0000136864 rev. 4.2 4
The pump:
When the regulation device must be stored, clean water should be pumped through the hose in order to rinse
it.
Then position the roller arm at 45°, turning it clockwise.
These two precautions will facilitate subsequent reactivation of the device.
Keep away from freezing conditions.
The probe:
Extract the pH probe from the relative probe holder. Replace it in the original bottle filled with tap water. If
necessary, close the probe holder using a plug the size of a 5 euro cent coin.
!
Keep away from freezing conditions, intense heat and direct sunlight.
!
DO NOT INSERT EXCESSIVE QUANTITIES of chemical product above the probe.
As the pH electrode is comprised of glass elements, handle it with care.
The alarms
OFA (overdose alarm):
First OFA alarm: the ALARM light flashes; activation after 3 consecutive dosing cycles in which the set value was
not reached; the system continues to guarantee the measurement and dosing functions.
Second OFA alarm: the ALARM light flashes and the Hold light comes on; the dosing pump is blocked. The alarm is
activated after 4 consecutive dosing cycles in which the set level was not reached; to return to normal operation,
press the : button: the device resets the alarms and returns to normal measuring and dosing mode.
Flow: presence of incoming flow (enslavement by the filter pump).
Level Alarm led Flashing
Hold led light ON Alarm Relay Close - Push Key to open
Alarm Relay
- Restore Product tank
Out Range measure Alarm led Flashing Alarm Relay Close - Push Key to open
Alarm Relay
- Replace Measure pH
OFA First Alarm (time >28 min) Alarm led Flashing Alarm Relay open - Push Key to reset
OFA Second Alarm (time >40 min) Alarm led Flashing
Hold led light ON Alarm Relay Close - Push Key to reset
Flow Rate Hold led light ON Alarm Relay open - Restore Flow Rate
Calibration Function Alarm led Flashing
Hold led light Flashing Alarm Relay open - Restore Probe or Buffer
solution and repet
calibration function
System Error All leds Flashing Alarm Relay open Push Key to retry
initialization
To restore Default parameters run Following steps:
•Power off Pool Basic unit
•Keep Press with Key and Power On
•The unit flashing all leds
•
Press Key to restore Default parameters.
•Set Point value= 7,4 pH
•Dosing Method = Acid
•Calibration = ON
•OFA = OFF
•Flow Rate= ON
•
Easy Calibration = OFF