Precision Flow Measurement
An ON
I
CON Brand
AG90 New Meter Installation
Quick Start Guide
253.872.0284 Page 1 of 4 seametrics.com
Recommended ToolsRecommended Tools
Recommended:
• 3/32” or 2.5mm (small) at head screw driver
• Crescent Wrench
• 5/32” Hex Key
Optional:
• Channel locks
• 6” Strap Wrench
WarningsWarnings
• Conrm that U-clip retainer is installed and never remove the
U-clip retainer when pipe is under pressure—may result in
serious injury.
• Saddle bolts must be tightened evenly. Do not over tighten.
• Install security seals during installation if regulations require.
• Improper sealing of cables or cable glands will void warranty.
• Ensure proper grounding when required.
• Programming pipe ID, pipe insertion hole size and straight pipe
conguration is required for the meter to read.
• To prevent tampering or changes to the program, after initial
set-up, either enter a security pass-code, or remove the display
lid and place a security tab over one of the arrows before
installing the housing security seal.
Refer to instruction manual for further details.
Check ComponentsCheck Components
OptionalOptional
Display Security Seal
Cable Gland
Security Clip & Seal
(Security Clip has small hole for seal wire)
Meter with Indicator
and hinged cover Saddle Assembly
(Packaged Separately)
Saddle Gasket
Power/Output Cable
Security Tab
LT-14494r2.5 20200130
1/30/2020
Seametrics • 19026 72nd Avenue South • Kent, Washington 98032 • USA
(P) 253.872.0284 • (F) 253.872.0285 • 1.800.975.8153 • seametrics.com
AG90 New Meter Installation
Quick Start Guide
Choose a position that will ensure a full pipe. Choose a position that will minimize ow distortion.
PositioningPositioning
*10X/5X 2X
Flow
Flow
Flow
Flow
Flow
Flow
BEFORE INSTALLING measure & record inside diameter (ID) of pipe.
Side (3 o’clock) and top
(12 o’clock) installations
are acceptable.
*Upstream straight pipe is selected during initial setup. Upstream
options are 5X or 10X the diameter and are based on the
amount of straight pipe available in either new or propeller meter
replacement installation. Downstream straight pipe requirement
is 2X the diameter. See programming setup for details.
1. If you have the bolt-in, 3-vane style they may be removed or left in place. They do not aect performance.
2. If you have the 6-vane, bolt-in style with a cone, it should be removed.
3. For 6-vane, weld-in style installed in a spool, turn the spool 180° and reinstall with the vanes behind the meter.
4. Orice, tube or Vortab style conditioners will not aect performance.
Flow Straightening Vanes/Conditioners
1. Clean the mounting surface, remove any
roughness from the area and cut a 1.75”
hole into pipe. Place gasket centered over
pipe opening.
2. Place saddle top over gasket. 3. Make sure saddle top covers entire gasket.
1st
4th
3rd
2nd
NOTE: Gasket will extendNOTE: Gasket will extend
outside of a 4” pipeoutside of a 4” pipe
saddle’s sides.saddle’s sides.
5. Place saddle plates over saddle
clamp threads. Attach nuts and
tighten as shown below. Torque to 75
ft-lb in cross pattern.
6. Insert the EX90 sensor into the saddle 6. Insert the EX90 sensor into the saddle
tting and secure with mounting clip or tting and secure with mounting clip or
attach security clip and seals if required.attach security clip and seals if required.
4. Place the saddle clamps under the
pipe and align with the clamp guides
on the saddle top.
LT-14494r2.5 20200130
1/30/2020
Seametrics • 19026 72nd Avenue South • Kent, Washington 98032 • USA
(P) 253.872.0284 • (F) 253.872.0285 • 1.800.975.8153 • seametrics.com
AG90 New Meter Installation
Quick Start Guide
WiringWiring
Unscrew the display lid and remove it from the meter. Unsnap the display assembly and remove it from the meter
exposing the internal wiring connector. Install the wires through the cable glands into the 2 pin screw connector.
(C1 = power/output cable)
EXTERNAL POWER WARNING!
Ribbon cable must be connected BEFORE external power is applied. See Instruction manual for details.
With No External Output - If not using external output, no wiring is required.
White (C1) Pulse-
Green (C1) Pulse+
C1
With Pulse Output Only Pulse with External DC Power Source
9 - 36 VDC at 250mA max, 30mA average9 - 36 VDC at 250mA max, 30mA average
White (C1) Pulse-
Green (C1) Pulse+
C1
Red (C1) DC+
Black (C1) DC-
Clamp cable with strain relief
clips. Attach drain wire lug to
bracket post.
Remove plug & o-ring. Insert
cable gland/strain relief. Feed
cable through cable gland.
CRITICAL!
Torque cable gland sealing
nut to 22 in-lbs.
WARNING: Improper sealing of glands or cables will invalidate any warranty.
If plugs or cable glands are removed, reinstall using Teon pipe sealant, or
tape, to ensure maximum moisture protection.
LT-14494r2.5 20200130
1/30/2020
Seametrics • 19026 72nd Avenue South • Kent, Washington 98032 • USA
(P) 253.872.0284 • (F) 253.872.0285 • 1.800.975.8153 • seametrics.com
AG90 New Meter Installation
Quick Start Guide
Menu SystemMenu System
1.2345678
100
TOTAL
CU FT
X1000
RATE
GPM
T UNIT R UNIT SET PDAMP
SET 4
SET 4
SET 20
SET 20
SET F
SET F
EXIT
TOTAL = GALLONS
PRESS + TO SET TOTAL
UNITS FOR DISPLAY
T UNIT R UNIT SET PDAMP
SET 4
SET 4
SET 20
SET 20
SET F
SET F
EXIT
TOTAL:
PRESS TO CHANGE
GALLONS
0.0000000
SET ID
TOTAL
ACRE
FEET
RATE
GPM
OK
FORWARD TOTALFORWARD TOTAL
The HOME Screen displays ow volume, direction of the ow total and ow RATE
along with status conditions such as Empty Pipe. Two buttons below the LCD display
are used to access menu screens for viewing and changing meter setup parameters.
These two buttons are light sensors which can detect when a nger is covering them
and activate upon release. Only three button touch actions are needed to control
navigation through the menus, settings changes and back to the home screen.
HORIZONTAL SCROLLING:
Tap right-hand button to
scroll horizontally through
menu tabs or move
horizontally within a tab
dialog when applicable.
TAP
SELECT: Tap left-hand button
to change a highlighted item
within a tab dialog.
TAP
ENTER/EXIT: Hold left
button while tapping right
button once to enter or exit
a tab dialog or to navigate
between the HOME and
other menu screens.
+
HOLD TAP
All menu screens consist of two rows of tabs surrounding a dialog box that lets you
view and change setup parameters. To enter the Menu System perform the hold
and tap sequence. +
HOLD TAP
Changing SettingsChanging Settings
T UNIT
View or change TOTAL volume units
R UNIT
View or change ow RATE units
SET P
View or change pulse output
scaling
DAMP
View or change # of sample
periods for rolling average*
DAMP default set to 15
EXIT
Return to HOME SCREEN or
Tap ve times, to enter a
SUBMENU screen from which
you can access the required
pipe setting functions
T UNIT R UNIT SET P DAMP
SET 20 SET FSAMP EXIT
TOTAL = GALLONS
PRESS + TO SET TOTAL
UNITS FOR DISPLAY
T UNIT R UNIT SET P DAMP
SET 20 SET FSAMP EXIT
FLOW RATE = GALLONS/MIN
PRESS + TO SET RATE
UNITS FOR DISPLAY
T UNIT R UNIT SET P DAMP
SET 20 SET FSAMP EXIT
00001.0 GALLONS
PRESS + TO SET
GALLONS TOTALIZED PER
PULSE SENT OUT PULSE1
T UNIT R UNIT SET P DAMP
SET 20 SET FSAMP EXIT
DAMPING = 1
PRESS + TO SET
DAMPING VALUE
T UNIT R UNIT SET P DAMP
SET 20 SET F EXIT
PRESS + EXIT MENU
AND RETURN TO FLOW
DISPLAY
Enter Pipe Settings (Required)
INFO: Meter model, serial number, rmware version
SAMP: Sample rate (Default is set to 5 seconds. Battery life with a 5-second
sampling rate is four years. Increase sampling interval to extend battery life.
(May vary depending on environmental factors. See instruction manual for more
details.)
SAMP SETEX EXIT
INFO
PRESS + TO SET OUTER
WALL DIAMETER AND WALL
THICKNESS
SETUP
SAMP SETUP EXIT
ID
INFO
PRESS AND TO CHANGE
SAMP SETUP EXIT
HOLE
INFO
PRESS AND TO CHANGE
SAMP SETUP EXIT
PRESS + TO EXIT MENU
AND RETURN TO FLOW
DISPLAY
INFO
SAMP SETUP EXIT
PIPE
INFO
PRESS AND TO CHANGE
10/2 ELB
SAMP
SAMP
SAMP
SAMP
SAMP
LARGE
04.500
IN.
SAMP SETEX EXIT
INFO
PRESS + TO SET OUTER
WALL DIAMETER AND WALL
THICKNESS
SETUP
SAMP SETUP EXIT
ID
INFO
PRESS AND TO CHANGE
SAMP SETUP EXIT
HOLE
INFO
PRESS AND TO CHANGE
SAMP SETUP EXIT
PRESS + TO EXIT MENU
AND RETURN TO FLOW
DISPLAY
INFO
SAMP SETUP EXIT
PIPE
INFO
PRESS AND TO CHANGE
10/2 ELB
SAMP
SAMP
SAMP
SAMP
SAMP
LARGE
04.500
IN.
1. SETUP
View or change meter conguration settings.
SAMP SETEX EXIT
INFO
PRESS + TO SET OUTER
WALL DIAMETER AND WALL
THICKNESS
SETUP
SAMP SETUP EXIT
ID
INFO
PRESS AND TO CHANGE
SAMP SETUP EXIT
HOLE
INFO
PRESS AND TO CHANGE
SAMP SETUP EXIT
PRESS + TO EXIT MENU
AND RETURN TO FLOW
DISPLAY
INFO
SAMP SETUP EXIT
PIPE
INFO
PRESS AND TO CHANGE
10/2 ELB
SAMP
SAMP
SAMP
SAMP
SAMP
LARGE
04.500
IN.
3. ID (Required)
View or change inner diameter of the pipe. Measurement
in inches.
4.HOLE (Required)
View or change installation pipe hole size. Tap then
to change setting from N/A (default) to small and large sizes.
Note: See instruction manual page 14 for hole size description.
5. PIPE (Required)
View or change pipe conguration based on installation. Tap
then to change setting from N/A (default) to 10/2,
5/2 or STRAIGHT (conditions with 10 diameters or more)
2. SETUP (Menu Functionality)
The highlighted value can be changed using the arrows. The value on the left is the
menu name. Cycle through these by pressing . Press to highlight and change
the value of the menus.
INITIAL SETUP OF ID, HOLE, AND PIPE IS REQUIRED
FOR THE METER TO OPERATE PROPERLY.
SAMP SETEX EXIT
INFO
PRESS + TO SET OUTER
WALL DIAMETER AND WALL
THICKNESS
SETUP
SAMP SETUP EXIT
ID
INFO
PRESS AND TO CHANGE
SAMP SETUP EXIT
HOLE
INFO
PRESS AND TO CHANGE
SAMP SETUP EXIT
PRESS + TO EXIT MENU
AND RETURN TO FLOW
DISPLAY
INFO
SAMP SETUP EXIT
PIPE
INFO
PRESS AND TO CHANGE
10/2 ELB
SAMP
SAMP
SAMP
SAMP
SAMP
LARGE
04.500
IN.
SMALL
1.2345678
100
TOTAL
CU FT
X1000
RATE
GPM
T UNIT R UNIT SET PDAMP
SET 4
SET 4
SET 20
SET 20
SET F
SET F
EXIT
TOTAL = GALLONS
PRESS + TO SET TOTAL
UNITS FOR DISPLAY
T UNIT R UNIT SET PDAMP
SET 4
SET 4
SET 20
SET 20
SET F
SET F
EXIT
TOTAL:
PRESS TO CHANGE
GALLONS
0.0000000
SET ID
TOTAL
ACRE
FEET
RATE
GPM
OK
FORWARD TOTALFORWARD TOTAL
ENTER PASSCODE
SET 20 SET FSAMP
000000
PRESS + TO CHANGE
HOME SCREEN
The Home Screen page on a new
meter will indicate for you to SET ID.
+
HOLD TAP
to move to the Passcode
Screen.
PASSCODE
Set passcode if required, or
+
HOLD TAP
to move to the Standard
Menu Options Screen.

Other manuals for EX90

2

This manual suits for next models

1

Other Seametrics Measuring Instrument manuals

Seametrics FT420 User manual

Seametrics

Seametrics FT420 User manual

Seametrics jWAVE User manual

Seametrics

Seametrics jWAVE User manual

Seametrics PD10 User manual

Seametrics

Seametrics PD10 User manual

Seametrics PE202 User manual

Seametrics

Seametrics PE202 User manual

Seametrics WP Series User manual

Seametrics

Seametrics WP Series User manual

Seametrics FT400 User manual

Seametrics

Seametrics FT400 User manual

Seametrics EX80 Series User manual

Seametrics

Seametrics EX80 Series User manual

Seametrics iMAG 4700p User manual

Seametrics

Seametrics iMAG 4700p User manual

Seametrics AG3000 Series User manual

Seametrics

Seametrics AG3000 Series User manual

Seametrics AG2000 User manual

Seametrics

Seametrics AG2000 User manual

Seametrics AG90 User manual

Seametrics

Seametrics AG90 User manual

Seametrics MDE User manual

Seametrics

Seametrics MDE User manual

Seametrics FT400 User manual

Seametrics

Seametrics FT400 User manual

Seametrics WMP Series User manual

Seametrics

Seametrics WMP Series User manual

Seametrics WMP Series User manual

Seametrics

Seametrics WMP Series User manual

Seametrics AG3000p User manual

Seametrics

Seametrics AG3000p User manual

Seametrics AG3000 Series User manual

Seametrics

Seametrics AG3000 Series User manual

Seametrics WMP-Series User manual

Seametrics

Seametrics WMP-Series User manual

Seametrics EX800 Series User manual

Seametrics

Seametrics EX800 Series User manual

Seametrics iMAG 4700 User manual

Seametrics

Seametrics iMAG 4700 User manual

Seametrics FT430 User manual

Seametrics

Seametrics FT430 User manual

Seametrics iMAG 4700 User manual

Seametrics

Seametrics iMAG 4700 User manual

Seametrics FT420 User manual

Seametrics

Seametrics FT420 User manual

Seametrics FT522 User manual

Seametrics

Seametrics FT522 User manual

Popular Measuring Instrument manuals by other brands

Burkert 8026 operating instructions

Burkert

Burkert 8026 operating instructions

Tektronix 1340 manual

Tektronix

Tektronix 1340 manual

Nitronic HRC3070 operating instructions

Nitronic

Nitronic HRC3070 operating instructions

GL McGAVIN ModieLamp manual

GL McGAVIN

GL McGAVIN ModieLamp manual

Onicon System-10 BTU Meter installation guide

Onicon

Onicon System-10 BTU Meter installation guide

GESTRA LRGT 16-1 Installation & operating instructions

GESTRA

GESTRA LRGT 16-1 Installation & operating instructions

Keller ECO 1 quick start guide

Keller

Keller ECO 1 quick start guide

Lutron Electronics DO-5512SD Operation manual

Lutron Electronics

Lutron Electronics DO-5512SD Operation manual

Milltronics AIRANGER DPL PLUS quick start guide

Milltronics

Milltronics AIRANGER DPL PLUS quick start guide

Seikom Electronic NLSW45-5 manual

Seikom Electronic

Seikom Electronic NLSW45-5 manual

Bosch BL 130I PROFESSIONAL operating instructions

Bosch

Bosch BL 130I PROFESSIONAL operating instructions

Bartington MS3 Operation manual

Bartington

Bartington MS3 Operation manual

KRAUSMANN 2200 Operation manual

KRAUSMANN

KRAUSMANN 2200 Operation manual

Sensepeek SP10 user guide

Sensepeek

Sensepeek SP10 user guide

Leica PINMASTER II PRO instructions

Leica

Leica PINMASTER II PRO instructions

horiba F-73 instruction manual

horiba

horiba F-73 instruction manual

OfiTE 120-53 instruction manual

OfiTE

OfiTE 120-53 instruction manual

Svantek SV 258PRO user manual

Svantek

Svantek SV 258PRO user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.