SD Biosensor STANDARD LipidoCare User manual

1
Thank you for your purchase of the STANDARD LipidoCare Analyzer!
This User Manual contains all of the information needed to use the
Analyzer. Please read this User Manual carefully before using the
Analyzer. Familiarize yourself with the required preparations and the
measurement procedure before performing the rst measurement.
Also read the package insert of the lipid test strips to be used for the
planned test.
If you cannot reach SD Biosensor Customer Care Service, please
contact your healthcare professional or local distributor. You can
also visit www.sdbiosensor.com for product demonstrations.
Thank you again for choosing the STANDARD LipidoCare Analyzer.

2
Abbreviations
TC Total Cholesterol
TG Triglycerides
HDL High Density Lipoprotein cholesterol
LDL Low Density Lipoprotein cholesterol
The packaging materials, labels and instruction of use for STANDARD LipidoCare Analyzer may
contain the following symbols or abbreviations which are listed below with their meaning:
Please consult instructions
for use
EC REP
Batch code
Caution, consult accompanying
documents
EC REP
To discard it separately from
other household waste
Consult instructions for use
EC REP
To indicate the temperature limitations in
which the transport package has to be
kept and handled
In Vitro diagnostic use
EC REP
Use by / Expiry date
Manufacturer
EC REP
Authorized Representative in the
European Community
Reference number
EC REP
Contains Sucient for <n> Tests
Date of manufacture Serial number of the analyzer
Do not reuse

3
CONTENTS
CHAPTER 1. Understanding Your New Analyzer 4
1. Indication for Use; Purpose of the Analyzer 4
2. Test principle 4
3. Before You Start Testing 4
4. The STANDARD LipidoCare Analyzer 5
5. Power supply 7
6. Test Strips 7
7. Accessories 8
CHAPTER 2. Testing the STANDARD LipidoCare Analyzer 8
1. Operating the Analyzer 8
2. Preparing for measurement 15
3. Running a test - Lipid 18
4. Running a test - Glucose 20
5. Running a test - Lipid/Glucose 21
6. Understanding Your Results 22
CHAPTER 3. Using the Analyzer Memory 24
1. Analyzer Memory 24
2. Displaying stored measured values 25
3. Displaying Glucose Averages 27
4. Delete Memory 28
CHAPTER 4. Printing 28
CHAPTER 5. Data transfer 29
CHAPTER 6. Bluetooth communication 29
1. Data Trasfer 29
2. Technical Information 30
CHAPTER 7. Control Solution Test 31
1. Control Solution Test 31
2. Control Solution Test for Lipid 31
3. Control Solution Test for Glucose 32
4. Troubleshooting control solution 33
CHAPTER 8. Check Strip Test 33
1. How to Use a Lipid Check Strip 33
2. How to Use a STANDARD Glucose Check Strip 34
CHAPTER 9. Maintenance and Troubleshooting 34
1. Cleaning your Analyzer 34
2. Maintenance, Testing and Transportation 36
CHAPTER 10. Screen Messages and Troubleshooting 36
1. Warning message 36
2. Error message 37
CHAPTER 11. Warnings, Precautions and Limitations 40
1. Test Strip for Lipid 40
2. Test Strip for Glucose 40
CHAPTER 12. Product Technical Information 40
1. Analyzer Specications 40
2. Test strip for Lipid 41
3. Test strip for Glucose 41
4. Electromagnetic Compatibility 41
Appendix 1: Information for Healthcare Professionals 42
Appendix 2: References 42

4
CHAPTER 1. Understanding New Analyzer
1. Indication for Use; Purpose of the Analyzer
The STANDARD LipidoCare Analyzer is designed to measure the quantitative measurement of the blood parameters: Total Cholesterol(‘TC’), Triglycerides(‘TG’),
High-Density Lipoprotein Cholesterol(‘HDL’), calculated ‘LDL’, ‘LDL/HDL’, ‘non-HDL’ and glucose. This system is intended for in vitro diagnostic use to test fresh
capillary whole blood, venous whole blood plasma or serum. The STANDARD LipidoCare Lipid Test system should be only used with the STANDARD LipidoCare
Lipid Test Strip and either the STANDARD LipidoCare Blood Glucose Test Strip or the SD Codefree Blood Glucose Test strip. The system is suitable for professional
use as well as for self-testing, and is a valuable tool in monitoring diabetes and cholesterol levels, but it is not intended to diagnose diabetes or cholesterol-
related cases.
2. Test principle
LIPID
By means of a codechip, the Analyzer reads the lot-specic characteristics of the lipid test strips currently in use. Then an unused test strip is inserted into the
Analyzer. While inserted, the application area of the test strip is illuminated by an LED (light-emitting diode) from below. Before the actual measurement is
performed, the reection behaviour of the test strip is determined by means of the light which is reected (from the application area).The blood sample is then
applied to the application area and the measurement chamber ap is closed. The constituent to be determined in the applied sample undergoes an enzymatic
reaction with formation of a dye.The amount of dye formed increases with the concentration of the substance to be determined. After a certain period of time,
the colour intensity is measured by illuminating the application area again from below using the LED. The intensity of the reected light is measured with a
detector (reectance photometry). The measured value is determined from the signal strength of the reected light, with the previously measured blank value
and the read lot-specic information (codechip) also being considered. Finally, the result is displayed and simultaneously stored in the memory.
GLUCOSE
The glucose test strip is designed with an electrode that measures glucose levels. Glucose in the blood sample mixes with reagent on the test strip that causes
a small electric current. The amount of current that is created depends on how much glucose is in the blood. The STANDARD LipidoCare Analyzer measures the
current that is created and converts the measurement to the amount of glucose that is in the blood. The blood glucose result is displayed on the LCD display. By
touching a drop of blood to the tip of the glucose test strip, the strip’s reaction chamber automatically draws the blood into the strip through capillary action.
When the chamber is full, the STANDARD LipidoCare Analyzer starts to measure the blood glucose level.
3. Before You Start Testing
Carefully read and follow the instructions in the measuring and package inserts for the lipid test strips and control solutions. It is very
important to follow the instructions in order to prevent an incorrect result or improper treatment.
Samples
Lipid
The STANDARD LipidoCare Analyzer for lipid testing is designed for measuring fresh capillary whole blood or venous blood (venous whole blood, serum or
plasma) sample.
* Professional use only for venous blood.
Glucose
The STANDARD LipidoCare Analyzer for glucose is designed for testing fresh capillary whole blood samples (for example, blood from your ngertip, palm, upper
arm, or forearm). Do not use the STANDARD LipidoCare Analyzer for glucose of serum or plasma or arterial, venous whole blood.
Refer to the Safety Information
There is a potential risk of infection. We recommend that healthcare professionals using the STANDARD LipidoCare Analyzer to perform measurements for more
than one patient use gloves and follow all other locally applicable health and safety regulations.
Operating conditions
To ensure proper function of your STANDARD LipidoCare Analyzer, observe the following guidelines:
• Operate the analyzer only with the acceptable environmental conditions; temperature: 5°C to 45°C, humidity: maximum relative humidity 80 % for

5
temperatures up to 31°C decreasing linearly to 39% relative humidity at 45°C, altitude: up to 3,776m.
• In order to perform a measurement, place the Analyzer on a level surface or hold it in your hand.
• Strong electromagnetic elds may impair the function of the Analyzer. Do not use the Analyzer close to sources of strong electromagnetic radiation.
4. The STANDARD LipidoCare Analyzer
The Analyzer includes:
* STANDARD LipidoCare Analyzer
* STANDARD Lipid Check Strip
* STANDARD Glucose Check Strip
* User Manual
* Quick Guide
* 4 AAA 1.5 V batteries
STANDARD LipidoCare Analyzer
F
A
B
C
E
D
ADisplay Shows a test result, messages, and the test
results stored in memory.
B
Arrow Buttons
Used for setup and review of memory moving to the
right and left.
CON/OFF Button Press to turn Analyzer ON or
OFF.
DLipid test strip slot Insert a lipid test strip here.
EMeasurement chamber flap Open to apply sample.
FGlucose Test Strip Slot Insert a glucose test strip
here.
L
JK
G
H
I
G
Battery compartment lid
Provide access to the battery compartment (4 AAA 1.5
V alkaline manganese batteries).
HCodechip Slot
Insert a codechip for lipid test strip.
IMeasurement chamber cover
You can remove this cover to clean the test strip guide.
J
Data port
Download test results to a personal computer if you
have the software. (SD LMS)
KPrinter port
Print test results to external thermal printer.
LSET/PRT Button
Press to set the Analyzer or print the test result.

6
Display of STANDARD LipidoCare Analyzer
BLSE
The symbols in the display have the following meaning.
BLSE
Indicate Codechip sign
BLSE
Warning when the battery is low or must be replaced
BLSE
Indicate printing
BLSE
Indicate index number of memory
BLSE
Codechip number or memory number stored in
memory
BLSE
Indicate during the analyzer setting
BLSE
Indicate blood checking
BLSE
Indicate beep setting
BLSE
Indicate if environmental
temperature is exceeded during testing
BLSE
Indicate the Lipid Prole and TCHDL strip
BLSE
Indicate open or close ap
BLSE
Unit of the test result
BLSE
Test result
BLSE
Indicate a test result stored in memory (Single parameter
mode)
BLSE
Indicate the average result
BLSE
Indicate a test result stored in memory (All parameter mode)
BLSE
Indicate the blood sample type.
BL: Capillary or Venous whole blood
SE: Serum or plasma
BLSE
Indicate a control
solution test result
BLSE
Indicate blood checking
BLSE
Indicate beep setting

7
BLSE
Glucose test strip
BLSE
Lipid test strip
BLSE
Indicate if the lipid test strip is exceeded the
expiration date
BLSE
Date
Indicate the Bluetooth ('bt') only for 02LA20G
BLSE
Indicate Measurement time
BLSE
Indicate Triglycerides
BLSE
Indicate Total cholesterol
BLSE
Indicate LDL cholesterol
BLSE
Indicate HDL cholesterol
BLSE
Indicate non-HDL cholesterol
BLSE
Indicate glucose
BLSE
Indicate LDL/HDL
5. Power supply
To save power, the Analyzer turns itself o after 5 minutes unless a button is pressed or a new test strip is inserted. When the Analyzer turns itself o, all test
results obtained so far remain in the memory. Battery life is expected at approximately 1000 measurements.When the battery warning is displayed for the rst
time, approximately 50 measurements can still be performed. In this case, replace the batteries as soon as possible.
When replacing the new batteries, you must re-enter the date and time. Use only AAA alkaline manganese batteries.
Measurement results including the related measurement date and time as well as all other Analyzer settings, remain stored even when no batteries are inserted.
Please respect the environment and dispose of used batteries according to your local regulations and laws.
Do not throw batteries onto an open re.There is a risk of explosion.
6. Test Strips
- Lipid strip -
Item Lipid Prole TC TG
Color Blue Red Yellow
Glucose strip Lipid Check Strip Glucose Check Strip
Glucose
Arming Knob Electrode

8
7. Accessories
Optional Accessories
STANDARD Micropipette 35µl
STANDARD Micropipette 10µl
STANDARD EziTube+ 35µl
STANDARD EziTube+ 10µl
STANDARD Lancets (30G)
STANDARD Lancing Device
STANDARD Safety Lancets (23G)
STANDARD Alchol Swab
STANDARDThermal Printer (model: MPT-700)
STANDARDThermal Printer (model: SPP-R200)
STANDARDThermal Printer Cable
STANDARDThermal Printer Paper (Nomal type)
STANDARDThermal Printer Paper (Label type)
STANDARD Lipid Manegement Software
Universal Mini USB Cable
CHAPTER 2. Testing the STANDARD LipidoCare Analyzer
1. Operating the Analyzer
Before using the Analyzer for the rst time, perform the following steps:
STEP 1. Insert batteries.
1. Ensure the Analyzer is switched o and turn it over.
2. Open the battery compartment lid by slightly pressing the tab towards the centre of the Analyzer.
3. Lift the lid upwards to remove it from the Analyzer.
4. Insert four batteries into the compartment according to the illustrations. Please note the orientation of the“+”(battery head) and“–”terminals (at
end). Use only alkaline manganese batteries (1.5V, AAA).

9
5. Close the battery compartment lid.
6. Turn the Analyzer on to test the function of the new batteries.
7. Check the display is functioning correctly to prevent misinterpretations due to defective display elements.
• If you think the display screen does not appear for long enough, you can hold down the button (the next time you switch
on the Analyzer).The display screen is then shown for the time the button is pressed.
• After inserting or replacing batteries, conrm that the time and date are set correctly. If they are not, reset the Analyzer before
testing.
• Always replace all four batteries at the same time because batteries with dierent capacities may impair the function of the
Analyzer. Do not use rechargeable batteries.
STEP 2. Set the Analyzer
Brief overview of the Analyzer settings
The following table provides an overview of the available settings.
Stage Setting Options
1 Beep On, O
2 Year YYYY
3 Date Format m-d, d-m
4 Date mm-dd, dd-mm
5 Time Format 12h, 24h
6 Time hh:mm
7 Sample. Type (for Lipid) SE, BL
8 Unit of the Test Result mg/dL, mmol/L
9 Auto Printing On, O
10 Printing Page P-1, P-2
11 Hypo warning (for glucose) O, 60, 70, 80
12 Alarm O, 4 Alarms
13 Bluetooth with PC* On, O
14 Bluetooth with Printer* On, O
15 Repairing Printer* On, O
* Only for 02LA20G
Stage 0) Entering set mode
1. Turn on the meter and then press and hold the ( ) button (on the
top side of the Analyzer) for 3 seconds to enter set mode. If you want to leave the
set mode, press the ( ) button.
2. Press the or button to change the currently ashing setting. You can
press the or button as many times as needed (or keep it pressed) until
the desired setting (value) is reached.

10
3. Press the button again to conrm (store) the current setting and go to the
next setting.
You can only move forward through the settings. Returning to the previous setting is not allowed. Corrections can only be made by repeating
the settings. The setting procedure can be terminated at any time by pressing the ( ) button. The settings made up to that
point will be stored.
4. Once the setting is completed, the Analyzer will be automatically turned o.
Stage 1) Setting - Beep
At the rst stage, you set the Beep function.
1. When you enter the set mode, the display for setting the beep will be the rst
step of the setting mode.
Ex) On
2. Set the beep mode on or off by pressing either or button and then
select the preferred feature by pressing the button. If you select the beep
on feature, a ‘beep’ sound is made at the same time; otherwise, if you select the
beep o feature, no sound is made.
Stage 2) Setting – Year
In the second stage, you set theYear in Date & Time setting.
1. After setting the Beep funtion, the display for setting the Year will appear. Set
the correct year by pressing
After setting the Beep funtion, the display for
or
After setting the Beep funtion, the display for
button and then select the correct year
by pressing the button.
Ex) 2011
Stage 3) Setting – Date Format
In stage 3, you set the Date Format in Date &Time setting.
1. Next will appear the setting display for Date Format. The Analyzer can display
either a Month-Day (m-d) format or a Day-Month (d-m) format. Set the
preferred format on the screen by pressing either
either a Month-Day (m-d) format or a Day-Month (d-m) format. Set the
or
either a Month-Day (m-d) format or a Day-Month (d-m) format. Set the
button and select
by pressing the button.

11
Stage 4) Setting – Date
In stage 4, you set the date
1. After setting the Date Format, set the correct month or day on the display by
pressing either
After setting the Date Format, set the correct month or day on the display by
or
After setting the Date Format, set the correct month or day on the display by
button and select by pressing the button.
Ex) 1- 1
Stage 5) Setting – Time Format
In stage 5, you set the Time Format in Date &Time setting.
1. Next will appear the display for setting the Time Format. The Analyzer can
display the time in either the 12h format or the 24h format. Set the preferred
format on the display by pressing either
display the time in either the 12h format or the 24h format. Set the preferred
or
display the time in either the 12h format or the 24h format. Set the preferred
button and select by
pressing the button.
Stage 6) Setting – Time
In stage 6, you set the Time, in Date &Time setting.
1. Next will appear the display for setting the current Time. Set the hour and minute on the display by pressing either or button and select by
pressing the button.
a. 12h Format
sub-step 1) set AM or PM,
sub-step 2) set hour(1 to 12),
sub-step 3) minute(1 to 59)
Ex) 12:00AM
b. 24h Format
sub-step 1) set hour(1 to 24),
sub-step 2) minute(1 to 59)
Stage 7) Setting – Sample Type
In stage 7, you set the Sample Type for lipid test.
1. After setting theTime, the display for setting the Sample Type (SPL) for the
lipid test will appear.
BL
- Capillary or Venous Whole Blood
SE
- Serum or Plasma
BL

12
2. Set the Sample type for lipid test,‘BL’or‘SE’, by pressing either or
button and then select the preferred feature by pressing the button.
SE
Stage 8) Setting – Unit of the Test Result
In stage 8, you set the Unit of the result.
1. After setting the SampleType, the display for setting the Unit of the result will
appear.
2. You can set the preferred unit by pressing or buttons and select by pressing buttons.
Stage 9) Setting – Auto Printing
In stage 9, you set the Auto Printing function.
1. After setting the Unit, the display for setting the Auto Printing function will
appear.
- Auto Printing
2.
Set the
Auto Printing
function,‘On’or‘O’, by pressing either or button and then select the preferred feature by pressing the button. If you set Auto
Printing to 'On', then your results will be automatically printed when a test is completed.
Stage 10) Setting – Printing Page
In stage 10, you set the Printing Page.
1. After the Auto Printing function setting, the display for setting the Printing
Page mode will appear.
2. Set the Printing Page mode, -‘P-1’(1 copy) or‘P-2’(2 copies) - by pressing
either or button and then select the preferred feature by pressing the
button.
Stage 11) Setting – Hypo Warning
In stage 11, you set the Hypo Warning function.
3. After the Printing Page setting, the display for setting the
Hypo Warning will appear.
Ex) O

13
4. Set the Hypo Warning ‘o’or the result you want to select
- 60, 70 or 80 mg/dL - by pressing either
‘o’ or the result you want to select
or
‘o’ or the result you want to select
button
and then selecting the preferred feature by pressing the
button.
• You can set the Analyzer to let you know when your glucose result indicates possible low blood glucose (hypoglycemia). You
can also select what blood glucose level you want this indicator to have - 60, 70 or 80 mg/dL.
• If your glucose results are lower than the selected hypo result, the hypo symbol will appear on the LCD with a ‘beep’sound. It
is very important to manage your hypoglycemia.
Stage 12) Setting – Alarm
In stage 12, you set the Alarm function. You can set a maximum
of 4 alarms to remind you to test your glucose.
1. After the HypoWarning setting, the display for setting the
Alarm will appear.
2. Set the rst Alarm function, ‘On’or ‘O’, by pressing either
the or button and then select the preferred feature
by pressing the button. If you select the Alarm‘O’
feature, the Analyzer automatically powers o.
3. If you select the Alarm‘on’feature in rst alarm mode, the
clock will blink. Set the desired hour and minute for the
alarm by pressing either the or button and then
select the preferred time by pressing the button.
4. The 2nd Alarm setting mode appears next. Set the Alarm in
the same way as above. [2, 3]
5. You can set the third and fourth Alarm modes in the same way as above. [2, 3]
6. When you nish setting the last alarm, the Analyzer automatically power o.
Stage 13) Setting – Bluetooth with PC *Only for 02LA20G
You can set Bluetooth communication with your PC. If the analyzer is paired with a PC, the test results are transferred automatically via Bluetooth.
1. After the alarm setting, the display for bluetooth with PC function will appear on
the screen.
2. Select ON or OFF by pressing the left or right button and then save it by pressing
the ON/OFF button.
BL BL

14
Stage 14) Setting – Bluetooth with Printer *Only for 02LA20G
You can set Bluetooth communication with a printer.
1. After the setting of Bluetooth with a PC, the display for bluetooth with printer
function will appear on the screen.
2. Select ON or OFF by pressing the left or right button and then save it by pressing
the ON/OFF button.
BL BL
If you set both auto-printing and Bluetooth with Printer as ON, you can get the test results printed automatically via Bluetooth
communication.
Stage 15) Setting – Re-Pairing Printer *Only for 02LA20G
You can set the analyzer to be automatically paired with the previously paired printer
whenever the printer is near the analyzer.
1. After the Bluetooth with printer setting, the display for re-pairing printer function
will appear on the screen.
2. Select ON or OFF by pressing the left or right button and then save it by pressing
the ON/OFF button.
BL BL
STEP 3. Insert a codechip
Lipid Test Strip
Coding
The codechip provides the Analyzer with important information on the production-specic characteristic of the respective lipid test strip to measure your results
accurately. The codechip is required when a new Lipid test strip package is opened. Before you use your Analyzer with a new box of lipid test strips, you should
set the Analyzer to“match”the lipid test strips. The number on the codechip should match the number on the outer box and the test strip foils.
TC
Red Case
TG
Yellow Case
Lipid Prole
Blue Case
Code Setting
1. Make sure the Analyzer is turned o. Remove the old codechip if one is installed.
2. Insert a new codechip until it clicks into place.
001
001
001
3. Turn the Analyzer ON. The 3-digit code number of the inserted codechip and the appropriate parameter name appears with the ashing lipid test strip

15
symbol.This number and parameter name must match the codechip. If it does not, repeat the steps above (1-2).
BL
BL BL BL
BL
BL
BL
BL BL BL
BL
BL
BL
BL BL
Lipid Profile TC TG
A codechip is included in each test strip package. Store the lipid test strips together with the codechip in the package.
Glucose Test Strip
A codechip is not necessary for glucose testing.
2. Preparing for measurement
Lipid Test System - components required:
• STANDARD LipidoCare Analyzer
• Lipid test strips for the desired parameter with the related codechip
• Lancets and Lancing device
• Tool for collecting the blood such as the STANDARD Ezi tube+ or micropipette
Only use STANDARD LipidoCare LipidTest Strip and either STANDARD LipidoCare Blood Glucose Test Strip or SD Codefree Blood Glucose test
strip.
Precaution
Do
• Read the test strip package insert.
• Operate the Analyzer at the acceptable test-specic temperatures.
- For Lipid testing 18-32 °C (64-90°F)
- For Glucose testing 10-45 °C (50-113°F)
• Place the Analyzer on a level surface or hold it steady in your hand.
• Keep the guidelines for cleaning.
• Make sure that all display elements are displayed.
• Keep the STANDARD Lipid Check Strip away from sunlight.
Do not
• Use lipid test strips beyond the expiration date, this may cause inaccurate results.
• Touch or remove the test strip during actual measurement.
• Subject the Analyzer to sudden movements during a measurement.

16
• Reuse the test strip.
• Store the Analyzer and strips at extreme temperatures.
• Store the Analyzer and lipid test strips under humid or damp conditions without suitable protection.
• Open the chamber ap during the measurment.
Not following these precautions can lead to inaccurate results.
Getting a Drop of Blood
Sample type
• For Lipid testing : Only use fresh capillary whole blood or venous blood in heparin or an EDTA treated vacuum tube.The venous blood sample should be
used within 6 hours after collection.
- Fresh Capillary Whole Blood
-Venous Whole Blood
- Venous Blood - Serum
- Venous Blood - Plasma
Only professional use for venous blood.
• For Glucose testing : Only use fresh capillary whole blood.
Capillary Whole Blood
1. Wash your hands in warm, soapy water. Rinse well and dry
completely. Warming ngers can increase blood ow.
2. Turn the lancet insert cap counterclockwise to remove it, insert
the lancet into the lancing device holder and push down rmly
until it is fully seated. Twist the lancet protective disk until it
separates from the lancet.
123
3
234
123
234
23
4
123
234
123
234
23
4
123
234
123
234
23
4
3. Replace the cap and turn it clockwise, until it is snug. Adjust the
puncture depth setting by turning the comfort dial. The dial has
1 to 5 steps, and the higher the step number, the stronger the
blood sampling pressure on the puncture site.
The comport tip oers 5 dierent levels of skin penetration.
• 1-2: for soft or thin skin
• 3: for average skin
• 4-5: for thick or callused skin
123
234
123
3
3
234
23
4
123
234
2
2
123
234
23
4
4. After cocking the lancing device back, hold the lancing device
rmly against the side of nger and then press the release
button.
123
234
123
234
23
4
123
234
123
234
23
4

17
How to use STANDARD EziTube+ *Only for lipid testing
1. Touch the tip of a STANDARD EziTube+ to the edge of the blood
drop. Blood will automatically be drawn up the tube by capillary
action. Keep the tube in contact with the blood until it is lled to
the black line.You should squeeze your nger if you need more
blood.
2. Transfer the collected sample to the sample well on the test strip
by gently squeezing the bulb of the STANDARD EziTube+.
3. Close the ap to start the test.
• Used STANDARD Ezi Tube+ can cause infection. Please discard the used STANDARD EziTube+ according to local regulations.
• Collect the blood sample carefully to try and avoid air bubbles.
How to use the micropipette *Only for lipid testing
1 2
123
234
123
234
23
4
3
Place the tip securely on the micropipette. Push the plunger down and hold it in this position. Touch the tip to the blood drop.
45 6
Slowly release the pluger and check the blood
is drawn into the tip. Open the ap and apply the blood sample. Remove the tip from the micropipette and dispose
of the tip in a biohazardous waste container.

18
3. Running a test - Lipid
Test Procedure
Check the following before performing a measurement:
1. Are date and time correct?
2. Does the battery symbol appear? If it appears, only a few more measurements can be performed. Replace the batteries as soon as possible.
3. Check the expiration date of the lipid test strip. Always use lipid test strips which are not expired.
4. Check the code number on the analyzer. Ensure it matches the code number on the codechip.
5. With lipid and glucose test strips, check the reaction area for discolouration before starting measurement. If you detect discolouration, this
test strip is unusable.
1. Remove a lipid test strip from the foil pouch and hold it by the
black handle with your thumb and index nger so that the
horizontal lines and triangle are facing upwards.
2. Open the Measurement Chamber Flap and insert the lipid test
strip into the ' LipidTest Strip Slot ' of the Analyzer. Make sure it
is fully inserted - you will know when it is fully inserted as the
meter will beep and the display will instruct you to close the
Measurement Chamber Flap.
BL
BL
3. Close the Measurement Chamber Flap and then re-open it when
instructed to do so on the display. When the Chamber Flap is
opened, the blood drop icon will ash on the screen.
BL
BL
4. Apply the blood sample from the pipette or capillary tube to the
sample application area of the lipid test strip. Do not touch the
application area.
5. When sucient blood has been added, the display will instruct
you to close the Measurement Chamber Flap. Once the ap is
closed, the meter will start counting down from 3 minutes.

19
6. When the countdown reaches 0, your result appears on the
screen. If using a lipid prole test strip, use the left and right
arrows to view the results for the dierent parameters.
BL
7. When the test is complete, pull out the used test strip and
discard in a biohazardous waste container. The Analyzer shuts o
automatically 3 seconds after you remove the test strip.
• Do not open the Measurment chamber ap during the test.
• If you do not close the measurement chamber ap immediately after applying blood, the analyzer display ashes the close
ap icon with countdown and beep.
• If you do not close the ap within 20 seconds of applying the blood, the Analyzer displays the ' E-7 ' message.
• The blood drop must ll the entire blood application hole.
Example of not
enough blood
Example of
enough blood
Example of not
enough blood

20
4. Running a test - Glucose
Test Procedure
1. Remove a new glucose test strip from its container. Be sure to tightly
replace the container cap immediately after removing a glucose test
strip.
2. Insert the glucose test strip into the 'Glucose Test Strip Slot' of the
Analyzer. The Analyzer will automatically enter the glucose testing
mode.
BL
3. When the blood drop symbol ashes, you are ready to perform a
glucose test.
BL
4. Obtain a blood sample using the lancet and lancing device.
0.9µL
actual size
5. Hold your nger to the tip of the glucose strip until the yellow window is completely lled with blood. Do not place the blood drop on top of the
glucose strip.
6. The blood will be drawn into the glucose strip automatically. If the beeper is turned on, the Analyzer will beep to let you know the glucose test is beginning.
7. When blood is applied to the glucose strip, the display counts down from 5 to 1 and your result appears on the display in just 5 seconds.
8. When the test is done, pull out the used glucose test strip.The Analyzer shuts o automatically 3 seconds after you remove the test strip.
Narrow Channel
STEP 1
STEP 2
Gently touch the glucose test strip to
the blood drop. Do not press the
glucose test strip against the finger.
Proper glucose test
strip appearance.
Table of contents