sizewise.com 800-814-9389
OPS-0005-1218 27650009, Rev. 3.0
Emergency Release
Slide red clip up for
emergency release.
Note: Emergency release
allows arm to be lowered
in case of patient
emergency or battery
power interruption.
Warning: Do not stand beneath Lift Arm, Spreader
Bar, or patient when using emergency release.
Digital Scale Operation
1. While sling is empty, press ZERO button to set
system to zero. The digit 0 will pan across display
for a few seconds, then display will read 0.0
Note: Always re-zero scale before weighing patient
2. Remove sling from spreader. Position patient in
sling, then reattach sling to spreader bar.
3. Carefully lift patient. When motion has stabilized,
press WEIGH to display patient weight.
Note: Display will automatically turn itself o after
60 seconds. If display turns o before you are able to
view weight data, press WEIGH again.