Contents
1.ScentinalOverview.......................................................................................2
2.ScentinalFeatures........................................................................................6
3.ScentinalDataServerandCommunicationProtocol..................................17
4.OdourMonitoring.......................................................................................21
5.TotalOdourManagement(TOMS).............................................................23
6.ScentinalAccessories..................................................................................25
7.ScentinalApplication(IndustriesandRecommendedSensors).................26
8.Installation,MaintenanceandOperationCost..........................................31
9.After‐SaleSupport......................................................................................34
10.Scentinal...................................................................................................35
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|2
1. Scentinal Overview
1.1AnIntelligentOdour&AirQualityMonitoringStation
Scentinal uses up to 20 sensing modules ranging from Photo-Ionization Detector, Non-Dispersive Infrared Detector,
Electro-Chemical Cell, Laser Scattered Counter and Metal Oxide Sensors. All data is stored locally and sent to the
central server accessible by the user. The Sensor Information Management System is used to not only store and
display monitoring results but also provide remote configuration, calibration, and diagnosis of multiple Scentinal
units.
Scentinal is a simultaneous ambient pollutant and odour emission monitoring system based on high accuracy (ppb
level) sensing technology. Scentinal can provide continuous monitoring of odorous gases such as H2S, Ammonia,
VOCs, and SO2 as well as non-odorous gases such as CO2 and Methane.
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|3
++++
Scentinal is easy to setup and use. Each Scentinal
comes pre-loaded with a SIM card so all you have
to do is mount it to a wall or a pole and power the
unit up. Scentinal will detect its location using built-
in GPS, and start transmitting to the central server.
All configuration and maintenance can be done on
the on-unit 7” touch screen monitor or remotely
through the SIMS software. (To know more about
SIMS software check page 17)
1.3TheMostEasytoUseAQStation
Scentinal is more than just a monitoring station, it can also act as an
intelligent process controller. Scentinal is equipped with up to three
optional programmable relays that can be used to activate alarms,
mitigation technology, and any other electrical device based on
monitored parameters. For example, the Scentinal can be used to
measure emissions from the exhaust of a bio-scrubber and bypass
carbon polishing filter if odour levels are below set limits, or activate
misting stations if odour levels at perimeter are exceeding 5 OU/m3.
The possibilities are endless!
1.2AnIntelligentProcessController
SETALARM ACTIVATE
FILTER
ACTIVATE
FANS SENDSMS
SPRAY
NEUTRALIZER
BYPASS
CARBON
FILTER
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|4
Scentinal is extremely affordable. At a fraction of the cost of a traditional
air quality station, Scentinal can provide pollutant and odour emission data
that would be critical in meeting your environmental objectives. In addition
to an affordable purchase price, Scentinal has minimal operating cost. Just
electricity and GPRS data (GPRS costs are less than $100 per year). Add to
that Scentroid’s unparalleled 2 year comprehensive warranty, covering all
aspects of the instrument including even sensors, and Scentinal becomes
the most affordable solution for continuous pollutant and odour monitoring
in the world!
1.4TheMostAffordableSolution
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|5
ProductNameScentinalSL50
Maximum#ofSensors20
TypeofsensorsPID,NDIR,EC,LaserParticulatecounter,andMOS
SamplingrateAdjustablefrom0.5to999min
#ofSamplingports1to3
Weight77 LB
Size24"x20"x8"
Powerrequirements100‐240V50/60HZ2A
PowerConsumption30WwithoutAC‐150WwithAC
Communication GPRS(Default),LAN(Default),WIFI(optional)
On‐BoardDatastorage64GB‐SDCard
CloudServerIncludedbyDefault
LocalServerOptional
On‐BoardServerIncludedbyDefault
UserInterface7”touchscreenonPaneldoorand
RemoteaccessSensorInformationManagementSystem
AmbientTemperature
range
0to35°CwithoutACsystem
‐50to+50°CwithHeatingandACsystem
Sampleconditions‐50to+50°Cand10–90%RHwithoutpre‐dilutionsystem
‐50to120°Cand0–100%RHwithpre‐dilutionsystem
Decontamination Automatedprocedureusingoxidizationtechnology
CalibrationManualusingcalibrationgasandOn‐boardscreen
Optionalautomaticcalibrationusingbuiltincalibrationgas
Warranty24Monthsfullwarrantyonallpartsincludingsensors
Sensorreplacement
frequency
Sensordependant–first2yearscoveredbywarranty
Software
SensorInformationmanagementSystem–freeaccessforlifeofproduct
CabinetNEMA4X
MountinghardwareWallandpolemountinghardwareincluded
1.5ScentinalSpecifications
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|6
2. Scentinal Features
2.1Reliabilitythrough300%Redundancy
Scentinal provides 3 levels of redundancy through:
a) Storage of all data on SD card
b) Transmission and storage of data on the on-board server
c) Transmission and storage of data on cloud/localized Server
2.2EachScentinalisalsoaServer
Standard on each Scentinal is a separate dedicated micro-
computer acting as an on-device server to run Scentroid’s Sensor
Information Management System. Through the 7” touchscreen,
users are able to view historical data, change system configuration,
recalibrate the instrument, and provide real-time diagnostics.
The Scentinal on-device server is capable of storing data locally for
up to 5 years. This data can be polled by the central station as
required and if the system communication is lost then the system
can recover without any data lost. In fact it is entirely possible to
run Scentinal with no external server. The system is password
protected to ensure only authorized users have access to critical
system parameters. The on-device server not only provides
additional redundancy but also makes the system extremely easy
to use.
SO
2
CO
H
2
S
CO
2
CH
4
OS
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|7
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|8
2.3ScentinalforProcessControl
In addition to email and SMS alarms, every Scentinal is equipped with 3 industrial relays of up to 20 Amps that can be used
to control a variety of equipment. For example Scentinal can be used to:
1. Provide visual and audible alarms
2. Engage odour control technologies such as misting systems when fence line pollutants levels exceed limits
3. Secondary polishing filter only when needed to reduce operating costs.
4. Activate external sampling pump for bag sampling.
The limits and conditions for engagement of each relay can be set based on pollutants or total odour units. All limits and
activation conditions are done through SIMS via remote server or on-device touch screen.
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|9
2.4AutomatedDecontaminationforMaintenancefreeOperation
Scentinal uses a new method of decontamination to ensure accurate reading even at ppb levels. Periodically
(interval is set through remote SIMS software or on-device server) the system injects carbon filtered air into the
sample line to measure contamination. If contamination is detected, Scentinal will start its ozone generator and
flush the lines, pumps, and sensors with ozone and hydroxyl. These reactive molecules will destroy all bacteria,
mold, and pollutants. Scentinal will then inject carbon filtered air again and ensure the contamination has been
dealt with. This means that once installed, Scentinal is virtually maintenance free.
2.5FlexibleSensingandModularDesign
The Scentinal product can be equipped with up to 20 sensors from
the extensive sensor list (Sensor list are on page 15). There are 5
categories of sensors including:
1. Photo-Ionization Detector,
2. Non-Dispersive Infrared Sensor,
3. Electro-Chemical Sensor,
4. Laser Scattered Counter (for PM1-10), and
5. Metal Oxide Sensor.
Each Scentinal can be customized with the exact sensors required for
the application at hand. Our flexible pricing means you pay for exactly
what you need.
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|10
2.6InstallationandConnectivity
Each Scentinal has a micro controller capable of recording the unit’s
exact GPS position. This position is sent to the central server with
each data transfer. At the time of the installation the technician
simply needs to place the Scentinal unit and power it on. The central
computer will automatically recognize the Scentinal unit and know
of its exact location. To reconfigure the network the physical sensor
can be moved and the system will learn and adapt to this change.
Multiple Scentinal units can be configured within one monitoring
area. The connectivity is flexible and secure using one of the
following options:
Encrypted transfer over GPRS
WIFI
LAN
Analog/SCADA
The system can either connect to a local server or Scentroid’s cloud
based SIMS server. Cloud based SIMS server is included and free
for the life of the Scentinal. It is even possible to operate Scentinal
with no centralized servers thanks to its on-board server. (To know
more about SIMS please check page 17)
2.7PrepaidGPRSSIMCARDIncluded
Each unit comes complete with a prepaid GPRS SIM card. The unit comes pre-configured with all
necessary information to transmit data to the central server. Data transfer costs are fully covered for the
first year. Once powered, the instrument will determine its location using a built in GPS receiver and start
transmitting data to the closest SIMS server.
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|11
2.8Portability
The small form factor and low
weight of the Scentinal makes it
easy to transport and install.
Scentinal weighs just 10 Kg and
can be setup in a matter of
minutes.
Easy to use
Portable
Quick to setup
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|12
2.9MultipleSamplingPort
Scentinal can be equipped with up to 3 sample
ports to allow the unit to measure pollutants
from different process points or locations. For
example Scentinal can be setup to record input
and output of a bio trickling filter to provide live
efficiency calculations. The ¼” sample ports
can be outfitted with ambient sampling hoods
or be directly connected to a PTFE line.
2.10NoiseMonitoring
In addition to gas and particulate monitoring,
Scentinal can also be equipped with outdoor
Class 1 noise sensor. No additional
equipment are required to measure and
record ambient noise. Integrated automatic
calibration make Scentinal convenient and
accurate. Measurement range is from 30 to
100 dB (A).
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|13
.
2.11WindSensor
2.12IntegrationwithTotalOdourManagementSystemSoftware(TOMS)
Scentinal can be equipped with on-board wind direction and wind speed sensor. This information can be used to determine
localized wind conditions such as turbulence and down drafts. Scentroid offers either ultrasonic or cup and vane wind sensors
to fit any application and budget. For gathering meteorological data, Scentroid provides an independent weather station that
can installed in accordance to USEPA guidelines. The Scentroid weather station is equipped with its own communication module
and will seamlessly integrate with Scentinal in the SIMS local or cloud based software.
The second stage of implementing a Scentinal sensing
system is to input the data into Scentroid’s real-time
dispersion modeling software TOMS. TOMS offers a
complete, integrated suite for odour management. The
system provides a perfect integration of real-time odour
impact estimation with management of odour complaints
from neighbouring residents. The simple to use software
uses Scentinal sensory data, in field-olfactometry and live
weather data to produce real time odour plumes showing
you the exact location and level of odour emissions.
Neighbouring complaints are automatically logged and
compared to odour emissions for fast and efficient
validation. The system will even provide forecasting of
when and where next odour episodes will be allowing the
plant to optimize its operation
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|14
2.13CreateAlarmsandNotifications
“Scentinal Information Management System“(SIMS) also
provides the platform to set up alarm levels. Alarm levels can
be setup based on individual pollutants or on the total
calculated odour units. Alarms will trigger SMS or emails to
plant operators for immediate action. Scentinal can also be
setup to provide localized visual and audible alarms.
The user can remotely configure each Scentinal, providing it
with the desired sampling rate, transmission rate, purging
frequency and much more. Scentinal can also transmit data
over WIFI or LAN network to a local server running a client SIMS
database for added security.
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|15
2.14ScentinalSensorList
Type
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|16
Type
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|17
3. Scentinal Data Server and Communication Protocol
3.1CloudBasedHosting
The central monitoring station is hosted on a secure cloud based
server allowing remote access via any internet based computer.
The access is restricted and the data is encrypted for maximum
security. Users are given a password and user name which will
define their permission level. Operator access can simply view
the results and create reports while administrator access can
reconfigure the system and set all parameters.
The monitoring station is designed to collect all data from the
sensors and present the sensor data in an easy to understand
graphical interface.
3.2LocalServer(Optional)
It is possible to hosting of SIMS (Scentinal Information
Management System) on a local server with secure Wi-Fi or LAN
connection. Scentroid will provide all necessarily hardware and
software to setup a local server.
This option includes:
• Computer hardware (including monitor, keyboard…)
• SIMS software
• Ethernet hub
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|18
3.3CommunicationProtocols
GPRS
Scentinal comes as default with GPRS module for wireless communication through existing cell towers. The communication
is encrypted and sent to Scentroid’s secured SIMS cloud server. The SIM card required is included and pre-paid for one
year. The SIM card is capable in operating in any country with no additional roaming charges. Local SIM cards can also be
used if required.
WIFI/LAN
Scentinal can also transmit data over WIFI or LAN network to Scentroid’s cloud server or a secured local server. LAN
connection is included by default and WIFI is included as an option when ordering.
Analog/SCADA
Scentinal is an open platform allowing interface to many other instruments and systems such as the plant SCADA. Scentinal
can be setup to transmit any one of the sensor outputs as 0-5V or 4-20mA to be connected to plant monitoring systems
such as SCADA.
3.4ScentinalInformationManagementSystem(SIMS)
Scentinal Information Management System, SIMS, is an all-inclusive software, used to view historical data, run diagnostics,
configure, and set alarm levels for Scentinal. Provided as part of the Scentinal package, the software is installed on:
1. On-board server (default)
2. Scentroid’s cloud based server (default)
3. Client’s localized server (optional).
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com

www.scentroid.comPage|19
(ES) Equipements Scientifiques SA - Département Bio-tests & Industries - 127 rue de Buzenval BP 26 - 92380 Garches
Tél. 01 47 95 99 90 - Fax. 01 47 01 16 22 - e-mail: bio@es-france.com - Site Web: www.es-france.com
Table of contents
Popular Measuring Instrument manuals by other brands

Hanna Instruments
Hanna Instruments HI83325 instruction manual

Vega
Vega VEGACAL 67 operating instructions

IFM Electronic
IFM Electronic SD8000 operating instructions

Acoustical Systems
Acoustical Systems UNI P2S owner's manual

Veris Industries
Veris Industries Hawkeye H749 installation guide

AFL
AFL M310 user guide