Responsive Respiratory Purify O3 Elite Series User manual

PORTABLE OZONE SANITIZER & DEODORIZER
USER MANUAL
TABLE OF CONTENTS:
Responsive Respiratory, Inc.
6244 Lemay Ferry Rd. • St. Louis, MO 63129
866-333-4030 • 866-333-4035 F
www.purifyo3.com • www.respondo2.com
Warnings & Specications 1
Parts & Components 2
Prior to Use, Important Notes, & Warranty 3
Charging the Battery 4
Mask, Tubing, Tees & Parts Sanitation & Deodorizing 5-7
CPAP/BiPAP Humidication Chamber Sanitation 8-9
Frequently Asked Questions 10
Para instrucciones en español - reérase a las páginas 11-21

Label Statement:
This device complies with Part 15 of the FCC Rules. Operaon is subject to the
following two condions: (1) this device may not cause harmful interference,
and (2) this device must accept any interference received, including interference
that may cause undesired operaon.
NOTE:
Please note that changes or modicaons not expressly approved by the party
responsible for compliance could void the user's authority to operate the
equipment.
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed
to provide reasonable protecon against harmful interference in a residenal
installaon. This equipment generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the instrucons, may
cause harmful interference to radio communicaons. However, there is no
guarantee that interference will not occur in a parcular installaon. If this
equipment does cause harmful interference to radio or television recepon,
which can be determined by turning the equipment o and on, the user is
encouraged to try to correct the interference by one or more of the following
measures: --Reorient or relocate the receiving antenna. --Increase the
separaon between the equipment and receiver. -- Connect the equipment into
an outlet on a circuit dierent from that to which the receiver is connected.
--Consult the dealer or an experienced radio/TV technician for help.
CALIFORNIA PROPOSITION 65 WARNING
WARNING: This product may expose you to chemicals known to the State of
California to cause cancer and birth defects or other reproducve harm.
ADVERTENCIA: Este producto puede exponerlo a sustancias químicas que el
Estado de Califonia reconoce como causantes de cáncer y defectos de nacimiento
u ostros daños reproducvos.

1
Thank you for purchasing Responsive Respiratory’s Purify O3 Elite Series
Ozone Sanitizer and Deodorizer. The Purify O3 Elite utilizes ozone technology
to sanitize and eliminate odors quickly and eectively without water, soap or
added chemicals. With proper use, the Purify O3 Elite Sanitizer is proven to
eectively kill 99.99% of germs and bacteria.
WARNINGS:
Read user manual carefully prior to operation for troubleshooting, maintenance
and safety information. Save for future reference.
Keep away from water. This device contains electronic components.
Use only in a well ventilated area.
Do not use near open ame, gas vapors or other combustible items.
Keep out of reach of children.
Do not breathe or ingest ozone release.
Allow the sanitizing case or container to remain closed, undisturbed for at least
90 minutes after the sanitation cycle is complete, to allow for the dissipation of
remaining ozone. Do not use equipment for at least 90 minutes after sanitation
cycle is nished.
SYMBOLS:
This device complies with part 15 of FCC rules. Operation is subject to the
following two conditions: (1) this device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that
may cause undesired operation.
Waste Electrical and Electronic Equipment (WEEE). It is the responsibility
of the user to dispose of this equipment at a designated collection
point.
Consult Manual
Class 3 Electronic Appliance
III
SPECIFICATIONS:
Model: Purify O3 Elite Series 190-6100 (XD200)
Size: 5.75” L x 2.5”W x 1.5” H
Weight: 0.49 lbs
Power Source: 3.7V, 2000mAh Rechargeable Lithium-ion battery
Input: AC100-240V, 50/60Hz
Rated Power: <- 15VA
Operating Temperature: 23° F ~ 113° F / -20° C ~ 55° C
Storage Temperature: 23° F ~ 113° F
FDA Registered

2
PARTS INCLUDED:
• Purify O3 Elite sanitizer with rechargeable battery
• USB cable and charging adapter
• Tubing adapter
• Dual purpose sanitizing/travel case
• User Manual
Charging Adapter
#190-6004
USB Cable
#190-6103
Dual Purpose Sanitizing/Travel Case
#190-6002
Tubing Adapter
#190-6005
Purify O3 Elite Sanitizer
Replacement component parts available for purchase separately, contact your local
retailer for availability and pricing.
USB Charging Port
Tubing Port
On/O Touch Sensor
Battery Life Indicator
Disinfection Progress Ring
SANITIZER COMPONENTS:

3
PRIOR TO USE:
Prior to use, please ensure:
• Electrical devices are turned OFF
• All water is emptied from the CPAP/BiPAP humidication chamber
• Visible residue on masks, nasal pillows or accessories is wiped clean using a
cleansing wipe or damp cloth
• Accessories to be sanitized are disconnected/removed from the device
For optimal sanitation, place both the Purify O3 Elite and the accessories inside
the sanitizing/travel case and zip closed. A plastic storage tote (12 qts or less)
with tight tting lid may also be used in place of the zippered sanitizing case to
disinfect a CPAP/BiPAP humidication chamber and/or mask/tubing assembly.
Always use the Purify O3 Elite sanitizer in a well ventilated area. Avoid breathing
in or ingesting ozone release. Allow Purify O3 Elite to sit for 90 minutes
following disinfection cycle to allow for residual ozone to dissipate.
IMPORTANT NOTES:
For health and safety reasons, do not use your equipment for at least 90
minutes after sanitizing to allow the residual ozone to dissipate back to
oxygen.
Ensure the Purify O3 Elite is operated in a well ventilated area. Do not
breathe or ingest the ozone release during/after operation.
Warranty is VOID if repairs/modications to the device are made by
unauthorized persons. To obtain service within the warranty period, please
contact the retailer where unit was purchased.
Purify O3 Elite is intended for use to sanitize and deoderize parts and
accessories. Please refer to your machine’s user manual for guidance on
cleaning and disinfection of PAP and nebulizer machines.
LIMITED WARRANTY:
The Purify O3 Elite Series sanitizer is covered by a limited warranty for a period
of two (2) years from date of purchase and covers defects in the product
resulting from faults in the material or workmanship. Damage due to improper
use, misuse or normal wear is not covered.
Please register your device online: www.respondo2.com/warranty and attach
your receipt to this booklet.
Warranty service and/or replacement is at the discretion of Responsive
Respiratory. For warranty requests, please contact your retailer or health care
dealer where Purify O3 Elite was purchased.

4
CHARGING THE BATTERY:
The Purify O3 Elite sanitizer contains a rechargeable Lithium-ion battery. For
best performance and battery life, run the sanitizer 4-5 disinfection cycles or
until the battery is fully drained before recharging. Please ensure the sanitizer is
powered o while charging.
To charge the battery, connect the USB cable to the charging adapter. Plug
the USB cable into the charging port on the side of the sanitizer as shown in
Figure 1 below. Then plug the charging adapter into an electrical outlet. During
the charging process, the Battery Life Indicator will illuminate with a series of
blinking white indicator lights. When the Purify O3 Elite is fully charged, the
Battery Life Indicator lights will change to solid white.
Charging time averages 2-3 hours and may vary depending on the battery’s
remaining capacity. If the sanitizer stops working during use after a full charge,
the battery may have reached the end of its serviceable life.
LOW POWER WARNING:
When low battery levels are reached, the Purify O3 Elite will alert the user with
both visual and auditory alerts. The Battery Life Indicator lights will blink rapidly
and a long buzz will sound. If in use, the sanitizer will attempt to continue
sanitizing for the remainder of the available battery life. It is recommended to
charge the sanitizer as soon as possible after the low power warning.
Figure 1

5
SANITIZING & DEODERIZING MASKS, TUBING & PARTS:
1. Disconnect masks, tubing, Tees and accessories from machine.
2. Using a cleansing wipe or damp, clean cloth, lightly surface clean mask,
nasal pillow or other surfaces of visible residue.
3. Place Purify O3 Elite sanitizer and accessories to be sanitized/deodorized
inside the supplied sanitizing/travel case. (For best results, connect
CPAP tubing directly to the Purify O3 Elite as shown in Figure 2.) If the
accessories are too large for the sanitizing case, a medium-sized storage
container with tightly tting lid (12 qts or less) may be used, refer to
Figure 3 below.
NOTE: Purify O3 Elite is intended for use to sanitize and deodorize accessory
items only. Please refer to your user manual for guidance on cleaning
and disinfection of the PAP or nebulizer machine.
WARNING:
Operate only in a well ventilated area.
Do not breathe or ingest the ozone.
Figure 2
Figure 3
CPAP Mask shown attached to Purify O3 Elite

6
4. To activate the Purify O3 Elite, place and hold nger on the On/O
Touch Sensor for 2 seconds as shown in Figure 4. The Purify O3 Elite will
beep once and the Disinfection Progress Ring shown in Figure 5 will
illuminate to indicate the disinfection process has begun. The Battery
Life Indicator will also light to show the estimated battery life remaining.
NOTE: The On/O Touch Sensor is activated by covering the button - it will not
depress if pushed.
5. Zip case fully closed (or secure lid, if utilizing container). The sanitizing/
deodorizing cycle for Purify O3 Elite is approximately 35 minutes in
length. During the cycle, the Progress Indicator Ring will darken to
indicate when the cycle is one fourth, halfway, and three-fourths
complete as shown in Figure 6. At the conclusion of disinfection, the
sanitizer will beep once and power o.
The sanitizer and accessories must remain in the closed case/container
for an additional 90 minutes to allow the residual ozone to dissipate.
Before use, run your PAP or nebulizer machine for 20-60 seconds with
mask and tubing attached to ush fresh air through accessories.
Your masks, tubing, Tees and accessories may be sanitized daily, as
desired.
Figure 6
Figure 4 Figure 5
On/O Touch Sensor Disinfection Progress Ring
1/4 cycle 1/2 cycle 3/4 cycle

7
NOTE: To stop the sanitation process mid-cycle, place nger on the On/O
Touch Sensor for 4 seconds. The sanitizer will signal the cancellation
with a long beep and power o. Reseal case/container and let sanitizer
and accessories sit in closed case/container for at least 90 minutes to
allow residual ozone to dissipate before use. Stopping a sanitizing cycle
before completion is not recommended.
IMPORTANT:
Do not use your mask, tubing or accessories for
at least 90 minutes after sanitation is complete to
allow remaining ozone to dissipate.

8
SANITIZING CPAP/BIPAP HUMIDIFICATION
CHAMBER & TUBING :
1. Ensure CPAP/BiPAP machine is turned o.
2. Empty water from humidication chamber.
3. Place humidication chamber and/or tubing inside the supplied
sanitizing/travel case. If the tubing and humidication chamber are too
large for the sanitizing case, a medium-sized storage container with
tightly tting lid (12 qts or less) may be used, refer to Figure 7 below.
NOTE: Purify O3 Elite is intended for use to sanitize CPAP/BiPAP accessories.
Please refer to your CPAP/BiPAP machine’s user manual for guidance on
cleaning and disinfection of the PAP machine.
4. To activate the Purify O3 Elite, place and hold nger on the On/O
Touch Sensor for 2 seconds as shown in Figure 8. The Purify O3 Elite will
beep once and the Disinfection Progress Ring shown in Figure 9 will
illuminate to indicate the disinfection process has begun. The Battery
Life Indicator will also light to show the estimated battery life remaining.
NOTE: The On/O Touch Sensor is activated by covering the button - it will not
depress if pushed.
Figure 7
Figure 8 Figure 9
On/O Touch Sensor Disinfection Progress Ring
CPAP humidication chamber shown with Purify O3 Elite

9
5. Zip case fully closed (or secure lid, if utilizing container). The sanitizing
cycle for Purify O3 Elite is approximately 35 minutes in length. During the
cycle, the Progress Indicator Ring will darken to indicate when the cycle
is one fourth, halfway, and three-fourths complete as shown in Figure 10.
At the conclusion of disinfection, the sanitizer will beep once and power
o.
The sanitizer and accessories must remain in the closed case/container
for an additional 90 minutes to allow the residual ozone to dissipate.
Before use, run your PAP machine for 20-60 seconds with mask and
tubing attached to ush fresh air through accessories.
Your CPAP/BiPAP humidication chamber may be sanitized weekly, as
desired.
NOTE: To stop the sanitation process mid-cycle, place nger on the On/O
Touch Sensor for 4 seconds. The sanitizer will signal the cancellation with
a long beep and power o. Reseal case/container and let sanitizer and
accessories sit in closed case/container for at least 90 minutes to allow
residual ozone to dissipate before use. Stopping a sanitizing cycle before
completion is not recommended.
IMPORTANT:
Do not use your CPAP/BiPAP accessories for at least
90 minutes after sanitation is complete to allow
remaining ozone to dissipate.
WARNING:
Operate only in a well ventilated area.
Do not breathe or ingest the ozone.
Figure 10
1/4 cycle 1/2 cycle 3/4 cycle

10
FREQUENTLY ASKED QUESTIONS:
The disinfection process leaves an odor:
- Before using, run your CPAP/BiPAP or nebulizer machine for 20-60 seconds to
allow any residual odor to ush from the device.
- Run the Purify O3 Elite disinfection cycle earlier in the day to allow time for
the residual odor to dissipate before use.
Charging indicator light does not turn on:
- Check the charging adapter and wire connection. Ensure battery has been
charged for at least one hour.
Purify O3 Elite does not turn on after charging:
- Ensure the On/O Touch Sensor is fully covered for 2 full seconds to power on.
- Connect the Purify O3 Elite to the charger and plug into an electrical outlet.
Cover the On/O Touch Sensor for 2 seconds to power on. Disconnect from
outlet and use as normal.
Battery is fully charged but Purify O3 Elite sanitizer is not working:
- Contact the retailer where sanitizer was purchased for warranty evaluation.
The Purify O3 Elite stops functioning:
- Battery may be depleted. Connect charger to sanitizer and plug into electrical
outlet to charge.
Purify O3 Elite stops during use after a full charge:
- Battery may have reached the end of its serviceable life. Charge sanitizer fully
and attempt to run a complete sanitizing cycle.
- If the problem persists and sanitizer is less than 2 years from date of purchase,
contact the retailer the sanitizer was purchased from for warranty evaluation.
If over 2 years from purchase, dispose of the product per electronic recycling
guidelines in your area.
Purify O3 Elite does not power o automatically after the sanitation cycle:
- Power the sanitizer o by covering the On/O Touch Sensor for 4 seconds.
- If the problem persists and sanitizer is less than 2 years from date of purchase,
contact the retailer the sanitizer was purchased from for warranty evaluation.
If over 2 years from purchase, dispose of the product per electronic recycling
guidelines in your area.
No ozone is generated:
- Verify the sanitizer has adequate battery life. Connect charger to sanitizer and
plug into electrical outlet. Power sanitizer on and conrm fan is working. If fan
is inoperable, contact the retailer where sanitizer was purchased for warranty
evaluation.
Sanitizing multiple accessories / components in one cycle:
- A medium-sized storage container with lid (12 qts or less) may be used to
sanitize masks, tubing and water humidication chamber in one cycle to
save time.
- Follow sanitizing directions as listed in instruction section of this booklet.
Sanitizing CPAP/BiPAP or nebulizer machines:
- Purify O3 Elite is intended for use to sanitize equipment and accessories only.
- Please refer to your CPAP/BiPAP or nebulizer machine’s user manual for
guidance on cleaning and disinfection of the PAP machine.

11
DESINFECTADOR PORTATIL DE OZONO &
DESINFECTANTE PARA EQUIPOS CPAP/BiPAP
MANUAL DE USO
Gracias por comprar la serie Purify O3 Elite de Responsive Respiratory
Desinfectante de ozono CPAP/BiPAP. Purify O3 Elite utiliza tecnología de ozono
para desinfectar equipos CPAP/BiPAP de forma rápida y efectiva sin agua, jabón
ni productos químicos añadidos. Con el uso adecuado, se ha demostrado que el
desinfectante Purify O3 Elite elimina ecazmente el 99.99% de los gérmenes y
bacterias.
PRECAUCIONES:
Lea cuidadosamente el manual de uso antes de utilizar el equipo, para
solución de problemas, información de mantenimiento y seguridad.
Guárdelo para futuras referencias.
Manténgalo alejado de agua. Este equipo tiene componentes electrónicos.
Uselo solamente en área bien ventilada.
No lo use cerca de llamas, vapor de gas u otros combustibles.
Manténgalo fuera del alcance de los niños.
No respire o ingiera ozono liberado por el equipo.
Permita que la estuche o contenedor de desinfección permanezca cerrado,
sin molestias durante al menos 90 minutos después de que se complete
el ciclo de desinfección, para permitir la disipación del ozono restante. No
utilice accesorios CPAP/BiPAP durante al menos 90 minutos después de
que nalice el ciclo de saneamiento.
ESPECIFICACIONES:
Model: Purify O3 Elite Series 190-6100 (XD200)
Tamaño: 5.75” L x 2.5”W x 1.5” H
Peso: 0.49 lbs
Energía Eléctrica: 3.7V, 2000mAh Batería recargable de ión-litio
Input: AC100-240V, 50/60Hz
Rated Power: <- 15VA
Temperatura de funcionamiento: 23° F ~ 113° F / -20° C ~ 55° C
Storage Temperature: 23° F ~ 113° F
FDA Registered

12
Adaptador de carga
#190-6004
Cable USB
#190-6103
Estuche de doble uso disenfector/viaje
#190-6002
Adaptador de tubo
#190-6005
Purify O3 Elite Unidad Desinfectante
Para comprar piezas de repuesto disponibles por separado, contacte a su vendedor
minorista local para disponibilidad y precios.
Entrada de carga USB
Entrada del tubo
El sensor táctil de
encendido/apagado
Indicador de carga
Anillo de progreso de desinfección
COMPONENTES DE LA UNIDAD:
PARTES INCLUIDAS:
• Purify O3 Elite unidad desinfectante con batería recargable
• Cable USB y adaptador de carga
• Adaptador de tubo
• Estuche de doble uso disenfector/viaje
• Manual de uso

13
ANTES DE USAR:
Antes de usar, por favor asegúrese que:
• El equipo CPAP/BiPAP esté apagado
• Toda el agua haya sido vaciada de la cámara humicadora
• Cualquier residuo visible en la máscara/almohadilla nasal sea limpiado
usando una toallita de limpieza o un paño húmedo
• Los accesorios a desinfectar se desconectan/retiran del máquina CPAP/BiPAP
Para un saneamiento óptimo de los accesorios CPAP/BiPAP, coloque los
accesorios CPAP/BiPAP y el Purify O3 Elite dentro del estuche de desinfección/
viaje y cierre con cremallera. También se puede usar una recipiente de plástico
(12 cuartos o menos) con tapa en lugar del estuche de desinfección para
desinfectar la cámara de humidicación o el conjunto de máscara/tubo.
Utilice siempre el desinfectante Purify O3 Elite en un área bien ventilada. Evite
respirar o ingerir la liberación de ozono. Permita que Purify O3 Elite repose
durante 90 minutos después del ciclo de desinfección para permitir que se
disipe el ozono residual.
NOTAS IMPORTANTES:
Por razones de salud y seguridad, no use su máquina CPAP/BiPAP y accesorios
hasta por lo menos depues de pasados 90 minutos de desinfección para permitir
que el ozono producido se dicipe de nuevo al oxígeno.
Asegúrese que el Purify O3 sea operado en un área bien ventilada. No respire o
ingiera el ozono disipado durante/después de la operación.
La garantía será ANULADA si reparaciones/modicaciones han sido realizadas por
personas no autorizadas. Para obetener servicio durante el período de garantía,
comuníquese con el vendedor donde compro la unidad.
Purify O3 Elite está diseñado para desinfectar los accesorios CPAP/BiPAP. Consulte
el manual de usuario de su máquina CPAP/BiPAP para obtener orientación sobre la
limpieza y desinfección de la máquina PAP.
LIMITE DE LA GARANTIA:
El Purify O3 Elite está cubierto por una garantía limitada de 2 años a partir de
la fecha de compra y cubre defectos por el material usado o de la fabricación.
Daños por uso impropio, mal uso o desgaste, no están cubiertos por la garantía.
Por favor registre su equipo en linea: www.respondo2.com/warranty
y adjunte su recibo a éste folleto.
El servicio y/o reemplazo de la garantía queda a discreción de Responsive
Respiratory. Para solicitudes de garantía, comuníquese con su vendedor o
distribuidor médica donde adquirió Purify O3 Elite.

14
CARGANDO LA BATERÍA:
La unidad desinfectadora Purify O3 Elite contiene una batería de ión-litio
recargable. Antes de recargar la batería, para un mejor rendimiento y vida útil
de la batería, use el desinfectante 4-5 ciclos de desinfección, o hasta que la
batería esté completamente descargada. Asegúrese de que el desinfectante
esté apagado mientras se carga.
Para cargar la batería, conecte el cable USB al adaptador de carga. Conecte
el cable a la entrada del USB a un lado de la unidad como muestra la Figura 1
debajo. Luego, conecte el adaptador de carga a una toma de corriente. Durante
el proceso de carga, el indicador de duración de la batería se iluminará con
una serie de luces indicadoras blancas parpadeantes. Cuando el Purify O3 Elite
está completamente cargado, las luces indicadoras de duración de la batería
cambiarán a blanco sólido.
El tiempo promedio de carga es de 2-3 horas y puede variar dependiendo en
la capacidad restante en la batería. Si la unidad deja de funcionar mientras
está en uso depués de haber cargado completamente la batería, puede haber
llegado al n de su vida útil.
ADVERTENCIA DE BAJA POTENCIA:
Cuando se alcanzan niveles bajos de batería, Purify O3 Elite alertará al usuario
con alertas visuales y auditivas. Las luces indicadoras de duración de la batería
parpadearán rápidamente y sonará un sonido prolongado. Si está en uso, el
desinfectante intentará continuar desinfectando por el resto de la vida útil de
la batería disponible. Se recomienda cargar el desinfectante lo antes posible
después de la advertencia de baja potencia.
Figura 1

15
DESINFECCIÓN MÁSCARA/TUBERÍA CPAP/BiPAP:
1. Desconecte la máscara y la tuberia del máquina CPAP/BiPAP.
2. Usando una toallita de limpieza o un paño húmedo, limpie ligeramente
la máscara/ almohadilla nasal de cualquier residuo.
3. Coloque la máscara, el tubo y el desinfectante Purify O3 Elite dentro
del estuche de desinfección / viaje suministrado. (Para obtener los
mejores resultados, conecte el tubo directamente al Purify O3 Elite
como se muestra en la Figura 2.) Si el conjunto del tubo y la máscara
es demasiado grande para la caja de desinfección, un contenedor de
almacenamiento tamaño mediano con tapa sólida (12 cuartos o menos)
puede usarse, consulte la Figura 3 a continuación. Para el saneamiento
solo con máscara, coloque la máscara y Purify O3 Elite dentro de la caja
de desinfección/almacenamiento.
NOTA: Purify O3 Elite está diseñado para desinfectar los accesorios CPAP/BiPAP.
Consulte el manual de usuario de su máquina CPAP/BiPAP para obtener
orientación sobre la limpieza y desinfección de la máquina PAP.
ADVERTENCIA:
Operar solo en un área bien
ventilada. No respire ni ingiera el ozono.
Figura 2
Figura 3

16
4. Para activar el Purify O3 Elite, coloque y mantenga el dedo en el sensor
táctil de encendido/apagado durante 2 segundos como se muestra en la
Figura 4. El Purify O3 Elite emitirá un pitido una vez y el anillo de progreso
de desinfección que se muestra en la Figura 5 se iluminará para indicar
el proceso de desinfección ha comenzado. El indicador de duración de la
batería también se iluminará para mostrar la duración restante estimada
de la batería.
NOTA: El sensor táctil de encendido / apagado se activa al cubrir el botón; no se
presionará si se presiona.
5. Cierre la estuche completamente cerrada (o asegure la tapa, si
utiliza contenedor). El ciclo de desinfección para Purify O3 Elite es de
aproximadamente 35 minutos de duración. Durante el ciclo, el Anillo
indicador de progreso se oscurecerá para indicar cuándo el ciclo está
completo en un cuarto, a la mitad y en tres cuartos, como se muestra en
la Figura 6. Al nalizar la desinfección, el desinfectante emitirá un sonido
y se apagará.
El desinfectante y los accesorios deben permanecer en la caja/recipiente
cerrado durante 90 minutos adicionales para permitir que el ozono
residual se disipe. Antes de usar, ejecute su máquina PAP durante 20-60
segundos con la máscara y el tubo conectados para purgar el aire fresco
a través de los accesorios.
Su máscara puede ser desinfectada diariamente, según lo deseado.
Figura 4 Figura 5
El sensor táctil de encendido/apagado Anillo de progreso de desinfección
Figura 6
1/4 ciclo 1/2 ciclo 3/4 ciclo

17
IMPORTANTE:
No use la máquina CPAP/BiPAP
hasta por lo menos 90 minutos después de que
el proceso de desinfección haya terminado para
permitir que el residuo de ozono se disipe.
NOTA: Para detener el proceso de saneamiento a mitad del ciclo, coloque el
dedo en el sensor táctil de encendido/apagado durante 4 segundos.
El desinfectante indicará la cancelación con un pitido largo y apagado.
Vuelva a sellar la caja/recipiente y deje que el desinfectante y los
accesorios se asienten en la caja/recipiente cerrado durante al menos 90
minutos para permitir que el ozono residual se disipe antes de usar. No
se recomienda detener un ciclo de desinfección antes de completarlo.

18
DESINFECCIÓN DE CÁMARA DE
HUMIDIFICACIÓN Y TUBERÍA CPAP/BIPAP:
1. Asegúrese que la máquina CPAP/BiPAP esté apagada y desconectada.
2. Vacíe el agua de la cámara de humidicación.
3. Coloque la cámara y/o tubo de humidicación dentro del estuche de
desinfección/viaje suministrado. Si el tubo y la cámara de humidicación
son demasiado grandes para la caja de desinfección, se puede usar un
recipiente de almacenamiento de tamaño mediano con tapa sólida (12
cuartos o menos), consulte la Figura 7 a continuación.
NOTA: Purify O3 Elite está diseñado para desinfectar los accesorios CPAP/BiPAP.
Consulte el manual de usuario de su máquina CPAP/BiPAP para obtener
orientación sobre la limpieza y desinfección de la máquina PAP.
4. Para activar el Purify O3 Elite, coloque y mantenga el dedo en el sensor
táctil de encendido/apagado durante 2 segundos como se muestra
en la Figura 8. El Purify O3 Elite emitirá un pitido una vez y el anillo de
progreso de desinfección que se muestra en la Figura 9 se iluminará
para indicar que El proceso de desinfección ha comenzado. El indicador
de duración de la batería también se iluminará para mostrar la duración
restante estimada de la batería.
NOTA: El sensor táctil de encendido/apagado se activa al cubrir el botón; no se
presionará si se presiona.
Figura 7
Figura 8 Figura 9
El sensor táctil de encendido/apagado Anillo de progreso de desinfección
Other manuals for Purify O3 Elite Series
1
This manual suits for next models
1
Table of contents
Languages:
Other Responsive Respiratory Accessories manuals