SAEHAN SH5002 User manual

SPRING HAND DYNAMOMETER(SH5002)
User Manual

INSTRUCTIONS FOR USE
SAEHAN SPRING HAND DYNAMOMETER
SH5002
GENERAL:
This device is to measure hand grip strength and used to secure an index of general upper
body strength adjust the hand grip size to exert the subject’s maximum grip. Move pointer
lever to zero with your finger allow three trials with each hand alternating. The device
should be checked for correct condition prior to each usage by a person familiar with proper
operation of the balance.
For continuous use for years, take care of it, keep in the protective carrying case to prevent
damage from water and accidental dropping. Wipe off any water or liquid to prevent the
spring from rusting.

INFORMATION ON COMPONENTS, PACKAGE CONTENT :
No.
Name
Function
1
Dial Face
Scale measures up to 100kg(220 lbs)
2
Cap
Keep spring securely
3
Pointer
Pointer on the dial indicates force exerted
4
Handle Grip
Handle for gripping, Adjustable for hand size
5
Base
Base for handle
6
Grip size Scale
Adjustable for hand size scale
7
Knob
Rotate the knob to adjust hand size

Scope of supply :
- Device
- Soft case
- Operating instructions

SAFETY PRECAUTION :
Carefully read this operation manual before setup, even if you are already familiar with SAEHAN
Spring dynamometer.
Precautions for use
- Do not set the handle grip to a position less than 4 cm as shown by the grip size
scale.
- Do not turn the cap.
- Do not drop the instrument, otherwise, it may be damaged, Handle it carefully.
- Do not use or keep the device in direct sunlight, in area of high temperature,
especially in places near heating equipment, in moist or dusty places, or where it will
be directly exposed to water.
- If the device is dirty, wipe it with a soft, dry cloth, if very dirty, use a cloth wet with a
neutral detergent. Do not use thinner, alcohol, or similar liquids. They will damage
the surface of the device.
- Do not test if you have arm or shoulder pain, or high, uncontrolled blood pressure.
- Use the proper technique as described below for accurate and consistent results. If
needed, ask for a demonstration.
- Adjust the handgrip dynamometer so the second joint of the fingers fit snugly on the
handle when gripping. Hold the dynamometer with arm extended down about thigh
level.
- Make a maximal squeeze. Don’t let the dynamometer touch any part of the body or
any other object. Also avoid sudden thrusts or jars, which can make the reading
inaccurate.
- When squeezing, breathe out or exhale during contraction to prevent high
intrathoracic pressure. Don’t hold your breath and strain.
- Measure both hands. Take three tries per hand. Combine the best score of each
hand to give a total grip strength reading. Record test results in kilograms.
OPERATING INSTRUCTION :
1. Have Subject hold the spring Hand Dynamometer with the scale facing away from his
face. Adjust the grip size until the joints of the Subject’s fingers are approximately on a
plane perpendicular to the scale plate. When the adjustment is optimal, the subject will
sense he can comfortably exert his maximum grip.
2. Move the pointer to zero with your finger.
3. Subject should stand straight with his weight evenly distributed on both feet. With his
arms hanging at his sides, he should squeeze the Spring Hand Dynamometer with as

much force as possible being careful to squeeze only once. Record the resultant scale
reading.
4. Allow three trials with each hand, right and left hand alternating. There should be a brief
pause of about 10 to 15 seconds between each trial to avoid excessive fatigue. Record
the amount registered at each trial, but use only the highest recorded for each hand.
Compare your findings with the averages shown in Table 1 and 2.
Measurement :
1. Holding device in one of your hands, turn the knob to adjust the grip range until the
second joint of your forefinger is bent through 90 degrees.
2. Stand upright and relaxed, extend your arms downward, and grip the device and exert full
force without letting your arms touch your body. During the measurement, do not wave
the device.
3. Read the value displayed on the dial face and record it.
l Turn the pointer to set the pointer zero again.
Methods :
1. Have the subject hold the spring hand dynamometer with scale facing away from his face.
2. Adjust the grip size until the joints of the subject’s finger same approximately on a plane
perpendicular to the scale plate.
3. When the adjustment is optimal, the subject will sense he can comfortably exert his
maximum grip.
4. Move the pointer lever to zero with your finger.
5. Force Range : Measures from zero to 100 kg (220 pound).
6. Resolution and precision
Measurement graduations are every five pounds (within the red inner ring) and every 0.5
kg (whthin the black outer ring).
STRENGTH NORMS :
Grip strength, measured with hand grip dynamometer, correlates with overall upper body strength
and is an easy, safe test to perform, A good grip is linked to higher lean body mass, lower risk
of disability in old age, and increased longevity. Compare your test results with the norms shown
below.

CHILDREN DATA FOR SPRING GRIP TESTER
TABLE 1
Averages of Strength of Grip, In kg. for childrn
ADULT DATA FOR SPRING GRIP TESTER
TABLE 2
Averages of Strength of Grip, In kg. for ADULT

WARNING :
Changes to the unit’s design are not permitted. This may lead to incorrect measuring results,
safety-related faults and destruction of the device.
The device may only be operated in accordance with the described default settings. Other areas
of use must be released by SAEHAN.
SPECIFICATIONS :
Measurement range
0 to 100kg / 0 to 220lbs
Minimum unit
0.5 kg (5 pound)
Accuracy
100kg ± 2%
Dimension
Approx. 200(W) x 138(L) x 40(H) mm
Weight
Approx. 0.684kg(1.5 pounds)
TRANSPORTATION & STORAGE CONDITION :
Environment :
1. General :
- It is not aimed to be used in households.
- It is used in physical therapy rooms, doctors’ offices and patients’ rooms, etc.
- Equipment can be damaged when dropped to the floor.
Impossible to expect to work properly any more
If it works properly, calibration / exactitude should be maintained.
2. Conditions of visibility :
- Ambient luminance range : 100 lux ~ 1,500 lux
- Viewing distance : 20 cm ~ 40 cm
- Viewing angle : normal to the display ± 20◦
3. Physical :
- Temperature range : 10℃ ~ 40℃
- Relative humidity range : 10 % ~ 90 %
- Ambient pressure range : 700 hPa ~ 1,060 hPa

SERVICE, MAINTENANCE, DISPOSAL :
Cleaning :
Do not use aggressive detergents (solvents etc.). Apply soapy water to moist cloth or use
household detergent. Prevent fluid from penetrating into device. Finish by polishing with dry
soft cloth. Remove dirt immediately.
Service, maintenance :
The appliance may only be opened by trained service technicians who are authorized by SAEHAN.
Annual Calibration and Repair :
It is recommended that this instrument be checked for accurate calibration annually. If the
instrument has been dropped or there is reason the suspect that calibration is required, the
instrument should be serviced immediately. Contact your supplier or call SAEHAN at +82-55-294-
0381, 24-hour phone support at +82-10-5282-0161, or send email to saehancorp@naver.com.
Additional product and service information can be obtained at www.saehanmedical.com.
Disposal :
Disposal of packaging and appliance must be carried out by operator according to valid national
or regional law of the location where the appliance is used.
Warranty : Warranty claims shall be voided in case
- Our conditions in the operation manual are ignored.
- The appliance is used outside the described uses
- The appliance is modified or opened
- Mechanical damage and damage caused by media, liguids
- Natural wear and tear
- Non-intended use or electric installation
- The measuring system is overloaded
- Dropping the balance
- One year warranty. Any defective unit caused by workmanship will be replaced with
new unit.

Symbol description :
Definition of using symbols
Symbols
Meaning
Manufacturer
Date of manufactured
Authorized representative
in the European
Community
Medical device
Serial number
Batch code
Catalogue number
Caution
Fragile, handle with care
Keep dry
Keep away from sunlight
Consult instructions for use
※ If there has been occurred any serious accident in relation to the our device
released in market, the quality management representative of our company will
notify the Notified Body & the competent authority of any serious accident in
accordance with the vigilance procedure.

[SH-UM-001, Rev.5, 2022.06.01]
Manufacturer
EC Representative
SAEHAN Corporation
23-29, Bongamgongdan 13-gil,
Masanhoewon-gu, Changwon-si
Gyeongsangnam-do,
51342
South Korea
Phone : +82-55-294-0381
Fax : +82-55-293-5019
E-mail : saehancorp@naver.com
Website : www.saehanmedical.com
MVS In Motion
Westdijk 150
2830 Tisselt,
Belgium – Europe
Phone : +32-3-866-40-76
E-mail : info@mvs-in-motion.com
Website : mvs-in-motion.com
Table of contents
Other SAEHAN Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

jenway
jenway 6500 Service manual

Signal Hound
Signal Hound USB-TG124A user manual

Drucker Diagnostics
Drucker Diagnostics QBC STAR System Operator's manual

Sea-Bird Scientific
Sea-Bird Scientific WQM-a user manual

YOKOGAWA
YOKOGAWA PR300 Initial setup

Reed Instruments
Reed Instruments SL-417 instruction manual

Photon Systems Instruments
Photon Systems Instruments AquaPen-C AP 110-P Manual and user guide

Delmhorst
Delmhorst F-6 owner's manual

Macherey-Nagel
Macherey-Nagel PF-3 manual

wattwatchers
wattwatchers Auditor 3R quick start guide

Hy-Gain
Hy-Gain 421A Installation and operation instruction

ACO
ACO 6226 instruction manual