Model ST-171
Temperature & Humidity
USB Data Logger
Instruction
Manual
Shop for REED products online at: www.DataLoggerStore.com 1.877.766.5412
reedinstruments
www com
2
Pantone 485 & 123
485 = (0/100/100/10)
123 = (0/18/100/0)
FEATURES
• Memory for 32,000 readings (16,000 temperature and 16,000 humidity
readings)
• Dew point indication
• Multi-mode to start logging
• Selectable measuring cycle: 2s, 5s, 10s, 30s, 1m, 5m, 10m, 30m, 1hr,
2hr, 3hr, 6hr, 12hr, 24hr
• User-selectable alarm thresholds for both temperature and humidity
• Status indication via red/yellow/green LEDs
• USB interface for set-up and data download
• Analysis software
• Long battery life
DESCRIPTION
1. Protective cover
2. USB connector to PC port
3. Start button
4. RH & temperature sensors
5. Alarm LED (red/yellow)
6. Record LED (green)
7. Mounting clip
Shop for REED products online at: www.DataLoggerStore.com 1.877.766.5412
reedinstruments
www com
3
Pantone 485 & 123
485 = (0/100/100/10)
123 = (0/18/100/0)
LED STATUS GUIDE
LEDs INDICATION ACTION
REC ALM Both LED lights OFF:
Logging not active OR
Low Battery
Start logging, replace
battery and download
the data
REC ALM Single green ash every
10s*: Logging, no alarm
condition** Double green
ash every 10s*: Delayed
start
To start, hold the start
button until Green and
Yellow LEDs ash
REC ALM Single red ash every 10s*:
Logging, low alarm for
RH*** Double red ash
every 10s*: Logging, high
alarm for RH*** Single red
ash every 60s:
Low battery****
Logging will stop
automatically. No data
will be lost. Replace
battery and download
data.
REC ALM Single yellow ash every
10s*: Logging, low alarm
for TEMP*** Double yellow
ash every 10s*: Logging,
high alarm for TEMP***
Single yellow ash every
60s: Logger memory full
Download data
Shop for REED products online at: www.DataLoggerStore.com 1.877.766.5412
reedinstruments
www com
4
Pantone 485 & 123
485 = (0/100/100/10)
123 = (0/18/100/0)
* To save power, the logger’s LED ashing-cycle can be changed to 20s
or 30s via the supplied software.
** To save power, alarm LEDs for temperature and humidity can be dis-
abled via the supplied software.
***When both temperature and relative humidity readings exceed alarm
level synchronously, LED status indication alternate every cycle. For
example: If there is only one alarm, the REC LED blinks for one cycle
and alarm LED will blink for next cycle. If there are two alarms, REC
LED will not blink. First alarm will blink for rst cycle and the next
alarm will blink for next cycle.
**** When the battery is low, all operations will be disabled automatically.
NOTE: Logging automatically stops when the battery weakens (logged
data will be retained). The supplied software is required to restart log-
ging and to download logged data.
***** To use the delay function. Run the datalogger graph software,
click on the computer icon on the menu bar (2nd from left) or select
LOGGER SET from the LINK pull-down menu. The Setup window will
appear, and you will see there are two options: Manual and Instant. If
you select the Manual option, after you click the Setup button, the log-
ger won't start logging immediately until you press the yellow button in
logger's housing.
SPECIFICATIONS
Logging Rate: Selectable sampling interval from 2 seconds up to
24 hours
Operating Temperature: -35 to 80°C or -31 to 176°F
Battery type: 3.6V lithium (1/2AA) (SAFT LS14250, Tadiran TL-5101
or equivalent)
Battery life: 1 year (typically) depending on logging rate, ambient
temperature and use of alarm LEDs
Dimensions/Weight: 101x25x23 mm (4x1x.9”)/172g (6oz)
Shop for REED products online at: www.DataLoggerStore.com 1.877.766.5412
reedinstruments
www com
5
Pantone 485 & 123
485 = (0/100/100/10)
123 = (0/18/100/0)
BATTERY REPLACEMENT
Only use 3.6V lithium batteries. Before replacing the battery, remove the
data logger from the PC. Follow steps 1 through 4 below:
1. Use a pointed object (eg. a small screwdriver) to open the casing.
Lever the casing off in the direction of the arrow.
2. Pull the data logger from the casing.
3. Replace/insert the battery into the battery compartment observing
the right polarity. The two displays briey light up for control purposes
(alternating green/yellow/green).
4. Slide the data logger back into the casing until it snaps into place.
Now the data logger is ready for programming.
Function Range Accuracy
Relative
Humidity
0 to 20% & 80 to 100%
20 to 40% & 60 to 80%
40 to 60%
±5.0%
±3.5%
±3.0%
Temperature -40 to -10°C & +40 to +70°C
10 to +40°C
-40 to +14°F & 104 to 158°F
+14 to +104°F
±2°C
±1°C
±3.6°F
±1.8°F
Dew point
Temperature
-40 to 70°C (-40 to 158°F) ±2.0°C (±4.0°F)
Shop for REED products online at: www.DataLoggerStore.com 1.877.766.5412
reedinstruments
www com
6
Pantone 485 & 123
485 = (0/100/100/10)
123 = (0/18/100/0)
NOTE:
Leaving the model plugged into the USB port for longer than necessary
will cause some of the battery capacity to be lost.
WARNING:
Handle lithium batteries carefully, observe warnings on battery casing.
Dispose of in accordance with local regulations.
SENSOR RECONDITIONING
Over time, the internal sensor may be compromised as a result of pol-
lutants, chemical vapors, and other environmental conditions which can
lead to inaccurate readings. To recondition the internal sensor, please
follow the procedure below:
Bake the Logger at 80°C (176°F) at<5%RH for 36 hours followed by
20-30°C (70-90°F) at>74%RH for 48 hour ( for re-hydration).
If permanent damage to the internal sensor is suspected, replace the
Logger immediately to insure accurate readings.
Shop for REED products online at: www.DataLoggerStore.com 1.877.766.5412
reedinstruments
www com
7
Pantone 485 & 123
485 = (0/100/100/10)
123 = (0/18/100/0)
Notes _________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
________________________________________________
Shop for REED products online at: www.DataLoggerStore.com 1.877.766.5412

Other REED Data Logger manuals

REED SD-2010 User manual

REED

REED SD-2010 User manual

REED R6020 User manual

REED

REED R6020 User manual

REED R6030 User manual

REED

REED R6030 User manual

REED SD-1128 User manual

REED

REED SD-1128 User manual

REED SD-1128 User manual

REED

REED SD-1128 User manual

REED SD-6100 User manual

REED

REED SD-6100 User manual

REED R6020 User manual

REED

REED R6020 User manual

REED C-322 User manual

REED

REED C-322 User manual

REED SD-6020 User manual

REED

REED SD-6020 User manual

REED SD-3007 User manual

REED

REED SD-3007 User manual

REED R6030 User manual

REED

REED R6030 User manual

REED R5003 User manual

REED

REED R5003 User manual

REED ST-172 User manual

REED

REED ST-172 User manual

REED ST-173 User manual

REED

REED ST-173 User manual

Popular Data Logger manuals by other brands

DeltaTRAK FlashLink 40528 quick start guide

DeltaTRAK

DeltaTRAK FlashLink 40528 quick start guide

OVARRO Eureka5 user manual

OVARRO

OVARRO Eureka5 user manual

Magna BLUEPIRAT Rapid Series user manual

Magna

Magna BLUEPIRAT Rapid Series user manual

Fortinet FortiSIEM 500F Configuration guide

Fortinet

Fortinet FortiSIEM 500F Configuration guide

process-informatik EtherSens Bridge Handling-Short Instruction

process-informatik

process-informatik EtherSens Bridge Handling-Short Instruction

Onset HOBO Pendant UA-001 Series manual

Onset

Onset HOBO Pendant UA-001 Series manual

Lascar EL-SIE-2+ manual

Lascar

Lascar EL-SIE-2+ manual

Agrowtek GrowControl INSTALLATION & USER OPERATION MANUAL

Agrowtek

Agrowtek GrowControl INSTALLATION & USER OPERATION MANUAL

Plura RUBIDIUM Series Installation & Systems Manual

Plura

Plura RUBIDIUM Series Installation & Systems Manual

Calex Excelog 6 Operator's guide

Calex

Calex Excelog 6 Operator's guide

Marathon microDL quick start guide

Marathon

Marathon microDL quick start guide

Computech DataMaxx Pro Dash 8 instruction manual

Computech

Computech DataMaxx Pro Dash 8 instruction manual

wireless Seismic RT System 2 Deployment guide

wireless Seismic

wireless Seismic RT System 2 Deployment guide

Inspired Energy ATA004A operating instructions

Inspired Energy

Inspired Energy ATA004A operating instructions

Omega OM-VIB-101 user guide

Omega

Omega OM-VIB-101 user guide

VOLTCRAFT Lux DL-230L operating instructions

VOLTCRAFT

VOLTCRAFT Lux DL-230L operating instructions

Capetti Elettronica WSD12-VW user manual

Capetti Elettronica

Capetti Elettronica WSD12-VW user manual

In-situ Tube 300R owner's manual

In-situ

In-situ Tube 300R owner's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.