Configuration
© Rice Lake Weighing Systems ● All Rights Reserved 9
Backlight Shutdown (only applies to battery powered BP-P 12 x 14 scale) – Conserves battery life; Select the amount of time of inactivity
after which the backlight shuts off; Settings:
• AUTO – off after 5 seconds no activity
• OFF – always off
• ON – always on
Protocol – Determines manufacturer output protocol or serial setting the scale is configured for; Check third party software to confirm correct
selection; Settings:
• SMA – Interface Protocol (Section 6.4.2 on page 18)
• Auto-1 – Automatically transmit after stable weight above zero is removed from the scale platform
• Auto-2 – Automatically transmit when the item is placed on the scale and the weight stabilizes
• Print – Weight is transmitted only when the F1 button on the display panel is pressed
• NCI – General Serial Communications Protocol (782X and 76XX family) (Section 6.4.1 on page 16)
NOTE: If the scale is connected to a PC via the BenchPro USB port, the USB HID protocol is automatically selected; USB HID
settings are 1C19,0002.
Print Out – displays if is set to , or
Where: <LF> represents the line feed character (ØA hex)
W represents a weight digit character
uu represents the unit of measure characters (lb)
<CR> represents the carriage return character (ØD hex)
Baud Rate – For RS-232 connection to PC; Settings: 9600, 19200, 38400, 57600, 1200, 2400, 4800
Data Bits and Parity – For RS-232 connection to PC; Settings: 8 none, 7 even, 7 odd, 7 none
Stop bits – For RS-232 connection to PC; Settings: 1, 2
Enable or disable the tare button; Settings: OFF, ON;
NOTE: Enabling tare could violate metrological approvals.
Gravity Compensation – See Section 4.3 on page 10 for detailed information; Settings:
• No – Deactivated, calibrate scale with known accurate calibration weights
• Yes – View original calibration gravity and modify local gravity settings
Manufacturing mode only; Do not use; Do not adjust; Settings: OFF, ON
Units of Measure – turn on and off units of measure; applicable settings are dependent on the model of scale purchased; most models have
a minimum of two units of measure turned on; to avoid incorrect weight being displayed or transmitted to the PC, only have the applicable
units of measure turned on; settings: lb, lb:oz, oz, kg, g
Capacity – defines the maximum capacity of the scale and determines the weight value to be used:
• When selecting lb, the calibration weight used must be in lb
• When selecting kg, the calibration weight used must be in kilograms
NOTE: Do not select capacities other than those indicated by the manufacturer.
Configures the unit for multi-range or multi-interval on select models; Settings:
• Off – Unit displays in single range, 2rnG, 2intvL, 3rnG, 3intvL
Divisions (multi-range) – Follows gross weight and whatever the largest value count-by/display division you reach that is what the scale
locks on and uses until the scale goes back to zero and resets; Range uses the W1, W2, W3 annunciators
• kg Triple Range – 0-5kg x 0.001kg, 0-10kg x 0.002kg, 0-35kg x 0.01kg
• lb/oz Dual Range – 0-7lb x 0.05oz, 0-70lb x 0.2oz
Divisions (multi-interval) – Follows net weight and uses the applicable count-by/display division based on the weight on the scale
• kg Triple Interval – 0-5kg x 0.001kg, 5-10kg x 0.002kg, 10-35kg x 0.01kg
• lb/oz Dual Interval – 0-7lb x 0.05oz, 7-70lb x 0.2oz
NOTE: For factory use only.
Calibration – See Section 5.0 on page 11 for detailed information
NOTE: To exit the calibration () parameter without saving changes, the unit must be powered off.
Diagnostic menu – used to troubleshoot scale operation; See Section 7.1 on page 22 for more information; Settings:
• RAM – displays if functioning properly; If anything else displays, contact RLWS for a new PCB
• ROM – displays if functioning properly; If anything else displays, contact RLWS for a new PCB
• div-A – Display internal counts after auto zero tracking
• div-O – Display internal counts
Done – exit the configuration menu, save settings and return to weigh mode
Parameter Description
Table 4-2. Service Menu Parameter Settings (Continued)
Output Data String Settings
Setting Formatted Output Data String
LFuuLF <LF> WWW.WW uu <CR> <LF>
LFuu-- <LF> WWW.WW uu <CR> --
LF- -LF <LF> WWW.WW -- <CR> <LF>
LF- --- <LF> WWW.WW -- <CR> --
-- uuLF -- WWW.WW uu <CR> <LF>
--uu-- -- WWW.WW uu <CR> --
-- - -LF -- WWW.WW -- <CR> <LF>
-- - --- -- WWW.WW -- <CR> --