OPERATOR’S MANUAL
VER1.0C615
Fingertip Pulse Oximeter
General Description
Oxygen Saturation is a percentage of Oxyhemoglobin (HbO2) capacity, compounded
with oxygen, by all combinative hemoglobin (Hb) capacity in blood. In other words, it
is consistency of Oxyhemoglobin in blood. It is a very important parameter for the
Respiratory Circulation System. Many respiratory diseases can result in oxygen
saturation being lowered in human blood. Additionally, the following factors can
reduce oxygen saturation: Automatic regulation of organ dysfunction caused by
Anesthesia, Intensive Postoperative Trauma, injuries caused by some medical
examinations. That situation might result in light-headedness, asthenia, and vomiting.
Therefore, it is very important to know the oxygen saturation of a patient so that
doctors can find problems in a timely manner.
The fingertip pulse Oximeter features small size, low power consumption, convenient
operation and portability. It is only necessary for a patient to put one of his fingers into
the fingertip photoelectric sensor for diagnosis, and a display screen will show oxygen
saturation. It has been proven in clinical experiments that it also features high
precision and repeatability.
Measurement Principle
Principle of the Oximeter is as follows: A mathematical formula is established making
use of Lambert Beer Law according to Spectrum Absorption Characteristics of
Reductive hemoglobin(RHb) and Oxyhemoglobin (HbO2) in glow and near-infrared
zones. Operation principle of the instrument: Photoelectric Oxyhemoglobin Inspection
Technology is adopted in accordance with Capacity Pulse Scanning and Recording
Technology, so that two beams of different wavelength of lights (660nm glow and
940nm near infrared light) can be focused onto a human nail tip through a clamping
finger-type sensor. A measured signal obtained by a photosensitive element, will be
shown on the Oximeter’s display through process in electronic circuits and
microprocessor shown on the Oximeter’s display through electronic circuits and a
microprocessor.
Diagram of Operation Principle
1. Red and Infrared-ray Emission Tube
2. Red and Infrared-ray Receipt Tube
Product Operation Scope
Fingertip Pulse Oximeter is a portable non-invasive, spot-check, oxygen saturation of
arterial hemoglobin (SpO2) and pulse rate of adult and pediatric patients at home, and
hospital (including clinical use in internal medicine/surgery, anesthesia, intensive care
and etc). The device is not intended for continuously monitoring.
The PULE OXIMETER requires no routine calibration or maintenance other than
replacement of batteries.
Technical Specifications
1. Display: LED
2. SpO2:
Measurement range:70~99%
Accuracy:70%~99%, ±3%; <70% no definition
3. Pulse Rate:
Measure range:30~235 bpm
Accuracy:30~100bpm ±2bpm; 101~235bpm, ±2%
Pulse Intensity: Bar graph Indicator
4. Power Requirements:
Two AAA alkaline Batteries
Power consumption: Less than 30mA
Low power indication:
Battery Life: Two AAA 1.5V, 800mAh alkaline batteries could be continuously
operated as long as 30 hours.
5. Dimension:
Length: 58mm
Width: 32mm
Height: 37mm
Weight: 33g (without batteries)
6. Environment Requirements:
Operation Temperature: 5~40℃
Storage Temperature: -20~55℃
Ambient Humidity: ≤80%, no condensation in operation;
≤93%, no condensation in storage
7.Measurement Performance in Low Perfusion Condition: The pulse wave is
available without failure when the simulation pulse wave amplitude is at 0.6% using
the test equipment (BIO-TEK INDEX Pulse Oximeter Tester).
8.Interference Resistance Capacity against Ambient Light: Device works
normally when mixed noise produced by BIO-TEK INDEX Pulse Oximeter Tester.
Precautions for use
1 Do not use the pulse oximeter in an MRI or CT environment
2 Do not use the pulse oximeter in situations where alarms are required. The
device has no alarms.
3 Explosion hazard: Do not use the pulse oximeter in an explosive atmosphere.
4 The pulse oximeter is intended only as an adjunct in patient assessment. It
must be used in conjunction with other methods of assessing clinical signs and
symptoms.
5 Check the pulse oximeter sensor application site frequently to determine the
positioning of the sensor and circulation and skin sensitivity of the patient.
6 Do not stretch the adhesive tape while applying the pulse oximeter sensor. This
may cause inaccurate readings or skin blisters.
7 Before use, carefully read the manual.
8 The pulse oximeter has no SpO2alarms; it is not for continuous monitoring, as
indicated by the symbol.
9 Prolonged use or the patient’s condition may require changing the sensor site
periodically. Change sensor site and check skin integrity, circulatory status, and
correct alignment at least every 4 hours.
10 Inaccurate measurements may be caused by autoclaving, ethylene oxide
sterilizing, or immersing the sensors in liquid may cause inaccurate readings.
11 Significant levels of dysfunctional hemoglobins(such as carbonxy-hemoglobin
or methemoglobin).
12 Intravascular dyes such as indocyanine green or methylene blue
13 SpO2measurements may be adversely affected in the presence of high
ambient light. Shield the sensor area (with a surgical towel, or direct sunlight,
for example) if necessary.
14 Excessive patient movement may cause accurate readings.
15 Venous pulsations may cause accurate readings.
16 Placement of a sensor on an extremity with a blood pressure cuff, arterial
catheter, or intravascular line may cause accurate readings.
17 The patient has hypotension, severe vasoconstriction, severe anemia, or
hypothermia.
18 The patient is in cardiac arrest or in shock.
19 Fingernail polish or false fingernails may cause inaccurate SpO2readings.
Follow local ordinances and recycling instructions regarding disposal or
recycling of the device and device components, including batteries.
Product Properties
1. Easy to use.
2. Small in volume, light in weight.
3. Lower power consumption.
4. Low voltage warning will be indicated in visual window when battery voltage is low.
5. The product will automatically be powered off when no signal is in the product for
longer than 8 seconds.
Operation Instructions
1. Install two AAAbatteries into battery cassette, and then close its cover.
2. Open the clamp as shown in the picture below.
3. Insert one finger into the flexible fingertip of the Oximeter fully.
4. Press the power switch once on the front panel.
5. The patient’s finger should remain still while the Oximeter is working.
6. Read corresponding data from display screen.
When you insert a finger into the Oximeter, your nail surface must be upward.
NOTE: Please use the medical alcohol to clean the rubber touching the finger inside of
Oximeter, and clean the test sensor using alcohol before and after operation. (The
rubber inside of the Oximeter is medical rubber, which has no toxin and no harm to the
skin of human being).
Product Accessories
1. One strap
2. Two batteries
3. One user’s manual
Strap Installation
1. Thread thinner end of the strap through the loop.
2. Thread thicker end of the strap through the threaded end before pulling it tightly.
Brief Description of Front Panel
The Pulse Bar graph displays the strength of the pulse rate signal. The height of the
bar graph shows the patient’s pulse strength.
Battery Installation
1. Put the twoAAAbatteries into battery cassette in correct polarities.
2. Press the button down along the arrow and push the battery cover shown as
below:
Notes:
Battery polarities must be correctly installed to avoid
damage to the device.
Please insert or remove batteries in right order, or
may cause damage to the device bracket.
Please remove batteries if the Oximeter will not be
used for a long time.
Maintenance and Storage
1. Replace batteries in time when low voltage lamp is lighted.
2. Clean surface of the fingertip Oximeter before it is used in diagnosis for patients.
3. Remove batteries inside the battery cassette if the Oximeter will not be operated
for a long time.
4. It is best to preserve the product in a place where ambient temperature –20~55
℃and humidity is <93%, no condensation.
5. It is recommended that the product should be kept in a dry environment anytime.
A wet ambient might affect its lifetime and even might damage the product.
6. Please follow the law of the local government to deal with used batteries.
Calibrating the Oximeter
1. The functional tester cannot be used to assess the accuracy of the oximeter.
2. The test method used to establish the SpO2accuracy is clinical testing. The
oximeter used to measure the arterial haemoglobin oxygen saturation levels and
these levels are to be compared to the levels determined from arterial blood
sampling with a CO-oximeter.
3. Index 2 that made by Bioteck company is a function tester. Set Tech to 1, R curve
to 2, and then users can use this particular calibration curve to measure the
oximeter.
Declaration
EMC of this product complies with IEC60601-1-2 standard.
The materials which the user can come into contact have no toxicity and no action on
tissues, comply with ISO10993-1, ISO10993-5 and ISO10993-10.
Guidance and manufacture’s declaration – electromagnetic
emissions-for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
The Pulse Oximeter is intended for use in the electromagnetic environment
specified below. The customer or the user of the Pulse Oximeter should assure
that it is used in such and environment.
Emission test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11 Group 1 The Pulse Oximeter uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emission
CISPR 11 Class B The Pulse Oximeter is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Possible Problems and resolutions
Problems Possible reason Solution
SpO2or PR can
not be shown
normally
1.Finger is not plugged correctly
2.Patient’s Oxyhemoglobin value is too
low to be measured.
1. Retry by plugging the finger.
2. Try some more times, If you can make
sure about no problem existing in the
product. Please go to a hospital for exact
diagnosis.
SpO2or PR is
shown unstably 1.Finger might not be plugged deeply
enough.
2.Finger is trembling or patient’s body
is in movement status.
1.Retry by plugging the finger.
2.Try not to move.
The Oximeter
can not be
powered on
1.Power of batteries might be
inadequate or not be there at all.
2.Batteries might be installed
incorrectly.
3.The Oximeter might be damaged.
1.Please replace batteries
2.Please reinstall the batteries
3.Please contact with local customer
service centre.
Indication lamps
are suddenly off 1.The product is automatically
powered off when no signal is
detected longer than 8 seconds
2. Lower power
1.Normal
2.Replace the batteries
“Error3”or
“Error4” is
displayed on
screen.
1 Low power
2 Receiving tube being shielded or
damaged together with broken
connector.
3.Mechanical malfunction for
receive-emission tube
4 Amp circuit malfunction.
1 Change new batteries.
2 Please contact with local customer
service center.
3 Please contact with local customer
service center.
4 Please contact with local customer
service center.
Symbol Definitions
Symbol Definition
Type BF applied part
Attention, consult accompanying documents.
SpO2﹪Oxygen saturation
Pulse rate (BPM)
Low power indication
NOT for continuous monitoring
SN Serial No.
Applicable models
SM-210 SM-220
Note: The illustrations used in this manual may differ slightly from the appearance of the
actual product.
Manufactured for:IRoams Network
Add:713 W. Duarte Rd #G810
Arcadia, CA 91007 USA
ALL RIGHTS RESERVED
Issue date: 2011-2-18

This manual suits for next models

1

Other SantaMedical Medical Equipment manuals

SantaMedical SM-165 User manual

SantaMedical

SantaMedical SM-165 User manual

SantaMedical SM-165 User manual

SantaMedical

SantaMedical SM-165 User manual

Popular Medical Equipment manuals by other brands

Otto Bock Genium X3 3B5-3 Instructions for use

Otto Bock

Otto Bock Genium X3 3B5-3 Instructions for use

Ossur Rebound Instructions for use

Ossur

Ossur Rebound Instructions for use

Leckey Splashy manual

Leckey

Leckey Splashy manual

Boston Scientific Artisan Directions for use

Boston Scientific

Boston Scientific Artisan Directions for use

Invacare Electric Portable Patient Lift RPA600-1E operating & maintenance manual

Invacare

Invacare Electric Portable Patient Lift RPA600-1E operating & maintenance manual

ACKERMANN 16-2000-700 Instructions for use

ACKERMANN

ACKERMANN 16-2000-700 Instructions for use

Motion Concepts F Series owner's manual

Motion Concepts

Motion Concepts F Series owner's manual

Instrumentarium ORTHOPANTOMOGRAPH OP300 quick guide

Instrumentarium

Instrumentarium ORTHOPANTOMOGRAPH OP300 quick guide

MX HEALTH MX86200GR user manual

MX HEALTH

MX HEALTH MX86200GR user manual

HEINE EN200 Series Instructions for use

HEINE

HEINE EN200 Series Instructions for use

inhealth Blom-Singer Instructions for use

inhealth

inhealth Blom-Singer Instructions for use

Orliman 3010-C INSTRUCTIONS FOR USE AND PRESERVATION

Orliman

Orliman 3010-C INSTRUCTIONS FOR USE AND PRESERVATION

DENTSPLY Cavitron Prophy-Jet Directions for use

DENTSPLY

DENTSPLY Cavitron Prophy-Jet Directions for use

Arjo Tornado Assembly and installation instructions

Arjo

Arjo Tornado Assembly and installation instructions

Inogen One G3 Equipment guide

Inogen

Inogen One G3 Equipment guide

Gima 45560 manual

Gima

Gima 45560 manual

Heaco ECG600G user manual

Heaco

Heaco ECG600G user manual

EKOM DK50 4VR/50 S/M user manual

EKOM

EKOM DK50 4VR/50 S/M user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.