User-filled ULTRA Passive Sampler
Cat. No. 690-200
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 • www.skcinc.com
Applications
SKC patented* user-filled ULTRA®Passive Sampler provides for low ppb to ppt
detection of organic vapors in ambient and indoor air. Filling the sampler is simple
and provides a wide variety of sorbent options and a longer shelf-life.
* U.S. Patent No. 6,607,581
Performance Profile
Sampling Rates: Go to the SKC Passive Sampling Guide at
www.skcinc.com/catalog/passive-guide.php.
Analysis: Thermal desorption; gas chromatography
See Analysis Instructions at www.skcinc.com/instructions/40125.pdf.
Shelf-life: Sorbent vials containing Tenax®TA, Chromosorb®106,
Carbopack X, or Anasorb®GCB1 have a 60-day shelf-life at
≤39.2 F (4 C).
Sorbent vials containing Charcoal have a one-year shelf-life
at ≤39.2 F (4 C).
www.skcinc.com
Sliding cover in open
(sampling) position
Sliding cover in
closed position
Back view
Clip
Diffusion holes
(sample sorbent
compartment)
Sliding
cover
Blank/
correction
sorbent
compartment
Clip
Sample
compartment
Sliding
cover
Blank/
correction
compartment
Clip
Sample sorbent
compartment
Plugs
Blank/
correction
sorbent
compartment
Filling the Sampler
1. Remove sampler from the resealable pouch.
2. With the back of the sampler facing upward and
the metal clip oriented to the left, remove the plug
from the sample sorbent compartment (on left
side, immediately above diffusion holes).
3. Align the funnel (Cat. No. 690-301) over the compartment access port on the
sampler and press down firmly.
Ensure funnel is securely installed on the access port.
If funnel does not sit properly, the sorbent will not fill the sampler.
4. Open one vial of selected sorbent.
a. Slowly pour small amounts of the sorbent
into the funnel. Note: If sorbent starts to back
up into the funnel, lift funnel and housing as one
unit and gently tap against a hard surface until
funnel clears.
Do not lift funnel until sorbent is completely
transferred into housing.
b. Repeat until all sorbent has been transferred from the vial into the hous-
ing compartment.
5. Once all sorbent has been transferred from the vial, remove the funnel, and
replace the plug on the compartment access port.
6. If a blank/correction is required, remove the plug
from the blank/correction compartment (on right
side, immediately above SKC cover).
7. Repeat Step 3.
8. Open second vial and repeat Steps 4a. and b.
9. Once the sampler has been filled, remove funnel, and replace plug on com-
partment access port.
6
2
Sample sorbent
compartment
Plug
Sliding cover
Clip
4a
Blank/
correction
compartment
Plug
Sliding cover
Clip
1a
1b
2
3
4
Clip
Diffusion holes
(sample sorbent
compartment)
Sliding cover
Blank/
correction
sorbent
compartment
Clip
Sample compartment
Sliding cover
Blank/
correction
compartment
Sampling
1. Clip sampler(s) in appropriate location.
a. Indoors: Clip sampler(s) to stand accessory
(Cat. No. 690-302) and locate at appropriate
height.
b. Outdoors: Clip sampler(s) inside shelter
(open on two ends and boom to allow air
to flow freely). Hang shelter securely at the
appropriate height.
Ensure diffusion holes in sampler are not
obstructed in any way.
2. Start sampling: Slide sampler cover to open
position (diffusion holes visible). Write sample
start time on pouch label.
3. Stop sampling: Slide sampler cover to closed
position (arrow and SKC visible). Write sample
stop time on pouch label.
Important: Measure and record ambient temperature
and atmospheric pressure. Include in information sent
to the laboratory.
4. a. Place sampler in its resealable pouch
immediately after sampling.
Or
b. Remove plug from sample sorbent
compartment. Remove lid from sorbent vial
and place opening of sorbent vial over the
compartment access port. Transfer sorbent
to vial by tipping the sampler upside down
while holding the vial in place securely beneath the sampler. Gently tap
the sampler to remove any remaining sorbent. Replace lid on sorbent
vial and place vial in resealable pouch. Repeat for blank/correction
sorbent.
Ensure sample ID on sampler or vial matches sample ID on pouch.
5. Seal pouch(es). Place pouch(es) in container with
cool pack.
6. Use expedited shipping to send sample(s),
pertinent information, and Chain of Custody
form(s) to a laboratory for analysis.
For further information on analysis, see Analysis Instructions 40125.
5
6
&
Storage
Before Sampling: • Vials of Tenax TA, Chromosorb 106, Carbopack X, or
Anasorb GCB1: Up to 60 days at ≤39.2 F (4 C) in a clean,
organic solvent-free environment
• Vials of Charcoal: Up to one year at ≤39.2 F (4 C) in a
clean, organic solvent-free environment
After Sampling: 21 days at ≤39.2 F (4 C) in a clean, organic solvent-free
environment
Ordering
User-filled ULTRA Passive Sampler Cat. No.
Empty Housing, packaged in reusable pouch, ea, requires sorbent
vials listed below; select sorbent appropriate for the application 690-200
Sorbent Vials, pk/2
Contain purged sorbent (< 25 ng typical background levels)
Anasorb GCB1,#‡ 370 mg 690-201
Chromosorb 106,#¥ 285 mg 690-203
Tenax TA,#253 mg 690-204
Charcoal,#Ω500 mg, for solvent extraction 690-205
Carbopack X,#400 mg 690-206
Sampling Accessories
Transfer Funnel, for ULTRA only 690-301
Stand for Indoor Sampling 690-302
Shelter for Outdoor Sampling 690-303
Analysis Accessory
Thermal Desorption Tube, Perkin Elmer, 0.25-inch OD x 3.5-inch
length, includes screens and end caps P226530
# Limited-shelf life; see Storage above
‡ Comparable to Carbopack B
¥ Go to osha.gov and search on ULTRA for additional sampling rates.
ΩCan be stored for up to one year at ≤39.2 F (4 C)
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which
providesSKC’ssoleliabilityandthe buyer’sexclusive remedy. Toviewthecomplete
SKC Limited Warranty and Return Policy, go to hp://www.skcinc.com/warranty.
Form 40128 Rev 1807

Other SKC Laboratory Equipment manuals

SKC Quick Take 30 User manual

SKC

SKC Quick Take 30 User manual

SKC ULTRA 690-101-NB User manual

SKC

SKC ULTRA 690-101-NB User manual

SKC 520 Series User manual

SKC

SKC 520 Series User manual

SKC GS-3 Cyclone User manual

SKC

SKC GS-3 Cyclone User manual

SKC ULTRA 690-105 User manual

SKC

SKC ULTRA 690-105 User manual

SKC biosampler 225-9595 User manual

SKC

SKC biosampler 225-9595 User manual

SKC ULTRA 690-101 User manual

SKC

SKC ULTRA 690-101 User manual

SKC 249-01 User manual

SKC

SKC 249-01 User manual

SKC AirLite Sample Pump User manual

SKC

SKC AirLite Sample Pump User manual

SKC ULTRA 690-105-NB User manual

SKC

SKC ULTRA 690-105-NB User manual

SKC Preloaded Disposable Parallel Particle... User manual

SKC

SKC Preloaded Disposable Parallel Particle... User manual

SKC 232 Series User manual

SKC

SKC 232 Series User manual

SKC Sioutas Cascade Impactor User manual

SKC

SKC Sioutas Cascade Impactor User manual

SKC 225-68 User manual

SKC

SKC 225-68 User manual

SKC 233 Series User manual

SKC

SKC 233 Series User manual

SKC UMEX-100 User manual

SKC

SKC UMEX-100 User manual

SKC Disposable Parallel Particle Impactor User manual

SKC

SKC Disposable Parallel Particle Impactor User manual

SKC 590-400 User manual

SKC

SKC 590-400 User manual

SKC 231 Series User manual

SKC

SKC 231 Series User manual

SKC Reusable Parallel Particle Impactor User manual

SKC

SKC Reusable Parallel Particle Impactor User manual

Popular Laboratory Equipment manuals by other brands

Delsys Trigno Avanti user guide

Delsys

Delsys Trigno Avanti user guide

Velp Scientifica REC F20700081 instruction manual

Velp Scientifica

Velp Scientifica REC F20700081 instruction manual

Opsytec Dr. Grobel BS-04 manual

Opsytec Dr. Grobel

Opsytec Dr. Grobel BS-04 manual

MO BIO Laboratories UltraClean GelSpin DNA instruction manual

MO BIO Laboratories

MO BIO Laboratories UltraClean GelSpin DNA instruction manual

TMC STACIS 2100 Installation & operation manual

TMC

TMC STACIS 2100 Installation & operation manual

Qiagen QIAreach eHub user manual

Qiagen

Qiagen QIAreach eHub user manual

Merck Millipore Scepter 3.0 overview

Merck

Merck Millipore Scepter 3.0 overview

Owen Mumford AutoSqueeze OP6100 instructions

Owen Mumford

Owen Mumford AutoSqueeze OP6100 instructions

Peak Scientific Precision SL A5 User maintenance guide

Peak Scientific

Peak Scientific Precision SL A5 User maintenance guide

Labnet Orbit S2030-LS-B instruction manual

Labnet

Labnet Orbit S2030-LS-B instruction manual

Johnson & Johnson Sterrad 100NX user guide

Johnson & Johnson

Johnson & Johnson Sterrad 100NX user guide

GE AKTA pilot 600 user manual

GE

GE AKTA pilot 600 user manual

EIGER EG-UV03 instruction manual

EIGER

EIGER EG-UV03 instruction manual

GL Sciences OPTIC-4 user guide

GL Sciences

GL Sciences OPTIC-4 user guide

UNAOHM AP 01 HD user manual

UNAOHM

UNAOHM AP 01 HD user manual

Thermo Scientific Applied Biosystems QuantStudio 6 Pro user guide

Thermo Scientific

Thermo Scientific Applied Biosystems QuantStudio 6 Pro user guide

Hanna Instruments HI97705 instruction manual

Hanna Instruments

Hanna Instruments HI97705 instruction manual

Selecta PRESOCLAVE III 50L manual

Selecta

Selecta PRESOCLAVE III 50L manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.