy („autostart“ option), is taken out of the power supply and
reconnected later, the device will switch on, independently, display the text „AUTOMODE“
and switch off again. The same display appears if a chip card is inserted into a switched-off
control unit.
3.4 TIMER MODE (ONLY QRS®-101)
QRS®-101 contain 2 independent „timers“. These programmable timing circuits serve to automatically start a
therapy. Here you are free to choose the starting time as well as all therapy parameters (therapy program, intensity,
duration and applicator).
The timers have to be programmed and activated before their first use. Afterwards the timers will start the selected
therapy every 24 hours, unless they have been deactivated explicitly.
Both of the QRS®-101 timers are equal and will be activated and programmed the identical
commands. The following example shows the commands for timer 1. Timer 2 is to be handled identically.
!NOTES
• To perform a therapy in timer mode the device may be turned off but must be
connected to a live power supply (see chapter 2.2, Power Supply).
• At the first use of the timers they are set as described in chapter 3.1.6, Restore
Basic Settings.
21

GB
TIMER 1: [OFF] . . .
TIME T1: 22:30
!NOTE
If a QRS®-101 control unit with activated timer is taken out of the power supply and reconnected
later, the device will switch on, independently, display the text „AUTOMODE“ and switch off
again.
Activation and deactivation of both timers start on a switched-off device. To program the timer 1 please press
the
1
-key keeping it pressed. Now turn the device on via
On/Off
-key. The display will then show
To activate the timer apply the
2
-key, to deactivate the timer apply the
1
-key. The display will
change, accordingly. This command must be confirmed via the
A/B
OK
-key.
If the timer is deactivated, the device will now start with the standard operating mode. If the timer is activated
you may now start with its programming.
As its first programming step you will set the time at which the therapy shall be started, automatically. The
display will show the following menu (exemplary time):
The blinking cursor is positioned on the time’s first digit „2“, as shown above. You may now change the time
by increasing the digit positioned below the cursor by pressing the
2
-key, or to lessen it via the
1
-key.
The
A/B
OK
-key confirms the current value and goes to the next position.
After confirming the last time digit you automatically reach the menu to set the therapy parameters. Alterna-
tely you can now see the settings programmed for this timer regarding therapy program, therapy intensity,
therapy duration and applicator. To change one or several of these settings you may press the corresponding
control key in any given order.
Should 2 applicators be connected to the control unit you may also change the respective applicator via
A/B
OK
-key. After the first click on this key the programmed applicator will be displayed for 1.5 seconds.
The respective applicator will be changed by pressing the
A/B
OK
-key a second time while the display is still
visible – i.e. no later than 1.5 seconds after the first click.
Programming the timer is concluded by pressing the
Start
Stop
or
On/Off
-key. In both cases all previously
set values will be saved for this timer. By pressing the Start-/Stop-key the device will be set to the standard
operating mode. The on/off-key will turn the device off.
22
Other manuals for QRS-101
1
This manual suits for next models
1
Table of contents
Other QRS Medical Equipment manuals