Dent ELITEpro XC Manual

Operator’s Guide
ELITEpro XC™
and
ELOG™15
Software
ELITEpro XC™/ELOG™15
March 1, 2017

ii
DENT Instruments
ELITEpro
Operator’s Guide: ELITEpro XC and ELOG™Software ©2016 DENT Instruments, Inc. All rights reserved.
This manual may not be reproduced or distributed without written permission from DENT Instruments.
Congratulations!
You now own one of the world’s most versatile, accurate, feature-rich, and lowest priced power/energy
loggers, the ELITEpro XC™. The ELITEpro XC logger contains many features, some of which enable you to
measure data across multiple platforms, view and graph real-time data, and retrieve data remotely.
In addition, this manual describes the features and functionality of the accompanying software
ELOG™15, which is designed to be user-friendly, intuitive, and an easy-to-use component of your
energy measuring projects.
Thank You for your business!
Trademarks
ELITEpro XC™, ELOG™, and PhaseChek™are trademarks of DENT Instruments, Inc.
Windows®, Windows® Vista, Windows® 7, Windows® 8, Windows® 10, and Notepad® are registered
trademarks of Microsoft Corporation.
Bluetooth® wireless technology is a registered trademark of the Bluetooth SIG. If your unit is equipped
with Bluetooth wireless technology, look for the following trademark symbol:
DENT Instruments | 925 SW Emkay Drive | Bend, Oregon 97702 USA
Phone 541.388.4774 | Fax 541.385.9333 | www.DENTinstruments.com

iii
DENT Instruments
ELITEpro
TABLE OF CONTENTS
Introduction....................................................................................................................... 7
ELITEpro XC SAFETY SUMMARY and SPECIFICATIONS.....................................................................8
ELITEpro XC RÉSUMÉ DE SÉCURITÉ ET SPÉCIFICATIONS................................................................11
ELITEpro XC TECHNICAL SPECIFICATIONS ......................................................................................14
ELITEpro XC CONNECTIONS ...........................................................................................................16
DENT INSTRUMENTS WARRANTY STATEMENT .............................................................................17
Prepping & Using the ELITEpro ......................................................................................... 19
Install the ELOG Software on the Computer..................................................................................19
Mount the ELITEpro XC..................................................................................................................20
ELITEpro XC Connections.........................................................................................................21
Complete the Wiring Connections.................................................................................................22
CT Connections........................................................................................................................22
Connecting Split-Core Style Millivolt CTs to a Load..........................................................23
Connecting RōCoil CTs to a Load ......................................................................................25
Voltage Connections ...............................................................................................................26
Analog Channel Connections...................................................................................................26
Current Loop Connection..................................................................................................28
Voltage Connection...........................................................................................................29
Power In/Out Connection .......................................................................................................30
Digital Channel Connections ...................................................................................................30
Connecting the USB Cable.......................................................................................................33
Other Communication Connections........................................................................................34
Communicating with the ELITEpro XC............................................................................................35
Terminating the ELITEpro XC to Computer Connection..........................................................37
Best Practices................................................................................................................... 38
Creating and Using the Logger Setup Tables .................................................................................38
Create a New Setup Table .......................................................................................................38
Open an Existing Setup Table on Your PC ...............................................................................39
Retrieve an Existing Setup Table from the ELITEpro XC..........................................................40
Configure the Setup Table.......................................................................................................41
Data Logging Controls .............................................................................................................42
Start Options.....................................................................................................................42
Stop Options .....................................................................................................................42

iv
DENT Instruments
ELITEpro
Table of Contents
Memory Capacity ....................................................................................................................42
Quick Setups............................................................................................................................43
Physical Power Channels.........................................................................................................45
Calculated Power Channels.....................................................................................................48
Combining Channels .........................................................................................................48
Analog Output Channels .........................................................................................................49
Send a Setup Table to the Logger............................................................................................50
Save a Setup Table on the Computer......................................................................................51
LED Diagnostics..............................................................................................................................52
PhaseChek™ LEDs....................................................................................................................52
COMM—Communications LED ...............................................................................................52
Power Measurement—Connecting Sensors & Configuring Channels...........................................53
Wiring Label Conventions........................................................................................................54
Current Transformer (CT) Selection ........................................................................................54
CT Selection Examples.............................................................................................................56
Connecting CTs and Configuring Channels..............................................................................56
ELOG Quick Setup Wiring Diagrams........................................................................................58
Using the ELOG Software...............................................................................................................62
ELOG Speed Buttons................................................................................................................62
Keyboard Shortcuts .................................................................................................................64
File Menu.................................................................................................................................64
Edit Menu ................................................................................................................................66
View Menu ..............................................................................................................................66
Logger Menu ........................................................................................................................... 67
Communication Sub-Menu...............................................................................................67
Retrieve Data From Logger ...............................................................................................72
Display Real-Time Values Sub-Menu ................................................................................74
Retrieve Logger Setup Table .............................................................................................80
Logger Clock Sub-Menu ....................................................................................................80
Edit Logger Description Line .............................................................................................81
Logging ON/OFF................................................................................................................81
Clear Logger Sub-Menu.....................................................................................................82
Tools Menu..............................................................................................................................83
PC Setup............................................................................................................................83
Update Logger Firmware (ELITEpro XC only)....................................................................84
Restore Hidden Messages.................................................................................................84

v
DENT Instruments
ELITEpro
Table of Contents
Data Menu...............................................................................................................................85
Create New Graph ............................................................................................................85
Graph Options...................................................................................................................88
Retrieve Data From Logger ...............................................................................................89
Open Data File ..................................................................................................................89
Export Data File.................................................................................................................89
Import CSV File..................................................................................................................90
Data File Statistics.............................................................................................................91
Analyze a Selected Data Range.........................................................................................93
Extract a Range of Data.....................................................................................................94
Check Data for Errors........................................................................................................95
Fill Missing Records...........................................................................................................95
Appendices ...................................................................................................................... 96
Appendix A—Installing Firmware Updates....................................................................................96
Firmware Updates...................................................................................................................96
Downloading and Installing Firmware ..............................................................................96
Appendix B—Unusual Metering Systems ......................................................................................97
3-Phase Delta with a “Wild” (Red) Leg....................................................................................97
2–½ Element Meters ...............................................................................................................98
Using The 2-Watt Element (2 Ct) Method On A 3-Wire Delta Service....................................99
Using The 3-Watt Element (3 Ct) Method On A 3-Wire Delta Service..................................100
Appendix C—Troubleshooting ELITEpro XC Driver Installation ...................................................102
Appendix D—AUTOPOLL: Unattended Data Collection...............................................................106
AutoPoll Menu Options.........................................................................................................106
File Menu ........................................................................................................................106
Edit Menu........................................................................................................................107
View Menu......................................................................................................................108
Communications Menu...................................................................................................108
Tools Menu .....................................................................................................................109
Log File ............................................................................................................................110
Understanding the Logger Lists.............................................................................................111
Configuring a Logger List.................................................................................................112
Entering AutoPoll Tasks ..................................................................................................113
Logger Tasks....................................................................................................................113
Call Interval .....................................................................................................................114

vi
DENT Instruments
ELITEpro
Table of Contents
Appendix E—Calculated Power Channel Formulas .....................................................................115
Wye Configuration L1-N, L2-N, L3-N .....................................................................................115
DELTA Configuration: L1-L2 and L3-L2 ..................................................................................116
Appendix F—Communication Options ........................................................................................117
USB (Standard) ......................................................................................................................117
Ethernet LAN (Standard) .......................................................................................................118
Wi-Fi and Wireless Access Point Connection (Optional).......................................................121
ELITEpro XC as its own Access Point ...............................................................................121
ELITEpro XC Connected to a Wireless Network ..............................................................123
Accessing the ELITEpro ...................................................................................................125
Troubleshooting Wi-Fi Connections ...............................................................................127
ELITEpro Mobile App (Optional)............................................................................................128
Bluetooth (Optional)..............................................................................................................128
Cell Modem Connection (Optional).......................................................................................131
Appendix G—Meter Installation Final Checklist..........................................................................132
Index ............................................................................................................................. 136

7
DENT Instruments
ELITEpro
INTRODUCTION
The ELITEpro XC is a powerful and versatile tool for pinpointing electric usage and quantifying energy
usage. It is capable of measuring, storing, and analyzing consumption data including Volts, Amps, Watts,
Volt-Amps (VA), Volt-Amps reactive (VAR), Kilowatts (kW), Kilowatt Hours (kWh), KVAh, kVARh, and
Power Factor (Pf) of an electrical load or an entire building. Analog input channels are especially helpful
when used in conjunction with power measurements to correlate the consumption of electricity with
environmental, HVAC plant performance or other process conditions. The ELITEpro XC also offers some
power quality features such as the ability to view in real-time voltage, current, and power waveforms
and calculate harmonics to the 63rd order.
The ELITEpro XC is line-powered from the service being measured, eliminating the need for external
power or space-consuming batteries at the job site. The measurements are stored in on-board,
non-volatile memory in a time series format at an interval selected by the user. The ELITEpro XC can be
mounted anywhere with its magnetic back and is small enough to be secured inside the electrical panel.
The Windows-based ELOG software package is used to set up the meter, display metered values, and
retrieve and analyze the collected data. ELOG graphically displays recorded data, performs analysis and
facilitates automatic remote data collection. Data is also easily exported to popular spreadsheet and
database programs for additional analysis.
The ELITEpro XC requires a computer and ELOG™software for setup, data retrieval, and data analysis.
Download and test the software prior to the field installation of the logger.
If your installation is using ELOG with legacy devices, contact DENT Instruments for a prior version of this
manual.
For further information or to provide comments and suggestions, contact us at:
800.388.0770 or info@DENTInstruments.com

8
DENT Instruments
ELITEpro
Introduction
ELITEpro XC SAFETY SUMMARY and SPECIFICATIONS
This general safety information is to be used by both
the Logger operator and servicing personnel. DENT
Instruments, Inc. assumes no liability for user’s
failure to comply with these safety guidelines.
4RH8
E186827
Conforms to UL Std 61010-1
Certified to CSA Std C22.2 No.
61010-1
The ELITEpro XC is an Over-Voltage Category III device. Use approved rubber gloves with mechanical
protection and goggles when operating the device.
CAUTION: This LOGGER may contain life-threatening voltages. QUALIFIED PERSONNEL MUST disconnect all high
voltage wiring before using or servicing the LOGGER.
Warning: Use of this device in a manner for which it is not intended may impair
its means of protection.
SYMBOLS ON EQUIPMENT
Denotes caution. See manual for a description of the meanings.
When connecting the ELITEpro XC to an AC load, follow these steps in sequence
to prevent a shock hazard.
1. If possible, de-energize the circuit to be monitored.
2. Connect the CTs to the phases being monitored.
3. Connect the voltage leads to the different phases. Use proper safety
equipment (gloves, mask, and protective clothing) as required for the
voltages monitored.
Risk of electric shock. Life-threatening voltages may be present. Qualified
personnel only.
DO NOT EXCEED 600V Phase to Phase. This logger is equipped to monitor loads
up to 600V. Exceeding this voltage will cause damage to the logger and danger to
the user. Always use a Potential Transformer (PT) for loads in excess of 600V. The
ELITEpro XC is a 600 volt Over-Voltage Category III device.
Dispose of properly.

9
DENT Instruments
ELITEpro
Introduction
IN: 6-10 VDC, 500 mA
OUT: 6 VDC, 200 mA maximum
USB port
SENSOR LIMITATIONS
USE ONLY SHUNTED CURRENT TRANSFORMERS (CTs).
Do not use other CTs. Only use shunted CTs with a 333mV maximum output only.
Serious shock hazard and logger damage can occur if unshunted CTs are used. The
UL listing covers the use of the following DENT Instruments CTs that are UL
Recognized and have been evaluated to IEC 61010-1:
CT-RGT12-XXXX (solid core), CT-SRS-XXX (split core), CT-HSC-020-X (20A Mini), CT-
HSC-050-X (50A Mini), CT-HMC-0100-X (100A Midi), CT-HMC-0200-X (200A Midi),
CT-RXX-1310-U (RōCoil), CTRXX-A4-U (RōCoil), CT-CON-1000-X, CT-CON-0150EZ-X
and the CT-SRL-XXX.
The use of any other CT will invalidate the UL Listing of the ELITEpro XC.
Pulse: Use only "dry contact" non-energized pulse inputs (ELITEpro SP only). Use
of energized pulse initiators can cause damage to the logger and a potential shock
hazard to the user. Wiring must meet 600V AC CAT III rating.
DC VOLTAGES
The meter is designed to measure AC voltages for installations up to 600 VAC. The
meter is also capable of measuring DC voltages up to 600 VDC with the following
caveats.
1) The internal fuse installed in meters with serial numbers before XC1703xxx are
not a suitable disconnect for voltages above 80 VDC. Customers using the meter for
high voltage DC systems must install a UL listed inline 600 volt DC fuse with a rating
of 1 amp or less. These fuses are commonly available in the solar power market.
2) UL testing of the meter was conducted on AC systems only. DC results are
comparable but not tested.
3) DENT Instruments ELOG software does not currently have a configuration that is
specific to DC systems. Values reported for voltage, current, KW, KWH will be
correct however it should be noted that the meter will also report values that
pertain to AC systems (such as VARs and Power Factor) that should be ignored.

10
DENT Instruments
ELITEpro
Introduction
The equipment is protected throughout by double insulation (IEC 536 Class II). CAT
III 80-600 VAC 125mA 50/60 Hz
FCC COMPLIANCE
This device has been tested and found to comply with the limits for a Class A digital device, pursuant to
part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment. This equipment generates,
uses, and can radiate radio frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the user will be
required to correct the interference at user’s own expense.
Operation is subject to the following two conditions: (1) This device may not cause harmful interference,
and (2) this device must accept any interference received, including interference that may cause
undesired operation.
MAINTENANCE
There is no required maintenance with the ELITEpro XC. Abide by the following items:
Cleaning: No cleaning agents, including water, shall be used on the ELITEpro XC.
Battery Life: The lithium battery is only used to maintain the date and clock settings during power
failure and has a life expectancy of greater than 10 years. Contact DENT Instruments for service.
No accessories are approved for use with the ELITEpro XC other than those specified in the DENT
Instruments product literature and price sheets.
If the logger appears damaged or defective, first disconnect all power and sensors. Call or email
technical support for assistance.
DENT Instruments
Bend, Oregon USA
Phone: 541.388.4774
DENTinstruments.com
Email: techhelp@DENTinstruments.com

11
DENT Instruments
ELITEpro
Introduction
ELITEpro XC RÉSUMÉ DE SÉCURITÉ ET SPÉCIFICATIONS
Cette information de sécurité est destinée à être
utilisée à la fois par l'opérateur de l'enregistreur et
le personnel de service. DENT Instruments, Inc
n'assume aucune responsabilité pour l'utilisateur qui
ne respecte pas les directives en matière de
sécurité.
4RH8
E186827
Conforme à UL Std 61010-1
Certifié CSA Std C22.2 No. 61010-1
Le ELITEpro XC est un appareil de surtension de catégorie III. Utiliser des gants en caoutchouc
approuvé avec protection mécanique et des lunettes lors de l'utilisation de l'appareil.
ATTENTION: Ce LOGGER peut contenir de hautes tensions qui peuvent être dangereuses. UN PERSONNEL QUALIFIÉ
DOIT débrancher tous les câbles à haute tension avant d’utiliser ou de réparer du LOGGER.
Attention: L'utilisation de cet appareil d'une manière pour laquelle il n'est pas destiné peut
annuler ses moyens de protection.
SYMBOLES DES EQUIPEMENTS
Signifie prudence. Voir le manuel pour une description de la signification.
En faisant la connexion du ELITEpro XC à une prise de courant alternatif, suivez ces étapes
en ordre pour empêcher un risque de choc.
1. Décharger le circuit à contrôler.
2. Connectez le TC aux phases à surveiller.
3. Connectez les fils de tension à des phases différentes. Utiliser des équipements de
sécurité (gants et des vêtements de protection) qui sont nécessaires pour les tensions
surveillées.
Indique haute tension. Risque de choc électrique. Hautes tensions peuvent être présentes
qui mettent la vie en danger. Personnel qualifié uniquement.
NE PAS DEPASSER 600V Phase à Phase. Ce compteur peut contrôler les charges jusqu'à
600V. Le dépassement de cette tension peut causer des dommages à l'appareil et du
danger pour l'utilisateur. Utiliser toujours le potentiel transformateur (PT) pour des charges
de plus de 600V. Le ELITEpro XC est un appareil à 600 V de surtension de catégorie III.
Se débarrasser de correctement.

12
DENT Instruments
ELITEpro
Introduction
IN: 6-10 VDC, 500 mA
OUT: 6 VDC, 200 mA maximum
USB port
LIMITATIONS DE DÉTECTEUR
UTILISEZ SEULEMENT TRANSFORMATEURS DE COURANT (TC) SHUNTÉE.
N’utilisez pas d'autres TC. Utilisez seulement des TC shuntée avec une puissance maximale
333mV. Un sérieux risque de décharge électrique et des dommages à l'enregistreur peut se
produire si des TC pas shuntée sont utilisés. Utiliser seulement les CTs des DENT
Instruments suivants qui sont énumérés jusqu'au 600V/CATIII.
CT-RGT12-XXXX (solid core), CT-SRS-XXX (split core), CT-HSC-020-X (20A Mini), CT-HSC-050-
X (50A Mini), CT-HMC-0100-X (100A Midi), CT-HMC-0200-X (200A Midi), CT-RXX-1310-U
(RōCoil), CTRXX-A4-U (RōCoil), CT-CON-1000-X, CT-CON-0150EZ-X et le CT-SRL-XXX.
Impulsion: n'utilisez que "contact sec" impulsion intrants non charger (ELITEpro SP
uniquement). Utilisation de impulsion initiateurs charger peut causer des dommages à
l'enregistreur et une décharge électrique potentielle pour l'utilisateur. Il faut que
l'installation électrique satisfasse l'évaluation de 600V AC CATIII.
TENSIONS C. C.
La ELITEpro est conçu pour mesurer tensions c. a. pour les installations jusqu'à 600 V c. a. Le ELITEpro
est également capable de mesurer les tensions c. c. jusqu'à 600 V c. c. avec les avertissements
suivants.
1) Le fusible interne installé dans le ELITEpro n'est pas un dispositif de désaccouplage adéquat pour
des tensions supérieures à 80 V c. c. Les clients utilisant la ELITEpro pour haute tension DC systèmes
doivent installer un UL inline 600 volts DC fusible avec une valeur nominale de 1 amp ou moins. Ces
fusibles sont couramment disponibles dans le marché des piles solaires. Serial #’s < XC1703xxx.
2) UL tests du ELITEpro a été menée sur systèmes CA uniquement. DC résultats sont comparables,
mais pas testé.
3) Impact Instruments logiciel ELOG ne dispose actuellement pas de configuration qui sont
spécifiques aux systèmes à courant continu. Les valeurs rapportées pour la tension, le courant, en
KW, KWH sera correct mais il convient néanmoins de noter que le compteur sera également rapport
valeurs qui se rapportent aux systèmes à courant alternatif (tels que VARs et facteur de puissance)
qui doit être ignoré.
L'équipement protégé en double isolation (IEC 536 Classe II). CAT III 80-600 VAC 125mA
50/60 Hz

13
DENT Instruments
ELITEpro
Introduction
Entretien
Il n'y a aucun entretien requis avec le ELITEpro XC. Respectez les points suivants:
Nettoyage: Aucun agents de nettoyage, y compris l'eau, doit être utilisé sur le ELITEpro XC.
Espérance de Vie de la Batterie: La pile au lithium est utilisée uniquement pour maintenir les
paramètres de date et d'heure en cas de coupure de le courant et a une espérance de vie de plus de 10
ans. Contactez DENT Instruments pour le service.
Pas d'accessoires approuvés pour une utilisation avec le ELITEpro XC sauf ceux spécifiés par DENT
Instruments dans ses documentations sur les produits et également sur les prix.
Si le compteur semble endommagé ou défectueux, tout d'abord déconnecter le pouvoir de l'appareil.
Alors s'il vous plaît appelez 541.388.4774 ou contacter par courriel l'assistance technique pour obtenir
de l'aide.
DENT Instruments
Bend, Oregon USA
Phone: 541.388.4774
DENTinstruments.com
Email: techhelp@DENTinstruments.com

14
DENT Instruments
ELITEpro
Introduction
ELITEpro XC TECHNICAL SPECIFICATIONS
Specification
Description
Memory
16 MB non-volatile
Service Types
Single Phase-Two Wire, Single Phase-Three Wire, Three Phase-Four Wire (WYE),
Three Phase-Three Wire (DELTA), DC Systems (solar/battery).
Voltage Channels
3 channels, CAT III, 0-600 VAC (line-to-line) or 600 VDC
Current Channels
4 channels, .67 VAC max, +/- 1 VDC max; ideal for 333 mV full scale CTs
Maximum Current Channel
Input Voltage
External transducer dependent; 200% of current transducer rating (mV CTs);
Measure up to 6000A with RōCoil CTs
Measurement Type
True RMS using high-speed digital signal processing (DSP)
Line Frequency
DC/50/60Hz
Waveform Sampling
12 kHz
200 samples/cycle at 60Hz
240 samples/cycle at 50Hz
Channel Sampling Rate
(internal sampling)
8 Hz or every 125 mSec.
Data Interval
The default integration period is fifteen minutes. The choices are 1, 3, 10, 15, 30
seconds; 1, 2, 5, 10, 15, 20 and 30 minutes; 1 and 12 hours; 1 day.
Energy Measurements
Volts, Amps, Amp-Hrs (Ah), kW, kWh, kVAR, kVARh, kVA, kVAh, Displacement
Power Factor (dPF). All parameters for each phase and for system total.
Analog Measurements
0–10 Volts, 0 or 4–20 ma current loop non-isolated, external power for current
loop. User scaling for sensor reporting in physical units.
Accuracy
Better than 1% (<0.2% typical) for V, A, kW, kVAR, kVA, PF
Resolution
0.01 Amp, 0.1 Volt, 1 Watt, 1 VAR, 1 VA, 0.01 Power Factor, 0.001 ANALOG
LED Indicators
Tri-colors (red, green, and blue): 1 LED to indicate communication, 4 LEDs for
correct phasing (green when voltage and current on the same phase; red when
incorrectly wired, blue for wireless, Bluetooth wireless technology, and
Ethernet).
Communication
USB (Standard)
USB standard (Type B). 1.8 M (6 FT) A-to-B USB Cable (included)
ETHERNET (Standard)
Standard RJ-45 Connector supports 10/100 MB Ethernet over Cat 5 or better.
Configure for DHCP or Static IP address.
BLUETOOTH® Wireless
Technology (Optional)
Internal antenna: typical conditions <10ft up to 100ft range. Requires Bluetooth
wireless technology adapter on Window PC. Adapter must support the Serial Port
Profile (SPP).

15
DENT Instruments
ELITEpro
Introduction
Specification
Description
Wi-Fi Adapter (Optional)
Internal Wi-Fi adapter with two separate configurations:
Internal antenna/External 5 dbi antenna: typical conditions <40ft up to 100ft
depending on environment and antenna choice.
Power
Line Power Supply
From L1 Phase to L2 Phase. 80-600V (AC or DC) CAT III DC/50/60Hz, 125 mA, 5 W,
or 10 VA max.
Non-user replaceable .5 Amp internal fuse protection.
Power In (optional)
6-10 VDC maximum, 500 mA minimum
Mechanical
Operating Temperature
-7 to + 60°C (20 to 140°F)
Humidity
5% to 95% non-condensing
Enclosure
ABS plastic, 94-V0 flammability rating
Weight
340 g (12 ounces, excluding sensors and leads)
Dimensions
69 x 58 x 203 mm (2.7” x 2.2” x 8.0”)
Minimum System Requirements
Operating System
Windows® 7 (32 or 64 bit), Windows® 8, Windows® 10 or Vista (32 or 64 bit)
Processor
Pentium Class 1 GHz or more recommended
Hard Drive
50 MB minimum available
Communications Port
One USB port required for logger connection and ELOG software installation

16
DENT Instruments
ELITEpro
Introduction
ELITEpro XC CONNECTIONS
ELITEpro XC Top End View
Power In/Out
Terminal Block for
Analog Input
Connections
Logging On/Comm LED
USB Connection
PhaseChek™
LEDs
Magnetic Mounting Feet
Ethernet connection
Optional Wi-Fi—External
5 dbi antenna
CT Connections
Voltage Lead
Connections

17
DENT Instruments
ELITEpro
Introduction
DENT INSTRUMENTS WARRANTY STATEMENT
Rev 4/9/2015
DENT Instruments, Inc. promises buyer that any product sold by DENT Instruments, Inc. shall be free from material defects in design, materials,
or manufacturing for the period shown on the current datasheet for that product beginning from the manufacture date; provided, however,
that the warranty shall not extend to ordinary wear and tear, or to normally replaceable components (e.g., batteries and humidity sensor
elements). During the warranty period, DENT Instruments, Inc. may repair or replace (at its sole discretion) any product suffering from a
warranty defect and returned freight prepaid by buyer, with no charge to buyer for any warranty repair or replacement. The warranty shall
remain in full force and effect for said period, provided that the product: (1) was installed, operated, and maintained properly; (2) has not been
abused or misused; (3) has not been repaired, altered, or modified outside of DENT Instruments’ authorized facilities; (4) has not been sold
subject to other warranty terms specified at the time of sale (e.g. the optional warranty periods available for CWVS series products). This
warranty provides specific legal rights that may be varied by local laws.
ELITEpro Series and Accessories
Seller warrants to Buyer that ELITEpro Series products and accessories or services are free from substantial defect in material and workmanship
under normal use given prior installation and maintenance for the period of 1 year (12 months) after delivery
LIMITATION OF WARRANTY
CORRECTION OF DEFECTS BY REPAIR, REPLACEMENT, SERVICE OR CREDIT WILL BE AT SELLER’S OPTION AND CONSTITUTE FULFILLMENT OF ALL
OBLIGATIONS TO BUYER FOR BREACH OF WARRANTY.
Seller assumes no warranty liability with respect to defects in the product caused by: (a) modification, repair, installation, operation or
maintenance of the product by anyone other than Seller or its agent, except as described in Seller’s documentation; or (b) the negligent or
other improper use of the product.
Other manufacturers’ equipment purchased by Seller and resold to Buyer will be limited to that manufacturers’ warranty. Seller assumes no
warranty liability for other manufacturers' equipment furnished by Buyer.
No agent, distributor, or representative is authorized to make any warranties on behalf of Seller or to assume for Seller any other liability in
connection with any Seller product or service.
DISCLAIMER OF WARRANTY
Buyer understands and agrees as follows:
THE ABOVE WARRANTY REPLACES ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, & ALL OTHER OBLIGATIONS OR LIABILITIES OF SELLER,
INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. ALL OTHER WARRANTIES ARE DISCLAIMED
AND EXCLUDED BY SELLER.THE FOREGOING WILL BE THE SOLE AND EXCLUSIVE REMEDY WHETHER IN CONTRACT, TORT, OR OTHERWISE, &
SELLER WILL NOT BE LIABLE FOR INJURIES OR DAMAGES CAUSED BY THE GROSS NEGLIGENCE OF SELLER. THIS LIMITATION APPLIES TO ALL
SERVICES AND PRODUCTS DURING AND AFTER THE WARRANTY PERIOD.
LIMITATION OF REMEDIES
IN NO EVENT WILL SELLER BE LIABLE FOR ANY SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR COMMERCIAL LOSSES EVEN IF SELLER
HAS BEEN ADVISED OF THE POSSIBILITY THEREOF.

18
DENT Instruments
ELITEpro
Introduction
GENERAL PROVISIONS
A. Seller reserves the right to subcontract any obligation hereunder.
B. No waiver will be valid unless in writing and no waiver granted will release Buyer from subsequent strict compliance herewith.
C. Seller is not liable for failure or delay in fulfilling its obligations under these Terms and Conditions due to causes beyond its reasonable
control.
D. The laws of the State of Oregon, U.S.A. will apply to all transactions hereunder. Any action hereunder will be brought in Deschutes County,
in the State of Oregon, U.S.A. Any claim, except for nonpayment, will be brought within one year of product shipment or completion of services
and Buyer will be liable for any collection costs or attorney fees.
E. Claims for non-conforming orders must be submitted within 30 days from shipment date.
F. These Terms & Conditions together with Seller’s Order Acknowledgment constitute the entire agreement between the parties with respect
to the subject matter hereof & supersede any prior of contemporaneous agreement or representation written or oral. Any amendment hereto
must be written and signed by Seller.
Service and Support for DENT Instruments Products
Before calling, remember that many questions are answered in the Operator's Guide for your product. Furthermore, there is extensive product
information on the website. In the event that you have a problem with your DENT product that you cannot resolve, please note the following
information. Before contacting DENT Tech Support, note the Model Name and Serial Number for your product. You should also be prepared to
list the hardware and software details of your PC and related equipment.
CONTACT DENT INSTRUMENTS TECHNICAL SUPPORT
541.388.4774 or 800.388.0770
TECHHELP@DENTINSTRUMENTS.COM
WWW.DENTINSTRUMENTS.COM

19
DENT Instruments
ELITEpro
PREPPING & USING THE ELITEPRO
IMPORTANT: DO NOT CONNECT THE ELITEpro XC TO YOUR COMPUTER UNTIL YOU HAVE INSTALLED ELOG 15
Install the ELOG Software on the Computer
Complete this step prior to installation and ascertain that communication between the ELITEpro XC and
PC is established before going into the field.
NOTE: The new path for Data files in ELOG 15 is: C:\Users\UserName\Documents\DENT\ELOG\Data
ELOG is compatible with Windows® 7, Windows® 8, Windows® 10 or Vista (32 or 64 bit).
1) Insert the ELOG CD into the CD/DVD drive.
The installer starts automatically. If it does not, browse to the CD and locate the ELOGInstaller.exe
program. Start the installer by double-clicking ELOGInstaller.exe.
-OR-
Insert the ELOG USB memory stick into the USB port.
Browse to the memory stick and locate the ELOGInstaller.exe program. Start the installer by double-
clicking ELOGInstaller.exe.
2) Follow the directions on the screen to install the software.
NOTE: When ELOG is installed, another program, AutoPoll, is also installed. AutoPoll may be used for
unattended and automatic data retrieval from a number of remote loggers where the computer calls the
loggers. See Appendix D—AUTOPOLL: Unattended Calling from the PC to the Logger.
3) Install the ELITEpro XC Driver on your computer.
a) Connect the USB cable to a USB port on your computer, inserting the other end in the USB port
on the ELITEpro XC (Hint: Look for this USB symbol: ).
NOTE: If a USB cable is used inside an electrical panel it must be rated to the appropriate voltage or
wrapped in an appropriately rated insulating sleeve. The DENT-supplied cable does not meet this
requirement without an insulating sleeve.
b) Perform the setup steps onscreen. You must allow ELOG to install the USB driver for the
ELITEpro XC to work correctly.
c) If the driver installation fails, see Appendix C: Troubleshooting ELITEpro XC Driver Installation.

20
DENT Instruments
ELITEpro
Prepping & Using the ELITEpro
Mount the ELITEpro XC
The ELITEpro XC must be installed in an approved electrical panel or enclosure, following proper
installation practices according to the local electrical codes.
WARNING! Remove the LOGGER from all sources of voltage before mounting.
AVERTISSEMENT! ENLEVEZ L'ENREGISTREUR DE TOUTES LES SOURCES DE TENSION AVANT DE
MONTER.
1) Mount the ELITEpro XC using the black tab on the end of the case containing the PhaseChek and
voltage lead connections. Use a screw to securely mount the ELITEpro XC near a dedicated circuit
disconnect breaker.
NOTE: Allow adequate space for accessing all ports and connectors. The mounting tab is intended for
short term metering (less than 6 months) and must be subject to monthly monitoring by qualified
personnel to verify the ELITEpro XC or its wiring have not shifted.
To remove the logger,remove the screw and roll the magnets off the cabinet it is mounted to.
NOTE: For temporary, spot load monitoring, the four magnets on the back of the ELITEpro XC case can
be utilized to hold the logger on unpainted metal. The maximum combined weight of the logger and any
attachments the magnets can support with a 4 times safety margin is 18oz (500 gm). The logger must be
attended by the electrician when it is temporarily mounted by the four magnets.
In order to prevent injury upon removal, it is recommended practice to roll the device off, due to the
strength of the magnets.