ATO M rotary installation guide
System calibration (CAL)
Calibration is an essential operation that completes readhead set-up, with the optimum incremental
and reference mark signal settings stored in the readhead’s non-volatile memory.
Before system calibration:
uClean the disc and readhead optical window
uIf reinstalling, restore factory defaults.
uEnsure AGC is switched off (readhead set-up LED is Green, Red or Orange).
uMaximise the signal strength for complete rotation around the disc. (readhead set-up LED is Green)
NOTE: CAL routine maximum speed <100 mm/s.
Step 1 – Incremental signal calibration
uInitiate calibration routine.
uThe readhead set-up LED will now periodically single-flash Blue to indicate that it is in the incremental
signal calibration routine. The set-up signal (VX) will be a nominal 0 V.
uRotate the readhead around the disc, ensuring you do not pass a reference mark, until the readhead
set-up LED starts double-flashing Blue. This indicates the incremental signal is now calibrated and
the new settings are stored in the readhead memory. The set-up signal (VX) will be a nominal 1.65 V.
uThe system is now ready for reference mark phasing.
uFor systems without reference mark, go to ‘Calibration routine - manual exit’
uIf the system does not automatically enter the reference mark phasing stage (no double-flashing Blue
of the readhead set-up LED) the calibration of the incremental signals has failed. After ensuring failure
is not due to overspeed (>100 mm/s), exit the calibration routine, restore factory defaults and check the
readhead installation and system cleanliness before repeating the calibration routine.
Ti interface
Press the CAL button
on the end of the
interface for
<2 seconds
using a
2 mm
allen key or
similar tool.
WARNING! Activating the CAL
switch only requires 2.5 N force.
Applying excess force may
permanently damage the switch.
No interface
Ground the
CAL pin for
<2 seconds
ACi interface
Connect the CAL pads
together or ground the
remote CAL line (pin 8)
for <2 seconds
Ri interface
Press the CAL button
on the side of the
interface for
<2 seconds
using a
2 mm
allen key or
similar tool.
WARNING! Activating the CAL
switch only requires 2.5 N force.
Applying excess force may
permanently damage the switch.
CAL
button
2 CAL
pads
Step 2 – Reference mark phasing
uRotate the readhead back and forth over the reference mark until the readhead set-up LED stops flashing
and remains Green. The reference mark is now phased. The set-up signal (VX) will be a nominal 3.3 V
depending upon system set-up (see ‘Interface output specifications’ for more details).
uThe system automatically exits the CAL routine and is ready for operation.
uIf the readhead set-up LED continues double-flashing Blue after passing the reference mark many times,
it is not detecting the reference mark. Ensure that the readhead orientation and lateral offset are correct.
Calibration routine – manual exit
uThe calibration routine can be exited at any stage. Depending upon the interface used follow the relevant
section in the table below to exit CAL mode.
uOn successful exit the readhead set-up LED will stop flashing Blue and remain Green or Red.
Restoring factory defaults
When realigning the readhead, reinstalling the system, or in the case of continued calibration failure,
factory defaults should be restored.
To restore factory defaults:
uSwitch system off, then switch on again using the following methods depending on interface used.
uThe readhead set-up LED will flash Blue four times on switch on.
uRelease CAL button, CAL pads link or CAL pin ground.
uCheck the ‘Readhead mounting/installation’ and recalibrate the system.
NOTE: System must be recalibrated after restoring factory defaults.
Switching Automatic Gain Control (AGC) on or off
AGC can be switched on or off via the interface or the CAL line.
uThe readhead set-up LED will be Green with the addition of Blue when AGC is active.
NOTE: The system must be calibrated before switching AGC on.
No interface
Ground and hold the
CAL pin whilst switching
the system on.
ACi interface
Connect the CAL pads
together or ground the
remote CAL line (pin 8)
whilst switching the
system on.
Ri interface
Press and hold the CAL
button on the side of the
interface whilst switching
the system on
Ti interface
Press and hold the
CAL button on the end
of the interface whilst
switching the system on
No interface
Ground the CAL pin for
<2 seconds.
ACi interface
Connect the CAL pads
together or ground the
remote CAL line (pin 8) for
<2 seconds.
Ri interface
Press and hold the CAL
button on the side of the
interface for <2 seconds.
Ti interface
Press the CAL button on
the end of the interface for
<2 seconds.
No interface
Ground the CAL pin for
>3 seconds then remove
ground link
Ri interface
Press and hold the CAL
button on the side of the
interface for >3 seconds
then release
ACi interface
Connect the CAL pads
together or ground the
remote CAL line (pin 8) for
>3 seconds then disconnect
Ti interface
Press and hold the CAL
button on the end of the
interface for >3 seconds
then release