RGB Omicrom FT User manual

Portable Vital Signs Monitor
Omicrom FT
Service Manual
RGB Medical Devices, S.A.
C/ Alfonso Gómez, 42 - 28037 MADRID (SPAIN)
Cat. I101-7305 - Rev. 21 September 1998

SALES AND SERVICE
SPAIN: RGB Medical Devices, S.A.
c/ Alfonso Gomez, 42
28037 MADRID
SPAIN
Phone No ++ 34 91 304 20 11
Fax No ++ 34 91 327 39 03
YOUR LOCAL
CONTACT
FOR SALES AND
SERVICE OF
RGB MEDICAL DEVICES
PRODUCTS:
TO BE ENTERED BY LOCAL COMPANY OR AGENT
COPYRIGHT ©
RGB MEDICAL
DEVICES, S.A.
1998
All rights reserved. The information contained in this
publication may not be used for any purpose other than that for
which it was originally supplied. The publication may not be
reproduced in parts or in whole without the written consent of
RGB MEDICAL DEVICES.
In order to maintain and improve standards of manufacturing,
methods of functioning and to increase reliability, RGB
MEDICAL DEVICES equipments are periodically reviewed. For
this reason, the contents of this publication are subject to
change without notice.


OMICROM FT
SERVICE MANUAL
21 September 1998 i
LIST OF CONTENTS
LIST OF CONTENTS.............................................................................. i
1. INTRODUCTION............................................................................ 1.1
1.1. Quality, Reliability and Safety. ................................................................1.1
1.2. General Precautions.................................................................................1.2
2. SERVICE INFORMATION............................................................. 2.1
2.1. Monitor and accessories cleaning..........................................................2.1
2.2. Technical specifications..........................................................................2.2
2.2.1. Classification according IEC601. ................................................................................2.2
2.2.2. General Specifications................................................................................................2.2
2.2.3. Environmental Specifications. ....................................................................................2.3
2.2.4. Electrical Specifications..............................................................................................2.3
2.2.5. Electrocardiogram (ECG) Specifications.....................................................................2.3
2.2.6. Respiration specifications:..........................................................................................2.4
2.2.7. Non Invasive Arterial Pressure (NIBP) Specifications.................................................2.4
2.2.8. Invasive Arterial Pressure (PRES) Specifications.......................................................2.5
2.2.10. Pulse Oximetry (SpO2) Specifications.......................................................................2.5
2.2.11. Temperature (TEMP) Specifications.........................................................................2.6
2.2.12. Output Signals Specifications...................................................................................2.6
2.2.13. Materials Specifications............................................................................................2.6
2.2.14. Operating Specifications...........................................................................................2.7
2.3. Verification routines.................................................................................2.9
2.3.1. Daily verification routine.............................................................................................2.9
2.3.2. Periodical verification routine. ...................................................................................2.9
2.4. Alarm messages. ....................................................................................2.17
2.4.1. Initial self-test...........................................................................................................2.17
2.4.2. ECG messages. .......................................................................................................2.17
2.4.3. SpO2messages........................................................................................................2.18
2.4.4. NIBP messages........................................................................................................2.19
2.4.5. Invasive pressure messages.....................................................................................2.19
2.4.6. Respiration messages. .............................................................................................2.20
2.4.7. System messages. ...................................................................................................2.20
3. MONITOR INSTALLATION AND ADJUSTMENTS...................... 3.1
3.1. Unpacking .................................................................................................3.1
3.2. Monitor installation. .................................................................................3.2
3.3. Adjustments and option jumpers............................................................3.3
3.3.1. Supply board (P000PLAC05 Internal code).................................................................3.3
3.3.2. Base board (P000PLAC07 Internal Code)...................................................................3.4
3.3.3. Display board (P000PLAC14 Internal Code)...............................................................3.5
3.3.4. Main board (P000PLAC04 Internal Code)...................................................................3.6
3.3.5. Pulsioximeter board (P000PLAC19 Internal Code) .....................................................3.7
3.3.6. ECG and respiration board (P000PLAC20 Internal Code)...........................................3.8
3.3.7. NIBP and IBP board (P000PLAC01 Internal Code).....................................................3.9

OMICROM FT
SERVICE MANUAL
21 September 1998 ii
4. DISASSEMBLY AND REASSEMBLY .......................................... 4.1
4.1. Back Cover Removal................................................................................4.1
4.2. Battery .......................................................................................................4.2
4.3. Front cover................................................................................................4.2
4.4. Front Panel................................................................................................4.4
4.5. Measurement boards fastening. .............................................................4.6
4.6. NIBP Board (P000PLAC01 Internal Code) ..............................................4.8
4.7. Pump (C000BOMB01 Internal Code).......................................................4.9
4.8. AC/DC Board (C000ACDC01 Internal Code)...........................................4.9
4.9. Supply Board (P000PLAC05 Internal Code).........................................4.10
4.10. Connector Board (P000PLAC07 Internal Code).................................4.10
4.11. Power Line Filter ..................................................................................4.11
5. SCHEMATICS AND SPARE PARTS LIST.................................... 5.1
5.1. Programmable devices. ...........................................................................5.1
5.2. Spare parts list. ........................................................................................5.4
5.2.1. General spare parts list...............................................................................................5.4
5.2.2. P000PLAC01 Part list.................................................................................................5.6
5.2.3. P000PLAC04 Part list.................................................................................................5.9
5.2.4. P000PLAC05 Part list...............................................................................................5.11
5.2.5. P000PLAC07 Part list...............................................................................................5.13
5.2.6. P000PLAC10 Part list...............................................................................................5.15
5.2.7. P000PLAC14 Part list...............................................................................................5.18
5.2.8. P000PLAC15 Part list...............................................................................................5.20
5.2.9. P000PLAC19 Part list...............................................................................................5.21
5.2.10. P000PLAC20 Part list.............................................................................................5.24
5.3. Board schematics. .................................................................................5.27
5.3.1. P000PLAC01 Schematics.........................................................................................5.28
5.3.2. P000PLAC04 Schematics.........................................................................................5.34
5.3.3. P000PLAC05 Schematics.........................................................................................5.37
5.3.4. P000PLAC07 Schematics.........................................................................................5.39
5.3.5. P000PLAC10 Schematics.........................................................................................5.42
5.3.6. P000PLAC14 Schematics.........................................................................................5.47
5.3.7. P000PLAC15 Schematics.........................................................................................5.51
5.3.8. P000PLAC19 Schematics.........................................................................................5.53
5.3.9. P000PLAC20 Schematics.........................................................................................5.60

OMICROM FT
SERVICE MANUAL
21 September 1998 1.1
1. INTRODUCTION
This Service Manual has been prepared for the maintenance and service of any of
the various models of the Omicrom FT Portable Vital Signs Monitor.
The technical qualified personnel must read carefully and thoroughly this
manual before performing any disassembly or maintenance procedure
The family of Omicrom FT monitors have several models that monitors different
physiological signs. All models have a common base that contains a metallic
chassis, an external plastic enclosure and two common boards that provide power
supply and communications for the rest of boards. Each model includes different
measurement boards as a function of the measured parameters. An essential
documentation to use this Service Manual is the Operation Manual of the monitor,
that provides complementary information.
To make any question or comment send your request to:
RGB Medical Devices, S.A.
c/ Alfonso Gómez, 42
28037 MADRID (SPAIN)
Tel. ++ 34 91 304 20 11
Fax. ++ 34 91 327 39 03
1.1. Quality, Reliability and Safety.
This equipment has been designed with an special emphasis on those aspects
related with quality, reliability and safety, but RGB MEDICAL DEVICES will accept
responsability for these aspects when the following conditions are met:
a)Electrical installations in which the monitor is to be used must
comply with all regulations specified by the country in which the
monitor is being used.
b)The monitor is used in accordance with the instructions for use
provided by RGB MEDICAL DEVICES.
c)All maintenance, modification or repairment operations are
carried out by qualified personnel authorised by RGB MEDICAL
DEVICES.
d)Equipment installation must be carried out in accordance with
local requirements regarding responsibility and warranty.

OMICROM FT
SERVICE MANUAL
21 September 1998 1.2
1.2. General Precautions.
Read the following precautions thoroughly before performing any of the
disassembly procedures, because serious damage to the personnel or the
equipment can result if the precautions are not strictly followed.
All service operations must be carried out by qualified personnel. Besides, these
operations can only be performed by authorised personnel by RGB MEDICAL
DEVICES.
Before cleaning the equipment, always switch it off and disconnect it from the AC
power supply. Do not allow any liquid to enter the equipment case. Do not pour
liquid on the equipment while cleaning. Never use any abrasive material to make
the cleaning. Keep plugs and connectors meticulously clean and dry.
Do not expose the equipment to high humidity or heat.
Before starting with the disassembly procedure, the monitor must be turned off. It
is necessary to remove the power cable and all signal cables.
The front panel of the monitor is vulnerable to scratching, so a rough handling can
damage it.
When the monitor is reconnected and the power turned on without the plastic
covers, it is necessary to be careful in order to avoid exposure to high voltage. In
this case, do not power up the chassis on a conductive surface such as a bare
metal table. Place the chassis on a non-conductive material.
In order to avoid an electrostatic discharge, it is necessary to use adequate
protective means, such as antistatic surfaces and earthing means for technical
personnel. In case of handling spare parts sensitive to discharges, handle these
parts into their original package.
Normally, the measurement boards have a floating input section which is isolated
from the remainder of the board. Never handle this section of the board, because
the oil, dirt or dust deposited on these sections can cause leakage paths which
can degrade floating circuit isolation and patient safety.
During the disassembly procedure, be sure to keep all screws, nuts, plastic
spacers and other accessories for the posterior assembly of the monitor.

OMICROM FT
SERVICE MANUAL
21 September 1998 2.1
2. SERVICE INFORMATION
This chapter include several sections where different type of information for
servicing the Omicrom FT monitors is included.
2.1. Monitor and accessories cleaning.
The Omicrom FT monitor and accessories must be kept clean on a routine basis,
in accordance to the instructions provided here. To clean the monitor or their
accessories it is necessary to disconnect the monitor from the AC Power and the
different accessories from the monitor. After cleaning of the different parts of the
equipment , always allow them dry thoroughly before use.
The outside of the monitor may be wiped clean with a cloth slightly dampened with
mild detergent or normal hospital bactericides. Liquid must not be poured inside
the equipment. It must not be cleaned with isopropyl alcohol or other dissolvents.
Do not use abrasive agents. Avoid liquid entering inside the NIBP connector.
As a general rule, do not immerse the different cables and probes into water or
any other liquid. Besides, the continued flexing of cables and wires during the use
of the monitor or during the cleaning procedure could break the internal wires
and, as a consequence, produce the failure of the monitor.
A very important aspect that must be considered is the cleaning procedure of the
cuffs and hoses supplied with the unit. Care must be taken not to immerse these
elements in any kind of liquid to avoid that the liquid enters them at any time. If
this happens accidentally, then it may be dried by passing air through the cuff.
The cuff can be cleaned by hand washing in warm soapy water.
Both the ECG cable and the electrode leads may be cleaned and disinfected by
wiping with a 10% solution of household bleach or any other similar commercial
product. Do not use alcohol, since it will attack the protective plastic cover of the
cables causing them to undergo a premature erosion.
For the cleaning of sensors or cables, hold the sensor in one hand at the tip and
wipe the sensor and cable towards the plug end. If excessive pressure is used,
the covering will be stratched, which may break the internal wires and destroy the
probe.
Do not boil or autoclave the monitor, the sensor or the cables, because it can be
destroyed.

OMICROM FT
SERVICE MANUAL
21 September 1998 2.2
2.2. Technical specifications.
This section includes the technical specifications of Omicrom FT monitors.
General specifications are valid for all models of Omicrom FT monitors.
Specifications related to each vital sign are applicable according to the specific
model. In this case, it is mentioned the code of the board that monitors the vital
sign, because the specifications can vary according to the board used. The cross
reference of the boards used into each model of Omicrom FT is included in
Chapter 5 (“Schematics and spare parts list”).
2.2.1. Classification according IEC601.
Type of Protection: Class I, internally powered.
Mode of Operation:Continuous.
Degree of Mobility: Portable.
Degree of Protection against Electric Shock: Types BF and CF.
Note. This monitor can not be used in the presence of flammable anaesthetics
gases.
2.2.2. General Specifications.
External enclosure: ABS flame resistant.
Dimensions:232 length x 208 width x 242 height (mm).
Weight:4.8 kg.
Display:Electroluminescent display with very high contrast.
Viewing angle 160º.
Resolution 320 x 240 pixels.
Active area 120 x 90 mm
Keyboard: Power on/Power off key.
4 specific function keys.
5 soft keys.
Battery: Free maintenance lead acid battery 12 V - 2.1 Ah.
Size: 178 mm x 34 mm x 60 mm
Internal battery charger.
Maximum two hours of autonomy (it depends on model and
active parameters).
Safety Standards:
• IEC601-1 (1988). Medical electrical equipment. General
requirements for safety.
• IEC601-1-2 (1993). Medical electrical equipment. Part 1:
General requirements for safety. Section 2: Collateral
standard: Electromagnetic compatibility. Requirements and
tests.
• IEC601-2-27 (1994). Medical electrical equipment. Part 2:
Particular requirements for the safety of
electrocardiographic monitoring equipment.
• IEC601-2-30 (1995). Medical electrical equipment. Part 2:
Particular requirements for the safety of automatic cycling
indirect blood pressure monitoring.

OMICROM FT
SERVICE MANUAL
21 September 1998 2.3
• IEC601-2-34(1994. Medical electrical equipment. Part 2:
Particular requirements for the safety of direct blood
pressure monitoring equipment.
2.2.3. Environmental Specifications.
Storage temperature:20°C to 60°C
Operating temperature:0°C to 40°C
Relative humidity:5 % to 90%, non condensing
Altitude: -300 m to 4,500 m
2.2.4. Electrical Specifications.
AC input voltage:100-120 V~ / 200-240 V~
Input voltage frequency:50 / 60 Hz
AC input power:15 to 55 VA
AC fuses:250 V/1.6 A Type T.
Earth leakage current:< 100 µA
2.2.5. Electrocardiogram (ECG) Specifications.
Board P000PLAC20 - “ECG-R 280598 Rev. 03”
Patient Safety:Class I, type CF, protected against
defibrillator (according IEC601).
Heart Rate:Range: 20-300 beats per minute
Resolution: 1 beat per minute
Accuracy: ±2 %
Specific requirements:Input impedance: > 5 MΩat 10Hz
DC offset voltage range: ±0.6 V
System noise: < 30 µV
Signal Features:Bandwidth:
Diagnostic mode: 0.05 Hz to 100 Hz
Monitor mode: 0.5 Hz to 25 Hz
O.R. mode: 0.5 Hz to 25 Hz
Programmable gain:1 to 8 mV
Pacemaker detection and rejection:
Pulse duration: 0.1 to 2 ms
Pulse amplitude : ±2 to ±700 mV
Auxiliary Circuits:Calibration circuit.
Detection circuit of lead disconnection.
Detection circuit of cable disconnection.
Probe requirements:Three leads cable.
Available alarms:Internal failure.
Cable disconnection.
Lead disconnection.
Signal saturated.

OMICROM FT
SERVICE MANUAL
21 September 1998 2.4
2.2.6. Respiration specifications:
Board P000PLAC20 - “ECG-R 280598 Rev. 03”
Patient Safety:Class I, type CF, protected against
defibrillator (according IEC601).
Respiratory Rate:Range: 0-150 breaths per minute.
Resolution: 1 breath per minute.
Accuracy: 2 breaths per minute.
Bandwidth: 0 to 3 Hz
Sensitive Range:0.2 Ωto 15 Ω
Maximum impedance: 4 kΩ
Excitation Current: < 400 µA at 62.5 kHz.
Auxiliary circuits:Calibration circuit.
Detection circuit of lead disconnection.
Probe requirements:Three leads cable.
Available alarms:Internal failure.
Cable disconnection.
Lead disconnection.
Apnea alarm.
2.2.7. Non Invasive Arterial Pressure (NIBP) Specifications.
Board P000PLAC01 and P000PLAC10 - “TENSIOMETRO 230594 Rev. 01”
Patient Safety:Class I, type BF (according IEC601).
Heart Rate:Range: 40-250 pulses per minute.
Resolution: 1 pulse per minute.
Accuracy: ±3 %
Measurement Time:Typical: 25 to 45 sec.
Maximum: 120 seconds for adults.
90 seconds for neonates
Adults Measurement:Systolic pressure range: 30 to 250 mmHg
Diastolic pressure range: 10 to 218 mmHg
Mean pressure range: 20 to 234 mmHg
Overpressure limit: 315 mmHg
Neonates Measurement:Systolic pressure range: 30 to 130 mmHg
Diastolic pressure range: 10 to 114 mmHg
Mean pressure range: 20 to 122 mmHg
Overpressure limit: 165 mmHg
Measurement Accuracy:Never superior to ±10 mmHg.
Auxiliary circuits:Calibration circuit.
Overpressure protection circuit.
Probe requirements:Cuffs with one tube.
Available alarms:Internal failure.
Excessive measurement time.
Excessive deflation time.
Excessive inflation time.
Excessive time at the same pressure step.
Overpressure.
Small air leakages.
Weak signal.

OMICROM FT
SERVICE MANUAL
21 September 1998 2.5
2.2.8. Invasive Arterial Pressure (PRES) Specifications.
Board P000PLAC01 - “TENSIOMETRO 2030594 Rev. 01”
Patient Safety:Class I, type CF, protected against defibrillator
(according IEC601).
Number of channels:One.
Measurement range:-15 to 250 mmHg
Bandwidth: 0 to 15 Hz
Transducer sensitivity: 5 µV / V / mmHg
Heart Rate:Range: 20-250 pulses per minute
Resolution: 1 pulse per minute
Accuracy: ±3%
Gain:Accuracy: ±1 %
Drift: 0.1 % per °C
Zero Adjustment Circuit:Range: ±100 mmHg
Accuracy: ±1 mmHg
Drift: ±0.1 mmHg per °C
Auxiliary circuits:Calibration circuit.
Gain selection circuit.
Zero adjustment circuit.
Probe requirements:Connection Cable with electrical shielding.
Available alarms:Internal failure.
Cable disconnection.
Change of the scale.
2.2.10. Pulse Oximetry (SpO2) Specifications.
Board P000PLAC19 - “PULAIS 060498 Rev. 02”
Patient Safety:Class I, type BF, protected against defibrillator
(according IEC601).
Saturation Measurement:Range: 40 -100 %
Resolution: 1 %
Accuracy: 85 to 100 % - ±1.5 %
70 to 85 % - ±2 %
Heart Rate:Range: 30-250 pulses per minute
Resolution: 1 pulse per minute
Accuracy: ±3 %
Auxiliary circuits:Calibration circuit.
Detection circuit of sensor disconnection.
Probe requirements:BCI sensor or compatible.
Available alarms:Internal failure.
Cable or sensor disconnection.
Sensor off patient.
Insufficient light received.
Interferences.
Weak pulsatile signal.
Patient movement.

OMICROM FT
SERVICE MANUAL
21 September 1998 2.6
2.2.11. Temperature (TEMP) Specifications.
Board P000PLAC04 - “CONTROTEMP 200695 Rev. 30”
Patient Safety:Class I, type CF, protected against defibrillator
(according IEC601).
Measurement Range:15 to 45 °C / 59 to 113 °F
Resolution:0.1 °C / 0.2 °F
Accuracy:±0.1 °C (it is not included the probe tolerance).
Auxiliary circuits:Calibration circuit.
Probe requirements:YSI Series 400 Probes or compatibles.
2.2.12. Output Signals Specifications.
Board P000PLAC07 - “CONECTOR 130598 Rev. 03”
RS-232 output:One channel.
DB9 Male Connector.
PC connection or External Printer connection.
Electrically isolated.
RS-485 output:Two optional channels.
Electrically isolated.
2.2.13. Materials Specifications.
This section includes those components or parts whose specifications are
essential to guarantee the patient safety from the electrical point of view.
AC Power Cord:Female plug CEE22l.
Male plug Schuko IEC.
It must comply VDE / IEC standards.
AC Monitor Plug:Module Filter according to IEC601.
Two fuses incorporated.
1 pole rocker switch incorporated.
AC/DC Converter:It must comply with IEC601 Standard.
Isolation Voltage Input/Output >4000 V~.
Universal Input Voltage 85-250 V~ / 48-64 Hz.
Output voltage 15 VDC.
Minimum Output Power 40 W.

OMICROM FT
SERVICE MANUAL
21 September 1998 2.7
2.2.14. Operating Specifications.
Sweep Speed:The sweep speeds available are 12.5 mm/s and 25
mm/s.
Information by display:
The display can monitor the waveforms of ECG, NIBP,
SpO2, invasive pressure and respiration. The display
shows the numeric value of each measured parameter.
It shows all relevant information about each monitored
sign.
Heart Rate Measurement:
From each pulsatile signal, the system detects the
presence of a pulse and provides an acoustic and
visual indication. The source of the heart rate showed
is selected dynamically as a function of the active signs
(if it is not in an alarm situation) according to the
following priority criteria:
1) ECG
2) PRES1
3) PRES2
4) SpO2
5) NIBP
Alarm detection system:
For each vital sign monitored there is an independent
alarm detection system, that detects several alarm
conditions.
The presence of an alarm situation is indicated by
visual and acoustic means.
It is possible to disable the warnings for all signals at
the same time.
It is possible to disable the alarms of each parameter in
an independent way:
It is possible to select the patient alarm levels for each
parameter. These levels can be stored to be used the
next time the monitor is turned on.
The procedural alarms that can be detected have been
specified in previous sections.
Trends Storage:All parameters measured are stored periodically. The
time period between consecutive storage is selectable
by the user (options: 1, 2, 3, 5, 7, 10 and 15 minutes).
According to the selected time period, the storage
capacity varies between a minimum of 5 hours and a
maximum of 72 hours.

OMICROM FT
SERVICE MANUAL
21 September 1998 2.8
Configuration Options:
Configurations options available depend on each
model and can be stored to be used the next time that
the monitor is turned on:
• Sweep speed: 12.5 or 25 mm/s.
• QRS beep: Enabled or disabled.
• Alarm sound: Enabled or disabled.
• Temperature measurement units: °C or °F.
• NIBP and ECG patient type: Child/adult or
Neonate.
• Pacemaker rejection capability: Enabled or
disabled.
• ECG operating mode: Diagnostic, monitor or O.R.
Battery autonomy indicator:
The level of charge of the battery is displayed at the
bottom right part of the screen. The level of charge is
measured each ten seconds. When there are
approximately 15 minutes left before the battery is
powerless, the 'Battery' LED indicator starts to flicker,
thus indicating that it is necessary to power the monitor
from the AC mains power. During this time period an
alarm sound is activated and a transitory message is
displayed in the screen, to show in a clear way this
alarm condition. If this period is consumed, the
equipment is self-disconnected to avoid a large
discharge level which might be harmful to the battery.

OMICROM FT
SERVICE MANUAL
21 September 1998 2.9
2.3. Verification routines.
In this section it is explained the verification procedures that must be followed to
ensure that the Omicrom FT monitor is performing well.
Any failure detected during the tests must be located and repaired. After this
repairment, the monitor must be checked again completely before using it.
2.3.1. Daily verification routine.
Daily or each time the monitor is used, the operator must review the next points:
1. To inspect the monitor and the accessories and, if signs of damaged are
detected, replace or not use the damaged items.
2. To clean the monitor and their accessories according the instructions
given in section 2.1.
3. To inspect the AC power cord and, if signs of damage are detected,
replace it with an original RGB part replacement.
4. To check the battery status. Power up the monitor from its internal
battery and check the level of charge at the indicator placed at the
bottom right part of the display. If the charge level is too low or if the
‘Battery’ LED is flashing, connect the monitor to the mains supply and
switch on the rear panel mains power switch. Leave the monitor
charging until the charge level reaches the maximum.
5. To check the monitor operation. The basic points to check are: graphical
presentation by the display, adjustment of the parameters measured,
selection of the configuration options and adjustment of patient alarm
levels for each parameter. After checking these points, return all
adjustments to the desired settings.
If any failure is detected during this daily verification, contact with qualified
technical personnel authorized by RGB MEDICAL DEVICES for servicing the
monitors.
2.3.2. Periodical verification routine.
This sections describes the verification routine that must be performed each six
months by a qualified technician. Previously it must be done the daily verification
routine describe at the previous section.
In order to carry out this verification, it is necessary to use the next auxiliary
equipments:
• BioTek Simulator Model LionHeart 3 or equivalent.
• Pressure manometer.
In this section it is described the verification associated with each signal. Only it is
necessary to carry out those verifications, according the Omicrom FT model (so
each model monitors several signals). To verify each signal only it is necessary to
activate the signal at the monitor.

OMICROM FT
SERVICE MANUAL
21 September 1998 2.10

OMICROM FT
SERVICE MANUAL
21 September 1998 2.11
Electrocardiogram signal verification.
Activate at the monitor the ECG signal and select a scale of 2 mV. Connect the
ECG simulator (Lionheart 3) to the Omicrom FT using the standard cable.
Signal filter checking:
a) Select at simulator a 2 Hz square waveform with an amplitude of 1 mV
and check that at monitor display the peak amplitude is 50% of
monitoring area (all ECG modes).
b) Select at simulator a 4 Hz sinusoidal waveform with an amplitude of 1
mV and check that at monitor display the peak amplitude is 50% of
monitoring area (all ECG modes).
c) Select at simulator a 10 Hz sinusoidal waveform with an amplitude of 1
mV and check that at monitor display the peak amplitude is 50% of
monitoring area (all ECG modes).
d) Select at simulator a 0.5 Hz sinusoidal waveform. Select at the Omicrom
FT the Diagnostic ECG mode and check that at monitor display the peak
amplitude is 50% of monitoring area. Select at the Omicrom FT the
Monitor or OR ECG mode and check that at monitor display the peak
amplitude is 25% of monitoring area.
e) Select at simulator a 50 Hz sinusoidal waveform (or 60 Hz depending on
AC mains supply frequency) with an amplitude of 1 mV and check that at
monitor display a flat line appears (all ECG modes).
ECG signal verification:
a) Select at simulator an ECG waveform with and amplitude of 1 mV and a
frequency of 80 bpm and check that at monitor display the QRS
amplitude is 50% of monitoring area (all ECG modes).
QRS detection:
a) Select at simulator QRS detection signal of 0,5 mV, 70 ms and 80 bpm
and check that the heart rate showed at monitor display is correct (all
ECG modes).
b) Select at simulator QRS detection signal of 0,5 mV, 40 ms and 80 bpm.
Select at Omicrom FT as Patient type neonates and check that the heart
rate showed at monitor display is correct (all ECG modes).
c) Select at simulator QRS detection signal of 0,15 mV, 100 ms and 80
bpm and check that monitor does not detect any QRS (all ECG modes).
Signal gain:
a) Select at simulator an ECG waveform of 1mV and 80 bpm.
b) Select at monitor an ECG scale of 4 mV.
c) Check that the amplitude of the waveform at the monitor display is 25%
of monitoring area.
d) Select at the monitor an ECG scale of 1 mV.
e) Check that the amplitude of the waveform at the monitor display is 100%
of monitoring area.
Lead selection.
a) Select at simulator an ECG waveform of 1mV and 80 bpm.

OMICROM FT
SERVICE MANUAL
21 September 1998 2.12
b) Change the lead selected at the monitor and check that the selection is
correct.
Pacer detection:
a) Enable at the Omicrom FT the pacemaker rejection capability. Select at
simulator an asynchronous pacemaker with an amplitude of 0 2 mV and
a duration of 0.2 ms.
b) Check at the monitor that the pacemaker indication appears on the
screen and the heart rate is correct.
c) Select at simulator the option “Non capture pacemaker”.
d) Check at the Omicrom FT that the pacemaker indication appears on the
screen but there is not QRS detection.
e) Select at the simulator a pulse amplitude of 200 mV and a pulse duration
of 2 ms and repeat the steps b) to d).
ECG noise level:
a) Connect all ECG leads together.
b) Check that a flat line appears on the display (all ECG modes).
ECG alarms:
a) Disconnect the ECG cable and check that Omicrom FT shows an alarm
message indicating “Cable off”.
b) With the ECG cable connected to the Omicrom FT and all ECG leads
placed at the simulator, disconnect each one of the leads independently
and check that Omicrom FT shows an alarm message indicating “Lead
off”.
c) Disconnect the three leads and check that the same message is shown.
d) With all cables and leads connected, select at simulator a normal QRS
with an amplitude of 1 mV and at the Omicrom FT the 1mV ECG scale.
Check that Omicrom FT shows an alarm message indicating “Signal
saturated”.
e) Select at simulator an ECG waveform with a heart rate of 120 bpm.
Change the alarm levels at Omicrom FT to check that the patient alarms
are being activated correctly.
Respiration signal verification.
Activate at the monitor the respiration signal and connect the respiration simulator
(Lionheart 3) to the Omicrom FT using the standard cable..
Respiratory rate:
a) Select at simulator a respiratory waveform of 15 rpm and with an
amplitude of 1 Ohm.
b) Select at monitor a signal size of 2X.
c) Check that respiratory rate showed at monitor display is correct.
Base impedance:
a) Select at simulator a respiratory waveform of 15 rpm with an amplitude of
1 Ohm.
b) Select at monitor a signal size of 2X.
c) Select at simulator a base impedance of 500 Ohms
d) Check that respiratory rate showed at monitor display is correct.

OMICROM FT
SERVICE MANUAL
21 September 1998 2.13
e) Select at simulator a base impedance of 1000 Ohms
f) Check that respiratory rate showed at monitor display is correct.
Signal gain:
a) Select at simulator a respiratory waveform of 15 rpm with an amplitude of
1 Ohm.
b) Select at monitor a signal size of 2X.
c) Check that at monitor display the signal amplitude is 50% of monitoring
area.
d) Select at monitor a signal size of 5X.
e) Check that at monitor display the signal amplitude is 20% of monitoring
area.
Invasive pressure signal verification.
Activate at the Omicrom FT the invasive pressure signal and connect the pressure
simulator (Lionheart 3) using the special cable supplied with the simulator.
Previously, it is necessary to calibrate the zero of the pressure measurement, so
you must select at simulator the zero calibration option and follow the zero
adjustment procedure described in the Omicrom FT Operation Manual.
Pressure measurement:
a) Select at simulator a static pressure of -10 mmHg.
b) Select at the Omicrom FT a negative pressure scale and check that the
pressure values showed at Omicrom FT display are correct.
c) Select the maximum scale at the Omicrom FT.
d) Select at simulator a static pressure of 50 mmHg.
e) Check that the pressure values showed at Omicrom FT display are
correct.
f) Select at simulator a static pressure of 150 mmHg.
g) Check that the pressure values showed at Omicrom FT display are
correct.
h) Select at simulator a static pressure of 240 mmHg.
i) Check that the pressure values showed at Omicrom FT display are
correct.
j) Select at simulator a pressure waveform of 120/0 mmHg.
k) Check that the pressure values showed at Omicrom FT display are
correct.
Cardiac rate measurement.
a) Select at the Omicrom FT the maximum pressure scale.
b) Select at simulator a pressure waveform of 120/0 mmHg and 80 bpm.
c) Check that the heart rate measurement showed at the Omicrom FT
display is correct.
d) Select at simulator a pressure waveform of 120/0 mmHg and 160 bpm.
e) Check that the heart rate measurement showed at the Omicrom FT
display is correct.
Invasive pressure alarms:
Table of contents
Popular Medical Equipment manuals by other brands

Getinge
Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics
Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care
Pressalit Care R1100 Mounting instruction

Denas MS
Denas MS DENAS-T operating manual

bort medical
bort medical ActiveColor quick guide

AccuVein
AccuVein AV400 user manual