PROGRAMMING
- 40 -
4.4.2 Span Calibration
This procedure determines the second point of the sensor
calibration curve for the sensor. A cylinder of standard
reference gas (span gas) fitted with a 500 cc/min. flow-
limiting regulator or a flow-matching regulator is needed to
perform this procedure. Choose the 500 cc/min. regulator
only because the flow rate matches the flow rate of the
pump inside. Alternatively, the span gas can first be filled
into a Tedlar Bag. Connect the calibration adapter to the
inlet port of the ppbRAE Monitor, and connect the tube to
the regulator or Tedlar bag.
1. Before executing a span calibration, make sure the span
value has been set correctly (see next sub-menu).
2. Make sure the monitor is connected to one of the span
gas sources described above.
3. Press the [Y/+] key at the “Span Cal?” to start the
calibration. The display shows the gas name and the
span value of the corresponding gas.
4. The display shows: “Apply gas now!” Turn on the
valve of the span gas supply.
5. Display shows “wait.... 30” with a count down timer
showing the number of remaining seconds while the
monitor performs the calibration.
6. To abort the calibration, press any key during the count
down. The display shows “Aborted!” and return to
“Span Cal?” sub-menu.
7. When the count-down timer reaches 0, continue to
apply gas. The display will read "Please wait" - continue
apply gas. The display will show the calibrated value.