QT Medical QT ECG Recorder Operating instructions

QT ECG
Troubleshooting
855-722-6006O /
310-755-3108
F /
Contact Us
Android iOS
0
Update Date:07/13/2021

1
Preventive Maintenance 2
1. QT ECG Recorder 3
1.1 QT ECG Recorder 3
1.2 Unable to turn on the QT ECG Recorder 3
1.3 Unable to turn off the QT ECG Recorder 3
1.4 Unable to connect the QT ECG Recorder to the mobile device 3
(iPad, Android tablet, etc) via Bluetooth
1.5 Connection Failed, No Recorders Found 4
1.6 4
5
2. APP 6
2.5 Can not connect to the device. 6
Table of content
3. Other Problems 12
3.1 12
3.2 13
Customer Service 14

2
Preventive Maintenance
QT ECG Recorder
To clean the QT ECG Recorder, dampen a soft cloth with lukewarm soapy
water or natural cleanser and gently wipe down the recorder. The QT ECG
Recorder should not be immersed in water under any circumstances. Take
care not to scratch the device with abrasive cleansers or excessive wiping on
the label. The QT ECG Recorder should be gently cleaned before every use. To
disinfect the QT ECG Recorder, dampen a soft cloth with isopropyl alcohol and
gently wipe the Recorder. Regularly inspect the QT ECG Recorder for damage
such as warping or cracking. If there is any damage, do not use the QT ECG
Recorder and contact QT Medical customer service.
The following sections describe how to clean the QT ECG system, including
recommended cleaning and disinfecting solutions and methods. This section
also covers routine preventive maintenance for the QT ECG Recorder.
The Electrode Strip is clean and ready to use straight out of the package.
There is no need to clean it. Electrode Strips are for one-time use only and
should be discarded after use.
The cleaning procedure for the wall charger is the same as that for the QT
ECG Recorder. Disconnect the wall charger from the power outlet before any
cleaning procedure.

3
1. QT ECG Recorder
1.1
1.2 Unable to turn on the QT ECG Recorder
1.3 Unable to turn off the QT ECG Recorder
1.4 Unable to connect the QT ECG Recorder to the mobile device
(iPad, Android tablet, etc) via Bluetooth
Solution : PWR LED indicator will turn off after the battery is fully charged.
If the PWR LED does not light up while charging, this may indicate that the
battery is fully charged.
Solution : Press and hold the power button for 30 seconds. If the Power LED
indicator still does not light up, try again after charging the device for 1 hour.
If above mentioned solution does not work, please contact QT Medical
customer service.
Solution : If Bluetooth is not connected when the QT ECG Recorder is on, the
blue CON LED indicator will blink. When Bluetooth is properly connected, the
CON LED indicator will be a solid light. If the CON LED indicator is off when
the Recorder is turned on, please contact QT Medical customer service.
Solution : Make sure the Micro-USB cable is not connected.
Solution : Press and hold power button for 30 seconds. If still unable to turn
off the QT ECG Recorder, please contact QT Medical customer service.

4
1.5 Connection Failed, No Recorders Found
Solution: (if PWR LED is solid green, CON LED is flashing blue), "connection
failed" appears, please reboot the recorder, and ensure the distance between the
recorder and mobile device is within 10 meters and no objects in between effecting
the connection.
Solution: Press "Cancel" after 3 minutes, reboot the QT ECG Recorder and try
pairing again, or refer to Step 2.5

5
1.7 (For Android) Pairing request does not appear
Solution: Swipe down to Settings and tap on "Pairing request" to manually start
the pairing process.
Please reboot the QT ECG Recorder and try again, or refer to step 2.5

6
2. APP
2.5 Cannot connect to the device
Solution : Make sure you run the latest version of the APP and the operating
system on your mobile device. Reboot and restart your mobile device. If the
problem persists, please contact QT Medical customer service.
Always ensure that the latest version of the app is running; if an older version
of the app is running, the app may not function properly (Please be advised,
we recommend that you turn off the automatic update functions for your
phone’s OS and app. Before proceeding with updates, please check our
compatibility list for the most updated OS/app combination that we currently
support.)
Solution : Please contact QT Medical customer service.
https://www.qtmedical.com/contact-us
Solution : Too many apps may be opened and running at the same time.
Reboot the mobile device and try again.
Solution : The problem may be caused by system overload. Disconnect the
QT ECG Recorder from the Electrode Strip then carefully remove the Strip
from the Patient’s chest. Restart the entire process by re-attaching the
Electrode Strip (or use a new Strip) on the patient. If the problem persists,
please contact QT Medical customer service.
Solution : If the connection fails over 3 times, please retry with the following
steps :
1. Reboot the device
2. Restart the APP
3. If connection fail remains, the possible reason is that too many Bluetooth
devices are connected to the mobile device, the problem may be solved by the
following steps:

7
(For )To remove connected Bluetooth device, please follow these steps
to disconnect, Setting > Bluetooth > Remove unused devices.
(For Android)Setting > Bluetooth > Remove unused devices or
Reset network settings.
Select Setting>Bluetooth>Select the SN of the QT ECG Recorder you wish to
connect.
Please force close the QT ECG APP.

8
Solution : Check the corresponding limb/chest electrode(s). Make sure each
electrode is attached properly to the skin and making good physical contact.
Firmly press on the area of the strip indicated in yellow to ensure good
electrode contact with the skin.
press firmly down on the strip to ensure it is making good skin contact.
If this does not resolve the problem, ensure the skin is clear. If there is
hair in the electrode strip area, shave along the indicated area. If the skin
is oily or dirty, try wiping down the skin under the electrode strip with a
clean cloth.
In a low humidity environment, electrodes may dry up, preventing
adhesion to the skin. If the electrodes have been stored in a dry
environment, and a "lead-off" alert is shown in the app during use,
please use a moist cloth to dampen the hydrogel circle(s) of unattached
electrode(s) indicated in the app alert, then reattach the electrode.
If lead-off remains, but the hydrogel circles has been dampened and strip
firmly contacted to skin already.
Please try the following steps:
1. Take off the strip and re-attached to skin,
2. Unplug the connector and plug in again.
3. then reboot the recorder and try again.
If lead-off remains still after all the process, please contact us.
1
2
3
4
5
See figure below :

9
*Common Lead off indication
A. Make sure each electrode is attached properly to the skin making good
physical contact, and the hydrogel circles on electrodes are moist and
conductive.
B. unplug the strip from the connector, and plug in again, then reboot the
recorder.
This usually happens when RL lead is off; make sure to press firmly on RL
to ensure proper attachment.
Follow the indications and check the corresponding leads, if lead is
attached properly but shows leads off, please use a moist cloth to
dampen the hydrogel circle(s) of these electrode(s) indicated in the app
alert, then reattach the electrode.

10
Solution : Review the User’s Manual to ensure all steps and instructions are
followed properly.
Solution : If the instructions are followed but the app problem persists, close
the app, reboot the mobile device and the QT ECG Recorder, and try again.
Note: For IOS system: Make sure the bluetooth in the "Settings" and "Control
center" are both opened.
Solution : If all the steps are followed but the recorder continues to fail to
record , contact QT Medical customer service.
Situation 1 : When the Electrode Strip first attaches to the recorder.
Situation 2 : When the mobile device first connects to the recorder through
Bluetooth.
Situation 3 : When the mobile device first comes back from sleep mode.
Situation 4 : When restarting recording after uploading successfully.
Solution : Wait for 5~10 seconds before recording. If the recorder continues
to fail, contact QT Medical customer service.
Solution : Disconnect the Recorder from the Electrode Strip and remove the
Electrode Strip from the patient. Charge the QT ECG Recorder using the wall
charger and Micro-USB cable provided. Do not attempt to use the Recorder
and charge the battery at the same time.

11
Wait for 5~10 seconds before recording. If the recorder
continues to fail, contact QT Medical customer service.
Reset network settings on the mobile device and record
again. If the problem persists, please contact QT Medical customer service.
Solution :
The weak signal message means there are lots of wireless interferences in the
environment that are affecting the Bluetooth transmission. The interferences
may come from smart phone or external wireless devices.
Please make sure your smart device has:
A. turned off WiFi Hotspot function.
B. disconnected all other Bluetooth devices such as wireless headphones,
wireless speakers, smart bands and smart watches.
Please make sure there are less than 10 smart devices (wireless
devices using WiFi and Bluetooth) around. If the external interference exists,
please try to find another place to do the measurement.
If the above mentioned steps do not resolve the issue, please try the
following:
Tip: To reset network settings,
open your phone’s Settings app and tap System >
Advanced > Reset options > Reset Wi-Fi, mobile & Bluetooth.
open the mobile device’s Settings app and tap General
management > Reset > Reset network settings.
G
S

12
3.1 Other conditions
3. Other Problems
1. Dense hair on the chest may cause weak signal, shaving the chest hair
along the adhesive area is suggested.
2. If the strips are loosely attached, causing leads off error, please use
surgical tape to securely attach the strip tightly.
3. Our strips have passed biocompatibility test. If you have a history of
sensitive skin or allergy, and your skin becomes red, itchy or inflamed
during the procedure, please remove and consult your physician.
4. For female, please avoid the breast; have the strip go around the
underside of the breast by moving the strip lower.

13
1
2. It is advised to remove the bra and
hold the left breast up to a "neutral
position", place the "blue box" in the
center of the chest between nipples.
1. Female patients should lie down ßat
and have someone on the side to help.
2
Neutral
Position
3. Wrap the tail of the strip around the
underside of the left breast. Press on
all parts of strip to ensure good adher-
ence to the skin.
3
4. In some women, it may be challeng-
ing to place the "blue box" between
the breasts. It may be necessary to
move the entire strip slightly lower to
allow more space for better place-
ment.
4
V1 V2
Instructions for Electrode Placement
on a Female Patient
3.2 Instructions for electrode placement on a female
patient.

14
1360 Valley Vista Dr., Suite 203
Diamond Bar, CA 91765 U.S.A.
TEL : +1 855-722-6006
FAX : +1 310-755-3108
www.qtmedical.com
For customer service support, please contact:
Other manuals for QT ECG Recorder
1
Table of contents
Other QT Medical Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Force Dimension
Force Dimension omega haptic device Series user manual

Gima
Gima 32579 user manual

Graham Field
Graham Field Lumex FR587W Series Assembly and operating instructions

Merit Medical
Merit Medical Q50X Instructions for use

Olympus
Olympus PSD-30 instructions

Meyra-Ortopedia
Meyra-Ortopedia AS-Basic Medical MY275101 manual