FCC INFORMATION:
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following conditions: (1)This
device may not cause harmful interference, and (2) this
device must accept any interference received, including
interference that may cause undesired operation.
MAINTENANCE:
Keep the contacts area clean. Periodically examine the
contacts for gel build-up around the perimeter of the
contacts, which may interfere with or decrease the
delivery of therapy. A moist cloth or cotton swab may
be used to remove any gel build-up. Avoid immersion
of the device in liquid. Reliefband® Premier must not
be opened or altered in any way.
ENVIRONMENTAL USE INFORMATION:
Reliefband® Premier should only be used within
the environmental ranges of 0°C to 38°C (32°F to
100°F) and 20% to 90% relative humidity (non-
condensing).
38
1. Turn the product to ON by pressing and holding
the Power On/O/Intensity Down Button (1) for
1.5 seconds
a. Upon powering on, the Intensity Level
Indicator LEDs (3) will illuminate in
sequence of 1-10, then back down to level 0.
b. The current battery charge is displayed for
1.5 seconds before fading away.
2. Press the Intensity Up Button (2) once to set
stimulation to the rst level. At this point
therapy is being delivered. If intensity level is
not set, Reliefband Premier will turn o.
a. Blue LED intensity lights will indicate
intensity level:
i. No LEDs illuminated = Level 0 =
Power ON but no therapy
ii. 1 LED illuminated = Level 1 Therapy
iii.10 LED’s illuminated = Level 10 Therapy
b. The Therapy On Indicator (5) will pulse
on and o indicating that the device is
delivering therapy.
Size Rectangular shape – max case
dimensions: 14.5mm H x 23.5mm W x
4.1mm L, with Smart Band and Simple
Band.
Weight Approximately 50 grams
User Controls Two Push Buttons: On/O/Intensity Down
and Intensity Up
User Display Blue LED logo ( ) pulsing means device
is providing therapy. Ten diagonal blue
LED bars indicate therapy power level.
Four circular blue LED segments indicate
battery charge level.
Low Battery When battery level registers 15% or less,
a single LED segment will switch from
static blue to ashing red.
Output
Channels
Two hypoallergenic surgical grade (316L)
stainless steel contacts.
Maximum
Output
40mA
USB Charging
Cable
Cable with custom magnetic connector
Battery Rechargeable Li-Ion Polymer - IEC62133
standard. Battery life is approximately
three years or 500 charge cycles.
Waveform
Characteristics
@500 ohm load
40mA, 22.5V, 31Hz, 350us per pulse
The Troubleshooting Chart below will help to assist
you in handling situations that may arise while you
use the product.
Situation Recommended Solution
Blue intensity
LEDs and
pulsing blue
Therapy On
LEDs are
illuminated but
no stimulation is
felt
Move the contacts around the underside
of the wrist to ensure proper placement.
Increase the intensity using the Intensity
Up Button (2). Re-apply a small amount of
Reliefband Conductivity Gel. Ensure gel
is rubbed into a sheen. Be sure the strap
is secure around the wrist. Be sure blue
Intensity LEDs (3) are illuminated at one
of the ten intensity settings.
No LEDs
Illuminated
Briey press the Intensity On/O/Down
button (1) once to check if device is
powered on. If power is on, the blue
Intensity LED bars (3), the Battery/Charging
Indicator segments (4) and the Therapy On
Indicator (5) will ash. If nothing ashes,
the device may be o. Press and hold the
Intensity On/O/Down button for 1.5
seconds to wake up the device. If no LEDs
ash, the battery may be drained. To charge
the battery - see the "Charging The Battery"
section.
Flashing red,
low battery LED
illuminated
This indicates a low battery and 15% (2.4
hours at level 5) or less of charge remains.
Recharge the battery as soon as possible
– see "Charging The Battery" section
Model: Reliefband® Premier
Reliefband® Technologies warrants each new Reliefband®
Premier to be free from material defect in workmanship
and materials for (1 year) from the date of purchase, the
“Warranty Period.”This limited warranty is subject to (a)
the customer notifying Reliefband® Technologies within
1 year of the date of purchase of any material defect
in Reliefband® Premier workmanship and materials
(“Defective Product”) during the Warranty Period,
and (b) the customer returning the Defective Product
to Reliefband® Technologies. Contact Reliefband®
Technologies at 877-735-2263 for return or replacement
of Defective Product.
Reliefband® Premier is indicated for nonprescription
(over-the- counter) use in relief of nausea and vomiting
associated with physician-diagnosed migraine, hangover,
anxiety, motion sickness, chemotherapy and morning
3. Adjust intensity level using the Intensity Up
Button (2) until a "tingling" sensation is felt in
the palm or middle ngers of your hand.
4. If no “tingling” is felt, increase the intensity using
the Intensity Up Button (2).
5. If "tingling" is still not felt, check the position
of the Hypoallergenic Contacts (6) and adjust
until you feel "tingling" in your palm and middle
ngers.
6. Press and hold the Power On/O/Intensity Down
Button (1) for 1.5 seconds to turn the device o.
7. If the battery charge has depleted, recharge
the device according to "Charging The Device"
instructions.
a. Once battery level registers 15% or below,
the single LED will switch from static blue
to ashing red.
b. A ashing red LED indicates the device will
require charging as soon as possible.
ENVIRONMENTAL STORAGE INFORMATION:
The device can be stored and transported in the
environmental ranges of 0°C to 27°C (32°F to
80°F) and 20% to 80% relative humidity (non-
condensing). If the device has been stored or
transported in conditions outside this range, keep
it within the normal ranges for at least 30 minutes
before using.
DISPOSAL INFORMATION
Reliefband Premier is not regulated as hazardous
goods. It can be recycled or disposed with normal
household trash.
TECHNICAL INFORMATION
Reliefband complies with:
• IEC 60601-1:2005+A1:2012/EN 60601-1:2006
Medical electrical equipment - Part 1: General
requirements for basic safety and essential
performance
• IEC 60601-1-2:2007/EN 60601-1-2:2007 Medical
electrical equipment - Part 1-2: General
requirements for safety - Collateral standard:
Electromagnetic compatibility - Requirements
and tests
TIPS FOR EFFECTIVE USE
• Reliefband® Premier can be worn on either wrist provided
tingling is felt in the palm and/or middle ngers.
• For some people, applying and rubbing in the gel twice assists
in nerve stimulation.
• Reliefband® Conductivity Gel should be re-applied every 2-3
hours during use, any time after washing, or if stimulation
decreases.
• Contact with skin for prolonged periods may contribute to skin
irritation particularly for users who have allergies. To reduce
irritation, follow three simple wear and care tips: (1) Keep it
clean; (2) keep it dry; and (3) after extended wear, give your
wrist a rest by removing the band for an hour.
• Regularly clean your Reliefband Premier and wrist—especially
if perspiring, or before reapplying conductivity gel – before
returning the Reliefband Premier to your wrist. Wipe the band
and contacts with a clean, damp wipe using water or a small
amount of rubbing alcohol. Do NOT use hand soap, body
soap, dish soap, hand sanitizers, cleaning wipes or household
cleaners which could get trapped beneath the band and
irritate skin. Always dry the band well before putting it back on.
• To remove build-up of lotions and oils (such as sunscreen,
insect repellent and moisturizers) that can be trapped beneath
the band, we recommend that you use a soap-free cleanser like
Cetaphil Gentle Skin Cleanser or Aquanil; wipe o thoroughly,
and dry well.
• While still on the wrist, turn Reliefband® Premier o periodically
to determine if you still have nausea and vomiting symptoms.
If you still have symptoms, then turn it back on immediately.
sickness from pregnancy. Reliefband is also indicated as an
adjunct to antiemetics in reducing postoperative nausea.
Reliefband® Premier may not be eective for every person
because treatment outcomes vary and are dependent
upon a myriad of patient proles and characteristics.
Reliefband® Technologies’ obligation under this Limited
Warranty is expressly limited solely and exclusively to the
replacement of the Defective Product.
THE LIMITED REPRESENTATIONS AND WARRANTIES SET
FORTH HEREIN ARE RELIEFBAND TECHNOLOGIES’SOLE
REPRESENTATIONS AND WARRANTIES WITH RESPECT
TO THE PREMIER OR ANY OTHER SUBJECT MATTER
HEREOF AND ARE GIVEN IN LIEU OF ANY AND ALL
OTHER REPRESENTATIONS OR WARRANTIES, WHETHER
EXPRESS OR IMPLIED, INCLUDING, WITHOUT LIMITATION,
ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR
SUITABILITY FOR A PARTICULAR PURPOSE. THE REMEDIES
AFFORDED TO CUSTOMER FOR ANY BREACH OF
WARRANTY WILL BE LIMITED TO THOSE SET FORTH HEREIN
TO THE EXCLUSION OF ANY AND ALL OTHER REMEDIES.
CUSTOMER WILL NOT BE ENTITLED TO PUNITIVE,
INCIDENTAL, CONSEQUENTIAL OR SPECIAL DAMAGES,
INCLUDING WITHOUT LIMITATION CUSTOMER’S DENTAL
AND MEDICAL COSTS, AS A RESULT OF A DEFECTIVE
PRODUCT. NO AGREEMENT VARYING OR EXTENDING
THE FOREGOING WARRANTIES, REMEDIES OR THIS
LIMITATION WILL BE BINDING ON EITHER PARTY UNLESS
MEMORIALIZED IN A WRITING SIGNED BY BOTH PARTIES.
Some jurisdictions do not allow the exclusion or limitation
of special, indirect, incidental or consequential damages,
so the above limitation or exclusion may not apply to you.
INSTRUCTIONS FOR USE
CLINICALLY PROVEN RELIEF
FOR THE PREVENTION AND TREATMENT OF
NAUSEA AND VOMITING
PREMIER
RB2.0
DETAILED OPERATION TECHNICAL DATA
TROUBLESHOOTING
STORAGE AND USAGE
INFORMATION
LIMITED WARRANTY
STATEMENT
(2)
INTENSITY UP
BUTTON
(1)
POWER ON/OFF/INTENSITY
DOWN BUTTON
(3)
INTENSITY LEVEL
INDICATOR
(5)
THERAPY ON
INDICATOR
(6)
CONTACTS
YOU SHOULD FEEL A
"TINGLING" SENSATION IN THE
PALM OR MIDDLE FINGERS.
SINGLE FLASHING
RED LED
- LOW BATTERY
Reliefband® Premier produces
physiological eects as described in this
guide.
Reliefband® Premier is classied as a
TYPE BF Applied Part
Latex Free
Reliefband® Premier is splash-resistant
ATTENTION
IPX4
For European Union: Environmental Protection
Waste electrical products should not be
disposed of with household waste. Please
recycle where facilities exist. Check with
your Local Authority or retailer for recycling
advice.
California Only: Percholorate Material -
Special handling may apply. See
www.dtsc.ca.gov/hazardouswaste/perchlorate
Read these Instructions before using
Reliefband® Premier
This warranty shall not apply to any Defective Product
that has been tampered with or repaired and/or altered
by someone other than a duly-authorized Reliefband
Technologies’ representative. Further, this warranty shall
not apply to any Defective Product where the defect has
been caused by customer negligence, mishandling or
any customer use contrary to the enclosed instructions or
customer use inconsistent with the stated purpose.
This warranty is expressly limited solely to the original
purchaser and does not extend or confer any rights to
any transferee, assignee, or subsequent purchaser or user
of a Defective Product. Reliefband® Technologies’liability
for all claims, (whether based on contract, tort, breach of
warranty, or otherwise), which may arise in connection
with the purchase and use of any Defective Product
is limited to the purchase price paid by the original
purchaser.
P/N 6106980AW Rev B
• IEC 60601-2-10:2012/EN 60601-2-
10:2000+A1:2001 Medical electrical equipment
- Part 2-10: Particular requirements for the safety
of nerve and muscle stimulators
• When nished using Reliefband® Premier take it o your wrist.
Remove the gel from your wrist and the device with a tissue.
• When not using Reliefband® Premier be sure to turn the power
o to extend battery life.
• Store Reliefband® Premier in a safe, dry place. Keep it away
from young children.
• Avoid an excessive amount of gel; it could decrease the ecacy
of the device.
.com

(3.)THERAPY INTENSITY
LEVEL INDICATOR
(6.) HYPOALLERGENIC
CONTACTS
(5.)THERAPY ON INDICATOR
(1.) POWER ON/OFF/INTENSITY
DOWN BUTTON
(4.) BATTERY/CHARGING
INDICATOR
(2.) INTENSITY UP
BUTTON
(8.) LATCH(9.) CINCH BAR
(12.) CONTACT
ASSEMBLY
(10.) SMART BAND
(7.) CHARGING PORT
(13.) MAGNETIC CONNECTOR
(14.) USB CHARGING CABLE
(11.) SIMPLE BAND
2
3
1
P6
2APPLY
Clean the area rst.
Apply a small drop of gel and spread in a circle
about the size of a large coin with an even sheen
(i.e., a thin layer with a shiny appearance).
1AREA
Find the starting area (P6) on the wrist.
Using either wrist, the correct area is in-between the
two tendons on the underside of the wrist – about
two nger widths above the wrist crease.
QUICK REFERENCE INTRODUCTION SAFETY INFORMATION PACKAGE CONTENTS CHARGING THE DEVICE
HOW RELIEFBAND® PREMIER
WORKS
GET TO KNOW
RELIEFBAND® PREMIER
INDICATIONS FOR USE
Reliefband® Premier is indicated for use in the
treatment of nausea and vomiting associated with
physician-diagnosed migraine, hangover, anxiety,
motion sickness, chemotherapy and morning
sickness from pregnancy.
Reliefband® Premier is also indicated as an adjunct
to antiemetics in reducing postoperative nausea. For
over the counter use.
220 Gibraltar Road, Suite 270, Horsham, PA 19044
Reliefband® Premier is a registered trademark of
Reliefband Technologies LLC.
Reliefband® technology is protected by US Patent
No. 6,076,018; 7,127,288. Japan Patent No. 4,812,810;
4,220,129. Canadian Patent No. 2,449,554; 2,342,873.
Australian Patent Nos. 766463, 2004200090 and
2002305786, and European Patent No. 1,109,595-UK,
France and Germany (DE 69940405.3).
Designed in USA. Made in Malaysia.
©2020 Reliefband Technologies
In the U.S.A., call 1-877-735-2263
9:00 AM - 5:00 PM Mon. - Fri. CST
P/N 6106980AW Rev B
DESCRIPTION OF THE PREMIER:
Reliefband® Premier is a rechargeable, digital
wearable device that is worn on the wrist. It consists
of a hypoallergenic band with a control panel
on top and two electrical contacts underneath.
The two contacts must be placed rmly against
the underside of your wrist at the P6 location. It
provides relief from nausea and vomiting associated
with physician-diagnosed migraine, hangover,
anxiety, motion sickness, chemotherapy and
morning sickness from pregnancy.
Reliefband® Premier is also indicated as an adjunct
to antiemetics in reducing postoperative nausea.
A fully charged battery will last for approximately
18 hours of continuous use at the mid-power level
or below.
1
Reliefband® Premier delivers
proprietary pulses from the
contacts on the Smart Band that
signal the median nerve at the P6
location on the underside of the
wrist.
2
These signals travel through
the body’s nervous system
to the part of the brain
which controls nausea and
vomiting.
3
The signals have a
rebalancing eect,
normalizing nerve
messages from the brain
to the stomach and
reducing symptoms of
nausea and vomiting.
Nausea and vomiting are probably some of the
worst sensations you can feel – the combination of
dizziness, lightheadedness, heightened sensitivity to
smells, bloated abdominal discomfort, retching and
the ultimate involuntary evacuation of the stomach.
Drug-free, Reliefband® Premier uses clinically-proven
technology to quickly and eectively relieve nausea
and vomiting. Reliefband® Premier is the result of
many years of clinical research and development and
millions of dollars of investment. It is the rst of a new
class of Digital Wearable Technologies.
– it’s a Wearable that does something other than
measure things
- a “Digital Wearable Treatable”
Reliefband® Premier uses patented technology that
leverages the body’s natural neural pathways to block
signals to the center in the brain responsible for the
sensation of nausea. It has been used to treat nausea
and vomiting in clinical settings and hospitals for
many years. Now it is available over-the-counter for
the treatment of nausea and vomiting associated with
physician-diagnosed migraine, hangover, anxiety,
motion sickness, chemotherapy and morning sickness
from pregnancy. Reliefband is also an adjunct to
antiemetic therapy in reducing postoperative nausea.
CAUTION
• Pacemaker users - Use this device on the
wrist and only as directed to prevent possible
interference with your pacemaker. Avoid placing
the electrical contacts directly on your chest or
near the pacemaker. Consult your doctor before
using device.
• Nausea and vomiting may be signs of a serious
health problem. Contact a doctor if your nausea
and vomiting persists.
• Reliefband® Premier should only be used on the
wrist. Do not use it on any other area of the body.
• Reliefband® Premier does not cure the underlying
causes of nausea and vomiting.
(1) Reliefband® Premier
(1) Reliefband® Conductivity Gel
(1) Instructions For Use
(1) Magnetic USB Charging Cable
INSTRUCTIONS FOR USE
Model:Reliefband® 2.0
DRUG-FREE RELIEF
FROMNAUSEA, RETCHING AND VOMITING
2.0
• The relief you get may vary depending on
your body’s individual characteristics, and any
medications that you may be taking could cause
your body to respond dierently.
• Reliefband® Premier should be kept out of reach
of young children.
• If you are unable to tolerate any food or liquids or
are experiencing weight loss you should contact
your doctor.
• Reliefband® Premier is IPX4 splash-resistant.
This product is resistant to splashing water (rain,
waves).
• Avoid splashing water while charging.
• Do not immerse Reliefband® Premier in water or
wear while bathing, showering or swimming.
SIDE EFFECTS
• There are no known side eects for Reliefband®
Premier.
• There are no contraindications for Reliefband®
Premier.
IMPORTANT
Positioning Reliefband® Premier properly on
your wrist is essential for relief of your nausea
and vomiting symptoms. Before you start using
Reliefband® Premier, please read this Instructions
For Use carefully. You must feel a“tingling”sensation
in your palm and/or middle ngers for Reliefband®
Premier to work.
Reliefband Premier must be charged before use.
APPLYING RELIEFBAND® CONDUCTIVITY GEL
Reliefband® Conductivity Gel is an essential
component for use and therapy. Hypoallergenic
Reliefband® Conductivity Gel makes the delivery of
pulses much more ecient.
Note: Reliefband® Conductivity Gel should be
re-applied every 2-3 hours, or any time after washing,
or if stimulation decreases.
1. Power On/O/Intensity Down Button – press and
hold for 1.5 seconds to power on the device. During
active therapy, briey pressing this button will
decrease the intensity. Press and hold for 1.5 seconds
to turn the device o.
2. Intensity Up Button – Once the device has been
turned on, press this button to increase the intensity.
3. Therapy Intensity Level Indicator – 10 levels indicated
by blue illuminated LEDs.
4. Battery/Charging Indicator – four blue illuminated LED
segments indicate charge level. A single ashing red
segment indicates 15% or less battery life remaining.
5. Therapy On Indicator – pulsing blue LEDs indicate the
device is delivering therapy.
6. Hypoallergenic Contacts – 316L Stainless Steel,
therapy is delivered here.
7. Charging Port – attach the magnetic charging cable
here to recharge the device.
8. Latch – locks the Simple Band (11) tight to hold the
device in position.
9. Cinch Bar – gently pull the Simple Band (11) against
the Cinch Bar to achieve a snug t on the wrist.
10. Smart Band – unique shape holds the device in place
and makes it easy to put on and take o.
11. Simple Band – slide through the latch to adjust the t
to your wrist.
12. Contact Assembly – houses the contacts that deliver
therapy.
13. Magnetic Connector – attaches charging cable to the
device.
14. USB Charging Cable
15. UL Approved (5V) USB Charging Port
1. Ensure Reliefband Premier is turned o prior to
charging. Do not wear while charging.
2. Plug the USB Charging Cable (14) into any standard
UL approved 5 Volt USB charging port (15).
3. Attach the Magnetic Connector (13) end of the USB
Charging Cable (14) to the Charging Port (7) on the
body of the Reliefband® Premier device.
4. Blue LEDs segment(s) will ash when in charging mode:
a. 1 LED segment = 1% - 25% charged
b. 2 LED segments = 26% - 50% charged
c. 3 LED segments = 51% - 75% charged
d. 4 LED segments = 76% - 100% charged
5. Full charge is indicated when all four blue LED
segments remain constantly illuminated. Typically,
it takes about 75 minutes to reach full charge.
A 15-minute charge will give about 6 hours of
continuous use at the mid-power level or below
6. When not connected to the Reliefband, unplug
the USB Charging Cable (14) from the standard UL
approved 5 Volt USB charging port (15).
7. When turning the device on, the Battery Charge
Indicator (4) will indicate how much charge is left in
the device.
8. A single ashing red LED segment indicates 15% or
less battery life remaining.
INTENSITY
UP BUTTON
POWER
ON/OFF/INTENSITY
DOWN BUTTON
3ADJUST
Holding the device by the Contact Assembly, place
the contacts onto the gel over the P6 location on the
underside of the wrist. Fasten device snugly.
4ACTIVATE**
Press and hold the Power ON/OFF/Intensity Down
Button (-) for 1.5 seconds to turn on or o.
Press the intensity Up Button (+) once to set
stimulation to level 1. Increase intensity until tingling
is felt in palm and middle ngers.
Set to the highest comfortable setting for maximum
benet.
**Please charge for at least 1.5 hours before initial use.
(13) MAGNETIC CONNECTOR
(14)
USB CHARGING CABLE
(15)
UL APPROVED (5V) USB CHARGING PORT
(4) BATTERY/CHARGING INDICATOR
(7) CHARGING PORT
(a.) 1 LED (b.) 2 LEDs
SINGLE FLASHING RED LED
- LOW BATTERY
(c.) 3 LEDs (d.) 4 LEDs - FULLY CHARGED
VOLTAGE RATING: 5±0.25V
CURRENT RATING: 500mA MAX
WARNING
• No modication of this device is allowed.
• Charging cord must be unplugged from power
source when not in use.
• Charging Cord:
Voltage Rating: 5±0.25V
Current Rating: 500mA Max
EMERGO EUROPE
Prinsessegracht
20 2514 AP The Hague
The Netherlands

(3.)THERAPY INTENSITY
LEVEL INDICATOR
(6.) HYPOALLERGENIC
CONTACTS
(5.)THERAPY ON INDICATOR
(1.) POWER ON/OFF/INTENSITY
DOWN BUTTON
(4.) BATTERY/CHARGING
INDICATOR
(2.) INTENSITY UP
BUTTON
(8.) LATCH(9.) CINCH BAR
(12.) CONTACT
ASSEMBLY
(10.) SMART BAND
(7.) CHARGING PORT
(13.) MAGNETIC CONNECTOR
(14.) USB CHARGING CABLE
(11.) SIMPLE BAND
2
3
1
P6
2APPLY
Clean the area rst.
Apply a small drop of gel and spread in a circle
about the size of a large coin with an even sheen
(i.e., a thin layer with a shiny appearance).
1AREA
Find the starting area (P6) on the wrist.
Using either wrist, the correct area is in-between the
two tendons on the underside of the wrist – about
two nger widths above the wrist crease.
QUICK REFERENCE INTRODUCTION SAFETY INFORMATION PACKAGE CONTENTS CHARGING THE DEVICE
HOW RELIEFBAND® PREMIER
WORKS
GET TO KNOW
RELIEFBAND® PREMIER
INDICATIONS FOR USE
Reliefband® Premier is indicated for use in the
treatment of nausea and vomiting associated with
physician-diagnosed migraine, hangover, anxiety,
motion sickness, chemotherapy and morning
sickness from pregnancy.
Reliefband® Premier is also indicated as an adjunct
to antiemetics in reducing postoperative nausea. For
over the counter use.
220 Gibraltar Road, Suite 270, Horsham, PA 19044
Reliefband® Premier is a registered trademark of
Reliefband Technologies LLC.
Reliefband® technology is protected by US Patent
No. 6,076,018; 7,127,288. Japan Patent No. 4,812,810;
4,220,129. Canadian Patent No. 2,449,554; 2,342,873.
Australian Patent Nos. 766463, 2004200090 and
2002305786, and European Patent No. 1,109,595-UK,
France and Germany (DE 69940405.3).
Designed in USA. Made in Malaysia.
©2020 Reliefband Technologies
In the U.S.A., call 1-877-735-2263
9:00 AM - 5:00 PM Mon. - Fri. CST
P/N 6106980AW Rev B
DESCRIPTION OF THE PREMIER:
Reliefband® Premier is a rechargeable, digital
wearable device that is worn on the wrist. It consists
of a hypoallergenic band with a control panel
on top and two electrical contacts underneath.
The two contacts must be placed rmly against
the underside of your wrist at the P6 location. It
provides relief from nausea and vomiting associated
with physician-diagnosed migraine, hangover,
anxiety, motion sickness, chemotherapy and
morning sickness from pregnancy.
Reliefband® Premier is also indicated as an adjunct
to antiemetics in reducing postoperative nausea.
A fully charged battery will last for approximately
18 hours of continuous use at the mid-power level
or below.
1
Reliefband® Premier delivers
proprietary pulses from the
contacts on the Smart Band that
signal the median nerve at the P6
location on the underside of the
wrist.
2
These signals travel through
the body’s nervous system
to the part of the brain
which controls nausea and
vomiting.
3
The signals have a
rebalancing eect,
normalizing nerve
messages from the brain
to the stomach and
reducing symptoms of
nausea and vomiting.
Nausea and vomiting are probably some of the
worst sensations you can feel – the combination of
dizziness, lightheadedness, heightened sensitivity to
smells, bloated abdominal discomfort, retching and
the ultimate involuntary evacuation of the stomach.
Drug-free, Reliefband® Premier uses clinically-proven
technology to quickly and eectively relieve nausea
and vomiting. Reliefband® Premier is the result of
many years of clinical research and development and
millions of dollars of investment. It is the rst of a new
class of Digital Wearable Technologies.
– it’s a Wearable that does something other than
measure things
- a “Digital Wearable Treatable”
Reliefband® Premier uses patented technology that
leverages the body’s natural neural pathways to block
signals to the center in the brain responsible for the
sensation of nausea. It has been used to treat nausea
and vomiting in clinical settings and hospitals for
many years. Now it is available over-the-counter for
the treatment of nausea and vomiting associated with
physician-diagnosed migraine, hangover, anxiety,
motion sickness, chemotherapy and morning sickness
from pregnancy. Reliefband is also an adjunct to
antiemetic therapy in reducing postoperative nausea.
CAUTION
• Pacemaker users - Use this device on the
wrist and only as directed to prevent possible
interference with your pacemaker. Avoid placing
the electrical contacts directly on your chest or
near the pacemaker. Consult your doctor before
using device.
• Nausea and vomiting may be signs of a serious
health problem. Contact a doctor if your nausea
and vomiting persists.
• Reliefband® Premier should only be used on the
wrist. Do not use it on any other area of the body.
• Reliefband® Premier does not cure the underlying
causes of nausea and vomiting.
(1) Reliefband® Premier
(1) Reliefband® Conductivity Gel
(1) Instructions For Use
(1) Magnetic USB Charging Cable
INSTRUCTIONS FOR USE
Model:Reliefband® 2.0
DRUG-FREE RELIEF
FROMNAUSEA, RETCHING AND VOMITING
2.0
• The relief you get may vary depending on
your body’s individual characteristics, and any
medications that you may be taking could cause
your body to respond dierently.
• Reliefband® Premier should be kept out of reach
of young children.
• If you are unable to tolerate any food or liquids or
are experiencing weight loss you should contact
your doctor.
• Reliefband® Premier is IPX4 splash-resistant.
This product is resistant to splashing water (rain,
waves).
• Avoid splashing water while charging.
• Do not immerse Reliefband® Premier in water or
wear while bathing, showering or swimming.
SIDE EFFECTS
• There are no known side eects for Reliefband®
Premier.
• There are no contraindications for Reliefband®
Premier.
IMPORTANT
Positioning Reliefband® Premier properly on
your wrist is essential for relief of your nausea
and vomiting symptoms. Before you start using
Reliefband® Premier, please read this Instructions
For Use carefully. You must feel a“tingling”sensation
in your palm and/or middle ngers for Reliefband®
Premier to work.
Reliefband Premier must be charged before use.
APPLYING RELIEFBAND® CONDUCTIVITY GEL
Reliefband® Conductivity Gel is an essential
component for use and therapy. Hypoallergenic
Reliefband® Conductivity Gel makes the delivery of
pulses much more ecient.
Note: Reliefband® Conductivity Gel should be
re-applied every 2-3 hours, or any time after washing,
or if stimulation decreases.
1. Power On/O/Intensity Down Button – press and
hold for 1.5 seconds to power on the device. During
active therapy, briey pressing this button will
decrease the intensity. Press and hold for 1.5 seconds
to turn the device o.
2. Intensity Up Button – Once the device has been
turned on, press this button to increase the intensity.
3. Therapy Intensity Level Indicator – 10 levels indicated
by blue illuminated LEDs.
4. Battery/Charging Indicator – four blue illuminated LED
segments indicate charge level. A single ashing red
segment indicates 15% or less battery life remaining.
5. Therapy On Indicator – pulsing blue LEDs indicate the
device is delivering therapy.
6. Hypoallergenic Contacts – 316L Stainless Steel,
therapy is delivered here.
7. Charging Port – attach the magnetic charging cable
here to recharge the device.
8. Latch – locks the Simple Band (11) tight to hold the
device in position.
9. Cinch Bar – gently pull the Simple Band (11) against
the Cinch Bar to achieve a snug t on the wrist.
10. Smart Band – unique shape holds the device in place
and makes it easy to put on and take o.
11. Simple Band – slide through the latch to adjust the t
to your wrist.
12. Contact Assembly – houses the contacts that deliver
therapy.
13. Magnetic Connector – attaches charging cable to the
device.
14. USB Charging Cable
15. UL Approved (5V) USB Charging Port
1. Ensure Reliefband Premier is turned o prior to
charging. Do not wear while charging.
2. Plug the USB Charging Cable (14) into any standard
UL approved 5 Volt USB charging port (15).
3. Attach the Magnetic Connector (13) end of the USB
Charging Cable (14) to the Charging Port (7) on the
body of the Reliefband® Premier device.
4. Blue LEDs segment(s) will ash when in charging mode:
a. 1 LED segment = 1% - 25% charged
b. 2 LED segments = 26% - 50% charged
c. 3 LED segments = 51% - 75% charged
d. 4 LED segments = 76% - 100% charged
5. Full charge is indicated when all four blue LED
segments remain constantly illuminated. Typically,
it takes about 75 minutes to reach full charge.
A 15-minute charge will give about 6 hours of
continuous use at the mid-power level or below
6. When not connected to the Reliefband, unplug
the USB Charging Cable (14) from the standard UL
approved 5 Volt USB charging port (15).
7. When turning the device on, the Battery Charge
Indicator (4) will indicate how much charge is left in
the device.
8. A single ashing red LED segment indicates 15% or
less battery life remaining.
INTENSITY
UP BUTTON
POWER
ON/OFF/INTENSITY
DOWN BUTTON
3ADJUST
Holding the device by the Contact Assembly, place
the contacts onto the gel over the P6 location on the
underside of the wrist. Fasten device snugly.
4ACTIVATE**
Press and hold the Power ON/OFF/Intensity Down
Button (-) for 1.5 seconds to turn on or o.
Press the intensity Up Button (+) once to set
stimulation to level 1. Increase intensity until tingling
is felt in palm and middle ngers.
Set to the highest comfortable setting for maximum
benet.
**Please charge for at least 1.5 hours before initial use.
(13) MAGNETIC CONNECTOR
(14)
USB CHARGING CABLE
(15)
UL APPROVED (5V) USB CHARGING PORT
(4) BATTERY/CHARGING INDICATOR
(7) CHARGING PORT
(a.) 1 LED (b.) 2 LEDs
SINGLE FLASHING RED LED
- LOW BATTERY
(c.) 3 LEDs (d.) 4 LEDs - FULLY CHARGED
VOLTAGE RATING: 5±0.25V
CURRENT RATING: 500mA MAX
WARNING
• No modication of this device is allowed.
• Charging cord must be unplugged from power
source when not in use.
• Charging Cord:
Voltage Rating: 5±0.25V
Current Rating: 500mA Max
EMERGO EUROPE
Prinsessegracht
20 2514 AP The Hague
The Netherlands

(3.)THERAPY INTENSITY
LEVEL INDICATOR
(6.) HYPOALLERGENIC
CONTACTS
(5.)THERAPY ON INDICATOR
(1.) POWER ON/OFF/INTENSITY
DOWN BUTTON
(4.) BATTERY/CHARGING
INDICATOR
(2.) INTENSITY UP
BUTTON
(8.) LATCH(9.) CINCH BAR
(12.) CONTACT
ASSEMBLY
(10.) SMART BAND
(7.) CHARGING PORT
(13.) MAGNETIC CONNECTOR
(14.) USB CHARGING CABLE
(11.) SIMPLE BAND
2
3
1
P6
2APPLY
Clean the area rst.
Apply a small drop of gel and spread in a circle
about the size of a large coin with an even sheen
(i.e., a thin layer with a shiny appearance).
1AREA
Find the starting area (P6) on the wrist.
Using either wrist, the correct area is in-between the
two tendons on the underside of the wrist – about
two nger widths above the wrist crease.
QUICK REFERENCE INTRODUCTION SAFETY INFORMATION PACKAGE CONTENTS CHARGING THE DEVICE
HOW RELIEFBAND® PREMIER
WORKS
GET TO KNOW
RELIEFBAND® PREMIER
INDICATIONS FOR USE
Reliefband® Premier is indicated for use in the
treatment of nausea and vomiting associated with
physician-diagnosed migraine, hangover, anxiety,
motion sickness, chemotherapy and morning
sickness from pregnancy.
Reliefband® Premier is also indicated as an adjunct
to antiemetics in reducing postoperative nausea. For
over the counter use.
220 Gibraltar Road, Suite 270, Horsham, PA 19044
Reliefband® Premier is a registered trademark of
Reliefband Technologies LLC.
Reliefband® technology is protected by US Patent
No. 6,076,018; 7,127,288. Japan Patent No. 4,812,810;
4,220,129. Canadian Patent No. 2,449,554; 2,342,873.
Australian Patent Nos. 766463, 2004200090 and
2002305786, and European Patent No. 1,109,595-UK,
France and Germany (DE 69940405.3).
Designed in USA. Made in Malaysia.
©2020 Reliefband Technologies
In the U.S.A., call 1-877-735-2263
9:00 AM - 5:00 PM Mon. - Fri. CST
P/N 6106980AW Rev B
DESCRIPTION OF THE PREMIER:
Reliefband® Premier is a rechargeable, digital
wearable device that is worn on the wrist. It consists
of a hypoallergenic band with a control panel
on top and two electrical contacts underneath.
The two contacts must be placed rmly against
the underside of your wrist at the P6 location. It
provides relief from nausea and vomiting associated
with physician-diagnosed migraine, hangover,
anxiety, motion sickness, chemotherapy and
morning sickness from pregnancy.
Reliefband® Premier is also indicated as an adjunct
to antiemetics in reducing postoperative nausea.
A fully charged battery will last for approximately
18 hours of continuous use at the mid-power level
or below.
1
Reliefband® Premier delivers
proprietary pulses from the
contacts on the Smart Band that
signal the median nerve at the P6
location on the underside of the
wrist.
2
These signals travel through
the body’s nervous system
to the part of the brain
which controls nausea and
vomiting.
3
The signals have a
rebalancing eect,
normalizing nerve
messages from the brain
to the stomach and
reducing symptoms of
nausea and vomiting.
Nausea and vomiting are probably some of the
worst sensations you can feel – the combination of
dizziness, lightheadedness, heightened sensitivity to
smells, bloated abdominal discomfort, retching and
the ultimate involuntary evacuation of the stomach.
Drug-free, Reliefband® Premier uses clinically-proven
technology to quickly and eectively relieve nausea
and vomiting. Reliefband® Premier is the result of
many years of clinical research and development and
millions of dollars of investment. It is the rst of a new
class of Digital Wearable Technologies.
– it’s a Wearable that does something other than
measure things
- a “Digital Wearable Treatable”
Reliefband® Premier uses patented technology that
leverages the body’s natural neural pathways to block
signals to the center in the brain responsible for the
sensation of nausea. It has been used to treat nausea
and vomiting in clinical settings and hospitals for
many years. Now it is available over-the-counter for
the treatment of nausea and vomiting associated with
physician-diagnosed migraine, hangover, anxiety,
motion sickness, chemotherapy and morning sickness
from pregnancy. Reliefband is also an adjunct to
antiemetic therapy in reducing postoperative nausea.
CAUTION
• Pacemaker users - Use this device on the
wrist and only as directed to prevent possible
interference with your pacemaker. Avoid placing
the electrical contacts directly on your chest or
near the pacemaker. Consult your doctor before
using device.
• Nausea and vomiting may be signs of a serious
health problem. Contact a doctor if your nausea
and vomiting persists.
• Reliefband® Premier should only be used on the
wrist. Do not use it on any other area of the body.
• Reliefband® Premier does not cure the underlying
causes of nausea and vomiting.
(1) Reliefband® Premier
(1) Reliefband® Conductivity Gel
(1) Instructions For Use
(1) Magnetic USB Charging Cable
INSTRUCTIONS FOR USE
Model:Reliefband® 2.0
DRUG-FREE RELIEF
FROMNAUSEA, RETCHING AND VOMITING
2.0
• The relief you get may vary depending on
your body’s individual characteristics, and any
medications that you may be taking could cause
your body to respond dierently.
• Reliefband® Premier should be kept out of reach
of young children.
• If you are unable to tolerate any food or liquids or
are experiencing weight loss you should contact
your doctor.
• Reliefband® Premier is IPX4 splash-resistant.
This product is resistant to splashing water (rain,
waves).
• Avoid splashing water while charging.
• Do not immerse Reliefband® Premier in water or
wear while bathing, showering or swimming.
SIDE EFFECTS
• There are no known side eects for Reliefband®
Premier.
• There are no contraindications for Reliefband®
Premier.
IMPORTANT
Positioning Reliefband® Premier properly on
your wrist is essential for relief of your nausea
and vomiting symptoms. Before you start using
Reliefband® Premier, please read this Instructions
For Use carefully. You must feel a“tingling”sensation
in your palm and/or middle ngers for Reliefband®
Premier to work.
Reliefband Premier must be charged before use.
APPLYING RELIEFBAND® CONDUCTIVITY GEL
Reliefband® Conductivity Gel is an essential
component for use and therapy. Hypoallergenic
Reliefband® Conductivity Gel makes the delivery of
pulses much more ecient.
Note: Reliefband® Conductivity Gel should be
re-applied every 2-3 hours, or any time after washing,
or if stimulation decreases.
1. Power On/O/Intensity Down Button – press and
hold for 1.5 seconds to power on the device. During
active therapy, briey pressing this button will
decrease the intensity. Press and hold for 1.5 seconds
to turn the device o.
2. Intensity Up Button – Once the device has been
turned on, press this button to increase the intensity.
3. Therapy Intensity Level Indicator – 10 levels indicated
by blue illuminated LEDs.
4. Battery/Charging Indicator – four blue illuminated LED
segments indicate charge level. A single ashing red
segment indicates 15% or less battery life remaining.
5. Therapy On Indicator – pulsing blue LEDs indicate the
device is delivering therapy.
6. Hypoallergenic Contacts – 316L Stainless Steel,
therapy is delivered here.
7. Charging Port – attach the magnetic charging cable
here to recharge the device.
8. Latch – locks the Simple Band (11) tight to hold the
device in position.
9. Cinch Bar – gently pull the Simple Band (11) against
the Cinch Bar to achieve a snug t on the wrist.
10. Smart Band – unique shape holds the device in place
and makes it easy to put on and take o.
11. Simple Band – slide through the latch to adjust the t
to your wrist.
12. Contact Assembly – houses the contacts that deliver
therapy.
13. Magnetic Connector – attaches charging cable to the
device.
14. USB Charging Cable
15. UL Approved (5V) USB Charging Port
1. Ensure Reliefband Premier is turned o prior to
charging. Do not wear while charging.
2. Plug the USB Charging Cable (14) into any standard
UL approved 5 Volt USB charging port (15).
3. Attach the Magnetic Connector (13) end of the USB
Charging Cable (14) to the Charging Port (7) on the
body of the Reliefband® Premier device.
4. Blue LEDs segment(s) will ash when in charging mode:
a. 1 LED segment = 1% - 25% charged
b. 2 LED segments = 26% - 50% charged
c. 3 LED segments = 51% - 75% charged
d. 4 LED segments = 76% - 100% charged
5. Full charge is indicated when all four blue LED
segments remain constantly illuminated. Typically,
it takes about 75 minutes to reach full charge.
A 15-minute charge will give about 6 hours of
continuous use at the mid-power level or below
6. When not connected to the Reliefband, unplug
the USB Charging Cable (14) from the standard UL
approved 5 Volt USB charging port (15).
7. When turning the device on, the Battery Charge
Indicator (4) will indicate how much charge is left in
the device.
8. A single ashing red LED segment indicates 15% or
less battery life remaining.
INTENSITY
UP BUTTON
POWER
ON/OFF/INTENSITY
DOWN BUTTON
3ADJUST
Holding the device by the Contact Assembly, place
the contacts onto the gel over the P6 location on the
underside of the wrist. Fasten device snugly.
4ACTIVATE**
Press and hold the Power ON/OFF/Intensity Down
Button (-) for 1.5 seconds to turn on or o.
Press the intensity Up Button (+) once to set
stimulation to level 1. Increase intensity until tingling
is felt in palm and middle ngers.
Set to the highest comfortable setting for maximum
benet.
**Please charge for at least 1.5 hours before initial use.
(13) MAGNETIC CONNECTOR
(14)
USB CHARGING CABLE
(15)
UL APPROVED (5V) USB CHARGING PORT
(4) BATTERY/CHARGING INDICATOR
(7) CHARGING PORT
(a.) 1 LED (b.) 2 LEDs
SINGLE FLASHING RED LED
- LOW BATTERY
(c.) 3 LEDs (d.) 4 LEDs - FULLY CHARGED
VOLTAGE RATING: 5±0.25V
CURRENT RATING: 500mA MAX
WARNING
• No modication of this device is allowed.
• Charging cord must be unplugged from power
source when not in use.
• Charging Cord:
Voltage Rating: 5±0.25V
Current Rating: 500mA Max
EMERGO EUROPE
Prinsessegracht
20 2514 AP The Hague
The Netherlands

(3.)THERAPY INTENSITY
LEVEL INDICATOR
(6.) HYPOALLERGENIC
CONTACTS
(5.)THERAPY ON INDICATOR
(1.) POWER ON/OFF/INTENSITY
DOWN BUTTON
(4.) BATTERY/CHARGING
INDICATOR
(2.) INTENSITY UP
BUTTON
(8.) LATCH(9.) CINCH BAR
(12.) CONTACT
ASSEMBLY
(10.) SMART BAND
(7.) CHARGING PORT
(13.) MAGNETIC CONNECTOR
(14.) USB CHARGING CABLE
(11.) SIMPLE BAND
2
3
1
P6
2APPLY
Clean the area rst.
Apply a small drop of gel and spread in a circle
about the size of a large coin with an even sheen
(i.e., a thin layer with a shiny appearance).
1AREA
Find the starting area (P6) on the wrist.
Using either wrist, the correct area is in-between the
two tendons on the underside of the wrist – about
two nger widths above the wrist crease.
QUICK REFERENCE INTRODUCTION SAFETY INFORMATION PACKAGE CONTENTS CHARGING THE DEVICE
HOW RELIEFBAND® PREMIER
WORKS
GET TO KNOW
RELIEFBAND® PREMIER
INDICATIONS FOR USE
Reliefband® Premier is indicated for use in the
treatment of nausea and vomiting associated with
physician-diagnosed migraine, hangover, anxiety,
motion sickness, chemotherapy and morning
sickness from pregnancy.
Reliefband® Premier is also indicated as an adjunct
to antiemetics in reducing postoperative nausea. For
over the counter use.
220 Gibraltar Road, Suite 270, Horsham, PA 19044
Reliefband® Premier is a registered trademark of
Reliefband Technologies LLC.
Reliefband® technology is protected by US Patent
No. 6,076,018; 7,127,288. Japan Patent No. 4,812,810;
4,220,129. Canadian Patent No. 2,449,554; 2,342,873.
Australian Patent Nos. 766463, 2004200090 and
2002305786, and European Patent No. 1,109,595-UK,
France and Germany (DE 69940405.3).
Designed in USA. Made in Malaysia.
©2020 Reliefband Technologies
In the U.S.A., call 1-877-735-2263
9:00 AM - 5:00 PM Mon. - Fri. CST
P/N 6106980AW Rev B
DESCRIPTION OF THE PREMIER:
Reliefband® Premier is a rechargeable, digital
wearable device that is worn on the wrist. It consists
of a hypoallergenic band with a control panel
on top and two electrical contacts underneath.
The two contacts must be placed rmly against
the underside of your wrist at the P6 location. It
provides relief from nausea and vomiting associated
with physician-diagnosed migraine, hangover,
anxiety, motion sickness, chemotherapy and
morning sickness from pregnancy.
Reliefband® Premier is also indicated as an adjunct
to antiemetics in reducing postoperative nausea.
A fully charged battery will last for approximately
18 hours of continuous use at the mid-power level
or below.
1
Reliefband® Premier delivers
proprietary pulses from the
contacts on the Smart Band that
signal the median nerve at the P6
location on the underside of the
wrist.
2
These signals travel through
the body’s nervous system
to the part of the brain
which controls nausea and
vomiting.
3
The signals have a
rebalancing eect,
normalizing nerve
messages from the brain
to the stomach and
reducing symptoms of
nausea and vomiting.
Nausea and vomiting are probably some of the
worst sensations you can feel – the combination of
dizziness, lightheadedness, heightened sensitivity to
smells, bloated abdominal discomfort, retching and
the ultimate involuntary evacuation of the stomach.
Drug-free, Reliefband® Premier uses clinically-proven
technology to quickly and eectively relieve nausea
and vomiting. Reliefband® Premier is the result of
many years of clinical research and development and
millions of dollars of investment. It is the rst of a new
class of Digital Wearable Technologies.
– it’s a Wearable that does something other than
measure things
- a “Digital Wearable Treatable”
Reliefband® Premier uses patented technology that
leverages the body’s natural neural pathways to block
signals to the center in the brain responsible for the
sensation of nausea. It has been used to treat nausea
and vomiting in clinical settings and hospitals for
many years. Now it is available over-the-counter for
the treatment of nausea and vomiting associated with
physician-diagnosed migraine, hangover, anxiety,
motion sickness, chemotherapy and morning sickness
from pregnancy. Reliefband is also an adjunct to
antiemetic therapy in reducing postoperative nausea.
CAUTION
• Pacemaker users - Use this device on the
wrist and only as directed to prevent possible
interference with your pacemaker. Avoid placing
the electrical contacts directly on your chest or
near the pacemaker. Consult your doctor before
using device.
• Nausea and vomiting may be signs of a serious
health problem. Contact a doctor if your nausea
and vomiting persists.
• Reliefband® Premier should only be used on the
wrist. Do not use it on any other area of the body.
• Reliefband® Premier does not cure the underlying
causes of nausea and vomiting.
(1) Reliefband® Premier
(1) Reliefband® Conductivity Gel
(1) Instructions For Use
(1) Magnetic USB Charging Cable
INSTRUCTIONS FOR USE
Model:Reliefband® 2.0
DRUG-FREE RELIEF
FROMNAUSEA, RETCHING AND VOMITING
2.0
• The relief you get may vary depending on
your body’s individual characteristics, and any
medications that you may be taking could cause
your body to respond dierently.
• Reliefband® Premier should be kept out of reach
of young children.
• If you are unable to tolerate any food or liquids or
are experiencing weight loss you should contact
your doctor.
• Reliefband® Premier is IPX4 splash-resistant.
This product is resistant to splashing water (rain,
waves).
• Avoid splashing water while charging.
• Do not immerse Reliefband® Premier in water or
wear while bathing, showering or swimming.
SIDE EFFECTS
• There are no known side eects for Reliefband®
Premier.
• There are no contraindications for Reliefband®
Premier.
IMPORTANT
Positioning Reliefband® Premier properly on
your wrist is essential for relief of your nausea
and vomiting symptoms. Before you start using
Reliefband® Premier, please read this Instructions
For Use carefully. You must feel a“tingling”sensation
in your palm and/or middle ngers for Reliefband®
Premier to work.
Reliefband Premier must be charged before use.
APPLYING RELIEFBAND® CONDUCTIVITY GEL
Reliefband® Conductivity Gel is an essential
component for use and therapy. Hypoallergenic
Reliefband® Conductivity Gel makes the delivery of
pulses much more ecient.
Note: Reliefband® Conductivity Gel should be
re-applied every 2-3 hours, or any time after washing,
or if stimulation decreases.
1. Power On/O/Intensity Down Button – press and
hold for 1.5 seconds to power on the device. During
active therapy, briey pressing this button will
decrease the intensity. Press and hold for 1.5 seconds
to turn the device o.
2. Intensity Up Button – Once the device has been
turned on, press this button to increase the intensity.
3. Therapy Intensity Level Indicator – 10 levels indicated
by blue illuminated LEDs.
4. Battery/Charging Indicator – four blue illuminated LED
segments indicate charge level. A single ashing red
segment indicates 15% or less battery life remaining.
5. Therapy On Indicator – pulsing blue LEDs indicate the
device is delivering therapy.
6. Hypoallergenic Contacts – 316L Stainless Steel,
therapy is delivered here.
7. Charging Port – attach the magnetic charging cable
here to recharge the device.
8. Latch – locks the Simple Band (11) tight to hold the
device in position.
9. Cinch Bar – gently pull the Simple Band (11) against
the Cinch Bar to achieve a snug t on the wrist.
10. Smart Band – unique shape holds the device in place
and makes it easy to put on and take o.
11. Simple Band – slide through the latch to adjust the t
to your wrist.
12. Contact Assembly – houses the contacts that deliver
therapy.
13. Magnetic Connector – attaches charging cable to the
device.
14. USB Charging Cable
15. UL Approved (5V) USB Charging Port
1. Ensure Reliefband Premier is turned o prior to
charging. Do not wear while charging.
2. Plug the USB Charging Cable (14) into any standard
UL approved 5 Volt USB charging port (15).
3. Attach the Magnetic Connector (13) end of the USB
Charging Cable (14) to the Charging Port (7) on the
body of the Reliefband® Premier device.
4. Blue LEDs segment(s) will ash when in charging mode:
a. 1 LED segment = 1% - 25% charged
b. 2 LED segments = 26% - 50% charged
c. 3 LED segments = 51% - 75% charged
d. 4 LED segments = 76% - 100% charged
5. Full charge is indicated when all four blue LED
segments remain constantly illuminated. Typically,
it takes about 75 minutes to reach full charge.
A 15-minute charge will give about 6 hours of
continuous use at the mid-power level or below
6. When not connected to the Reliefband, unplug
the USB Charging Cable (14) from the standard UL
approved 5 Volt USB charging port (15).
7. When turning the device on, the Battery Charge
Indicator (4) will indicate how much charge is left in
the device.
8. A single ashing red LED segment indicates 15% or
less battery life remaining.
INTENSITY
UP BUTTON
POWER
ON/OFF/INTENSITY
DOWN BUTTON
3ADJUST
Holding the device by the Contact Assembly, place
the contacts onto the gel over the P6 location on the
underside of the wrist. Fasten device snugly.
4ACTIVATE**
Press and hold the Power ON/OFF/Intensity Down
Button (-) for 1.5 seconds to turn on or o.
Press the intensity Up Button (+) once to set
stimulation to level 1. Increase intensity until tingling
is felt in palm and middle ngers.
Set to the highest comfortable setting for maximum
benet.
**Please charge for at least 1.5 hours before initial use.
(13) MAGNETIC CONNECTOR
(14)
USB CHARGING CABLE
(15)
UL APPROVED (5V) USB CHARGING PORT
(4) BATTERY/CHARGING INDICATOR
(7) CHARGING PORT
(a.) 1 LED (b.) 2 LEDs
SINGLE FLASHING RED LED
- LOW BATTERY
(c.) 3 LEDs (d.) 4 LEDs - FULLY CHARGED
VOLTAGE RATING: 5±0.25V
CURRENT RATING: 500mA MAX
WARNING
• No modication of this device is allowed.
• Charging cord must be unplugged from power
source when not in use.
• Charging Cord:
Voltage Rating: 5±0.25V
Current Rating: 500mA Max
EMERGO EUROPE
Prinsessegracht
20 2514 AP The Hague
The Netherlands

(3.)THERAPY INTENSITY
LEVEL INDICATOR
(6.) HYPOALLERGENIC
CONTACTS
(5.)THERAPY ON INDICATOR
(1.) POWER ON/OFF/INTENSITY
DOWN BUTTON
(4.) BATTERY/CHARGING
INDICATOR
(2.) INTENSITY UP
BUTTON
(8.) LATCH(9.) CINCH BAR
(12.) CONTACT
ASSEMBLY
(10.) SMART BAND
(7.) CHARGING PORT
(13.) MAGNETIC CONNECTOR
(14.) USB CHARGING CABLE
(11.) SIMPLE BAND
2
3
1
P6
2APPLY
Clean the area rst.
Apply a small drop of gel and spread in a circle
about the size of a large coin with an even sheen
(i.e., a thin layer with a shiny appearance).
1AREA
Find the starting area (P6) on the wrist.
Using either wrist, the correct area is in-between the
two tendons on the underside of the wrist – about
two nger widths above the wrist crease.
QUICK REFERENCE INTRODUCTION SAFETY INFORMATION PACKAGE CONTENTS CHARGING THE DEVICE
HOW RELIEFBAND® PREMIER
WORKS
GET TO KNOW
RELIEFBAND® PREMIER
INDICATIONS FOR USE
Reliefband® Premier is indicated for use in the
treatment of nausea and vomiting associated with
physician-diagnosed migraine, hangover, anxiety,
motion sickness, chemotherapy and morning
sickness from pregnancy.
Reliefband® Premier is also indicated as an adjunct
to antiemetics in reducing postoperative nausea. For
over the counter use.
220 Gibraltar Road, Suite 270, Horsham, PA 19044
Reliefband® Premier is a registered trademark of
Reliefband Technologies LLC.
Reliefband® technology is protected by US Patent
No. 6,076,018; 7,127,288. Japan Patent No. 4,812,810;
4,220,129. Canadian Patent No. 2,449,554; 2,342,873.
Australian Patent Nos. 766463, 2004200090 and
2002305786, and European Patent No. 1,109,595-UK,
France and Germany (DE 69940405.3).
Designed in USA. Made in Malaysia.
©2020 Reliefband Technologies
In the U.S.A., call 1-877-735-2263
9:00 AM - 5:00 PM Mon. - Fri. CST
P/N 6106980AW Rev B
DESCRIPTION OF THE PREMIER:
Reliefband® Premier is a rechargeable, digital
wearable device that is worn on the wrist. It consists
of a hypoallergenic band with a control panel
on top and two electrical contacts underneath.
The two contacts must be placed rmly against
the underside of your wrist at the P6 location. It
provides relief from nausea and vomiting associated
with physician-diagnosed migraine, hangover,
anxiety, motion sickness, chemotherapy and
morning sickness from pregnancy.
Reliefband® Premier is also indicated as an adjunct
to antiemetics in reducing postoperative nausea.
A fully charged battery will last for approximately
18 hours of continuous use at the mid-power level
or below.
1
Reliefband® Premier delivers
proprietary pulses from the
contacts on the Smart Band that
signal the median nerve at the P6
location on the underside of the
wrist.
2
These signals travel through
the body’s nervous system
to the part of the brain
which controls nausea and
vomiting.
3
The signals have a
rebalancing eect,
normalizing nerve
messages from the brain
to the stomach and
reducing symptoms of
nausea and vomiting.
Nausea and vomiting are probably some of the
worst sensations you can feel – the combination of
dizziness, lightheadedness, heightened sensitivity to
smells, bloated abdominal discomfort, retching and
the ultimate involuntary evacuation of the stomach.
Drug-free, Reliefband® Premier uses clinically-proven
technology to quickly and eectively relieve nausea
and vomiting. Reliefband® Premier is the result of
many years of clinical research and development and
millions of dollars of investment. It is the rst of a new
class of Digital Wearable Technologies.
– it’s a Wearable that does something other than
measure things
- a “Digital Wearable Treatable”
Reliefband® Premier uses patented technology that
leverages the body’s natural neural pathways to block
signals to the center in the brain responsible for the
sensation of nausea. It has been used to treat nausea
and vomiting in clinical settings and hospitals for
many years. Now it is available over-the-counter for
the treatment of nausea and vomiting associated with
physician-diagnosed migraine, hangover, anxiety,
motion sickness, chemotherapy and morning sickness
from pregnancy. Reliefband is also an adjunct to
antiemetic therapy in reducing postoperative nausea.
CAUTION
• Pacemaker users - Use this device on the
wrist and only as directed to prevent possible
interference with your pacemaker. Avoid placing
the electrical contacts directly on your chest or
near the pacemaker. Consult your doctor before
using device.
• Nausea and vomiting may be signs of a serious
health problem. Contact a doctor if your nausea
and vomiting persists.
• Reliefband® Premier should only be used on the
wrist. Do not use it on any other area of the body.
• Reliefband® Premier does not cure the underlying
causes of nausea and vomiting.
(1) Reliefband® Premier
(1) Reliefband® Conductivity Gel
(1) Instructions For Use
(1) Magnetic USB Charging Cable
INSTRUCTIONS FOR USE
Model:Reliefband® 2.0
DRUG-FREE RELIEF
FROMNAUSEA, RETCHING AND VOMITING
2.0
• The relief you get may vary depending on
your body’s individual characteristics, and any
medications that you may be taking could cause
your body to respond dierently.
• Reliefband® Premier should be kept out of reach
of young children.
• If you are unable to tolerate any food or liquids or
are experiencing weight loss you should contact
your doctor.
• Reliefband® Premier is IPX4 splash-resistant.
This product is resistant to splashing water (rain,
waves).
• Avoid splashing water while charging.
• Do not immerse Reliefband® Premier in water or
wear while bathing, showering or swimming.
SIDE EFFECTS
• There are no known side eects for Reliefband®
Premier.
• There are no contraindications for Reliefband®
Premier.
IMPORTANT
Positioning Reliefband® Premier properly on
your wrist is essential for relief of your nausea
and vomiting symptoms. Before you start using
Reliefband® Premier, please read this Instructions
For Use carefully. You must feel a“tingling”sensation
in your palm and/or middle ngers for Reliefband®
Premier to work.
Reliefband Premier must be charged before use.
APPLYING RELIEFBAND® CONDUCTIVITY GEL
Reliefband® Conductivity Gel is an essential
component for use and therapy. Hypoallergenic
Reliefband® Conductivity Gel makes the delivery of
pulses much more ecient.
Note: Reliefband® Conductivity Gel should be
re-applied every 2-3 hours, or any time after washing,
or if stimulation decreases.
1. Power On/O/Intensity Down Button – press and
hold for 1.5 seconds to power on the device. During
active therapy, briey pressing this button will
decrease the intensity. Press and hold for 1.5 seconds
to turn the device o.
2. Intensity Up Button – Once the device has been
turned on, press this button to increase the intensity.
3. Therapy Intensity Level Indicator – 10 levels indicated
by blue illuminated LEDs.
4. Battery/Charging Indicator – four blue illuminated LED
segments indicate charge level. A single ashing red
segment indicates 15% or less battery life remaining.
5. Therapy On Indicator – pulsing blue LEDs indicate the
device is delivering therapy.
6. Hypoallergenic Contacts – 316L Stainless Steel,
therapy is delivered here.
7. Charging Port – attach the magnetic charging cable
here to recharge the device.
8. Latch – locks the Simple Band (11) tight to hold the
device in position.
9. Cinch Bar – gently pull the Simple Band (11) against
the Cinch Bar to achieve a snug t on the wrist.
10. Smart Band – unique shape holds the device in place
and makes it easy to put on and take o.
11. Simple Band – slide through the latch to adjust the t
to your wrist.
12. Contact Assembly – houses the contacts that deliver
therapy.
13. Magnetic Connector – attaches charging cable to the
device.
14. USB Charging Cable
15. UL Approved (5V) USB Charging Port
1. Ensure Reliefband Premier is turned o prior to
charging. Do not wear while charging.
2. Plug the USB Charging Cable (14) into any standard
UL approved 5 Volt USB charging port (15).
3. Attach the Magnetic Connector (13) end of the USB
Charging Cable (14) to the Charging Port (7) on the
body of the Reliefband® Premier device.
4. Blue LEDs segment(s) will ash when in charging mode:
a. 1 LED segment = 1% - 25% charged
b. 2 LED segments = 26% - 50% charged
c. 3 LED segments = 51% - 75% charged
d. 4 LED segments = 76% - 100% charged
5. Full charge is indicated when all four blue LED
segments remain constantly illuminated. Typically,
it takes about 75 minutes to reach full charge.
A 15-minute charge will give about 6 hours of
continuous use at the mid-power level or below
6. When not connected to the Reliefband, unplug
the USB Charging Cable (14) from the standard UL
approved 5 Volt USB charging port (15).
7. When turning the device on, the Battery Charge
Indicator (4) will indicate how much charge is left in
the device.
8. A single ashing red LED segment indicates 15% or
less battery life remaining.
INTENSITY
UP BUTTON
POWER
ON/OFF/INTENSITY
DOWN BUTTON
3ADJUST
Holding the device by the Contact Assembly, place
the contacts onto the gel over the P6 location on the
underside of the wrist. Fasten device snugly.
4ACTIVATE**
Press and hold the Power ON/OFF/Intensity Down
Button (-) for 1.5 seconds to turn on or o.
Press the intensity Up Button (+) once to set
stimulation to level 1. Increase intensity until tingling
is felt in palm and middle ngers.
Set to the highest comfortable setting for maximum
benet.
**Please charge for at least 1.5 hours before initial use.
(13) MAGNETIC CONNECTOR
(14)
USB CHARGING CABLE
(15)
UL APPROVED (5V) USB CHARGING PORT
(4) BATTERY/CHARGING INDICATOR
(7) CHARGING PORT
(a.) 1 LED (b.) 2 LEDs
SINGLE FLASHING RED LED
- LOW BATTERY
(c.) 3 LEDs (d.) 4 LEDs - FULLY CHARGED
VOLTAGE RATING: 5±0.25V
CURRENT RATING: 500mA MAX
WARNING
• No modication of this device is allowed.
• Charging cord must be unplugged from power
source when not in use.
• Charging Cord:
Voltage Rating: 5±0.25V
Current Rating: 500mA Max
EMERGO EUROPE
Prinsessegracht
20 2514 AP The Hague
The Netherlands

(3.)THERAPY INTENSITY
LEVEL INDICATOR
(6.) HYPOALLERGENIC
CONTACTS
(5.)THERAPY ON INDICATOR
(1.) POWER ON/OFF/INTENSITY
DOWN BUTTON
(4.) BATTERY/CHARGING
INDICATOR
(2.) INTENSITY UP
BUTTON
(8.) LATCH(9.) CINCH BAR
(12.) CONTACT
ASSEMBLY
(10.) SMART BAND
(7.) CHARGING PORT
(13.) MAGNETIC CONNECTOR
(14.) USB CHARGING CABLE
(11.) SIMPLE BAND
2
3
1
P6
2APPLY
Clean the area rst.
Apply a small drop of gel and spread in a circle
about the size of a large coin with an even sheen
(i.e., a thin layer with a shiny appearance).
1AREA
Find the starting area (P6) on the wrist.
Using either wrist, the correct area is in-between the
two tendons on the underside of the wrist – about
two nger widths above the wrist crease.
QUICK REFERENCE INTRODUCTION SAFETY INFORMATION PACKAGE CONTENTS CHARGING THE DEVICE
HOW RELIEFBAND® PREMIER
WORKS
GET TO KNOW
RELIEFBAND® PREMIER
INDICATIONS FOR USE
Reliefband® Premier is indicated for use in the
treatment of nausea and vomiting associated with
physician-diagnosed migraine, hangover, anxiety,
motion sickness, chemotherapy and morning
sickness from pregnancy.
Reliefband® Premier is also indicated as an adjunct
to antiemetics in reducing postoperative nausea. For
over the counter use.
220 Gibraltar Road, Suite 270, Horsham, PA 19044
Reliefband® Premier is a registered trademark of
Reliefband Technologies LLC.
Reliefband® technology is protected by US Patent
No. 6,076,018; 7,127,288. Japan Patent No. 4,812,810;
4,220,129. Canadian Patent No. 2,449,554; 2,342,873.
Australian Patent Nos. 766463, 2004200090 and
2002305786, and European Patent No. 1,109,595-UK,
France and Germany (DE 69940405.3).
Designed in USA. Made in Malaysia.
©2020 Reliefband Technologies
In the U.S.A., call 1-877-735-2263
9:00 AM - 5:00 PM Mon. - Fri. CST
P/N 6106980AW Rev B
DESCRIPTION OF THE PREMIER:
Reliefband® Premier is a rechargeable, digital
wearable device that is worn on the wrist. It consists
of a hypoallergenic band with a control panel
on top and two electrical contacts underneath.
The two contacts must be placed rmly against
the underside of your wrist at the P6 location. It
provides relief from nausea and vomiting associated
with physician-diagnosed migraine, hangover,
anxiety, motion sickness, chemotherapy and
morning sickness from pregnancy.
Reliefband® Premier is also indicated as an adjunct
to antiemetics in reducing postoperative nausea.
A fully charged battery will last for approximately
18 hours of continuous use at the mid-power level
or below.
1
Reliefband® Premier delivers
proprietary pulses from the
contacts on the Smart Band that
signal the median nerve at the P6
location on the underside of the
wrist.
2
These signals travel through
the body’s nervous system
to the part of the brain
which controls nausea and
vomiting.
3
The signals have a
rebalancing eect,
normalizing nerve
messages from the brain
to the stomach and
reducing symptoms of
nausea and vomiting.
Nausea and vomiting are probably some of the
worst sensations you can feel – the combination of
dizziness, lightheadedness, heightened sensitivity to
smells, bloated abdominal discomfort, retching and
the ultimate involuntary evacuation of the stomach.
Drug-free, Reliefband® Premier uses clinically-proven
technology to quickly and eectively relieve nausea
and vomiting. Reliefband® Premier is the result of
many years of clinical research and development and
millions of dollars of investment. It is the rst of a new
class of Digital Wearable Technologies.
– it’s a Wearable that does something other than
measure things
- a “Digital Wearable Treatable”
Reliefband® Premier uses patented technology that
leverages the body’s natural neural pathways to block
signals to the center in the brain responsible for the
sensation of nausea. It has been used to treat nausea
and vomiting in clinical settings and hospitals for
many years. Now it is available over-the-counter for
the treatment of nausea and vomiting associated with
physician-diagnosed migraine, hangover, anxiety,
motion sickness, chemotherapy and morning sickness
from pregnancy. Reliefband is also an adjunct to
antiemetic therapy in reducing postoperative nausea.
CAUTION
• Pacemaker users - Use this device on the
wrist and only as directed to prevent possible
interference with your pacemaker. Avoid placing
the electrical contacts directly on your chest or
near the pacemaker. Consult your doctor before
using device.
• Nausea and vomiting may be signs of a serious
health problem. Contact a doctor if your nausea
and vomiting persists.
• Reliefband® Premier should only be used on the
wrist. Do not use it on any other area of the body.
• Reliefband® Premier does not cure the underlying
causes of nausea and vomiting.
(1) Reliefband® Premier
(1) Reliefband® Conductivity Gel
(1) Instructions For Use
(1) Magnetic USB Charging Cable
INSTRUCTIONS FOR USE
Model:Reliefband® 2.0
DRUG-FREE RELIEF
FROMNAUSEA, RETCHING AND VOMITING
2.0
• The relief you get may vary depending on
your body’s individual characteristics, and any
medications that you may be taking could cause
your body to respond dierently.
• Reliefband® Premier should be kept out of reach
of young children.
• If you are unable to tolerate any food or liquids or
are experiencing weight loss you should contact
your doctor.
• Reliefband® Premier is IPX4 splash-resistant.
This product is resistant to splashing water (rain,
waves).
• Avoid splashing water while charging.
• Do not immerse Reliefband® Premier in water or
wear while bathing, showering or swimming.
SIDE EFFECTS
• There are no known side eects for Reliefband®
Premier.
• There are no contraindications for Reliefband®
Premier.
IMPORTANT
Positioning Reliefband® Premier properly on
your wrist is essential for relief of your nausea
and vomiting symptoms. Before you start using
Reliefband® Premier, please read this Instructions
For Use carefully. You must feel a“tingling”sensation
in your palm and/or middle ngers for Reliefband®
Premier to work.
Reliefband Premier must be charged before use.
APPLYING RELIEFBAND® CONDUCTIVITY GEL
Reliefband® Conductivity Gel is an essential
component for use and therapy. Hypoallergenic
Reliefband® Conductivity Gel makes the delivery of
pulses much more ecient.
Note: Reliefband® Conductivity Gel should be
re-applied every 2-3 hours, or any time after washing,
or if stimulation decreases.
1. Power On/O/Intensity Down Button – press and
hold for 1.5 seconds to power on the device. During
active therapy, briey pressing this button will
decrease the intensity. Press and hold for 1.5 seconds
to turn the device o.
2. Intensity Up Button – Once the device has been
turned on, press this button to increase the intensity.
3. Therapy Intensity Level Indicator – 10 levels indicated
by blue illuminated LEDs.
4. Battery/Charging Indicator – four blue illuminated LED
segments indicate charge level. A single ashing red
segment indicates 15% or less battery life remaining.
5. Therapy On Indicator – pulsing blue LEDs indicate the
device is delivering therapy.
6. Hypoallergenic Contacts – 316L Stainless Steel,
therapy is delivered here.
7. Charging Port – attach the magnetic charging cable
here to recharge the device.
8. Latch – locks the Simple Band (11) tight to hold the
device in position.
9. Cinch Bar – gently pull the Simple Band (11) against
the Cinch Bar to achieve a snug t on the wrist.
10. Smart Band – unique shape holds the device in place
and makes it easy to put on and take o.
11. Simple Band – slide through the latch to adjust the t
to your wrist.
12. Contact Assembly – houses the contacts that deliver
therapy.
13. Magnetic Connector – attaches charging cable to the
device.
14. USB Charging Cable
15. UL Approved (5V) USB Charging Port
1. Ensure Reliefband Premier is turned o prior to
charging. Do not wear while charging.
2. Plug the USB Charging Cable (14) into any standard
UL approved 5 Volt USB charging port (15).
3. Attach the Magnetic Connector (13) end of the USB
Charging Cable (14) to the Charging Port (7) on the
body of the Reliefband® Premier device.
4. Blue LEDs segment(s) will ash when in charging mode:
a. 1 LED segment = 1% - 25% charged
b. 2 LED segments = 26% - 50% charged
c. 3 LED segments = 51% - 75% charged
d. 4 LED segments = 76% - 100% charged
5. Full charge is indicated when all four blue LED
segments remain constantly illuminated. Typically,
it takes about 75 minutes to reach full charge.
A 15-minute charge will give about 6 hours of
continuous use at the mid-power level or below
6. When not connected to the Reliefband, unplug
the USB Charging Cable (14) from the standard UL
approved 5 Volt USB charging port (15).
7. When turning the device on, the Battery Charge
Indicator (4) will indicate how much charge is left in
the device.
8. A single ashing red LED segment indicates 15% or
less battery life remaining.
INTENSITY
UP BUTTON
POWER
ON/OFF/INTENSITY
DOWN BUTTON
3ADJUST
Holding the device by the Contact Assembly, place
the contacts onto the gel over the P6 location on the
underside of the wrist. Fasten device snugly.
4ACTIVATE**
Press and hold the Power ON/OFF/Intensity Down
Button (-) for 1.5 seconds to turn on or o.
Press the intensity Up Button (+) once to set
stimulation to level 1. Increase intensity until tingling
is felt in palm and middle ngers.
Set to the highest comfortable setting for maximum
benet.
**Please charge for at least 1.5 hours before initial use.
(13) MAGNETIC CONNECTOR
(14)
USB CHARGING CABLE
(15)
UL APPROVED (5V) USB CHARGING PORT
(4) BATTERY/CHARGING INDICATOR
(7) CHARGING PORT
(a.) 1 LED (b.) 2 LEDs
SINGLE FLASHING RED LED
- LOW BATTERY
(c.) 3 LEDs (d.) 4 LEDs - FULLY CHARGED
VOLTAGE RATING: 5±0.25V
CURRENT RATING: 500mA MAX
WARNING
• No modication of this device is allowed.
• Charging cord must be unplugged from power
source when not in use.
• Charging Cord:
Voltage Rating: 5±0.25V
Current Rating: 500mA Max
EMERGO EUROPE
Prinsessegracht
20 2514 AP The Hague
The Netherlands

(3.)THERAPY INTENSITY
LEVEL INDICATOR
(6.) HYPOALLERGENIC
CONTACTS
(5.)THERAPY ON INDICATOR
(1.) POWER ON/OFF/INTENSITY
DOWN BUTTON
(4.) BATTERY/CHARGING
INDICATOR
(2.) INTENSITY UP
BUTTON
(8.) LATCH(9.) CINCH BAR
(12.) CONTACT
ASSEMBLY
(10.) SMART BAND
(7.) CHARGING PORT
(13.) MAGNETIC CONNECTOR
(14.) USB CHARGING CABLE
(11.) SIMPLE BAND
2
3
1
P6
2APPLY
Clean the area rst.
Apply a small drop of gel and spread in a circle
about the size of a large coin with an even sheen
(i.e., a thin layer with a shiny appearance).
1AREA
Find the starting area (P6) on the wrist.
Using either wrist, the correct area is in-between the
two tendons on the underside of the wrist – about
two nger widths above the wrist crease.
QUICK REFERENCE INTRODUCTION SAFETY INFORMATION PACKAGE CONTENTS CHARGING THE DEVICE
HOW RELIEFBAND® PREMIER
WORKS
GET TO KNOW
RELIEFBAND® PREMIER
INDICATIONS FOR USE
Reliefband® Premier is indicated for use in the
treatment of nausea and vomiting associated with
physician-diagnosed migraine, hangover, anxiety,
motion sickness, chemotherapy and morning
sickness from pregnancy.
Reliefband® Premier is also indicated as an adjunct
to antiemetics in reducing postoperative nausea. For
over the counter use.
220 Gibraltar Road, Suite 270, Horsham, PA 19044
Reliefband® Premier is a registered trademark of
Reliefband Technologies LLC.
Reliefband® technology is protected by US Patent
No. 6,076,018; 7,127,288. Japan Patent No. 4,812,810;
4,220,129. Canadian Patent No. 2,449,554; 2,342,873.
Australian Patent Nos. 766463, 2004200090 and
2002305786, and European Patent No. 1,109,595-UK,
France and Germany (DE 69940405.3).
Designed in USA. Made in Malaysia.
©2020 Reliefband Technologies
In the U.S.A., call 1-877-735-2263
9:00 AM - 5:00 PM Mon. - Fri. CST
P/N 6106980AW Rev B
DESCRIPTION OF THE PREMIER:
Reliefband® Premier is a rechargeable, digital
wearable device that is worn on the wrist. It consists
of a hypoallergenic band with a control panel
on top and two electrical contacts underneath.
The two contacts must be placed rmly against
the underside of your wrist at the P6 location. It
provides relief from nausea and vomiting associated
with physician-diagnosed migraine, hangover,
anxiety, motion sickness, chemotherapy and
morning sickness from pregnancy.
Reliefband® Premier is also indicated as an adjunct
to antiemetics in reducing postoperative nausea.
A fully charged battery will last for approximately
18 hours of continuous use at the mid-power level
or below.
1
Reliefband® Premier delivers
proprietary pulses from the
contacts on the Smart Band that
signal the median nerve at the P6
location on the underside of the
wrist.
2
These signals travel through
the body’s nervous system
to the part of the brain
which controls nausea and
vomiting.
3
The signals have a
rebalancing eect,
normalizing nerve
messages from the brain
to the stomach and
reducing symptoms of
nausea and vomiting.
Nausea and vomiting are probably some of the
worst sensations you can feel – the combination of
dizziness, lightheadedness, heightened sensitivity to
smells, bloated abdominal discomfort, retching and
the ultimate involuntary evacuation of the stomach.
Drug-free, Reliefband® Premier uses clinically-proven
technology to quickly and eectively relieve nausea
and vomiting. Reliefband® Premier is the result of
many years of clinical research and development and
millions of dollars of investment. It is the rst of a new
class of Digital Wearable Technologies.
– it’s a Wearable that does something other than
measure things
- a “Digital Wearable Treatable”
Reliefband® Premier uses patented technology that
leverages the body’s natural neural pathways to block
signals to the center in the brain responsible for the
sensation of nausea. It has been used to treat nausea
and vomiting in clinical settings and hospitals for
many years. Now it is available over-the-counter for
the treatment of nausea and vomiting associated with
physician-diagnosed migraine, hangover, anxiety,
motion sickness, chemotherapy and morning sickness
from pregnancy. Reliefband is also an adjunct to
antiemetic therapy in reducing postoperative nausea.
CAUTION
• Pacemaker users - Use this device on the
wrist and only as directed to prevent possible
interference with your pacemaker. Avoid placing
the electrical contacts directly on your chest or
near the pacemaker. Consult your doctor before
using device.
• Nausea and vomiting may be signs of a serious
health problem. Contact a doctor if your nausea
and vomiting persists.
• Reliefband® Premier should only be used on the
wrist. Do not use it on any other area of the body.
• Reliefband® Premier does not cure the underlying
causes of nausea and vomiting.
(1) Reliefband® Premier
(1) Reliefband® Conductivity Gel
(1) Instructions For Use
(1) Magnetic USB Charging Cable
INSTRUCTIONS FOR USE
Model:Reliefband® 2.0
DRUG-FREE RELIEF
FROMNAUSEA, RETCHING AND VOMITING
2.0
• The relief you get may vary depending on
your body’s individual characteristics, and any
medications that you may be taking could cause
your body to respond dierently.
• Reliefband® Premier should be kept out of reach
of young children.
• If you are unable to tolerate any food or liquids or
are experiencing weight loss you should contact
your doctor.
• Reliefband® Premier is IPX4 splash-resistant.
This product is resistant to splashing water (rain,
waves).
• Avoid splashing water while charging.
• Do not immerse Reliefband® Premier in water or
wear while bathing, showering or swimming.
SIDE EFFECTS
• There are no known side eects for Reliefband®
Premier.
• There are no contraindications for Reliefband®
Premier.
IMPORTANT
Positioning Reliefband® Premier properly on
your wrist is essential for relief of your nausea
and vomiting symptoms. Before you start using
Reliefband® Premier, please read this Instructions
For Use carefully. You must feel a“tingling”sensation
in your palm and/or middle ngers for Reliefband®
Premier to work.
Reliefband Premier must be charged before use.
APPLYING RELIEFBAND® CONDUCTIVITY GEL
Reliefband® Conductivity Gel is an essential
component for use and therapy. Hypoallergenic
Reliefband® Conductivity Gel makes the delivery of
pulses much more ecient.
Note: Reliefband® Conductivity Gel should be
re-applied every 2-3 hours, or any time after washing,
or if stimulation decreases.
1. Power On/O/Intensity Down Button – press and
hold for 1.5 seconds to power on the device. During
active therapy, briey pressing this button will
decrease the intensity. Press and hold for 1.5 seconds
to turn the device o.
2. Intensity Up Button – Once the device has been
turned on, press this button to increase the intensity.
3. Therapy Intensity Level Indicator – 10 levels indicated
by blue illuminated LEDs.
4. Battery/Charging Indicator – four blue illuminated LED
segments indicate charge level. A single ashing red
segment indicates 15% or less battery life remaining.
5. Therapy On Indicator – pulsing blue LEDs indicate the
device is delivering therapy.
6. Hypoallergenic Contacts – 316L Stainless Steel,
therapy is delivered here.
7. Charging Port – attach the magnetic charging cable
here to recharge the device.
8. Latch – locks the Simple Band (11) tight to hold the
device in position.
9. Cinch Bar – gently pull the Simple Band (11) against
the Cinch Bar to achieve a snug t on the wrist.
10. Smart Band – unique shape holds the device in place
and makes it easy to put on and take o.
11. Simple Band – slide through the latch to adjust the t
to your wrist.
12. Contact Assembly – houses the contacts that deliver
therapy.
13. Magnetic Connector – attaches charging cable to the
device.
14. USB Charging Cable
15. UL Approved (5V) USB Charging Port
1. Ensure Reliefband Premier is turned o prior to
charging. Do not wear while charging.
2. Plug the USB Charging Cable (14) into any standard
UL approved 5 Volt USB charging port (15).
3. Attach the Magnetic Connector (13) end of the USB
Charging Cable (14) to the Charging Port (7) on the
body of the Reliefband® Premier device.
4. Blue LEDs segment(s) will ash when in charging mode:
a. 1 LED segment = 1% - 25% charged
b. 2 LED segments = 26% - 50% charged
c. 3 LED segments = 51% - 75% charged
d. 4 LED segments = 76% - 100% charged
5. Full charge is indicated when all four blue LED
segments remain constantly illuminated. Typically,
it takes about 75 minutes to reach full charge.
A 15-minute charge will give about 6 hours of
continuous use at the mid-power level or below
6. When not connected to the Reliefband, unplug
the USB Charging Cable (14) from the standard UL
approved 5 Volt USB charging port (15).
7. When turning the device on, the Battery Charge
Indicator (4) will indicate how much charge is left in
the device.
8. A single ashing red LED segment indicates 15% or
less battery life remaining.
INTENSITY
UP BUTTON
POWER
ON/OFF/INTENSITY
DOWN BUTTON
3ADJUST
Holding the device by the Contact Assembly, place
the contacts onto the gel over the P6 location on the
underside of the wrist. Fasten device snugly.
4ACTIVATE**
Press and hold the Power ON/OFF/Intensity Down
Button (-) for 1.5 seconds to turn on or o.
Press the intensity Up Button (+) once to set
stimulation to level 1. Increase intensity until tingling
is felt in palm and middle ngers.
Set to the highest comfortable setting for maximum
benet.
**Please charge for at least 1.5 hours before initial use.
(13) MAGNETIC CONNECTOR
(14)
USB CHARGING CABLE
(15)
UL APPROVED (5V) USB CHARGING PORT
(4) BATTERY/CHARGING INDICATOR
(7) CHARGING PORT
(a.) 1 LED (b.) 2 LEDs
SINGLE FLASHING RED LED
- LOW BATTERY
(c.) 3 LEDs (d.) 4 LEDs - FULLY CHARGED
VOLTAGE RATING: 5±0.25V
CURRENT RATING: 500mA MAX
WARNING
• No modication of this device is allowed.
• Charging cord must be unplugged from power
source when not in use.
• Charging Cord:
Voltage Rating: 5±0.25V
Current Rating: 500mA Max
EMERGO EUROPE
Prinsessegracht
20 2514 AP The Hague
The Netherlands

FCC INFORMATION:
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following conditions: (1)This
device may not cause harmful interference, and (2) this
device must accept any interference received, including
interference that may cause undesired operation.
MAINTENANCE:
Keep the contacts area clean. Periodically examine the
contacts for gel build-up around the perimeter of the
contacts, which may interfere with or decrease the
delivery of therapy. A moist cloth or cotton swab may
be used to remove any gel build-up. Avoid immersion
of the device in liquid. Reliefband® Premier must not
be opened or altered in any way.
ENVIRONMENTAL USE INFORMATION:
Reliefband® Premier should only be used within
the environmental ranges of 0°C to 38°C (32°F to
100°F) and 20% to 90% relative humidity (non-
condensing).
38
1. Turn the product to ON by pressing and holding
the Power On/O/Intensity Down Button (1) for
1.5 seconds
a. Upon powering on, the Intensity Level
Indicator LEDs (3) will illuminate in
sequence of 1-10, then back down to level 0.
b. The current battery charge is displayed for
1.5 seconds before fading away.
2. Press the Intensity Up Button (2) once to set
stimulation to the rst level. At this point
therapy is being delivered. If intensity level is
not set, Reliefband Premier will turn o.
a. Blue LED intensity lights will indicate
intensity level:
i. No LEDs illuminated = Level 0 =
Power ON but no therapy
ii. 1 LED illuminated = Level 1 Therapy
iii.10 LED’s illuminated = Level 10 Therapy
b. The Therapy On Indicator (5) will pulse
on and o indicating that the device is
delivering therapy.
Size Rectangular shape – max case
dimensions: 14.5mm H x 23.5mm W x
4.1mm L, with Smart Band and Simple
Band.
Weight Approximately 50 grams
User Controls Two Push Buttons: On/O/Intensity Down
and Intensity Up
User Display Blue LED logo ( ) pulsing means device
is providing therapy. Ten diagonal blue
LED bars indicate therapy power level.
Four circular blue LED segments indicate
battery charge level.
Low Battery When battery level registers 15% or less,
a single LED segment will switch from
static blue to ashing red.
Output
Channels
Two hypoallergenic surgical grade (316L)
stainless steel contacts.
Maximum
Output
40mA
USB Charging
Cable
Cable with custom magnetic connector
Battery Rechargeable Li-Ion Polymer - IEC62133
standard. Battery life is approximately
three years or 500 charge cycles.
Waveform
Characteristics
@500 ohm load
40mA, 22.5V, 31Hz, 350us per pulse
The Troubleshooting Chart below will help to assist
you in handling situations that may arise while you
use the product.
Situation Recommended Solution
Blue intensity
LEDs and
pulsing blue
Therapy On
LEDs are
illuminated but
no stimulation is
felt
Move the contacts around the underside
of the wrist to ensure proper placement.
Increase the intensity using the Intensity
Up Button (2). Re-apply a small amount of
Reliefband Conductivity Gel. Ensure gel
is rubbed into a sheen. Be sure the strap
is secure around the wrist. Be sure blue
Intensity LEDs (3) are illuminated at one
of the ten intensity settings.
No LEDs
Illuminated
Briey press the Intensity On/O/Down
button (1) once to check if device is
powered on. If power is on, the blue
Intensity LED bars (3), the Battery/Charging
Indicator segments (4) and the Therapy On
Indicator (5) will ash. If nothing ashes,
the device may be o. Press and hold the
Intensity On/O/Down button for 1.5
seconds to wake up the device. If no LEDs
ash, the battery may be drained. To charge
the battery - see the "Charging The Battery"
section.
Flashing red,
low battery LED
illuminated
This indicates a low battery and 15% (2.4
hours at level 5) or less of charge remains.
Recharge the battery as soon as possible
– see "Charging The Battery" section
Model: Reliefband® Premier
Reliefband® Technologies warrants each new Reliefband®
Premier to be free from material defect in workmanship
and materials for (1 year) from the date of purchase, the
“Warranty Period.”This limited warranty is subject to (a)
the customer notifying Reliefband® Technologies within
1 year of the date of purchase of any material defect
in Reliefband® Premier workmanship and materials
(“Defective Product”) during the Warranty Period,
and (b) the customer returning the Defective Product
to Reliefband® Technologies. Contact Reliefband®
Technologies at 877-735-2263 for return or replacement
of Defective Product.
Reliefband® Premier is indicated for nonprescription
(over-the- counter) use in relief of nausea and vomiting
associated with physician-diagnosed migraine, hangover,
anxiety, motion sickness, chemotherapy and morning
3. Adjust intensity level using the Intensity Up
Button (2) until a "tingling" sensation is felt in
the palm or middle ngers of your hand.
4. If no “tingling” is felt, increase the intensity using
the Intensity Up Button (2).
5. If "tingling" is still not felt, check the position
of the Hypoallergenic Contacts (6) and adjust
until you feel "tingling" in your palm and middle
ngers.
6. Press and hold the Power On/O/Intensity Down
Button (1) for 1.5 seconds to turn the device o.
7. If the battery charge has depleted, recharge
the device according to "Charging The Device"
instructions.
a. Once battery level registers 15% or below,
the single LED will switch from static blue
to ashing red.
b. A ashing red LED indicates the device will
require charging as soon as possible.
ENVIRONMENTAL STORAGE INFORMATION:
The device can be stored and transported in the
environmental ranges of 0°C to 27°C (32°F to
80°F) and 20% to 80% relative humidity (non-
condensing). If the device has been stored or
transported in conditions outside this range, keep
it within the normal ranges for at least 30 minutes
before using.
DISPOSAL INFORMATION
Reliefband Premier is not regulated as hazardous
goods. It can be recycled or disposed with normal
household trash.
TECHNICAL INFORMATION
Reliefband complies with:
• IEC 60601-1:2005+A1:2012/EN 60601-1:2006
Medical electrical equipment - Part 1: General
requirements for basic safety and essential
performance
• IEC 60601-1-2:2007/EN 60601-1-2:2007 Medical
electrical equipment - Part 1-2: General
requirements for safety - Collateral standard:
Electromagnetic compatibility - Requirements
and tests
TIPS FOR EFFECTIVE USE
• Reliefband® Premier can be worn on either wrist provided
tingling is felt in the palm and/or middle ngers.
• For some people, applying and rubbing in the gel twice assists
in nerve stimulation.
• Reliefband® Conductivity Gel should be re-applied every 2-3
hours during use, any time after washing, or if stimulation
decreases.
• Contact with skin for prolonged periods may contribute to skin
irritation particularly for users who have allergies. To reduce
irritation, follow three simple wear and care tips: (1) Keep it
clean; (2) keep it dry; and (3) after extended wear, give your
wrist a rest by removing the band for an hour.
• Regularly clean your Reliefband Premier and wrist—especially
if perspiring, or before reapplying conductivity gel – before
returning the Reliefband Premier to your wrist. Wipe the band
and contacts with a clean, damp wipe using water or a small
amount of rubbing alcohol. Do NOT use hand soap, body
soap, dish soap, hand sanitizers, cleaning wipes or household
cleaners which could get trapped beneath the band and
irritate skin. Always dry the band well before putting it back on.
• To remove build-up of lotions and oils (such as sunscreen,
insect repellent and moisturizers) that can be trapped beneath
the band, we recommend that you use a soap-free cleanser like
Cetaphil Gentle Skin Cleanser or Aquanil; wipe o thoroughly,
and dry well.
• While still on the wrist, turn Reliefband® Premier o periodically
to determine if you still have nausea and vomiting symptoms.
If you still have symptoms, then turn it back on immediately.
sickness from pregnancy. Reliefband is also indicated as an
adjunct to antiemetics in reducing postoperative nausea.
Reliefband® Premier may not be eective for every person
because treatment outcomes vary and are dependent
upon a myriad of patient proles and characteristics.
Reliefband® Technologies’ obligation under this Limited
Warranty is expressly limited solely and exclusively to the
replacement of the Defective Product.
THE LIMITED REPRESENTATIONS AND WARRANTIES SET
FORTH HEREIN ARE RELIEFBAND TECHNOLOGIES’ SOLE
REPRESENTATIONS AND WARRANTIES WITH RESPECT
TO THE PREMIER OR ANY OTHER SUBJECT MATTER
HEREOF AND ARE GIVEN IN LIEU OF ANY AND ALL
OTHER REPRESENTATIONS OR WARRANTIES, WHETHER
EXPRESS OR IMPLIED, INCLUDING, WITHOUT LIMITATION,
ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR
SUITABILITY FOR A PARTICULAR PURPOSE. THE REMEDIES
AFFORDED TO CUSTOMER FOR ANY BREACH OF
WARRANTY WILL BE LIMITED TO THOSE SET FORTH HEREIN
TO THE EXCLUSION OF ANY AND ALL OTHER REMEDIES.
CUSTOMER WILL NOT BE ENTITLED TO PUNITIVE,
INCIDENTAL, CONSEQUENTIAL OR SPECIAL DAMAGES,
INCLUDING WITHOUT LIMITATION CUSTOMER’S DENTAL
AND MEDICAL COSTS, AS A RESULT OF A DEFECTIVE
PRODUCT. NO AGREEMENT VARYING OR EXTENDING
THE FOREGOING WARRANTIES, REMEDIES OR THIS
LIMITATION WILL BE BINDING ON EITHER PARTY UNLESS
MEMORIALIZED IN A WRITING SIGNED BY BOTH PARTIES.
Some jurisdictions do not allow the exclusion or limitation
of special, indirect, incidental or consequential damages,
so the above limitation or exclusion may not apply to you.
INSTRUCTIONS FOR USE
CLINICALLY PROVEN RELIEF
FOR THE PREVENTION AND TREATMENT OF
NAUSEA AND VOMITING
PREMIER
RB2.0
DETAILED OPERATION TECHNICAL DATA
TROUBLESHOOTING
STORAGE AND USAGE
INFORMATION
LIMITED WARRANTY
STATEMENT
(2)
INTENSITY UP
BUTTON
(1)
POWER ON/OFF/INTENSITY
DOWN BUTTON
(3)
INTENSITY LEVEL
INDICATOR
(5)
THERAPY ON
INDICATOR
(6)
CONTACTS
YOU SHOULD FEEL A
"TINGLING" SENSATION IN THE
PALM OR MIDDLE FINGERS.
SINGLE FLASHING
RED LED
- LOW BATTERY
Reliefband® Premier produces
physiological eects as described in this
guide.
Reliefband® Premier is classied as a
TYPE BF Applied Part
Latex Free
Reliefband® Premier is splash-resistant
ATTENTION
IPX4
For European Union: Environmental Protection
Waste electrical products should not be
disposed of with household waste. Please
recycle where facilities exist. Check with
your Local Authority or retailer for recycling
advice.
California Only: Percholorate Material -
Special handling may apply. See
www.dtsc.ca.gov/hazardouswaste/perchlorate
Read these Instructions before using
Reliefband® Premier
This warranty shall not apply to any Defective Product
that has been tampered with or repaired and/or altered
by someone other than a duly-authorized Reliefband
Technologies’ representative. Further, this warranty shall
not apply to any Defective Product where the defect has
been caused by customer negligence, mishandling or
any customer use contrary to the enclosed instructions or
customer use inconsistent with the stated purpose.
This warranty is expressly limited solely to the original
purchaser and does not extend or confer any rights to
any transferee, assignee, or subsequent purchaser or user
of a Defective Product. Reliefband® Technologies’liability
for all claims, (whether based on contract, tort, breach of
warranty, or otherwise), which may arise in connection
with the purchase and use of any Defective Product
is limited to the purchase price paid by the original
purchaser.
P/N 6106980AW Rev B
• IEC 60601-2-10:2012/EN 60601-2-
10:2000+A1:2001 Medical electrical equipment
- Part 2-10: Particular requirements for the safety
of nerve and muscle stimulators
• When nished using Reliefband® Premier take it o your wrist.
Remove the gel from your wrist and the device with a tissue.
• When not using Reliefband® Premier be sure to turn the power
o to extend battery life.
• Store Reliefband® Premier in a safe, dry place. Keep it away
from young children.
• Avoid an excessive amount of gel; it could decrease the ecacy
of the device.
.com

FCC INFORMATION:
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following conditions: (1)This
device may not cause harmful interference, and (2) this
device must accept any interference received, including
interference that may cause undesired operation.
MAINTENANCE:
Keep the contacts area clean. Periodically examine the
contacts for gel build-up around the perimeter of the
contacts, which may interfere with or decrease the
delivery of therapy. A moist cloth or cotton swab may
be used to remove any gel build-up. Avoid immersion
of the device in liquid. Reliefband® Premier must not
be opened or altered in any way.
ENVIRONMENTAL USE INFORMATION:
Reliefband® Premier should only be used within
the environmental ranges of 0°C to 38°C (32°F to
100°F) and 20% to 90% relative humidity (non-
condensing).
38
1. Turn the product to ON by pressing and holding
the Power On/O/Intensity Down Button (1) for
1.5 seconds
a. Upon powering on, the Intensity Level
Indicator LEDs (3) will illuminate in
sequence of 1-10, then back down to level 0.
b. The current battery charge is displayed for
1.5 seconds before fading away.
2. Press the Intensity Up Button (2) once to set
stimulation to the rst level. At this point
therapy is being delivered. If intensity level is
not set, Reliefband Premier will turn o.
a. Blue LED intensity lights will indicate
intensity level:
i. No LEDs illuminated = Level 0 =
Power ON but no therapy
ii. 1 LED illuminated = Level 1 Therapy
iii.10 LED’s illuminated = Level 10 Therapy
b. The Therapy On Indicator (5) will pulse
on and o indicating that the device is
delivering therapy.
Size Rectangular shape – max case
dimensions: 14.5mm H x 23.5mm W x
4.1mm L, with Smart Band and Simple
Band.
Weight Approximately 50 grams
User Controls Two Push Buttons: On/O/Intensity Down
and Intensity Up
User Display Blue LED logo ( ) pulsing means device
is providing therapy. Ten diagonal blue
LED bars indicate therapy power level.
Four circular blue LED segments indicate
battery charge level.
Low Battery When battery level registers 15% or less,
a single LED segment will switch from
static blue to ashing red.
Output
Channels
Two hypoallergenic surgical grade (316L)
stainless steel contacts.
Maximum
Output
40mA
USB Charging
Cable
Cable with custom magnetic connector
Battery Rechargeable Li-Ion Polymer - IEC62133
standard. Battery life is approximately
three years or 500 charge cycles.
Waveform
Characteristics
@500 ohm load
40mA, 22.5V, 31Hz, 350us per pulse
The Troubleshooting Chart below will help to assist
you in handling situations that may arise while you
use the product.
Situation Recommended Solution
Blue intensity
LEDs and
pulsing blue
Therapy On
LEDs are
illuminated but
no stimulation is
felt
Move the contacts around the underside
of the wrist to ensure proper placement.
Increase the intensity using the Intensity
Up Button (2). Re-apply a small amount of
Reliefband Conductivity Gel. Ensure gel
is rubbed into a sheen. Be sure the strap
is secure around the wrist. Be sure blue
Intensity LEDs (3) are illuminated at one
of the ten intensity settings.
No LEDs
Illuminated
Briey press the Intensity On/O/Down
button (1) once to check if device is
powered on. If power is on, the blue
Intensity LED bars (3), the Battery/Charging
Indicator segments (4) and the Therapy On
Indicator (5) will ash. If nothing ashes,
the device may be o. Press and hold the
Intensity On/O/Down button for 1.5
seconds to wake up the device. If no LEDs
ash, the battery may be drained. To charge
the battery - see the "Charging The Battery"
section.
Flashing red,
low battery LED
illuminated
This indicates a low battery and 15% (2.4
hours at level 5) or less of charge remains.
Recharge the battery as soon as possible
– see "Charging The Battery" section
Model: Reliefband® Premier
Reliefband® Technologies warrants each new Reliefband®
Premier to be free from material defect in workmanship
and materials for (1 year) from the date of purchase, the
“Warranty Period.”This limited warranty is subject to (a)
the customer notifying Reliefband® Technologies within
1 year of the date of purchase of any material defect
in Reliefband® Premier workmanship and materials
(“Defective Product”) during the Warranty Period,
and (b) the customer returning the Defective Product
to Reliefband® Technologies. Contact Reliefband®
Technologies at 877-735-2263 for return or replacement
of Defective Product.
Reliefband® Premier is indicated for nonprescription
(over-the- counter) use in relief of nausea and vomiting
associated with physician-diagnosed migraine, hangover,
anxiety, motion sickness, chemotherapy and morning
3. Adjust intensity level using the Intensity Up
Button (2) until a "tingling" sensation is felt in
the palm or middle ngers of your hand.
4. If no “tingling” is felt, increase the intensity using
the Intensity Up Button (2).
5. If "tingling" is still not felt, check the position
of the Hypoallergenic Contacts (6) and adjust
until you feel "tingling" in your palm and middle
ngers.
6. Press and hold the Power On/O/Intensity Down
Button (1) for 1.5 seconds to turn the device o.
7. If the battery charge has depleted, recharge
the device according to "Charging The Device"
instructions.
a. Once battery level registers 15% or below,
the single LED will switch from static blue
to ashing red.
b. A ashing red LED indicates the device will
require charging as soon as possible.
ENVIRONMENTAL STORAGE INFORMATION:
The device can be stored and transported in the
environmental ranges of 0°C to 27°C (32°F to
80°F) and 20% to 80% relative humidity (non-
condensing). If the device has been stored or
transported in conditions outside this range, keep
it within the normal ranges for at least 30 minutes
before using.
DISPOSAL INFORMATION
Reliefband Premier is not regulated as hazardous
goods. It can be recycled or disposed with normal
household trash.
TECHNICAL INFORMATION
Reliefband complies with:
• IEC 60601-1:2005+A1:2012/EN 60601-1:2006
Medical electrical equipment - Part 1: General
requirements for basic safety and essential
performance
• IEC 60601-1-2:2007/EN 60601-1-2:2007 Medical
electrical equipment - Part 1-2: General
requirements for safety - Collateral standard:
Electromagnetic compatibility - Requirements
and tests
TIPS FOR EFFECTIVE USE
• Reliefband® Premier can be worn on either wrist provided
tingling is felt in the palm and/or middle ngers.
• For some people, applying and rubbing in the gel twice assists
in nerve stimulation.
• Reliefband® Conductivity Gel should be re-applied every 2-3
hours during use, any time after washing, or if stimulation
decreases.
• Contact with skin for prolonged periods may contribute to skin
irritation particularly for users who have allergies. To reduce
irritation, follow three simple wear and care tips: (1) Keep it
clean; (2) keep it dry; and (3) after extended wear, give your
wrist a rest by removing the band for an hour.
• Regularly clean your Reliefband Premier and wrist—especially
if perspiring, or before reapplying conductivity gel – before
returning the Reliefband Premier to your wrist. Wipe the band
and contacts with a clean, damp wipe using water or a small
amount of rubbing alcohol. Do NOT use hand soap, body
soap, dish soap, hand sanitizers, cleaning wipes or household
cleaners which could get trapped beneath the band and
irritate skin. Always dry the band well before putting it back on.
• To remove build-up of lotions and oils (such as sunscreen,
insect repellent and moisturizers) that can be trapped beneath
the band, we recommend that you use a soap-free cleanser like
Cetaphil Gentle Skin Cleanser or Aquanil; wipe o thoroughly,
and dry well.
• While still on the wrist, turn Reliefband® Premier o periodically
to determine if you still have nausea and vomiting symptoms.
If you still have symptoms, then turn it back on immediately.
sickness from pregnancy. Reliefband is also indicated as an
adjunct to antiemetics in reducing postoperative nausea.
Reliefband® Premier may not be eective for every person
because treatment outcomes vary and are dependent
upon a myriad of patient proles and characteristics.
Reliefband® Technologies’ obligation under this Limited
Warranty is expressly limited solely and exclusively to the
replacement of the Defective Product.
THE LIMITED REPRESENTATIONS AND WARRANTIES SET
FORTH HEREIN ARE RELIEFBAND TECHNOLOGIES’ SOLE
REPRESENTATIONS AND WARRANTIES WITH RESPECT
TO THE PREMIER OR ANY OTHER SUBJECT MATTER
HEREOF AND ARE GIVEN IN LIEU OF ANY AND ALL
OTHER REPRESENTATIONS OR WARRANTIES, WHETHER
EXPRESS OR IMPLIED, INCLUDING, WITHOUT LIMITATION,
ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR
SUITABILITY FOR A PARTICULAR PURPOSE. THE REMEDIES
AFFORDED TO CUSTOMER FOR ANY BREACH OF
WARRANTY WILL BE LIMITED TO THOSE SET FORTH HEREIN
TO THE EXCLUSION OF ANY AND ALL OTHER REMEDIES.
CUSTOMER WILL NOT BE ENTITLED TO PUNITIVE,
INCIDENTAL, CONSEQUENTIAL OR SPECIAL DAMAGES,
INCLUDING WITHOUT LIMITATION CUSTOMER’S DENTAL
AND MEDICAL COSTS, AS A RESULT OF A DEFECTIVE
PRODUCT. NO AGREEMENT VARYING OR EXTENDING
THE FOREGOING WARRANTIES, REMEDIES OR THIS
LIMITATION WILL BE BINDING ON EITHER PARTY UNLESS
MEMORIALIZED IN A WRITING SIGNED BY BOTH PARTIES.
Some jurisdictions do not allow the exclusion or limitation
of special, indirect, incidental or consequential damages,
so the above limitation or exclusion may not apply to you.
INSTRUCTIONS FOR USE
CLINICALLY PROVEN RELIEF
FOR THE PREVENTION AND TREATMENT OF
NAUSEA AND VOMITING
PREMIER
RB2.0
DETAILED OPERATION TECHNICAL DATA
TROUBLESHOOTING
STORAGE AND USAGE
INFORMATION
LIMITED WARRANTY
STATEMENT
(2)
INTENSITY UP
BUTTON
(1)
POWER ON/OFF/INTENSITY
DOWN BUTTON
(3)
INTENSITY LEVEL
INDICATOR
(5)
THERAPY ON
INDICATOR
(6)
CONTACTS
YOU SHOULD FEEL A
"TINGLING" SENSATION IN THE
PALM OR MIDDLE FINGERS.
SINGLE FLASHING
RED LED
- LOW BATTERY
Reliefband® Premier produces
physiological eects as described in this
guide.
Reliefband® Premier is classied as a
TYPE BF Applied Part
Latex Free
Reliefband® Premier is splash-resistant
ATTENTION
IPX4
For European Union: Environmental Protection
Waste electrical products should not be
disposed of with household waste. Please
recycle where facilities exist. Check with
your Local Authority or retailer for recycling
advice.
California Only: Percholorate Material -
Special handling may apply. See
www.dtsc.ca.gov/hazardouswaste/perchlorate
Read these Instructions before using
Reliefband® Premier
This warranty shall not apply to any Defective Product
that has been tampered with or repaired and/or altered
by someone other than a duly-authorized Reliefband
Technologies’ representative. Further, this warranty shall
not apply to any Defective Product where the defect has
been caused by customer negligence, mishandling or
any customer use contrary to the enclosed instructions or
customer use inconsistent with the stated purpose.
This warranty is expressly limited solely to the original
purchaser and does not extend or confer any rights to
any transferee, assignee, or subsequent purchaser or user
of a Defective Product. Reliefband® Technologies’liability
for all claims, (whether based on contract, tort, breach of
warranty, or otherwise), which may arise in connection
with the purchase and use of any Defective Product
is limited to the purchase price paid by the original
purchaser.
P/N 6106980AW Rev B
• IEC 60601-2-10:2012/EN 60601-2-
10:2000+A1:2001 Medical electrical equipment
- Part 2-10: Particular requirements for the safety
of nerve and muscle stimulators
• When nished using Reliefband® Premier take it o your wrist.
Remove the gel from your wrist and the device with a tissue.
• When not using Reliefband® Premier be sure to turn the power
o to extend battery life.
• Store Reliefband® Premier in a safe, dry place. Keep it away
from young children.
• Avoid an excessive amount of gel; it could decrease the ecacy
of the device.
.com

FCC INFORMATION:
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following conditions: (1)This
device may not cause harmful interference, and (2) this
device must accept any interference received, including
interference that may cause undesired operation.
MAINTENANCE:
Keep the contacts area clean. Periodically examine the
contacts for gel build-up around the perimeter of the
contacts, which may interfere with or decrease the
delivery of therapy. A moist cloth or cotton swab may
be used to remove any gel build-up. Avoid immersion
of the device in liquid. Reliefband® Premier must not
be opened or altered in any way.
ENVIRONMENTAL USE INFORMATION:
Reliefband® Premier should only be used within
the environmental ranges of 0°C to 38°C (32°F to
100°F) and 20% to 90% relative humidity (non-
condensing).
38
1. Turn the product to ON by pressing and holding
the Power On/O/Intensity Down Button (1) for
1.5 seconds
a. Upon powering on, the Intensity Level
Indicator LEDs (3) will illuminate in
sequence of 1-10, then back down to level 0.
b. The current battery charge is displayed for
1.5 seconds before fading away.
2. Press the Intensity Up Button (2) once to set
stimulation to the rst level. At this point
therapy is being delivered. If intensity level is
not set, Reliefband Premier will turn o.
a. Blue LED intensity lights will indicate
intensity level:
i. No LEDs illuminated = Level 0 =
Power ON but no therapy
ii. 1 LED illuminated = Level 1 Therapy
iii.10 LED’s illuminated = Level 10 Therapy
b. The Therapy On Indicator (5) will pulse
on and o indicating that the device is
delivering therapy.
Size Rectangular shape – max case
dimensions: 14.5mm H x 23.5mm W x
4.1mm L, with Smart Band and Simple
Band.
Weight Approximately 50 grams
User Controls Two Push Buttons: On/O/Intensity Down
and Intensity Up
User Display Blue LED logo ( ) pulsing means device
is providing therapy. Ten diagonal blue
LED bars indicate therapy power level.
Four circular blue LED segments indicate
battery charge level.
Low Battery When battery level registers 15% or less,
a single LED segment will switch from
static blue to ashing red.
Output
Channels
Two hypoallergenic surgical grade (316L)
stainless steel contacts.
Maximum
Output
40mA
USB Charging
Cable
Cable with custom magnetic connector
Battery Rechargeable Li-Ion Polymer - IEC62133
standard. Battery life is approximately
three years or 500 charge cycles.
Waveform
Characteristics
@500 ohm load
40mA, 22.5V, 31Hz, 350us per pulse
The Troubleshooting Chart below will help to assist
you in handling situations that may arise while you
use the product.
Situation Recommended Solution
Blue intensity
LEDs and
pulsing blue
Therapy On
LEDs are
illuminated but
no stimulation is
felt
Move the contacts around the underside
of the wrist to ensure proper placement.
Increase the intensity using the Intensity
Up Button (2). Re-apply a small amount of
Reliefband Conductivity Gel. Ensure gel
is rubbed into a sheen. Be sure the strap
is secure around the wrist. Be sure blue
Intensity LEDs (3) are illuminated at one
of the ten intensity settings.
No LEDs
Illuminated
Briey press the Intensity On/O/Down
button (1) once to check if device is
powered on. If power is on, the blue
Intensity LED bars (3), the Battery/Charging
Indicator segments (4) and the Therapy On
Indicator (5) will ash. If nothing ashes,
the device may be o. Press and hold the
Intensity On/O/Down button for 1.5
seconds to wake up the device. If no LEDs
ash, the battery may be drained. To charge
the battery - see the "Charging The Battery"
section.
Flashing red,
low battery LED
illuminated
This indicates a low battery and 15% (2.4
hours at level 5) or less of charge remains.
Recharge the battery as soon as possible
– see "Charging The Battery" section
Model: Reliefband® Premier
Reliefband® Technologies warrants each new Reliefband®
Premier to be free from material defect in workmanship
and materials for (1 year) from the date of purchase, the
“Warranty Period.”This limited warranty is subject to (a)
the customer notifying Reliefband® Technologies within
1 year of the date of purchase of any material defect
in Reliefband® Premier workmanship and materials
(“Defective Product”) during the Warranty Period,
and (b) the customer returning the Defective Product
to Reliefband® Technologies. Contact Reliefband®
Technologies at 877-735-2263 for return or replacement
of Defective Product.
Reliefband® Premier is indicated for nonprescription
(over-the- counter) use in relief of nausea and vomiting
associated with physician-diagnosed migraine, hangover,
anxiety, motion sickness, chemotherapy and morning
3. Adjust intensity level using the Intensity Up
Button (2) until a "tingling" sensation is felt in
the palm or middle ngers of your hand.
4. If no “tingling” is felt, increase the intensity using
the Intensity Up Button (2).
5. If "tingling" is still not felt, check the position
of the Hypoallergenic Contacts (6) and adjust
until you feel "tingling" in your palm and middle
ngers.
6. Press and hold the Power On/O/Intensity Down
Button (1) for 1.5 seconds to turn the device o.
7. If the battery charge has depleted, recharge
the device according to "Charging The Device"
instructions.
a. Once battery level registers 15% or below,
the single LED will switch from static blue
to ashing red.
b. A ashing red LED indicates the device will
require charging as soon as possible.
ENVIRONMENTAL STORAGE INFORMATION:
The device can be stored and transported in the
environmental ranges of 0°C to 27°C (32°F to
80°F) and 20% to 80% relative humidity (non-
condensing). If the device has been stored or
transported in conditions outside this range, keep
it within the normal ranges for at least 30 minutes
before using.
DISPOSAL INFORMATION
Reliefband Premier is not regulated as hazardous
goods. It can be recycled or disposed with normal
household trash.
TECHNICAL INFORMATION
Reliefband complies with:
• IEC 60601-1:2005+A1:2012/EN 60601-1:2006
Medical electrical equipment - Part 1: General
requirements for basic safety and essential
performance
• IEC 60601-1-2:2007/EN 60601-1-2:2007 Medical
electrical equipment - Part 1-2: General
requirements for safety - Collateral standard:
Electromagnetic compatibility - Requirements
and tests
TIPS FOR EFFECTIVE USE
• Reliefband® Premier can be worn on either wrist provided
tingling is felt in the palm and/or middle ngers.
• For some people, applying and rubbing in the gel twice assists
in nerve stimulation.
• Reliefband® Conductivity Gel should be re-applied every 2-3
hours during use, any time after washing, or if stimulation
decreases.
• Contact with skin for prolonged periods may contribute to skin
irritation particularly for users who have allergies. To reduce
irritation, follow three simple wear and care tips: (1) Keep it
clean; (2) keep it dry; and (3) after extended wear, give your
wrist a rest by removing the band for an hour.
• Regularly clean your Reliefband Premier and wrist—especially
if perspiring, or before reapplying conductivity gel – before
returning the Reliefband Premier to your wrist. Wipe the band
and contacts with a clean, damp wipe using water or a small
amount of rubbing alcohol. Do NOT use hand soap, body
soap, dish soap, hand sanitizers, cleaning wipes or household
cleaners which could get trapped beneath the band and
irritate skin. Always dry the band well before putting it back on.
• To remove build-up of lotions and oils (such as sunscreen,
insect repellent and moisturizers) that can be trapped beneath
the band, we recommend that you use a soap-free cleanser like
Cetaphil Gentle Skin Cleanser or Aquanil; wipe o thoroughly,
and dry well.
• While still on the wrist, turn Reliefband® Premier o periodically
to determine if you still have nausea and vomiting symptoms.
If you still have symptoms, then turn it back on immediately.
sickness from pregnancy. Reliefband is also indicated as an
adjunct to antiemetics in reducing postoperative nausea.
Reliefband® Premier may not be eective for every person
because treatment outcomes vary and are dependent
upon a myriad of patient proles and characteristics.
Reliefband® Technologies’ obligation under this Limited
Warranty is expressly limited solely and exclusively to the
replacement of the Defective Product.
THE LIMITED REPRESENTATIONS AND WARRANTIES SET
FORTH HEREIN ARE RELIEFBAND TECHNOLOGIES’ SOLE
REPRESENTATIONS AND WARRANTIES WITH RESPECT
TO THE PREMIER OR ANY OTHER SUBJECT MATTER
HEREOF AND ARE GIVEN IN LIEU OF ANY AND ALL
OTHER REPRESENTATIONS OR WARRANTIES, WHETHER
EXPRESS OR IMPLIED, INCLUDING, WITHOUT LIMITATION,
ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR
SUITABILITY FOR A PARTICULAR PURPOSE. THE REMEDIES
AFFORDED TO CUSTOMER FOR ANY BREACH OF
WARRANTY WILL BE LIMITED TO THOSE SET FORTH HEREIN
TO THE EXCLUSION OF ANY AND ALL OTHER REMEDIES.
CUSTOMER WILL NOT BE ENTITLED TO PUNITIVE,
INCIDENTAL, CONSEQUENTIAL OR SPECIAL DAMAGES,
INCLUDING WITHOUT LIMITATION CUSTOMER’S DENTAL
AND MEDICAL COSTS, AS A RESULT OF A DEFECTIVE
PRODUCT. NO AGREEMENT VARYING OR EXTENDING
THE FOREGOING WARRANTIES, REMEDIES OR THIS
LIMITATION WILL BE BINDING ON EITHER PARTY UNLESS
MEMORIALIZED IN A WRITING SIGNED BY BOTH PARTIES.
Some jurisdictions do not allow the exclusion or limitation
of special, indirect, incidental or consequential damages,
so the above limitation or exclusion may not apply to you.
INSTRUCTIONS FOR USE
CLINICALLY PROVEN RELIEF
FOR THE PREVENTION AND TREATMENT OF
NAUSEA AND VOMITING
PREMIER
RB2.0
DETAILED OPERATION TECHNICAL DATA
TROUBLESHOOTING
STORAGE AND USAGE
INFORMATION
LIMITED WARRANTY
STATEMENT
(2)
INTENSITY UP
BUTTON
(1)
POWER ON/OFF/INTENSITY
DOWN BUTTON
(3)
INTENSITY LEVEL
INDICATOR
(5)
THERAPY ON
INDICATOR
(6)
CONTACTS
YOU SHOULD FEEL A
"TINGLING" SENSATION IN THE
PALM OR MIDDLE FINGERS.
SINGLE FLASHING
RED LED
- LOW BATTERY
Reliefband® Premier produces
physiological eects as described in this
guide.
Reliefband® Premier is classied as a
TYPE BF Applied Part
Latex Free
Reliefband® Premier is splash-resistant
ATTENTION
IPX4
For European Union: Environmental Protection
Waste electrical products should not be
disposed of with household waste. Please
recycle where facilities exist. Check with
your Local Authority or retailer for recycling
advice.
California Only: Percholorate Material -
Special handling may apply. See
www.dtsc.ca.gov/hazardouswaste/perchlorate
Read these Instructions before using
Reliefband® Premier
This warranty shall not apply to any Defective Product
that has been tampered with or repaired and/or altered
by someone other than a duly-authorized Reliefband
Technologies’ representative. Further, this warranty shall
not apply to any Defective Product where the defect has
been caused by customer negligence, mishandling or
any customer use contrary to the enclosed instructions or
customer use inconsistent with the stated purpose.
This warranty is expressly limited solely to the original
purchaser and does not extend or confer any rights to
any transferee, assignee, or subsequent purchaser or user
of a Defective Product. Reliefband® Technologies’liability
for all claims, (whether based on contract, tort, breach of
warranty, or otherwise), which may arise in connection
with the purchase and use of any Defective Product
is limited to the purchase price paid by the original
purchaser.
P/N 6106980AW Rev B
• IEC 60601-2-10:2012/EN 60601-2-
10:2000+A1:2001 Medical electrical equipment
- Part 2-10: Particular requirements for the safety
of nerve and muscle stimulators
• When nished using Reliefband® Premier take it o your wrist.
Remove the gel from your wrist and the device with a tissue.
• When not using Reliefband® Premier be sure to turn the power
o to extend battery life.
• Store Reliefband® Premier in a safe, dry place. Keep it away
from young children.
• Avoid an excessive amount of gel; it could decrease the ecacy
of the device.
.com

FCC INFORMATION:
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following conditions: (1)This
device may not cause harmful interference, and (2) this
device must accept any interference received, including
interference that may cause undesired operation.
MAINTENANCE:
Keep the contacts area clean. Periodically examine the
contacts for gel build-up around the perimeter of the
contacts, which may interfere with or decrease the
delivery of therapy. A moist cloth or cotton swab may
be used to remove any gel build-up. Avoid immersion
of the device in liquid. Reliefband® Premier must not
be opened or altered in any way.
ENVIRONMENTAL USE INFORMATION:
Reliefband® Premier should only be used within
the environmental ranges of 0°C to 38°C (32°F to
100°F) and 20% to 90% relative humidity (non-
condensing).
38
1. Turn the product to ON by pressing and holding
the Power On/O/Intensity Down Button (1) for
1.5 seconds
a. Upon powering on, the Intensity Level
Indicator LEDs (3) will illuminate in
sequence of 1-10, then back down to level 0.
b. The current battery charge is displayed for
1.5 seconds before fading away.
2. Press the Intensity Up Button (2) once to set
stimulation to the rst level. At this point
therapy is being delivered. If intensity level is
not set, Reliefband Premier will turn o.
a. Blue LED intensity lights will indicate
intensity level:
i. No LEDs illuminated = Level 0 =
Power ON but no therapy
ii. 1 LED illuminated = Level 1 Therapy
iii.10 LED’s illuminated = Level 10 Therapy
b. The Therapy On Indicator (5) will pulse
on and o indicating that the device is
delivering therapy.
Size Rectangular shape – max case
dimensions: 14.5mm H x 23.5mm W x
4.1mm L, with Smart Band and Simple
Band.
Weight Approximately 50 grams
User Controls Two Push Buttons: On/O/Intensity Down
and Intensity Up
User Display Blue LED logo ( ) pulsing means device
is providing therapy. Ten diagonal blue
LED bars indicate therapy power level.
Four circular blue LED segments indicate
battery charge level.
Low Battery When battery level registers 15% or less,
a single LED segment will switch from
static blue to ashing red.
Output
Channels
Two hypoallergenic surgical grade (316L)
stainless steel contacts.
Maximum
Output
40mA
USB Charging
Cable
Cable with custom magnetic connector
Battery Rechargeable Li-Ion Polymer - IEC62133
standard. Battery life is approximately
three years or 500 charge cycles.
Waveform
Characteristics
@500 ohm load
40mA, 22.5V, 31Hz, 350us per pulse
The Troubleshooting Chart below will help to assist
you in handling situations that may arise while you
use the product.
Situation Recommended Solution
Blue intensity
LEDs and
pulsing blue
Therapy On
LEDs are
illuminated but
no stimulation is
felt
Move the contacts around the underside
of the wrist to ensure proper placement.
Increase the intensity using the Intensity
Up Button (2). Re-apply a small amount of
Reliefband Conductivity Gel. Ensure gel
is rubbed into a sheen. Be sure the strap
is secure around the wrist. Be sure blue
Intensity LEDs (3) are illuminated at one
of the ten intensity settings.
No LEDs
Illuminated
Briey press the Intensity On/O/Down
button (1) once to check if device is
powered on. If power is on, the blue
Intensity LED bars (3), the Battery/Charging
Indicator segments (4) and the Therapy On
Indicator (5) will ash. If nothing ashes,
the device may be o. Press and hold the
Intensity On/O/Down button for 1.5
seconds to wake up the device. If no LEDs
ash, the battery may be drained. To charge
the battery - see the "Charging The Battery"
section.
Flashing red,
low battery LED
illuminated
This indicates a low battery and 15% (2.4
hours at level 5) or less of charge remains.
Recharge the battery as soon as possible
– see "Charging The Battery" section
Model: Reliefband® Premier
Reliefband® Technologies warrants each new Reliefband®
Premier to be free from material defect in workmanship
and materials for (1 year) from the date of purchase, the
“Warranty Period.”This limited warranty is subject to (a)
the customer notifying Reliefband® Technologies within
1 year of the date of purchase of any material defect
in Reliefband® Premier workmanship and materials
(“Defective Product”) during the Warranty Period,
and (b) the customer returning the Defective Product
to Reliefband® Technologies. Contact Reliefband®
Technologies at 877-735-2263 for return or replacement
of Defective Product.
Reliefband® Premier is indicated for nonprescription
(over-the- counter) use in relief of nausea and vomiting
associated with physician-diagnosed migraine, hangover,
anxiety, motion sickness, chemotherapy and morning
3. Adjust intensity level using the Intensity Up
Button (2) until a "tingling" sensation is felt in
the palm or middle ngers of your hand.
4. If no “tingling” is felt, increase the intensity using
the Intensity Up Button (2).
5. If "tingling" is still not felt, check the position
of the Hypoallergenic Contacts (6) and adjust
until you feel "tingling" in your palm and middle
ngers.
6. Press and hold the Power On/O/Intensity Down
Button (1) for 1.5 seconds to turn the device o.
7. If the battery charge has depleted, recharge
the device according to "Charging The Device"
instructions.
a. Once battery level registers 15% or below,
the single LED will switch from static blue
to ashing red.
b. A ashing red LED indicates the device will
require charging as soon as possible.
ENVIRONMENTAL STORAGE INFORMATION:
The device can be stored and transported in the
environmental ranges of 0°C to 27°C (32°F to
80°F) and 20% to 80% relative humidity (non-
condensing). If the device has been stored or
transported in conditions outside this range, keep
it within the normal ranges for at least 30 minutes
before using.
DISPOSAL INFORMATION
Reliefband Premier is not regulated as hazardous
goods. It can be recycled or disposed with normal
household trash.
TECHNICAL INFORMATION
Reliefband complies with:
• IEC 60601-1:2005+A1:2012/EN 60601-1:2006
Medical electrical equipment - Part 1: General
requirements for basic safety and essential
performance
• IEC 60601-1-2:2007/EN 60601-1-2:2007 Medical
electrical equipment - Part 1-2: General
requirements for safety - Collateral standard:
Electromagnetic compatibility - Requirements
and tests
TIPS FOR EFFECTIVE USE
• Reliefband® Premier can be worn on either wrist provided
tingling is felt in the palm and/or middle ngers.
• For some people, applying and rubbing in the gel twice assists
in nerve stimulation.
• Reliefband® Conductivity Gel should be re-applied every 2-3
hours during use, any time after washing, or if stimulation
decreases.
• Contact with skin for prolonged periods may contribute to skin
irritation particularly for users who have allergies. To reduce
irritation, follow three simple wear and care tips: (1) Keep it
clean; (2) keep it dry; and (3) after extended wear, give your
wrist a rest by removing the band for an hour.
• Regularly clean your Reliefband Premier and wrist—especially
if perspiring, or before reapplying conductivity gel – before
returning the Reliefband Premier to your wrist. Wipe the band
and contacts with a clean, damp wipe using water or a small
amount of rubbing alcohol. Do NOT use hand soap, body
soap, dish soap, hand sanitizers, cleaning wipes or household
cleaners which could get trapped beneath the band and
irritate skin. Always dry the band well before putting it back on.
• To remove build-up of lotions and oils (such as sunscreen,
insect repellent and moisturizers) that can be trapped beneath
the band, we recommend that you use a soap-free cleanser like
Cetaphil Gentle Skin Cleanser or Aquanil; wipe o thoroughly,
and dry well.
• While still on the wrist, turn Reliefband® Premier o periodically
to determine if you still have nausea and vomiting symptoms.
If you still have symptoms, then turn it back on immediately.
sickness from pregnancy. Reliefband is also indicated as an
adjunct to antiemetics in reducing postoperative nausea.
Reliefband® Premier may not be eective for every person
because treatment outcomes vary and are dependent
upon a myriad of patient proles and characteristics.
Reliefband® Technologies’ obligation under this Limited
Warranty is expressly limited solely and exclusively to the
replacement of the Defective Product.
THE LIMITED REPRESENTATIONS AND WARRANTIES SET
FORTH HEREIN ARE RELIEFBAND TECHNOLOGIES’ SOLE
REPRESENTATIONS AND WARRANTIES WITH RESPECT
TO THE PREMIER OR ANY OTHER SUBJECT MATTER
HEREOF AND ARE GIVEN IN LIEU OF ANY AND ALL
OTHER REPRESENTATIONS OR WARRANTIES, WHETHER
EXPRESS OR IMPLIED, INCLUDING, WITHOUT LIMITATION,
ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR
SUITABILITY FOR A PARTICULAR PURPOSE. THE REMEDIES
AFFORDED TO CUSTOMER FOR ANY BREACH OF
WARRANTY WILL BE LIMITED TO THOSE SET FORTH HEREIN
TO THE EXCLUSION OF ANY AND ALL OTHER REMEDIES.
CUSTOMER WILL NOT BE ENTITLED TO PUNITIVE,
INCIDENTAL, CONSEQUENTIAL OR SPECIAL DAMAGES,
INCLUDING WITHOUT LIMITATION CUSTOMER’S DENTAL
AND MEDICAL COSTS, AS A RESULT OF A DEFECTIVE
PRODUCT. NO AGREEMENT VARYING OR EXTENDING
THE FOREGOING WARRANTIES, REMEDIES OR THIS
LIMITATION WILL BE BINDING ON EITHER PARTY UNLESS
MEMORIALIZED IN A WRITING SIGNED BY BOTH PARTIES.
Some jurisdictions do not allow the exclusion or limitation
of special, indirect, incidental or consequential damages,
so the above limitation or exclusion may not apply to you.
INSTRUCTIONS FOR USE
CLINICALLY PROVEN RELIEF
FOR THE PREVENTION AND TREATMENT OF
NAUSEA AND VOMITING
PREMIER
RB2.0
DETAILED OPERATION TECHNICAL DATA
TROUBLESHOOTING
STORAGE AND USAGE
INFORMATION
LIMITED WARRANTY
STATEMENT
(2)
INTENSITY UP
BUTTON
(1)
POWER ON/OFF/INTENSITY
DOWN BUTTON
(3)
INTENSITY LEVEL
INDICATOR
(5)
THERAPY ON
INDICATOR
(6)
CONTACTS
YOU SHOULD FEEL A
"TINGLING" SENSATION IN THE
PALM OR MIDDLE FINGERS.
SINGLE FLASHING
RED LED
- LOW BATTERY
Reliefband® Premier produces
physiological eects as described in this
guide.
Reliefband® Premier is classied as a
TYPE BF Applied Part
Latex Free
Reliefband® Premier is splash-resistant
ATTENTION
IPX4
For European Union: Environmental Protection
Waste electrical products should not be
disposed of with household waste. Please
recycle where facilities exist. Check with
your Local Authority or retailer for recycling
advice.
California Only: Percholorate Material -
Special handling may apply. See
www.dtsc.ca.gov/hazardouswaste/perchlorate
Read these Instructions before using
Reliefband® Premier
This warranty shall not apply to any Defective Product
that has been tampered with or repaired and/or altered
by someone other than a duly-authorized Reliefband
Technologies’ representative. Further, this warranty shall
not apply to any Defective Product where the defect has
been caused by customer negligence, mishandling or
any customer use contrary to the enclosed instructions or
customer use inconsistent with the stated purpose.
This warranty is expressly limited solely to the original
purchaser and does not extend or confer any rights to
any transferee, assignee, or subsequent purchaser or user
of a Defective Product. Reliefband® Technologies’liability
for all claims, (whether based on contract, tort, breach of
warranty, or otherwise), which may arise in connection
with the purchase and use of any Defective Product
is limited to the purchase price paid by the original
purchaser.
P/N 6106980AW Rev B
• IEC 60601-2-10:2012/EN 60601-2-
10:2000+A1:2001 Medical electrical equipment
- Part 2-10: Particular requirements for the safety
of nerve and muscle stimulators
• When nished using Reliefband® Premier take it o your wrist.
Remove the gel from your wrist and the device with a tissue.
• When not using Reliefband® Premier be sure to turn the power
o to extend battery life.
• Store Reliefband® Premier in a safe, dry place. Keep it away
from young children.
• Avoid an excessive amount of gel; it could decrease the ecacy
of the device.
.com

FCC INFORMATION:
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following conditions: (1)This
device may not cause harmful interference, and (2) this
device must accept any interference received, including
interference that may cause undesired operation.
MAINTENANCE:
Keep the contacts area clean. Periodically examine the
contacts for gel build-up around the perimeter of the
contacts, which may interfere with or decrease the
delivery of therapy. A moist cloth or cotton swab may
be used to remove any gel build-up. Avoid immersion
of the device in liquid. Reliefband® Premier must not
be opened or altered in any way.
ENVIRONMENTAL USE INFORMATION:
Reliefband® Premier should only be used within
the environmental ranges of 0°C to 38°C (32°F to
100°F) and 20% to 90% relative humidity (non-
condensing).
38
1. Turn the product to ON by pressing and holding
the Power On/O/Intensity Down Button (1) for
1.5 seconds
a. Upon powering on, the Intensity Level
Indicator LEDs (3) will illuminate in
sequence of 1-10, then back down to level 0.
b. The current battery charge is displayed for
1.5 seconds before fading away.
2. Press the Intensity Up Button (2) once to set
stimulation to the rst level. At this point
therapy is being delivered. If intensity level is
not set, Reliefband Premier will turn o.
a. Blue LED intensity lights will indicate
intensity level:
i. No LEDs illuminated = Level 0 =
Power ON but no therapy
ii. 1 LED illuminated = Level 1 Therapy
iii.10 LED’s illuminated = Level 10 Therapy
b. The Therapy On Indicator (5) will pulse
on and o indicating that the device is
delivering therapy.
Size Rectangular shape – max case
dimensions: 14.5mm H x 23.5mm W x
4.1mm L, with Smart Band and Simple
Band.
Weight Approximately 50 grams
User Controls Two Push Buttons: On/O/Intensity Down
and Intensity Up
User Display Blue LED logo ( ) pulsing means device
is providing therapy. Ten diagonal blue
LED bars indicate therapy power level.
Four circular blue LED segments indicate
battery charge level.
Low Battery When battery level registers 15% or less,
a single LED segment will switch from
static blue to ashing red.
Output
Channels
Two hypoallergenic surgical grade (316L)
stainless steel contacts.
Maximum
Output
40mA
USB Charging
Cable
Cable with custom magnetic connector
Battery Rechargeable Li-Ion Polymer - IEC62133
standard. Battery life is approximately
three years or 500 charge cycles.
Waveform
Characteristics
@500 ohm load
40mA, 22.5V, 31Hz, 350us per pulse
The Troubleshooting Chart below will help to assist
you in handling situations that may arise while you
use the product.
Situation Recommended Solution
Blue intensity
LEDs and
pulsing blue
Therapy On
LEDs are
illuminated but
no stimulation is
felt
Move the contacts around the underside
of the wrist to ensure proper placement.
Increase the intensity using the Intensity
Up Button (2). Re-apply a small amount of
Reliefband Conductivity Gel. Ensure gel
is rubbed into a sheen. Be sure the strap
is secure around the wrist. Be sure blue
Intensity LEDs (3) are illuminated at one
of the ten intensity settings.
No LEDs
Illuminated
Briey press the Intensity On/O/Down
button (1) once to check if device is
powered on. If power is on, the blue
Intensity LED bars (3), the Battery/Charging
Indicator segments (4) and the Therapy On
Indicator (5) will ash. If nothing ashes,
the device may be o. Press and hold the
Intensity On/O/Down button for 1.5
seconds to wake up the device. If no LEDs
ash, the battery may be drained. To charge
the battery - see the "Charging The Battery"
section.
Flashing red,
low battery LED
illuminated
This indicates a low battery and 15% (2.4
hours at level 5) or less of charge remains.
Recharge the battery as soon as possible
– see "Charging The Battery" section
Model: Reliefband® Premier
Reliefband® Technologies warrants each new Reliefband®
Premier to be free from material defect in workmanship
and materials for (1 year) from the date of purchase, the
“Warranty Period.”This limited warranty is subject to (a)
the customer notifying Reliefband® Technologies within
1 year of the date of purchase of any material defect
in Reliefband® Premier workmanship and materials
(“Defective Product”) during the Warranty Period,
and (b) the customer returning the Defective Product
to Reliefband® Technologies. Contact Reliefband®
Technologies at 877-735-2263 for return or replacement
of Defective Product.
Reliefband® Premier is indicated for nonprescription
(over-the- counter) use in relief of nausea and vomiting
associated with physician-diagnosed migraine, hangover,
anxiety, motion sickness, chemotherapy and morning
3. Adjust intensity level using the Intensity Up
Button (2) until a "tingling" sensation is felt in
the palm or middle ngers of your hand.
4. If no “tingling” is felt, increase the intensity using
the Intensity Up Button (2).
5. If "tingling" is still not felt, check the position
of the Hypoallergenic Contacts (6) and adjust
until you feel "tingling" in your palm and middle
ngers.
6. Press and hold the Power On/O/Intensity Down
Button (1) for 1.5 seconds to turn the device o.
7. If the battery charge has depleted, recharge
the device according to "Charging The Device"
instructions.
a. Once battery level registers 15% or below,
the single LED will switch from static blue
to ashing red.
b. A ashing red LED indicates the device will
require charging as soon as possible.
ENVIRONMENTAL STORAGE INFORMATION:
The device can be stored and transported in the
environmental ranges of 0°C to 27°C (32°F to
80°F) and 20% to 80% relative humidity (non-
condensing). If the device has been stored or
transported in conditions outside this range, keep
it within the normal ranges for at least 30 minutes
before using.
DISPOSAL INFORMATION
Reliefband Premier is not regulated as hazardous
goods. It can be recycled or disposed with normal
household trash.
TECHNICAL INFORMATION
Reliefband complies with:
• IEC 60601-1:2005+A1:2012/EN 60601-1:2006
Medical electrical equipment - Part 1: General
requirements for basic safety and essential
performance
• IEC 60601-1-2:2007/EN 60601-1-2:2007 Medical
electrical equipment - Part 1-2: General
requirements for safety - Collateral standard:
Electromagnetic compatibility - Requirements
and tests
TIPS FOR EFFECTIVE USE
• Reliefband® Premier can be worn on either wrist provided
tingling is felt in the palm and/or middle ngers.
• For some people, applying and rubbing in the gel twice assists
in nerve stimulation.
• Reliefband® Conductivity Gel should be re-applied every 2-3
hours during use, any time after washing, or if stimulation
decreases.
• Contact with skin for prolonged periods may contribute to skin
irritation particularly for users who have allergies. To reduce
irritation, follow three simple wear and care tips: (1) Keep it
clean; (2) keep it dry; and (3) after extended wear, give your
wrist a rest by removing the band for an hour.
• Regularly clean your Reliefband Premier and wrist—especially
if perspiring, or before reapplying conductivity gel – before
returning the Reliefband Premier to your wrist. Wipe the band
and contacts with a clean, damp wipe using water or a small
amount of rubbing alcohol. Do NOT use hand soap, body
soap, dish soap, hand sanitizers, cleaning wipes or household
cleaners which could get trapped beneath the band and
irritate skin. Always dry the band well before putting it back on.
• To remove build-up of lotions and oils (such as sunscreen,
insect repellent and moisturizers) that can be trapped beneath
the band, we recommend that you use a soap-free cleanser like
Cetaphil Gentle Skin Cleanser or Aquanil; wipe o thoroughly,
and dry well.
• While still on the wrist, turn Reliefband® Premier o periodically
to determine if you still have nausea and vomiting symptoms.
If you still have symptoms, then turn it back on immediately.
sickness from pregnancy. Reliefband is also indicated as an
adjunct to antiemetics in reducing postoperative nausea.
Reliefband® Premier may not be eective for every person
because treatment outcomes vary and are dependent
upon a myriad of patient proles and characteristics.
Reliefband® Technologies’ obligation under this Limited
Warranty is expressly limited solely and exclusively to the
replacement of the Defective Product.
THE LIMITED REPRESENTATIONS AND WARRANTIES SET
FORTH HEREIN ARE RELIEFBAND TECHNOLOGIES’ SOLE
REPRESENTATIONS AND WARRANTIES WITH RESPECT
TO THE PREMIER OR ANY OTHER SUBJECT MATTER
HEREOF AND ARE GIVEN IN LIEU OF ANY AND ALL
OTHER REPRESENTATIONS OR WARRANTIES, WHETHER
EXPRESS OR IMPLIED, INCLUDING, WITHOUT LIMITATION,
ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR
SUITABILITY FOR A PARTICULAR PURPOSE. THE REMEDIES
AFFORDED TO CUSTOMER FOR ANY BREACH OF
WARRANTY WILL BE LIMITED TO THOSE SET FORTH HEREIN
TO THE EXCLUSION OF ANY AND ALL OTHER REMEDIES.
CUSTOMER WILL NOT BE ENTITLED TO PUNITIVE,
INCIDENTAL, CONSEQUENTIAL OR SPECIAL DAMAGES,
INCLUDING WITHOUT LIMITATION CUSTOMER’S DENTAL
AND MEDICAL COSTS, AS A RESULT OF A DEFECTIVE
PRODUCT. NO AGREEMENT VARYING OR EXTENDING
THE FOREGOING WARRANTIES, REMEDIES OR THIS
LIMITATION WILL BE BINDING ON EITHER PARTY UNLESS
MEMORIALIZED IN A WRITING SIGNED BY BOTH PARTIES.
Some jurisdictions do not allow the exclusion or limitation
of special, indirect, incidental or consequential damages,
so the above limitation or exclusion may not apply to you.
INSTRUCTIONS FOR USE
CLINICALLY PROVEN RELIEF
FOR THE PREVENTION AND TREATMENT OF
NAUSEA AND VOMITING
PREMIER
RB2.0
DETAILED OPERATION TECHNICAL DATA
TROUBLESHOOTING
STORAGE AND USAGE
INFORMATION
LIMITED WARRANTY
STATEMENT
(2)
INTENSITY UP
BUTTON
(1)
POWER ON/OFF/INTENSITY
DOWN BUTTON
(3)
INTENSITY LEVEL
INDICATOR
(5)
THERAPY ON
INDICATOR
(6)
CONTACTS
YOU SHOULD FEEL A
"TINGLING" SENSATION IN THE
PALM OR MIDDLE FINGERS.
SINGLE FLASHING
RED LED
- LOW BATTERY
Reliefband® Premier produces
physiological eects as described in this
guide.
Reliefband® Premier is classied as a
TYPE BF Applied Part
Latex Free
Reliefband® Premier is splash-resistant
ATTENTION
IPX4
For European Union: Environmental Protection
Waste electrical products should not be
disposed of with household waste. Please
recycle where facilities exist. Check with
your Local Authority or retailer for recycling
advice.
California Only: Percholorate Material -
Special handling may apply. See
www.dtsc.ca.gov/hazardouswaste/perchlorate
Read these Instructions before using
Reliefband® Premier
This warranty shall not apply to any Defective Product
that has been tampered with or repaired and/or altered
by someone other than a duly-authorized Reliefband
Technologies’ representative. Further, this warranty shall
not apply to any Defective Product where the defect has
been caused by customer negligence, mishandling or
any customer use contrary to the enclosed instructions or
customer use inconsistent with the stated purpose.
This warranty is expressly limited solely to the original
purchaser and does not extend or confer any rights to
any transferee, assignee, or subsequent purchaser or user
of a Defective Product. Reliefband® Technologies’liability
for all claims, (whether based on contract, tort, breach of
warranty, or otherwise), which may arise in connection
with the purchase and use of any Defective Product
is limited to the purchase price paid by the original
purchaser.
P/N 6106980AW Rev B
• IEC 60601-2-10:2012/EN 60601-2-
10:2000+A1:2001 Medical electrical equipment
- Part 2-10: Particular requirements for the safety
of nerve and muscle stimulators
• When nished using Reliefband® Premier take it o your wrist.
Remove the gel from your wrist and the device with a tissue.
• When not using Reliefband® Premier be sure to turn the power
o to extend battery life.
• Store Reliefband® Premier in a safe, dry place. Keep it away
from young children.
• Avoid an excessive amount of gel; it could decrease the ecacy
of the device.
.com

FCC INFORMATION:
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following conditions: (1)This
device may not cause harmful interference, and (2) this
device must accept any interference received, including
interference that may cause undesired operation.
MAINTENANCE:
Keep the contacts area clean. Periodically examine the
contacts for gel build-up around the perimeter of the
contacts, which may interfere with or decrease the
delivery of therapy. A moist cloth or cotton swab may
be used to remove any gel build-up. Avoid immersion
of the device in liquid. Reliefband® Premier must not
be opened or altered in any way.
ENVIRONMENTAL USE INFORMATION:
Reliefband® Premier should only be used within
the environmental ranges of 0°C to 38°C (32°F to
100°F) and 20% to 90% relative humidity (non-
condensing).
38
1. Turn the product to ON by pressing and holding
the Power On/O/Intensity Down Button (1) for
1.5 seconds
a. Upon powering on, the Intensity Level
Indicator LEDs (3) will illuminate in
sequence of 1-10, then back down to level 0.
b. The current battery charge is displayed for
1.5 seconds before fading away.
2. Press the Intensity Up Button (2) once to set
stimulation to the rst level. At this point
therapy is being delivered. If intensity level is
not set, Reliefband Premier will turn o.
a. Blue LED intensity lights will indicate
intensity level:
i. No LEDs illuminated = Level 0 =
Power ON but no therapy
ii. 1 LED illuminated = Level 1 Therapy
iii.10 LED’s illuminated = Level 10 Therapy
b. The Therapy On Indicator (5) will pulse
on and o indicating that the device is
delivering therapy.
Size Rectangular shape – max case
dimensions: 14.5mm H x 23.5mm W x
4.1mm L, with Smart Band and Simple
Band.
Weight Approximately 50 grams
User Controls Two Push Buttons: On/O/Intensity Down
and Intensity Up
User Display Blue LED logo ( ) pulsing means device
is providing therapy. Ten diagonal blue
LED bars indicate therapy power level.
Four circular blue LED segments indicate
battery charge level.
Low Battery When battery level registers 15% or less,
a single LED segment will switch from
static blue to ashing red.
Output
Channels
Two hypoallergenic surgical grade (316L)
stainless steel contacts.
Maximum
Output
40mA
USB Charging
Cable
Cable with custom magnetic connector
Battery Rechargeable Li-Ion Polymer - IEC62133
standard. Battery life is approximately
three years or 500 charge cycles.
Waveform
Characteristics
@500 ohm load
40mA, 22.5V, 31Hz, 350us per pulse
The Troubleshooting Chart below will help to assist
you in handling situations that may arise while you
use the product.
Situation Recommended Solution
Blue intensity
LEDs and
pulsing blue
Therapy On
LEDs are
illuminated but
no stimulation is
felt
Move the contacts around the underside
of the wrist to ensure proper placement.
Increase the intensity using the Intensity
Up Button (2). Re-apply a small amount of
Reliefband Conductivity Gel. Ensure gel
is rubbed into a sheen. Be sure the strap
is secure around the wrist. Be sure blue
Intensity LEDs (3) are illuminated at one
of the ten intensity settings.
No LEDs
Illuminated
Briey press the Intensity On/O/Down
button (1) once to check if device is
powered on. If power is on, the blue
Intensity LED bars (3), the Battery/Charging
Indicator segments (4) and the Therapy On
Indicator (5) will ash. If nothing ashes,
the device may be o. Press and hold the
Intensity On/O/Down button for 1.5
seconds to wake up the device. If no LEDs
ash, the battery may be drained. To charge
the battery - see the "Charging The Battery"
section.
Flashing red,
low battery LED
illuminated
This indicates a low battery and 15% (2.4
hours at level 5) or less of charge remains.
Recharge the battery as soon as possible
– see "Charging The Battery" section
Model: Reliefband® Premier
Reliefband® Technologies warrants each new Reliefband®
Premier to be free from material defect in workmanship
and materials for (1 year) from the date of purchase, the
“Warranty Period.”This limited warranty is subject to (a)
the customer notifying Reliefband® Technologies within
1 year of the date of purchase of any material defect
in Reliefband® Premier workmanship and materials
(“Defective Product”) during the Warranty Period,
and (b) the customer returning the Defective Product
to Reliefband® Technologies. Contact Reliefband®
Technologies at 877-735-2263 for return or replacement
of Defective Product.
Reliefband® Premier is indicated for nonprescription
(over-the- counter) use in relief of nausea and vomiting
associated with physician-diagnosed migraine, hangover,
anxiety, motion sickness, chemotherapy and morning
3. Adjust intensity level using the Intensity Up
Button (2) until a "tingling" sensation is felt in
the palm or middle ngers of your hand.
4. If no “tingling” is felt, increase the intensity using
the Intensity Up Button (2).
5. If "tingling" is still not felt, check the position
of the Hypoallergenic Contacts (6) and adjust
until you feel "tingling" in your palm and middle
ngers.
6. Press and hold the Power On/O/Intensity Down
Button (1) for 1.5 seconds to turn the device o.
7. If the battery charge has depleted, recharge
the device according to "Charging The Device"
instructions.
a. Once battery level registers 15% or below,
the single LED will switch from static blue
to ashing red.
b. A ashing red LED indicates the device will
require charging as soon as possible.
ENVIRONMENTAL STORAGE INFORMATION:
The device can be stored and transported in the
environmental ranges of 0°C to 27°C (32°F to
80°F) and 20% to 80% relative humidity (non-
condensing). If the device has been stored or
transported in conditions outside this range, keep
it within the normal ranges for at least 30 minutes
before using.
DISPOSAL INFORMATION
Reliefband Premier is not regulated as hazardous
goods. It can be recycled or disposed with normal
household trash.
TECHNICAL INFORMATION
Reliefband complies with:
• IEC 60601-1:2005+A1:2012/EN 60601-1:2006
Medical electrical equipment - Part 1: General
requirements for basic safety and essential
performance
• IEC 60601-1-2:2007/EN 60601-1-2:2007 Medical
electrical equipment - Part 1-2: General
requirements for safety - Collateral standard:
Electromagnetic compatibility - Requirements
and tests
TIPS FOR EFFECTIVE USE
• Reliefband® Premier can be worn on either wrist provided
tingling is felt in the palm and/or middle ngers.
• For some people, applying and rubbing in the gel twice assists
in nerve stimulation.
• Reliefband® Conductivity Gel should be re-applied every 2-3
hours during use, any time after washing, or if stimulation
decreases.
• Contact with skin for prolonged periods may contribute to skin
irritation particularly for users who have allergies. To reduce
irritation, follow three simple wear and care tips: (1) Keep it
clean; (2) keep it dry; and (3) after extended wear, give your
wrist a rest by removing the band for an hour.
• Regularly clean your Reliefband Premier and wrist—especially
if perspiring, or before reapplying conductivity gel – before
returning the Reliefband Premier to your wrist. Wipe the band
and contacts with a clean, damp wipe using water or a small
amount of rubbing alcohol. Do NOT use hand soap, body
soap, dish soap, hand sanitizers, cleaning wipes or household
cleaners which could get trapped beneath the band and
irritate skin. Always dry the band well before putting it back on.
• To remove build-up of lotions and oils (such as sunscreen,
insect repellent and moisturizers) that can be trapped beneath
the band, we recommend that you use a soap-free cleanser like
Cetaphil Gentle Skin Cleanser or Aquanil; wipe o thoroughly,
and dry well.
• While still on the wrist, turn Reliefband® Premier o periodically
to determine if you still have nausea and vomiting symptoms.
If you still have symptoms, then turn it back on immediately.
sickness from pregnancy. Reliefband is also indicated as an
adjunct to antiemetics in reducing postoperative nausea.
Reliefband® Premier may not be eective for every person
because treatment outcomes vary and are dependent
upon a myriad of patient proles and characteristics.
Reliefband® Technologies’ obligation under this Limited
Warranty is expressly limited solely and exclusively to the
replacement of the Defective Product.
THE LIMITED REPRESENTATIONS AND WARRANTIES SET
FORTH HEREIN ARE RELIEFBAND TECHNOLOGIES’ SOLE
REPRESENTATIONS AND WARRANTIES WITH RESPECT
TO THE PREMIER OR ANY OTHER SUBJECT MATTER
HEREOF AND ARE GIVEN IN LIEU OF ANY AND ALL
OTHER REPRESENTATIONS OR WARRANTIES, WHETHER
EXPRESS OR IMPLIED, INCLUDING, WITHOUT LIMITATION,
ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR
SUITABILITY FOR A PARTICULAR PURPOSE. THE REMEDIES
AFFORDED TO CUSTOMER FOR ANY BREACH OF
WARRANTY WILL BE LIMITED TO THOSE SET FORTH HEREIN
TO THE EXCLUSION OF ANY AND ALL OTHER REMEDIES.
CUSTOMER WILL NOT BE ENTITLED TO PUNITIVE,
INCIDENTAL, CONSEQUENTIAL OR SPECIAL DAMAGES,
INCLUDING WITHOUT LIMITATION CUSTOMER’S DENTAL
AND MEDICAL COSTS, AS A RESULT OF A DEFECTIVE
PRODUCT. NO AGREEMENT VARYING OR EXTENDING
THE FOREGOING WARRANTIES, REMEDIES OR THIS
LIMITATION WILL BE BINDING ON EITHER PARTY UNLESS
MEMORIALIZED IN A WRITING SIGNED BY BOTH PARTIES.
Some jurisdictions do not allow the exclusion or limitation
of special, indirect, incidental or consequential damages,
so the above limitation or exclusion may not apply to you.
INSTRUCTIONS FOR USE
CLINICALLY PROVEN RELIEF
FOR THE PREVENTION AND TREATMENT OF
NAUSEA AND VOMITING
PREMIER
RB2.0
DETAILED OPERATION TECHNICAL DATA
TROUBLESHOOTING
STORAGE AND USAGE
INFORMATION
LIMITED WARRANTY
STATEMENT
(2)
INTENSITY UP
BUTTON
(1)
POWER ON/OFF/INTENSITY
DOWN BUTTON
(3)
INTENSITY LEVEL
INDICATOR
(5)
THERAPY ON
INDICATOR
(6)
CONTACTS
YOU SHOULD FEEL A
"TINGLING" SENSATION IN THE
PALM OR MIDDLE FINGERS.
SINGLE FLASHING
RED LED
- LOW BATTERY
Reliefband® Premier produces
physiological eects as described in this
guide.
Reliefband® Premier is classied as a
TYPE BF Applied Part
Latex Free
Reliefband® Premier is splash-resistant
ATTENTION
IPX4
For European Union: Environmental Protection
Waste electrical products should not be
disposed of with household waste. Please
recycle where facilities exist. Check with
your Local Authority or retailer for recycling
advice.
California Only: Percholorate Material -
Special handling may apply. See
www.dtsc.ca.gov/hazardouswaste/perchlorate
Read these Instructions before using
Reliefband® Premier
This warranty shall not apply to any Defective Product
that has been tampered with or repaired and/or altered
by someone other than a duly-authorized Reliefband
Technologies’ representative. Further, this warranty shall
not apply to any Defective Product where the defect has
been caused by customer negligence, mishandling or
any customer use contrary to the enclosed instructions or
customer use inconsistent with the stated purpose.
This warranty is expressly limited solely to the original
purchaser and does not extend or confer any rights to
any transferee, assignee, or subsequent purchaser or user
of a Defective Product. Reliefband® Technologies’liability
for all claims, (whether based on contract, tort, breach of
warranty, or otherwise), which may arise in connection
with the purchase and use of any Defective Product
is limited to the purchase price paid by the original
purchaser.
P/N 6106980AW Rev B
• IEC 60601-2-10:2012/EN 60601-2-
10:2000+A1:2001 Medical electrical equipment
- Part 2-10: Particular requirements for the safety
of nerve and muscle stimulators
• When nished using Reliefband® Premier take it o your wrist.
Remove the gel from your wrist and the device with a tissue.
• When not using Reliefband® Premier be sure to turn the power
o to extend battery life.
• Store Reliefband® Premier in a safe, dry place. Keep it away
from young children.
• Avoid an excessive amount of gel; it could decrease the ecacy
of the device.
.com
Table of contents
Popular Personal Care Product manuals by other brands

Buki
Buki Professional Studio Nail Art manual

Kenneth Cole
Kenneth Cole Wellness Watch Instruction Manual and Warranty Booklet

Beurer
Beurer EM 32 Instructions for use

SunFX
SunFX PRO Series Training manual

Johnson & Johnson Vision
Johnson & Johnson Vision ACUVUE 2 Instruction guide

KLAPP
KLAPP Skin Analyser instruction manual