Sleep8 has been tested and found to comply within the limits for
a Class B digital device, pursuant to Part 15 of FCC rules.
The included adapter is ETL and RoHS certified. Patent Pending.
Model No: SLP82043
Environmental Characteristics:
Operating: 50˚F to 100˚F, 15% to 70% humidity
Storage: 32˚F to 131˚F, 15% to 70% humidity
Electrical Characteristics:
Adapter input: AC 100~240V, 50/60HZ
Adapter output: DV 5V, 1A
Power consumption: 4W Max
Battery: 1 lithium ion battery, 2200 mAh. UL listed,
Passed UN 38.3
Indications of Use, Cautions,
& Specifications
Quick Start Guide
For sanitizing all CPAP supplies and accessories
SLP82043
Indications of Use: Sleep8 is designed to sanitize a Positive Airway Pressure (PAP)
machine’s supplies and accessories. Individuals who use a PAP machine for Obstructive
Sleep Apnea (OSA) can use Sleep8 to sanitize all PAP supplies and accessories on a daily
basis, or when the user wants the added benefit of thorough sanitization.
Always wipe your PAP mask clean to remove any visible dirt or oils prior to use. The
tubing must be disconnected from the PAP machine, then connected to the Sleep8 bag’s
internal port. All supplies and accessories zipped inside the bag will be disinfected
during the sanitization process.
DANGER: Keep away from water sources including rain, bathtubs, sinks, and pools.
DANGER: Do not start Sleep8 if the bag is not zipped closed.
DANGER: Keep away from open flame or fire.
DANGER: Keep away from children. It is not a toy.
DANGER: Do not use in explosive air environments, near gas vapors, or other
flammables, etc.
DANGER: Do not place any items on top of your Sleep8 or bag, particularly
flammables or ignitable items.
DANGER: Never place any living thing inside the Sleep8 bag.
WARNING: Do not dismantle the Sleep8 device.
WARNING: Do not run the Sleep8 device for more than 2 hours consecutively.
WARNING: There are sensitive electronics inside the Sleep8 device. If your Sleep8
device or bag is soiled, wipe it clean with a dry cloth. Do not submerge your Sleep8
or the bag, and do not use any chemical cleaners to clean them.
WARNING: Do not use Sleep8 while your PAP machine is in use. Sleep8 is not a PAP
machine.
REASONS TO STOP USE: Sleep8 should not be used to clean masks that have come
in contact with lotions, creams, oils, or fragrances. Discontinue use of the CPAP if
patient develops any acne or rash along mask contact area.
Have questions or need replacements?
Visit MySleep8.com or call 888-549-9799.
The Charger
Power up your Sleep8 device using the
micro USB charger.
The Bag
Where the magic happens. Connect
your Sleep8 device to the filter bag
for simple and safe sanitization of
all CPAP components.
The Sleep8 Device
Your Sleep8 uses activated
oxygen technology to completely
sanitize your CPAP supplies and
accessories.
What’s Inside
1
2
3 Easy Steps
Ensure your mask has been
wiped clean to remove any
visible dirt or oils, prior to use.
Unzip the bag and find the round white valve inside. Detach the
tubing from your CPAP machine and attach it to this valve. Put all
CPAP tubing, mask, cushions, and dry water chamber* inside the
bag. Then zip the bag closed.
Press the power button until you
hear a beep and see the Sleep8
screen light up. The screen will
count down until the sanitization is
complete. It’s normal to hear the
fan oscillate and pause for short
periods. When done, your Sleep8 will automatically turn off.
The bag should be left closed for 1 hour after the cleaning is
completed. Afterward, unzip the bag to find your sanitized CPAP
accessories ready to use.
When the device is plugged in using the
Sleep8 charger, the device will beep once, the
screen will be dim, and the battery icon will
show the charging progress. Once the icon
shows 4 full bars, your Sleep8 is fully charged
and ready for action. Note: You can use
Sleep8 to clean while it’s charging.
Slide your Sleep8’s triangular
port onto the triangular
valve on the outside of the
bag. Make sure to match the
blue dots on the device and
the bag’s port. They should fit snuggly together. The device will
only work once there’s a secure connection. Again, make sure the
bag is completely zipped closed.
Connect Sleep8
outside the bag.
3
Press start & relax.
Charging your Sleep8 device.
When you notice the battery is low, plug
it in using the Sleep8 charger included.
You cannot overcharge your Sleep8.
O3
X
Alarm Cues
Battery
The battery icon indicates the power
level of the device. Depending on the
environment, the battery should last
about 2-3 uses per charge. When
this icon is low, be sure to charge
your Sleep8.
The Screen
Timer
The 2-digit countdown
indicates how many minutes
remain. Once fully sanitized,
the device will beep 5 times,
then “OK” will display in
place of the numbers.
Fan
The fan icon will
pulse during the
sanitization
process.
X
A red “X” will display
next to O₃ when the
bag is not properly
connected.
O3
The O₃ icon will display to indicate
the bag is correctly connected and
when sanitization is in process.
Device not connected properly
“X” will display next to “O₃” and device will beep 30 for seconds.
Press the power button to cancel the alarm.
X
Low battery
“E1” will pulse and device will
beep. Recharge your Sleep8.
Note: The device will function
while being recharged.
E1
E2 Fan failure
“E2” will pulse and device
will beep. Contact customer
support at MySleep8.com
for assistance.
Wipe CPAP accessories.
Put them in the bag.
Zip it closed.
For patients using a CPAP with a water chamber: always dry your water chamber thoroughly
before placing it in the bag.
In rare instances, certain proprietary tubes require special instructions. Please visit the FAQ
section under MySleep8.com for more information.
*
Visit MySleep8.com for a video
tutorial and other resources.

Popular Medical Equipment manuals by other brands

Powerex Medical FirstCall MFLD-LIQ-NFPA-GG-PSI operating & maintenance manual

Powerex

Powerex Medical FirstCall MFLD-LIQ-NFPA-GG-PSI operating & maintenance manual

DMT 630MA user guide

DMT

DMT 630MA user guide

Roscoe Medical TENS 7000 To Go Quick reference instructions

Roscoe Medical

Roscoe Medical TENS 7000 To Go Quick reference instructions

Jenx Multiseat Instructions for use

Jenx

Jenx Multiseat Instructions for use

Mindray WATO EX-55 Service manual

Mindray

Mindray WATO EX-55 Service manual

Coviden Kangaroo user manual

Coviden

Coviden Kangaroo user manual

Omron HJA-306 CaloriScan instruction manual

Omron

Omron HJA-306 CaloriScan instruction manual

Lowenstein Medical JOYCEeasy Instructions for use

Lowenstein Medical

Lowenstein Medical JOYCEeasy Instructions for use

Otto Bock Malleo Immobil Instructions for use

Otto Bock

Otto Bock Malleo Immobil Instructions for use

BIOTRONIK ICS 3000 Technical manual

BIOTRONIK

BIOTRONIK ICS 3000 Technical manual

Zimmer Biomet Arcos Modular FemoralRevision System SURGICAL TECHNIQUE

Zimmer Biomet

Zimmer Biomet Arcos Modular FemoralRevision System SURGICAL TECHNIQUE

GERATHERM UniqueResc+ instruction manual

GERATHERM

GERATHERM UniqueResc+ instruction manual

Emfit QS IP-9360 quick guide

Emfit

Emfit QS IP-9360 quick guide

Jumper JPD-ES100 manual

Jumper

Jumper JPD-ES100 manual

ISKOmed Belladorm Max Instructions for use

ISKOmed

ISKOmed Belladorm Max Instructions for use

TPC PM25 user manual

TPC

TPC PM25 user manual

Dorel Juvenile Group Safety 1st Hospital's Choice IH417 instructions

Dorel Juvenile Group

Dorel Juvenile Group Safety 1st Hospital's Choice IH417 instructions

CORFLEX Ranger II GS Application Instructions

CORFLEX

CORFLEX Ranger II GS Application Instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.