
Page 3
Version: 001
1. Choose tab "SYSTEM"
2. Press " START SYSTEM CALIBRATION". The robot arm is unlocked and can be moved freely.
3. Move robot arm by hand to a straight forward position at 90 ° in relation to the frame and in the lowest
position.
4. Move tube disk arm number 1 at buffer position.
5. Make sure the fingers on the motors are in a horizontal position.
6. Press "CALIBRATE"
The tube disk makes one complete turn and ends again with tube arm number 1 at buffer position.
7. Press "CLOSE" and the pop-up screen disappears.
8. Move the robot head by hand to approximately home position (top corner at backside).
9. Press "RESET" and System calibration is done.
Robot arm calibration:
All positions of the robot arm can be calibrated. All positions within the instrument are calibrated at the
factory for testing, but has to be checked/re-calibrated after mounting the robot arm on site. Also the track
position has to be re-calibrated on location before first use.
If the gripper picks tubes not correctly (e.g. if other tubes are used), the gripper and all positions have to re-
calibrated before first use. The "HOME"-position cannot be adjusted and is already defined through the "SYSTEM
CALIBRATION".
Other positions can be calibrated individually if needed and each new calibrated position can be checked
with the the manual movement functions on the SERVICE-TLAX SETTINGS SCREEN. (If necessary with all used
sample tube types).
Start with Go to SERVICE -TLAX SETTINGS -SERVICE and press CALIBRATE.
Calibration of Track position:
1. Choose "Track".
2. Place a tube transporter with tube to the track position.
3. Press "CALIBRATE" (the robot arm is set free and can be moved manually).
4. Move robot arm on the correct horizontal position: centered just above the tube cap.
5. Move robot arm downwards until the fingers from the gripper are just below the tube cap.
6. Press "SET LOCATION".