3
Performance Prole
Flow range: 10 to 30 L/min
Flow compensation
accuracy: ± 5% of set ow
Compensating ow 10 L/min at 90 inches water back pressure
back pressure range: 20 L/min at 50 inches water back pressure
30 L/min at 15 inches water back pressure
Power: Rechargeable 7.2-volt lithium-ion (Li-Ion) battery (see Li-Ion
Battery Shipment on p. 4) or battery with AC charger/adapter
(100-240 V)
Battery recharge time: Approximately 5 hrs
Typical cumulative
Spore Trap* (e.g., VersaTrap
®
): 5 hrs at 15 L/min (battery only)
run time†:
BioStage viable cascade impactor: 4 hrs at 28.3 L/min (battery only)
25-mm, 0.8 µm MCE lter: 1.5 hrs at 10 L/min (battery only)
25-mm, 0.8 µm MCE lter: 2.5 hrs at 10 L/min (fully charged
battery and AC adapter/charger)
25-mm, 1.2 µm MCE lter: 9 hrs at 10 L/min (fully charged
battery and AC adapter/charger)
37-mm, 0.8 µm MCE lter: >14 hrs at 10 L/min (battery only)
* Sampling times when using spore traps are usually ≤ 10 min. SKC
recommends reducing length of sample time when using a 30 L/min
ow rate to prevent overloading the media.
† Results obtained using a new pump and new fully charged baeries. Pump
performance may vary.
The AC charger/adapter can be used to extend baery run
time, but it is not a baery eliminator. Therefore, it will
not provide indenite run times.
Temperature: Operating: 32 to 113 F (0 to 45 C)
Storage: -4 to 113 F (-20 to 45 C)
Charging: 32 to 113 F (0 to 45 C)
Protect sample pump from weather when in use outdoors.
Compare the information in this table to pump compensation range to determine appropriate applications.
Typical Back Pressure of Sampling Media (inches water)
Flow Rate (L/min) 10 12 15 20
Filter/Pore Size (µm)
25-mm MCE, 0.8 65 80 103 148
25-mm MCE, 0.45 138 172 > 200 —
37-mm MCE, 0.8 22 28 36 51
37-mm PVC, 5.0 9 11 15 21