To verify a sample count:
1. After counting colonies using one pen color, switch to other pen color.
(We recommend starting with red and switching to black)
2. Press Button 1 one time. The Up arrow will disappear and the Down arrow will appear.
Note: If BOTH arrows appear and flash in the display, do NOT press Button 1 again until the arrows
STOP flashing. Pressing Button 1 again while both arrows are flashing will clear the memory for that
sample identification number.
3. Recount colonies in plate. During recount:
Pen tip will mark surface of plate cover
Audible beep will sound (if sound is ON)
LED will flash once
Counter total will DECREASE
Note: The original sample count (1st) will be continuously shown in the upper right hand corner of the
display. The TOTAL count (2nd) will be shown in the lower right hand corner. The TOTAL
count is the sum of the original count plus the recount.
Changing SAMPLE ID number:
To advance the sample ID number: Press Button 2
Audible sound will indicate change
Sample ID number will INCREASE (e.g. 01->02->03->…35)
To review the sample ID numbers:
1. Press and hold BOTH Button 2 and the Back Light button. Audible beep will sound
2. Press Button 2
Audible sound will indicate change
Sample ID number will DECREASE (e.g. 35->34->33->…1)
To clear a specific SAMPLE ID memory:
1. Scroll through the Sample Identification numbers by pressing Button 2 repeatedly until
the correct Sample ID number is shown in the display.
2. Press and hold Button 1 until both Up and Down Arrows flash in the display.
3. While both arrows are flashing, press Button 1.
LED will flash once
Audible Beep will sound (if sound is ON)
Counter display will show +0000
To clear ALL Sample ID memories:
1. Press and hold Power button for 3 seconds. Display becomes blank. Unit is turned off.
2. Press Power button to turn unit back on. Memory has been cleared.
Battery Replacement:
1. Locate and loosen retaining screw on battery cover on the underside of the unit.
2. Depress battery retaining clip to release battery from holder
3. Insert new battery with + side facing UP.
4. Slide battery cover back into place and tighten retaining screw to secure cover in place.
Cleaning:
To clean unit, wet paper towel with 70% ethanol solution and wipe down unit.
Helpful Hints:
1. Press firmly to make sure pen depresses with each count. The counter will emit an
audible sound and the LED will flash each time a count is registered.
2. To preserve pen life, cap the pens when they are not in use.
3. Operating Temperature: 5˚C to 40˚C @ 50-80% RH.
4. Storage Temperature: -10˚C to 40˚C @ 80% RH max.
Related Products:
F37864-2000
Catalog No. Description
F37864-2000 Mini Light Box II
F37865-0000 Mini Magnifier
F37736-0006 Small Cell Spreader for 6cm plates with 16mm (.62") wide spreading bar
F37736-0009 Large Cell Spreader for 9cm plates with 25mm (.62”) wide spreading bar
F40913-0000 Inoculating turn Table
F37865-0000