Rayleigh Instruments RI-F200 User manual

RI-F200-IM-V01
1
RI-F200-B-MB / RI-F200-G-MB
RI-F200-B-C / RI-F200-G-C
RI-F200 Instrucon Manual
X 4
RTU
Pulse SO
Safety related noficaon, symbols and instrucons that appear in this
operang manual or on the equipment must be strictly followed to
ensure the safety of personnel as well as the instrument. If the
equipment is not used in a manner specified by the manufacturer it may
impair the protecon provided by the equipment
· Do not use the equipment if there are mechanical damage
· Do not exceed the stated maximum rangs of the device
· No repairs, maintenance or adjustments are possible
· Read the complete instrucon manual prior to installaon or
operang the unit
· The equipment in its installed state must not come into close
proximity to any heang sources, oils, steam, causc vapours or
other unwanted process by-products
· Do not use in hazardous or classified locaon where explosion or
other dangers can be triggered by the device
PRODUCT SAFETY INSTALLATION PRECAUTIONS
Risk of electric shock!
Only to be installed by a competent person
· To prevent the risk of electrocuon, always isolate and lock-off the
power supply to the equipment prior to undertaking any work
· Always confirm absence of electricity prior to starng work using
appropriate voltage detecon equipment
· Wiring shall be done strictly according to the terminal layout
· Confirm that all connecons are correct before energizing the
equipment
· Roung of cables shall be way from any internal EMI source
· Copper cable should be used
· All wiring to be in accordance with applicable local standards
Specificaons Accuracy
Wiring Input 3Ø 4 wire, 3Ø 3 wire, 2Ø 3 wire, 1Ø 2 wire Voltage V L-N and V L-L ±0.5% of full scale
Rated Input Voltage 3x 11...300V AC (L-N), 19...519V AC (L-L) Current ±0.5% of full scale
Frequency Range 45...65Hz Frequency for L-N > 20V, L-L > 35V ±0.1% of full scale
CT Primary 1A/5A...10,000A configurable Acve, Reacve and Apparent Power 1%
CT Secondary 1A or 5A (max rang x1.2) Power Factor ±0.01 of Unity
VT Primary 100...500kV configurable Acve Energy Class 1 (IEC/EN 62053-21)
VT Secondary 100...500V AC (L-L) configurable Reacve Energy Class 2 (IEC/EN 62053-23)
Auxiliary / Power consumpon 100...240V AC (±15%) 45...65Hz / < 8VA
Display Update Rate 1 sec all parameters
Operang / Storage Temperature -10…55°C / -20…70°C Wh Resoluon and Default Pulse Weight
Humidity 0...85% non-condensing
Example
If CT = 100/5A (CT rao = 20) & VT = 350/350V (VT Rao = 1)
Wh resoluon = 0.1kWh (20 x 1 = <150) | Pulse O/P default = 0.1kWh/pulse
Protecon Degree (IEC/EN60529) IP54 Front only (rubber gasket fied)
Pulse Output External 5...27V DC / 100mA max
Pulse Resoluon / Duraon 0.01...99.99kWh per imp / 0.1...2 sec
Communicaon Modbus RTU over RS485
MBus (EN13757)
CT Rao x VT Rao <15 <150 <1500 <15k <150k
Wh / VArh / VAh 0.01k 0.1k 1k 0.01M 0.1M
>150k
1M
INT 0.01k 0.1k 1k 0.01M 0.1M 0.1M
MECHANICAL INSTALLATION
EO&E: Document subject to change
Installaon & Environmental
Panel mounted, indoor used only.
Installaon category: III (300V L-N)
Altude: up to 2000 m
Protecon Class: II
Polluon degree: II
All terminal covers must be fied aer wiring
RI-F200-IM-V01
2
WIRING
Modbus / MBus
Wiring Topology
Daisy Chain Star Network
Modbus ü û
MBus ü ü
A | B
+ | -
1 | 2
Typical Modbus
configuraon shown
For MBus interface
follow Wiring
Topology below
500m Max, ≤ 32 meters
Fuse class CC UL / fast acng 600V
Rang 0.5A
For ’Volt-free’ PLC or digital input,
voltage must be provided by the addion
of a DC PSU
Single Core
0.5 > 4mm2
Ø 2.5mm max
Stranded
0.5 > 2.5mm2 +
7mm
Ø 2.5 mm
0.8 Nm
Max
3P4W
3 CTs
1P2W
1 CT
2P3W
2 CTs
2 VTs
2P3W
2 CTs
3P3W
2 CTs
2 VTs
3P3W
2 CTs
3P4W
3 CTs
3 VTs
Voltage
and current
must be
from same
phase
RI-F200-IM-V01
3
F200-x-C
F200-x-MB
Seng Default Adjustment Range Network
& CT
Must
configure
VT
Adjust if
Using VT
Comms
Modbus /
MBus
Pulse O/P
Adjust if
ulised
System
Sengs
Oponal
1 1 Change Password 1000 NO / YES (0000 - 9998) ü
2 2 Phase Network Selecon 3P4W 3P4W
3P3W
1P2W-P1
ü
3 3 CT Secondary (see CT Label) 5 1A / 5A ü
4 4 CT Primary (see CT Label) 5 1A/5A > 10000A ü
5 5 PT Secondary 350 100V > 500V ü
6 6 PT Primary 350 100V > 500kV ü
7
7
Slave ID Modbus:
MBus (Primary ID):
1
1
1 > 255
1 > 250 ü
8
8
Baud rate Modbus:
MBus:
9600
2400
300 > 19200 bps
1200 > 9600 ü
9
9
Parity Modbus:
MBus:
None
Even
None / Odd / Even
Even ü
10
10
Stop Bit Modbus:
MBus:
1
1
1 / 2
1 ü
11 11 Back Light Off (0000 = Never) 0000 0 > 7200 Sec ü
12 12 Demand interval method Sliding Sliding / Fixed ü
13 13 Demand interval duraon 15 1 > 30 ü
14 14 Demand interval length 1 1 > 30 min ü
15 15 Max Auto Display Pages 21 1 > 21 ü
16 16 Change Page Sequence 1 > 21 No / YES (1 > 21) ü
17 17 Pulse Weight 0.01 00.01 > 99.99 kWh/imp ü
18 18 Pulse Duraon 0.1 0.1 > 2.0 Sec ü
X 19 MBus Secondary ID Serial # 0000 0000 > 9999 9999 ü
19 20 Factory Default No No / Yes Does not reset energy & demand values ü
20 21 Reset Energy No No / Yes (Password +1) ü
Once entered, reset each value individually
21 22 Reset Demand No No / Yes ü
22 23 Reset Auxiliary Switch ON No No / Yes ü
CONFIGURATION
Step A: Enter Configuraon Menu Step B: Configure each seng, as required, referring to
Sengs Table below, using the buons as follow:
1 Press either buon once to make digit or opon flash,
press again to move flashing cursor
2 Press to change digit or opon, press or
to move cursor posion as required
3 Press to save and move to next seng opon, Exit
menu once all sengs are configured (see Step 1)
Hold 3 Secs to
Enter Config
(same to Exit)
Password = 1000
(Refer to Step 2)
2
1
RI-F200-IM-V01
4
OPERATION
1 Current level bar graph (% of CT current rang)
2 Funcons & displayed measurement indicators:
RS485 communicaon in progress
Integraon of energy (blinks every 5 sec)
∑ Sum of 3-phase
AVG Average of 3-phase
DMD Max/Min Demand
TH Total Harmonic Distoron (THD)
IP Imported Energy (posive value)
EP Exported Energy (negave value)
Total Sum of 3-phase Energy (IP or EP)
Net Sum of IP + EP Energy
3 Funcon buons and funcon symbols
4 Phase & total or average instantaneous measurements
(V, A, PF, Hz, kW, kVAr, kVA) >> V/I/VAF/PF/P buons
5 Energy readings (kWh/kVArh/kVAh) >> E buon
Voltage Phase Sequence
Hold 3 sec
OK-CLK: L1 → L2 → L3
ANTI-CK: Incorrect Order
INVAL Id: Missing Phase û
ü
Rayleigh Instruments Limited (Head Office)
Raytel House, Cutlers Road, South Woodham Ferrers, Chelmsford, Essex. CM3 5WA
W: www.rayleigh.com E: sales@rayleigh.com
Rayleigh Instruments Limited SP. Z o.o. (European Office)
Ul. Aleje Jerozolimskie 214, Warsaw, Poland, 02-486
W: www.rayleigh.pl E: sales@rayleigh.pl
E
x1 > 9 Acve Energy - kWh L1 IP L2 IP L3 IP L1 EP L2 EP L3 EP ∑ IP ∑ EP Net (IP + EP)
x10 > 18 Reacve Energy - kVArh L1 IP L2 IP L3 IP L1 EP L2 EP L3 EP ∑ IP ∑ EP Net (IP + EP)
x19 > 22 Apparent Energy - kVAh L1 L2 L3 ∑ Hold 10 sec - change Page Scroll
E Auto <-> Manual
x23 > 24 Run Hour (0.01 hr = 36 sec) Number of Auxiliary Switch ON
No. of
Presses V I VAF PF P
x1 Voltage (L-N) Current L1: V/A/PF/Hz Power Factor Acve Power - kW
x2 Voltage (L-L) Current Max DMD L2: V/A/PF/Hz
Hold 10 sec
displays Serial #
Reacve Power - kVAr
x3 % THD (L-N) % THD L3: V/A/PF/Hz Apparent Power - kVA
x4 % THD (L-L) Avg: V/A/PF/Hz L1: kW/kVAr/kVA/PF
x5 L2: kW/kVAr/kVA/PF
x6 L3: kW/kVAr/kVA/PF
x7 ∑: kW/kVAr/kVA/PF
x8 Max DMD: kW/kVAr/kVA
x9 Min DMD: kW/kVAr
Hold 10 sec
Reverse Indicaon
0 = None
1, 2 or 3 = Reversed 1. L-N parameters are not
displayed For 3P3W
2. Parameters in BOLD not
displayed for 1P2W

Other Rayleigh Instruments Measuring Instrument manuals

Rayleigh Instruments RI-D384-C User manual

Rayleigh Instruments

Rayleigh Instruments RI-D384-C User manual

Rayleigh Instruments RI-F300 User manual

Rayleigh Instruments

Rayleigh Instruments RI-F300 User manual

Rayleigh Instruments RI-F220 User manual

Rayleigh Instruments

Rayleigh Instruments RI-F220 User manual

Rayleigh Instruments RI-D440 User manual

Rayleigh Instruments

Rayleigh Instruments RI-D440 User manual

Popular Measuring Instrument manuals by other brands

TSI Instruments Dickey-John MINI GAC 2500 quick start guide

TSI Instruments

TSI Instruments Dickey-John MINI GAC 2500 quick start guide

Quantel EverGreen2 Quick setup guide

Quantel

Quantel EverGreen2 Quick setup guide

LaserLiner StarFinder quick start guide

LaserLiner

LaserLiner StarFinder quick start guide

3M SCOTT SEMS II PRO Operation and maintenance instructions

3M

3M SCOTT SEMS II PRO Operation and maintenance instructions

Katronic KATflow 210 operating instructions

Katronic

Katronic KATflow 210 operating instructions

Watts HF Scientific CSM Installation, operation and maintenance manual

Watts

Watts HF Scientific CSM Installation, operation and maintenance manual

3M Attest 1295 manual

3M

3M Attest 1295 manual

Veeder-Root EMR3 XStream-PKG installation guide

Veeder-Root

Veeder-Root EMR3 XStream-PKG installation guide

DeFelsko PosiTector SPG Guide

DeFelsko

DeFelsko PosiTector SPG Guide

KaWe 12.20600.001 user manual

KaWe

KaWe 12.20600.001 user manual

4tech MiniCom Guardian Quick start manual

4tech

4tech MiniCom Guardian Quick start manual

Evoqua V10K instructions

Evoqua

Evoqua V10K instructions

Maxwell Digital Multimeters 25 602B user manual

Maxwell Digital Multimeters

Maxwell Digital Multimeters 25 602B user manual

ATAGO PAL- HIKARi 51 instruction manual

ATAGO

ATAGO PAL- HIKARi 51 instruction manual

Rinstrum M4201 quick start guide

Rinstrum

Rinstrum M4201 quick start guide

Geist Watchdog 100 Quick setup guide

Geist

Geist Watchdog 100 Quick setup guide

Horizon Fitness P-1000 Installation instructions manual

Horizon Fitness

Horizon Fitness P-1000 Installation instructions manual

ATAGO PAL-67S instruction manual

ATAGO

ATAGO PAL-67S instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.