Seluxit SLX Heartbeat User manual

Installation Guide

Purpose of this Document
This is the installation guide for SLX Heartbeat. The purpose of this document is to guide an
electrician through the process of installing the SLX Heartbeat AED Status Monitoring Kit.
Introduction
The steps, preconditions and requirements described in this document need to be fulfilled in
order for the SLX Heartbeat to be installed.
Preconditions
- AED model Zoll AED Plus
- AED cabinet, model CAH55100M or Cabix Outdoor
- Access to a 4G/3G network w/ sufficient signal strength
Unboxing
SLX Heartbeat Main Unit
Communication unit with magnetic
backplate.
Main Unit is used for:
1. Collection of sensor data
2. Sending of data to the cloud via
4G/3G/2G modem
The Main Unit contains several sensors
measuring sound, temperature and
humidity.
NB: The QR code sticker (the monitoring unit’s unique
ID) is now on the Main Unit casing itself, and not on
the backplate as shown in the pictures.
SLX Heartbeat for Zoll AED Plus, Installation Guide (version 1.4, September 2020) 3 of 15

SLX Heartbeat Main Unit - backplate
4x magnets + 4x screws are included.
The backplate is affixed to the Main Unit
using the 4 screws. The 4 magnets
(washers) ensure that the backplate can be
mounted correctly in the AED cabinet.
NB: The QR code sticker (the monitoring unit’s unique
ID) is now on the Main Unit casing itself, and not on
the backplate as shown in the pictures.
SLX Heartbeat Eye
Sensor Eye and Eye Socket
for Zoll AED Plus
The Sensor Eye is used to connect the Main
Unit to the AED. It has built in sensors for
measuring temperature and color.
The Eye Socket makes it possible to attach
the eye to the AED.
SLX Heartbeat Cable-Tie Mount
Cable-Tie Mount for installation
2x strips are included.
The included cable-tie mount and 2 strips
are used for neat installation of the
Sensor Eye.
SLX Heartbeat for Zoll AED Plus, Installation Guide (version 1.4, September 2020) 4 of 15

SLX Heartbeat Plate
Metal plate with double-sided tape
The plate is necessary in cabinets that don’t
have magnetic sides. In this case, the plate
is affixed to the cabinet before the Main unit,
which then attaches magnetically to this
plate.
SLX Heartbeat Battery
Battery for the Main Unit
With all of these components accounted for, you are now ready to begin installation of SLX
Heartbeat.
SLX Heartbeat for Zoll AED Plus, Installation Guide (version 1.4, September 2020) 5 of 15

Installation
1. Install battery in Main Unit
The 3 LEDs blink three times when
the battery is installed. After that, the
Main Unit is automatically set to
“service mode”.
Good to know:
5 minutes after the battery has been installed, the Main Unit automatically wakes
from “service mode” and starts the onboarding process. If the Main Unit is not
installed in the AED cabinet, it sends an error message to the Seluxit Cloud .
1
There is no cause for alarm in this case, as the Main Unit is reset during the
onboarding process which you will perform later, when the Main Unit is installed in
the AED cabinet.
2. Use the screws (including magnets)
to fasten the backplate to the Main
Unit. Magnets will thereby be placed
on all four corners.
NB: The QR code sticker (the monitoring unit’s unique
ID) is now on the Main Unit casing itself, and not on
the backplate as shown in the pictures.
1 Seluxit Cloud is called “Viasens by Seluxit: Secure IoT Infrastructure”
SLX Heartbeat for Zoll AED Plus, Installation Guide (version 1.4, September 2020) 6 of 15

3. Take a picture of the unique ID on
the Main Unit. It is located on the
back of the Main Unit.
NB: The QR code sticker (the monitoring unit’s unique
ID) is now on the Main Unit casing itself, and not on
the backplate as shown in the pictures.
The picture of the unique ID on the
Main Unit should be sent together
with the installation address to the
service company monitoring the
AEDs after the installation has been
completed. Alternative or additional
steps may be required depending on
the service company.
4. Open the cabinet and place the
cable-tie mount on the right side of
the cabinet, centered and about 12
cm from the top.
NB: When the cabinet is opened
you’ll hear a high-pitched alarm for
about 30 seconds. You can stop the
alarm by using a magnet placed on
the door contact. You need to bring
your own magnet for that purpose.
SLX Heartbeat for Zoll AED Plus, Installation Guide (version 1.4, September 2020) 7 of 15

5. If the cabinet HAS magnetic sides,
skip this step.
If the cabinet DOES NOT HAVE
magnetic sides, mount the SLX
Heartbeat Plate on the left-hand side
of the cabinet.
NB: To ensure the best possible
signal, the SLX Heartbeat Plate
should be mounted on the left-hand
side of the cabinet. The lower-left
corner should be about 2 cm from
the door and 4 cm up from the
bottom of the cabinet.
6. Connect the Sensor Eye in the Main
Unit.
7. Use the magnets on the back side of
the Main Unit to attach it to the
inside of the AED cabinet.
NB: To ensure the best possible
signal, the SLX Heartbeat Plate
should be mounted on the left-hand
side of the cabinet. The lower-left
corner should be about 2 cm from
the door and 4 cm up from the
bottom of the cabinet.
Make sure that the Main Unit is
installed so the cable jack for the
Sensor Eye points downwards.
SLX Heartbeat for Zoll AED Plus, Installation Guide (version 1.4, September 2020) 8 of 15

8. Attach the Sensor Eye cable to the
cable-tie mount using the included
strips. Ensure that there are 12-14
cm of cable from the Sensor Eye to
the cable-tie mount.
9. Place the Eye Socket on the AED
using the double-sided tape, already
affixed to the Eye Socket.
NB: For optimal hold, clean the
surface (for example with alcohol)
before the Eye Socket is affixed to
the AED. Be careful when you attach.
The glue is strong, so you have only
one chance to properly mount the
Eye Socket.
10. Place the AED in the cabinet.
SLX Heartbeat for Zoll AED Plus, Installation Guide (version 1.4, September 2020) 9 of 15

11. Click the Sensor Eye to the Eye
Socket.
The installation is now finished, and you are ready to continue on to the onboarding process.
SLX Heartbeat for Zoll AED Plus, Installation Guide (version 1.4, September 2020) 10 of 15

Onboarding
1. Press and hold the button in on the
Main Unit. The 3 LEDs will begin to
blink in a certain order as you press
the button. When the LEDs stop this
blinking pattern, you can release the
button.
2. All 3 LEDs will now blink
synchronously to show that the
Main Unit is in “service mode”.
3. Press and release the button once
to start the test phase.
SLX Heartbeat for Zoll AED Plus, Installation Guide (version 1.4, September 2020) 11 of 15

4. The system will now test that the
Sensor Eye works as expected. The
Sensor-Eye LED will blink for about
10 seconds. After that, this LED will
light up constantly, indicating that
the test was successful.
5. The system will now find the GPS
position of the Main Unit. The GPS
LED blinks for about 40 seconds.
After that, this LED will light up
constantly, indicating that the GPS
location was found.
6. The last step is to establish a
connection to the Seluxit cloud. The
cloud LED will blink until the Main
Unit is connected. After that, this
LED will light up constantly,
indicating that the test was
successful.
SLX Heartbeat for Zoll AED Plus, Installation Guide (version 1.4, September 2020) 12 of 15

7. When all 3 LEDs light up constantly
for 10 seconds, the onboarding has
been successful.
If one or more of the LEDs continues
to blink, the onboarding process has
failed.
Please refer to the troubleshooting
section for additional help.
8. The Main Unit now switches off all
LEDs and goes into sleep mode.
The onboarding process is now finished, and SLX Heartbeat is running. If you experience
problems, you should read the troubleshooting section below. In this section, you can get
input to how you can find and solve common problems.
SLX Heartbeat for Zoll AED Plus, Installation Guide (version 1.4, September 2020) 13 of 15

Troubleshooting
- Problem: I didn’t finish the installation or the onboarding process was started
before I finished the installation.
-Action:
- Finish the installation and then restart the onboarding process from the
first step
- Problem: The status indicator on the AED has a red cross, meaning that the AED is
not functioning correctly.
-Action:
- Report the problem to the service company monitoring the AEDs
-Problem: No LEDs blink during the onboarding process.
- Action:
- Remove battery.
- Put the battery back in.
- Start the onboarding process from the first step.
- If that doesn’t work, replace the battery with a new battery.
- Start the onboarding process again.
- Problem: The Sensor Eye LED doesn’t stop blinking.
-Action:
- Make sure that the Sensor Eye is placed correctly. Adjust and start the
onboarding process again from the first step.
- Change the Sensor Eye with a new Sensor Eye, and start the onboarding
process again from the first step.
- Problem: GPS LED doesn’t stop blinking.
-Investigate:
-Place a mobile telephone in the AED cabinet, close the door and try to
ring to the phone. If you can do so, the mobile signal is okay.
- Action:
- If the mobile signal is okay
→
start the onboarding sequence again
from the first step.
- If the mobile signal isn’t okay
→
call the service company monitoring
the AEDs to report the problem.
- Problem: Cloud LED doesn’t stop blinking.
- Investigate:
-Does the GPS LED light up?
- Action:
SLX Heartbeat for Zoll AED Plus, Installation Guide (version 1.4, September 2020) 14 of 15

-If the GPS LED lights up
→
start the onboarding sequence again from
the first step.
-If the GPS LED doesn’t light up, this can be a sign that the mobile signal
is not strong enough or is unstable.
-Place a mobile telephone in the AED cabinet, close the
door and try to ring to the phone. If you can do so, the
mobile signal is okay.
-If the mobile signal isn’t okay
→
call the service
company monitoring the AEDs to report the problem.
SLX Heartbeat for Zoll AED Plus, Installation Guide (version 1.4, September 2020) 15 of 15
Table of contents
Popular Medical Equipment manuals by other brands

Armstrong Medical
Armstrong Medical AquaVENT FD140 Instructions for use

Fresenius Medical Care
Fresenius Medical Care 2008K2 manual

Stryker
Stryker EOLE Operation & maintenance manual

laerdal
laerdal SimPad quick start guide

Graham Field
Graham Field Lumex FR577RG Series Assembly and operating instructions

Kernel
Kernel Flow 50 Quick reference guide