CAUTION: IF YOUR PUSH-IN PRESSURE CELL DOES NOT HAVE A PIEZOMETER,DO
NOT PROCEED WITH SATURATION.
The following steps outline the procedure for saturating the filter:
1Remove the filter set screw (Figure 3-1).
FIGURE 3-1 FILTER SET SCREW AND FILTER
CAUTION: TAKE CARE TO REMOVE ONLY THE FILTER SET SCREW.REMOVING THE
GLYCOL SET SCREW WILL DAMAGE THE PUSH-IN PRESSURE CELL.
2Submerge the Push-In Pressure Cell in a bucket of flat water (i.e. water that has
been sitting for 24 hours). This will fill the space above the diaphragm with water.
3Let the pressure cell sit in the water to completely saturate the filter.
4With the pressure cell submerged, slowly replace the filter set screw.
5The Push-In Pressure Cell should remain submerged until installation.
4 INSTALLATION
The following instructions outline the steps required to successfully install a Push-In
Pressure Cell:
1Drill a pilot hole slightly shorter than the planned installation depth.
2Connect the Push-In Pressure Cell to a readout. Take an initial reading.
3Carefully push the Pressure Cell into the pilot hole. Monitor the readout to ensure
the cell is not over-pressured.
CAUTION: MONITOR THE READOUT WHILE THE PRESSURE CELL IS BEING
INSTALLED TO PREVENT OVER-PRESSURING THE CELL.