Salutron ComBi D1000 User manual

7. CONSIDERATIONS
1. In order to weaken the influence of the measured material on the accu-
racy of measurement, it is recommended that the calibrations should be
done on the uncoated material to be measured.
2. Probes will eventually wear. Probe life will depend on the number of
measurements taken and how abrasive the coating is. Replacement of a
probe can be fitted by qualified persons only.
8. RESTORE FACTURY SETTINGS
1. When to restore?
It is recommended to restore factory settings in the one of following cases.
A. The gauge does not measure any more.
B. Measurement accuracy is degraded caused by the abraded probe or by
environmental conditions changed greatly.
2. How to restore?
Restore factory settings includes `Fe` setting and `NFe` setting. You can restore
one of them or both of them respectively. Please follow procedures below to
restore factory settings.
1. Please note the symbol on the display is `Fe` or `NFe`. If `Fe` is on the display,
the operation below is restoring the factory setting for `Fe` type. And if `NFe`
on the display, the operation below is restoring the factory setting for `NFe` type.
2. Depress Power key and not release it till `CAL` appears on the display. It is
about 6 seconds from starting depressing Power key.
3. When F:H or nF:H is on display, lift the probe to more than 5 centimetres.
Then press the Zero key again and the gauge returns to measurement state.
The factory setting is restored.
9. HOW TO IMPROVE THE ACCURACY OF FULL RANGE BY LN
1. It is strongly recommended that no changes should be made to the value of Ln
which will seriously affect the accuracy. Its value can be adjusted by professional
person only.
2. To call the Ln, just depressing the Power key and not release it until Ln shows on
the display. It takes about 13 seconds from starting depressing Power key. After
Releasing the Power key, the Ln value shows on the display.
3. The Ln value can be changed by Up/Down key in Fig. 2. Generally, the larger the
Value of Ln, the smaller the reading on a same thickness. A little variation of value
of Ln will cause a great change in reading at high end (e.g. at 500 µm/20mil).
The rules to adjust the value of Ln are as follow:
A. Reading at low end can be adjusted to the exact value by the Up or Down key.
B. To enlarge the Ln, if readings at low end (e.g. at 51 µm) is ok but reading at high
end (e.g. at 432 µm) is too large. On the contrary, to decrease the Ln, if reading at
low end (e.g. at 51 µm) is ok but reading at high end (e.g. at 432 µm) is too small.
C. Repeat procedures from A to B till the readings on every standard foil are
satisfying the accuracy.
Operation Manual
SaluTronComBi D1000
This Coating Thickness Gauge is small in size, light in weight, easy to
carry. Although complex and advanced, it is convenient to use and
operate. Its ruggedness will allow many years of use if proper operating
techniques are followed.
Please read the following instructions carefully and always keep this manual w
ithin easy reach.
1. FEATURES
•It meets the standards of both ISO 2178 and ISO 2360 as well as
DIN, ASTM and BS. Suitable for the laboratory and for use in
harsh field conditions.
•The Fe probes measure the thickness of nonmagnetic materials
(e.g. paint, plastic, porcelain enamel, copper, zinc, aluminium, chrome
etc.) on magnetic materials ( e.g. iron, nickel etc.). Often used to
measure the thickness of galvanizing layer, phosphate layer, copper
tile, aluminium tile, some alloy tile, paper etc.
•The NFe probes measure the thickness of nonmagnetic coatings on
non-magnetic metals. It is used on anodizing, varnish, paint, enamel,
plastic coatings, powder etc. applied to aluminium, brass, non-
magnetic stainless steel etc.
•Automatic substrate recognition
•Manual or automatic shut down
•Two measurement mode:
Single and Continuous
•Wide measuring range and high resolution
•Metric/imperial conversion
•Digital display gives exact reading with no guessing or errors
2. SPECIFICATIONS
Display: 4 digits, 10 mm LCD
Range: 0 ∼1250 µm / 0 ∼50 mil
other ranges may be specified
Resolution: 0.1 µm (0 ∼99.9 µm)
1 µm (over 100 µm)
Accuracy:
Power supply:
Operating condition:
Size:
Weight:
±<6% or 3.0 µm or 0.12mil
(Whichever is the larger)
4 x 1.5 AAA(UM-4) battery
Temp. 0°C ∼50°C, Humidity <80%
126 x 65 x 27 mm (5.0x2.6x1.1 inch)
about 81 g (not including battery)

3. FRONT PANEL DESCRIPTIONS
4. MEASURING PROCEDURE
1. Press the power key (3-3) to switch on the Gauge and ´0` shows up on
the Display (3-2). The gauge will restore the state of last operation
itself, with a symbol `Fe` or `NFe` indicating on the Display. The gauge
enters the auto mode which can automatically recognize the substrate
itself.
2. Place the probe (3-1) onto a coating layer to be measured. The reading
on the Display is the thickness of the coating layer.
3. To take the next measurement, just lift the probe (3-1) to more than
1 centimetre and then repeat the step 2.
4. If you suspect the accuracy of measurement, you should calibrate be-
fore taking the measurements. For the calibration procedures, please
refer to the calibration part 5.
5. The gauge can be switched off by pressing the power key (3-3). On
the other side, the gauge will power itself off about 50 sec. after the
last operation.
6. To change the measurement unit `µm` or `mil` by depressing the
Power key and not releasing it till `UNIT` on the display appears.
And the unit changed after releasing the power key. It takes about
6 sec. from starting pressing the Power key.
7. To change measurement mode from the single to continuous or vice
visa, just depressing the Power key and not releasing it still `SC`
on the Display. The measurement mode changes after releasing the
Power key. It takes about 8 sec. from starting depressing the Power key.
The symbol ((•)) represents continuous mode and `S` represents single
Mode. It is about 8 sec. from starting depressing Power key.
5. CALIBRATION
1. Zero adjustment for `Fe` and `NFe` should be carried out separately.
Take the iron substrate if `Fe` on Display, and take the aluminium
substrate if `NFe` on the Display. Place the probe (3-1) on the sub-
strate steadily. Press the Zero key (3-3) and `0` will be on the Display
before lifting the probe. If pressing the ZERO key but the probe is
not placed on the substrate or an uncoated standard. The zero
calibration is invalid
.
2. Select an appropriate calibration foil according to your measurement range.
3. Place the standard foil selected onto the substrate or the uncoated standard.
4. Place the sensor (3-1) in the middle of the standard foil and lift. The reading on the
display is the value measured. The displayed reading can be corrected by
pressing the UP button A or down button B (which are located in the battery
compartment) while the probe is away from the substrate or the measured body.
See Fig. 2
5. Repeat step 4 until the result is correct.
6. BATTERY REPLACEMENT
1. When it is necessary to replace the battery, i.e. battery voltage
less than approx. 4.5v, the battery symbol will appear on the
display.
2. Slide the Battery Cover (Fig. 1, 3-4) away from the instrument and
remove the batteries.
3. Install the batteries (4 x 1.5v AAA/UM-4) correctly in the case.
4. If the instrument is not to be used for any extended period, remove
batteries.
−
−−
−
+
++
+
+
++
+−
−−
−
−
−−
−
+
++
+
+
++
+−
−−
−
B
Down
A
UP
3 - 2
3
-
4
3 - 3
3-1 Probes inbuilt
3-2 Display
3-3 Zero/Power Key
3-4 Battery
Compartment/Cover
3 - 1
Fig. 1
Accessories:
Carrying case
Operation manual
Fe probe inbuilt
Nfe probe inbuilt
Calibration foil-set
Substrate (Iron)
Substrate
Other Salutron Measuring Instrument manuals