Rohde & Schwarz UMS100 User guide

3030.3059.72-02.00
System Manual
Division
Radiomonitoring and Radiolocation
Monitoring System
R&SUMS100/120
with ARGUS-UMS Software
3030.3013.02 3035.1025K02 3034.0090.02

3030.3059.72-02.00
R&S®ist eingetragenes Warenzeichen der Fa. Rohde & Schwarz GmbH & Co. KG.
Eigennamen sind Warenzeichen der jeweiligen Eigentümer.
R&S®is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
Proper names are trademarks of the respective owners.
Printed in the Federal
Republic of Germany
Rohde & Schwarz GmbH & Co. KG
Mühldorfstraße 15
D-81671 München
www.rohde-schwarz.com

Brand Names UMS100/120/120 System Manual
A3030.3059.72-02.00
Dear Customer,
The following abbreviations are used throughout this manual:
R&S®UMS100/120 is abbreviated with UMS100/120.
R&S®UMS100/120G is abbreviated with UMS100/120G.
R&S®UMS100/120HF is abbreviated with UMS100/120HF.
R&S®UMS100/120SHF is abbreviated with UMS100/120SHF.
R&S®ARGUS-UMS is abbreviated with ARGUS-UMS.
R&S®ARGUS is abbreviated with ARGUS.
R&S®is a registered trademark of Rohde & Schwarz GmbH & Co. KG and its subsidiaries.
CAUTION
This document is only applicable for ARGUS-UMS software of version 1.1 or higher.
(The controlled UMS100/120 therefore require firmware of version 1.1 or higher.) Older SW
versions have a different MMI and are not covered by this manual.


Contents UMS100/120/120 System Manual
33030.3059.72-02.00
Contents
Technical information ................................................................................................................................
1Operation preparation.........................................................................................................................7
1.1 Foreword........................................................................................................................................7
1.2 Installation/preparation ..................................................................................................................8
1.3 Web-based Service Tool in UMS100/120 .....................................................................................9
1.4 UMS100/120 configuration for LAN/WAN connections ...............................................................10
1.5 UMS100/120 configuration for dial-up connections using a mobile network...............................21
1.6 Configuring the control PC for a LAN/WAN connection ..............................................................21
1.7 Configuring the control PC for a GSM connection ......................................................................22
1.7.1 Outgoing connections (to the UMS100/120).......................................................................22
1.7.2 Incoming connections (from the UMS100/120) ..................................................................23
1.8 Installing ARGUS-UMS for communicating with UMS100/120 ...................................................29
2Operating with ARGUS-UMS ............................................................................................................31
2.1 Interface overview........................................................................................................................31
2.2 Connecting to a UMS100/120 .....................................................................................................32
2.3 Manual operation in "Direct Measurement Mode" DMM .............................................................35
2.3.1 Gain Control ........................................................................................................................36
2.3.2 Overload Indicator...............................................................................................................36
2.3.3 Getting an overview of the spectrum using Scan ...............................................................37
2.3.4 2D Waterfall Diagram in Scan Mode...................................................................................38
2.3.5 Max Hold in Scan Mode......................................................................................................39
2.3.6 Listen Mode in Scan ...........................................................................................................39
2.3.7 Bandwidth Measurement in Scan .......................................................................................40
2.3.8 Scan with handover to (FFM)..............................................................................................41
2.3.9 Fixed Frequency Measurement ..........................................................................................42
2.4 Definition of automatic measurements (AMM) ............................................................................45
2.4.1 Memory required for different measurements and audio recordings..................................45
2.4.2 Creating automatic measurements with AMM Wizard........................................................46
2.4.3 Other AMM settings ............................................................................................................51
2.4.4 Creating limit lines...............................................................................................................57
2.4.5 Creating frequency lists ......................................................................................................58
2.5 Analyzing alarms and measurement results................................................................................60
2.5.1 Audio playback....................................................................................................................63
2.6 ORM_ADC interface to other SW user interfaces .......................................................................63
3Maintenance.......................................................................................................................................64
3.1 Firmware update..........................................................................................................................64
3.2 Cleaning the hard drive and restarting UMS100/120 ..................................................................65
3.3 System support............................................................................................................................66
Appendix A

UMS100/120/120 System Manual Contents
3030.3059.72-02.00 4

Figures UMS100/120/120 System Manual
53030.3059.72-02.00
Figures
Figure 1-1: Main UMS100/120 connections to the central control PC (star network) ............................... 7
Figure 1-2: UMS100/120 connection to the main office (via LAN/WAN (red) or mobile network (blue)) .. 9
Figure 1-3: Service Tool, security certificate ........................................................................................... 11
Figure 1-4: Service Tool UMS100, Login window ................................................................................... 12
Figure 1-5: Service Tool UMS120, Login window ................................................................................... 12
Figure 1-6: Service Tool UMS120 Home Page ....................................................................................... 13
Figure 1-7: Service Tool, date and time entries incl. time zone .............................................................. 14
Figure 1-8: Service Tool, changing the Login parameters ...................................................................... 15
Figure 1-9: Service Tool, details via the control PC in use ..................................................................... 16
Figure 1-10: Service Tool, setting the communication port ..................................................................... 17
Figure 1-11: Options for UMS100............................................................................................................ 18
Figure 1-12: Options for UMS120............................................................................................................ 19
Figure 1-13: Antenna configuration for UMS120..................................................................................... 20
Figure 1-14: User administration ............................................................................................................. 24
Figure 1-15: Creating a user for dial-up .................................................................................................. 24
Figure 1-16: User properties, Member of .. window ................................................................................ 25
Figure 1-17: Group selection ................................................................................................................... 25
Figure 1-18: Group selection expanded .................................................................................................. 26
Figure 1-19: “name” Properties with entry............................................................................................... 26
Figure 1-20: Dial-in user permissions...................................................................................................... 27
Figure 1-21: Assigning an IP address for dial-up networking.................................................................. 28
Figure 1-22: System View with some UMS100 ....................................................................................... 30
Figure 2-1: ARGUS-UMS window ........................................................................................................... 31
Figure 2-2: Connection set up with configuration transfer....................................................................... 32
Figure 2-3: Select UMS100/120-window with activated connection ....................................................... 33
Figure 2-4: ARGUS-UMS connected to a UMS100/120 ......................................................................... 34
Figure 2-5: Antenna information .............................................................................................................. 35
Figure 2-6: Example of a fast scan for a wide frequency range.............................................................. 37
Figure 2-7: Example: Scan for high resolution ........................................................................................ 38
Figure 2-8: Scan with Max Hold .............................................................................................................. 39
Figure 2-9: Listen Mode in Scan.............................................................................................................. 40
Figure 2-10: Bandwidth Measurement .................................................................................................... 41
Figure 2-11: Settings for transferring configurations and marker on frequency of interest ..................... 42
Figure 2-12: Frequency measurement in FFM........................................................................................ 43
Figure 2-13: Frequency and offset measurement for AM carrier ............................................................ 44
Figure 2-14: Preconditions for AMM........................................................................................................ 45
Figure 2-15: Accessing AMM Wizard ...................................................................................................... 46
Figure 2-16: AMM Wizard Step 1, Timing Definition ............................................................................... 47
Figure 2-17: AMM Wizard Step 2, Range Definition ............................................................................... 48
Figure 2-18: AMM Wizard Step 3, Measurement Definition.................................................................... 49
Figure 2-19: AMM Wizard result analysis and alarm function................................................................. 51
Figure 2-20: AMM window, other functions ............................................................................................. 52
Figure 2-21: AMM Timing Definition window........................................................................................... 53
Figure 2-22: AMM Measurement Definition window................................................................................ 53
Figure 2-23: AMM Range Definition window ........................................................................................... 54
Figure 2-24: AMM Further Settings … window ....................................................................................... 55
Figure 2-25: AMM window with continuous measurement and the General Log File............................. 56
Figure 2-26: Creating a Limit Line from the Scan graphic....................................................................... 57
Figure 2-27: Offset setting for the Limit Line ........................................................................................... 57
Figure 2-28: Creating a frequency list ..................................................................................................... 58
Figure 2-29: Take over frequencies into the frequency list ..................................................................... 59
Figure 2-30: Disk capacity shortage on UMS100/120............................................................................. 60
Figure 2-31: Measurement result files on UMS100/120.......................................................................... 61
Figure 2-32: Corresponding File Info....................................................................................................... 61
Figure 2-33: File transfer to the Control PC ............................................................................................ 62

UMS100/120/120 System Manual Figures
3030.3059.72-02.00 6
Figure 2-34: Analyzing and processing measurement data.................................................................... 62
Figure 2-35: Audio playback.................................................................................................................... 63
Figure 3-1: Service Tool Firmware Update ............................................................................................. 64
Figure 3-2: Service Tool - Maintenance .................................................................................................. 65

Operation preparation UMS100/120/120 System Manual
73030.3059.72-02.00
1Operation preparation
1.1 Foreword
This manual describes the interconnection of one or more UMS100/120 with a control PC to a high-
capacity, extensively automatic radio monitoring network.
In addition, steps are described required for setting-up measurements which run automatically on the
UMS100/120.
The system operator is then only required if a UMS100/120 detects a deviation and has triggered an
alarm on the control PC in the main office. Manual tests must then be performed interactively to obtain
further information. This procedure is also outlined as an example.
Figure 1-1: Main UMS100/120 connections to the central control PC (star network)

UMS100/120/120 System Manual Operation preparation
3030.3059.72-02.00 8
1.2 Installation/preparation
The UMS100/120 components and control PC are also referred to as network elements in the following,
in accordance with ARGUS nomenclature. Before the actual configuration of these network elements
(UMS100/120 and control PC) can begin, the type of connection must be determined. Therefore it is
strongly recommended to enter the names, IP addresses, modem telephone numbers and other access
parameters such as user names and passwords, etc into a corresponding table. Only then can the
configuration of the network elements take place smoothly and enable communication.
Allocating IP addresses is essential for the network elements to be able to communicate clearly in the
network.
The definition of IP addresses in the private address range will not be discussed here in any detail.
There are explanations in the corresponding literature (e.g. using the Internet at Wikipedia:
http://de.wikipedia.org/wiki/Hauptseite or according to RFC 1918, go directly to
http://www.ietf.org/rfc/rfc1918.txt).
If more than one UMS100/120 is operating in the network, the configuration must take place before it is
connected to the network (LAN/WAN), as all UMS100/120 are set with a factory default address
(192.168.0.31).
One such configuration table is attached in appendix A which is filled in with several stations and the
accompanying parameters as an example.
This or the UMS100/120 to be connected is/are to be installed and connected to the power supply in
accordance with the "Mechanical Installation Manual".In the case of a LAN/WAN connection, the
UMS100/120 must be configured before with regards to the address.
The main office must be accessible at all times and supplied with the appropriate equipment for set-up
and operating a standard PC with keyboard, mouse and a printer.
Depending on the type of UMS100/120 connection, suitable telecommunication lines must be available.
This includes LAN/WAN access and/or analogue telephone lines for at least one, but usually two
modems. GSM module as modem can also be used alternatively. In this way the main office also has a
"wireless" connection.
Astandard PC is suitable as a control PC equipped with a Pentium IV processor, 512 MB memory, at
least 40 GB hard drive, CD or DVD-RW drive and an adequate number of RS232 ports (for external
modems) or PCI card slots (for internal modems). At least one free USB connector must be available
for the USB hardlock.

Operation preparation UMS100/120/120 System Manual
93030.3059.72-02.00
Figure 1-2: UMS100/120 connection to the main office (via LAN/WAN (red) or mobile network (blue))
When using a LAN/WAN connection via e.g. SHDSL connections, that are available in many countries
now as point-to-point connections should be completely end-to-end tested by the telephone network
operator, including modems.
1.3 Web-based Service Tool in UMS100/120
To carry out the minimal service work when up and running and for the initial configuration, a user
interface can be used which is web-based and can be started on each UMS100/120. A web browser
such as MS Internet Explorer from version 5.0 (better 6.0) can be used or e.g. Mozilla Firefox 1.4 on
any control or service PC. The control or service PC is connected to the system that has to to be
configured via a LAN/WAN or mobile network connection. Once the IP address has been entered the
configuration tool starts up. The user interface of the configuration tool is only available in English.
Furthermore, a service PC does not require any other special programs.

UMS100/120/120 System Manual Operation preparation
3030.3059.72-02.00 10
1.4 UMS100/120 configuration for LAN/WAN
connections
CAUTION
The configuration must take place before installing the device within the LAN as they all
have the same IP address (192.168.0.31) from the factory by default. Without previous
configuration, the devices would not be able to communicate individually in the network and
there would be an IP address conflict.
The configuration is either done in a central delivery store before they are distributed to the site or it is
done by a technician on site after the mechanical installation using a service notebook.
Astandard PC with a LAN interface and a web browser (e.g. MS Internet Explorer, see chapter 1.3) is
required for the configuration.
Connect the PC to the UMS100/120 in one of the two following ways:
•Connect the RJ45 socket on the PC to the corresponding socket on the bottom of the UMS100/120
to be configured directly using a cross wired LAN cable.
•Connect the RJ45 socket on the PC to the corresponding socket on the bottom of the UMS100/120
to be configured using an uncrossed cable by interconnecting a simple LAN switch or hub.
Then start the UMS100/120 by connecting it to the power supply.
Now start up the PC and set the network card to an IP address within the address range 192.168.0.xxx,
Subnet. 255.255.255.0, e.g. 192.168.0.11. The 192.168.0.31 IP address may not be used as this is the
UMS100/120 address (IP address conflict).
Then start up the web browser.
To open the service tool in UMS100/120, enter the following address in the browser window.
CAUTION
As this is a matter of a secure connection it is essential to enter "https:…..".
https://192.168.0.31
The following security message appears:

Operation preparation UMS100/120/120 System Manual
11 3030.3059.72-02.00
Figure 1-3: Service Tool, security certificate
As the communication between the control PC and UMS100/120 is processed in SSL (encrypted /
secure sockets layer), the UMS100/120 provides a certificate. A warning always appears to inform that
the name does not match the name of the site (see Figure 1-3). Acknowledge this certificate by clicking
on <Yes>.
UMS120 incorporates the modular version of the UMS100 complete system. Users therefore can
connect to the UMS120 a large range of monitoring antennas from Rohde & Schwarz (transducers are
interpreted) as well as from other suppliers (transducers are not interpreted) and configure these within
the service tool.
The displayed pages of the service tool are slightly different to UMS100 having an additional menu
entry for antenna configuration with UMS120 (refer to Figure 1-7).

UMS100/120/120 System Manual Operation preparation
3030.3059.72-02.00 12
Then the following Login window appears:
Figure 1-4: Service Tool UMS100, Login window
Figure 1-5: Service Tool UMS120, Login window

Operation preparation UMS100/120/120 System Manual
13 3030.3059.72-02.00
To log-in, enter the default user name and the corresponding password:
<Customer>
<21kd&woe>
In the following logged on status, the Home Page of the Service Tool appears with general information
and for currently installed versions of Firmware components.
CAUTION
For reasons of security, a timer runs which triggers an automatic logout after 30 minutes. In
this case the Login window appears again automatically.
Now the actual configuration can begin.
Figure 1-6: Service Tool UMS120 Home Page
For a better distinction between different UMS100/120 within the network it is possible to give the unit
an individual name under Measurement Unit.
Enter the name in the appropriate input field and click on <Set> to permanently take over the name.

UMS100/120/120 System Manual Operation preparation
3030.3059.72-02.00 14
As the first page Date and Time is now opened. Check or correct the settings.
•The same time zone must be set for all the UMS100/120s otherwise results between the
UMS100/120s cannot be correlated or automated measurements (AMMs) cannot start because
start time has passed already.
•The time must be the actual time in this time zone and not UTC time. Check the data again and
click on <Submit> to send.
In a later operation the UMS100/120s synchronize their system time with the control PC.
Figure 1-7: Service Tool, date and time entries incl. time zone
Now enter the parameters from the configuration table in the corresponding masks of the Service Tool.

Operation preparation UMS100/120/120 System Manual
15 3030.3059.72-02.00
On the Web Login Account page, the default login parameters can be replaced with customer-specific
entries.
•Overwrite <Customer> at Current User Name.
•Enter <21kd&woe> at Current Password.
•Enter a new password at New Password.
•Repeat the new entry in the sections underneath.
•Then check the data again and click on <Submit> to send.
CAUTION
Make a detailed note of the new access data and store it in a safe place. If you forget your
details or they get lost, access to the UMS100/120 may no longer be possible.
If this is the case, you will be obliged to pay for the factory customer services.
For UMS100 with Ser. Numbers above 100075 and for UMS120 it is possible to perform a
factory preset by using the internal Preset Button (for further details please refer to the
UMS100/120 installation manual).
Figure 1-8: Service Tool, changing the Login parameters

UMS100/120/120 System Manual Operation preparation
3030.3059.72-02.00 16
Now enter the details for the Control PC.Go to the relevant page.
•For LAN/WAN connections, enter the IP Address for the LAN interface.
•For GSM connections, enter the address as it is specified for the incoming dial-up networking.
•In the Control Unit field, always enter the "ARGUS Name" from the standard ARGUS installation.
For ARGUS-UMS - this is always "ARGUS-UMS".
•The Service to be selected is always “ARGUS_MT1” if the UMS100/120 is controlled by ARGUS-
UMS software, for standard ARGUS this may be different.
•When all the fields are filled in and the Connection Type <LAN> or <Dial-up Connection> has
been selected, send the data by clicking on <Submit>.
Figure 1-9: Service Tool, details via the control PC in use
The settings are necessary to route the alarm messages back to the (correct) control PC within the
network as well as to send back the results from automatic measurements that have been started by
ORM (order report module).
In case a GSM/CDMA mobile phone link is used for the connection, further settings as e.g. phone No.
etc. are necessary (please refer to next mask).

Operation preparation UMS100/120/120 System Manual
17 3030.3059.72-02.00
Now open the Communication window and enter the following parameters:
•The parameters for the LAN/WAN connection to be configured
•The parameters for the dial-up connection via e.g. GSM or CDMA (UMS120)
It is recommended that also these details are entered in case a wireless connection is necessary in the
future. But they can also be entered at a later date.
CAUTION
If you click on <Submit> once you have entered a new IP Address,Subnet Mask and
Default Gateway (only necessary with a router), the data is sent and a connection cannot
be kept open because the UMS100/120 is now only available via the new address
(https://<new address>).
If you want to open the service tool again, put the IP address as well as the Subnet mask of the service
or control PC into the new address range of the UMS100/120. Then connect to the Service Tool using
the new IP address.
Figure 1-10: Service Tool, setting the communication port

UMS100/120/120 System Manual Operation preparation
3030.3059.72-02.00 18
Options Management
Type depending different options for UMS100 and UMS120 are available.
Installed options are “removed” by activating of “remove” and clicking on <submit>.
In case new options are to be installed, the following instructions should be followed:
•Put the options-CD into the drive of the “Control-PC”.
•After selection of <Browse …> mark the CD-drive and the options file in there.
•Activate “install”.
•And finally click on <Submit>.
Figure 1-11: Options for UMS100
CAUTION
The configured options in UMS100 not only extend the selectable frequency range but also
set the appropriate transducer tables for the supplied antennas (HF or SHF). In case that
antennas from other suppliers are connected to the UMS100, the displayed field strength
levels in the ARGUS control SW will be not correct.
This manual suits for next models
4
Table of contents
Other Rohde & Schwarz Analytical Instrument manuals

Rohde & Schwarz
Rohde & Schwarz Hameg HZO51 User manual

Rohde & Schwarz
Rohde & Schwarz DDF205 User manual

Rohde & Schwarz
Rohde & Schwarz SMW-K40 User manual

Rohde & Schwarz
Rohde & Schwarz SMM-K507 User manual

Rohde & Schwarz
Rohde & Schwarz R&S RT-ZM15 User manual

Rohde & Schwarz
Rohde & Schwarz ZNH Series User manual