2
MODEL 1054A pH SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
The Model 1054A pH Analyzer requires a preamplifier to convert
the high impedance pH glass electrode signal to a low impedance
signal. The preamplifier may be located in one of three areas; in the
pH sensor for best performance, in a remote junction box when
process temperatures exceed 80°C (176°F) in submersion applica-
tions, or in the analyzer when the distance between the pH sensor
and the analyzer is 4.5 meters (15 feet) or less. The result is that
the pH signal may then be reliably transmitted from the sensor to
the analyzer using standard shielded 4-wire instrument cable.
The Model 1054A pH measures over the full range of 0-14 pH. The
current output may be calibrated to represent any 1 to 14 pH span.
A two-point calibration is made by immersing the sensor in two dif-
ferent buffer solutions and entering the pH values. When two
buffers are used, the microprocessor automatically calculates the
electrode slope which is used for self-diagnostics. This electrode
slope can be read on the display and manually adjusted for longer
electrode life. A one-point process standardization is easily accom-
plished by entering the pH value of a grab sample.
1.3 INSTRUMENT SPECIFICATIONS @ 25ÞC
Measurement Range: 0 to 14 pH
Output Scale Expansion: Zero suppression: up to 13 pH units
Span: Any pH from 1 to 14
Accuracy: ±0.01 pH
Repeatability: ±0.01 pH
Stability: ±0.01 pH/month, non-cumulative
Temperature Coefficient: Input: ±0.003 pH/°C
Output: ±0.006 pH/°C
Temperature Compensation: PT 100 RTD, Automatic or Manual
-15 to 100°C (5 to 212°F)
1.2 PHYSICAL SPECIFICATIONS - GENERAL
Panel Mount Enclosure: Black, ABS, NEMA 4X, IP65,
CSA Enclosure 4.
144 X 144 X 192mm, DIN size.
(5.7 X 5.7 X 7.6 inches).
Wall Mount Enclosure: Weatherproof, Heavy duty fiberglass
reinforced thermoplastic.
300 X 330 X 190mm
(11.75 X 13 X 7.5 inches).
Front Panel: Membrane keyboard with tactile feedback and user
selectable security. Black and white on grey.
Digital Display: LCD, black on grey
Optional, red LED
Character Height: 18mm (0.7 inch)
Electrical Classification:
Group I Panel Mount Enclosure:
FM Class I, Div. 2, Group A thru D.
28 Vdc relays - 6.0 amps resistive only
150 mA - Groups A & B; 400 mA - Group C;
540 mA - Group D; Ci = 0; Li = 0
CSA Class I, Div. 2, Group A thru D.
28 Vdc, 110 Vac & 230 Vac relays
6.0 Amps resistive only
Group II Wall Mount Enclosure: General Purpose
Power: 115 Vac, ± 10%, 50/60 Hz ± 6%, 4.0 W
230 Vac, ± 10%, 50/60 Hz ± 6%, 4.0 W
Current Output: Isolated, 0-20 mA or 4-20 mA
into 600 ohms maximum load, Direct or Reverse
Output Dampening: 0-255 seconds.
AmbientTemperature: -10 to 65°C (14 to 149°F)
Ambient Humidity: LCD: 0-85% RH
LED: 0-95% RH
Alarms: Dual, field selectable High/Low, High/High, Low/Low
Alarm 2 configurable as a fault alarm
Time Delay 0 to 255 seconds
Dual Setpoints, continuously adjustable
Hysteresis is adjustable up to 2 pH units
for low side/High Alarm and high side/Low Alarm
IntervalTimer: Interval: Minimum 10 minutes
On Counts: 1 to 60
On Duration: 1 to 299 seconds
Off Duration: 1 to 299 seconds
Wait Duration: 1 to 299 seconds
Controls dedicated relay
Relay Contacts: Epoxy Sealed Form A contacts, SPST, Normally
Open. Resistive Inductive
28 Vdc 6.0 Amps 3.0 Amps
115 Vac 6.0 Amps 3.0 Amps
230 Vac 6.0 Amps 1.5 Amps
Weight/Shipping Weight: 1.1 kg/1.6 kg (2.5 lbs./3.5 lbs.)
1.4 RECOMMENDED SENSORS:
Model 300 Retractable pH Sensor
Model 320B FlowThrough pH Sensor
Model 320HP High Purity pH Sensor *
Model 328A Steam Sterilizable pH Sensor
Model 381 Insertion/Submersion/Flow pH Sensor
Model 385 Retractable pH Sensor
Model 399 Disposable pH Sensor
Model 389 Disposable pH Sensor
* Model 1054pH includes programmable temperature corrections
required for the presence of ammonia used in treating high puri-
ty water