Settingthe clock (initial operation or after battery removal for longer periods)
Sampling
Hints for battery and mains power operation
Selection of the right place of measurement
Further processing of measured results
Download the software Radon Vision and the USB driver from our website (www.sarad.de) and install both on your computer.
Connect the USB cable to the PC and the USB interface (C) of the instrument. Start Radon Vision and click onto button to obtain
the COM port assigned to the unit. Open the setup dialog by clicking onto button. Set clock using the button.
he manual of Radon Vision on our website.
Turn on the display by button (B) and set the slide switch (D) to the right. The first measured results will be presented after four
hours (Avg = Average of the whole measurement, 24h = average of the last 24 hours). The display will alternate between
temperature/humidity and Radon automatically.
The Radon Scout Home can be operated several months by battery. Here are some tips for a long-lasting battery operation:
Use batteries of best quality brand that are especially optimized for long term use and low power consumption.
Reduce usage of display and backlight.
In case of low battery the red indicator (A) starts blinking. The measurement continues for a few days but it is impossible to start a
new measurement.
The internal clock will be sustained for a couple of minutes
during battery replacing. For mains power operation, use always a wall adapter with USB connector. Do not use the USB port of a
computer for power supply.
Do not expose the instrument to permanent shocks or vibrations. Place the unit stationary in that room where you will stay most
frequently. Avoid heat sources or radiofrequency emitting devices (e.g. mobile phones, microwave stoves, DECT phones or base
stations) in the immediate surrounding of the instrument. The distance from the wall should be more than 25cm (10inches)
because the wall itself could be the reason for an increasedRadon concentration.
Measured values of the previous eight years are stored in the instrument (circular memory structure). The measured data can be
read t shown, exported or printed by Radon Vision. Refer software manual (SARAD website) to find
out the complete functionality of Radon Vision.
The unit records the readings every four-hours. When operated from battery,
the display turns off after 30 seconds. Use button (B) to turn it on again. Another keystroke turns on the backlight. To cancel a
measurement, shift the slide switch (D) into the left position.
hrough the USB port (C) and
?
?
Remove discharged batteries immediately. Do not dispose batteries in domestic waste. You are legally
obliged to return them to the appropriate waste collection centres.
you will find
t
Manual Radon Scout Home
Thank you for the purchase of a Radon Scout Home. The instrument has
been designed for long-term indoor measurements of the Radon
concentration. The manufacturer cannot guarantee for damages that will
result from use other than intended. Please read the manual carefully
before operating the unit.
Inserting/replacing the batteries
Set the slide switch (D) into the
left position (sampling stopped)
Unscrew both screws (E) at
the rear of the instrument
Remove rear panel by
slightly tilting on the
upper rounding
Insert/replace the batteries
(AAA/LR6) on the left and right
(bottom first – check polarity)
12/2016 - SARAD GmbH www.sarad.de
1 2 3 4
E
D
C
AB
EN