ENGLISH – 5
4. WARNINGS TO PATIENT
The patient should be warned of the potential risks of defibrillator
malfunction if he is exposed to external magnetic, electrical, or
electromagnetic signals.
These potential interference sources may cause conversion to
inhibited mode (because of noise detection), erratic delivery of VT or
VF therapies, nominal programming, or much more rarely,
irreversible damage to the device’s circuits.
The main sources of high magnitude electromagnetic interference
are: powerful radiofrequency equipment (radar), industrial motors
and transformers, arc-welding equipment, high power loudspeakers.
Electrical equipment: Household electrical appliances do not affect
the functioning of the defibrillator, providing they are insulated to
current standards. However, patients should avoid using induction
ovens and cookers.
Antitheft gates: Since antitheft devices at the entrance to stores are
not subject to any safety standards, it is advisable to spend as little
time as possible in their vicinity.
Airport detection systems: Since airport detection systems are not
subject to any safety standards, it is advisable to spend as little time
as possible in their vicinity.
Work environment: The patient's work environment may be an
important source of interference. In that case, specific
recommendations may be required.
Cellular phones: Radiofrequency signals can interfere with the
functioning of PARADYM VR 8250 if the handset is placed too close
to the defibrillator. It is advisable to maintain a minimum distance of
15 cm (6 inches) between the cellular telephone and the implanted
device, when the telephone is turned on.
Sun exposure: Avoid prolonged sun exposure, since it may
overheat the defibrillator casing and its electronic components.
6– ENGLISH
Audible alarms: The patient should be warned about possible
audible alarms emitted by the device.
CAUTION: Do not tap sharply on the ICD can after implant, because
the ICD's sensing circuits can detect this as R-waves, and such
oversensing could result in inappropriate pacing, inhibition, or
therapy. Normal activities after implant do not result in such
oversensing.
4.1. RISKS RELATED TO MEDICAL ENVIRONMENT
It is advisable to carefully monitor defibrillator operation prior to and
after any medical treatment during which an electrical current from an
external source passes through the patient's body.
Magnetic Resonance Imaging: MRI is strictly contraindicated in
cardiac defibrillator patients.
Electrocautery or diathermy device: Diathermy and electrocautery
equipment should not be used. If such devices must be used:
1. Before procedure, deactivate ATP and shock therapies. 2. During
the procedure, keep the electrocautery device as far as possible from
the cardiac defibrillator. Set it at minimum intensity. Use it briefly.
3. After the procedure, check for proper implant function. The device
should never be exposed directly to the diathermy source.
External defibrillation: PARADYM VR 8250 is protected from
external defibrillation shocks. Before external defibrillation, deactivate
ATP and shock therapies. During external defibrillation, it is advisable
to avoid placing the defibrillating paddles directly over the casing or
over the lead. The defibrillating paddles should preferably be placed
in an anteroposterior position. Avoid any direct contact between the
defibrillation paddles and the conductive parts of the implanted leads
or casing of the implanted device. After external defibrillation, check
for proper device function.
Internal pacemaker: Use of the defibrillator is contraindicated in
implantable pacemaker patients.
ENGLISH – 7
Radiation therapy: Avoid exposure to ionizing radiation. Betatrons
are contraindicated. If high doses of radiation therapy cannot be
avoided, the defibrillator should be protected from direct exposure
with a screen. ATP and shock therapies should be disabled during
exposure and proper device function should be checked regularly
afterwards. Resulting damage may not be immediately detectable. If
irradiation of tissues close to the implantation site is necessary, it is
recommended that the cardiac defibrillator be moved. As a safety
measure, an external defibrillator should be immediately available.
Ultrasound therapy (lithotripsy): It is advisable not to administer
ultrasound therapy (or lithotripsy) to a patient with a defibrillator
implanted in the abdominal position. Concentrating the ultrasonic
field over the device could harm the patient and cause damage to the
defibrillator.
Diagnostic ultrasound (echography): The defibrillator is not
affected by ultrasound imaging devices.
Transcutaneous Electrical Nerve Stimulation (TENS): TENS may
interfere with defibrillator function. If necessary, the following
measures may reduce interference: 1. Place the TENS electrodes as
close together as possible and as far as possible from the pulse
generator and leads. 2. Monitor cardiac activity during TENS use.
Scales with body fat monitors and electronic muscle
stimulators: A patient with an implanted PARADYM VR 8250 should
not use these devices.
Surgical procedure: For safety reasons, it is preferable to not
program the Rate Response function before any surgical procedure
on the defibrillator patient.
4.2. STORAGE
The defibrillator is packaged in sterile packaging inside a cardboard
storage box. It is recommended that the device be stored at a
temperature ranging from 0 °C to 50 °C.
8– ENGLISH
Devices subjected to excessive impact, such as those that have
been dropped on a hard floor, should not be implanted. Any device
subjected to such an impact should be returned to your Sorin Group
representative for examination.
Devices MUST NOT be interrogated and programmed within the
vicinity of other devices.
4.3. CHARACTERISTICS OF THE STERILE PACKAGING
The sterile packaging contains:
―the defibrillator
―a ratcheting screwdriver
―a DF-1 defibrillating connector insulating plug
All of this equipment is ethylene oxide sterilized and hermetically
sealed in two-ply clear packaging compliant international standards.
5. IMPLANT PROCEDURE
5.1. NECESSARY EQUIPMENT
Implantation of PARADYM VR 8250 requires the following
equipment:
―Sorin Group dedicated programmer, equipped with the smartView
software interfaced with the programming head,
―pacing system analyser, as well as its sterile connecting cables,
to evaluate the pacing and sensing thresholds,
―a complete set of leads with corresponding introducers,
―physiological signal monitor capable of displaying simultaneously
the surface ECG and arterial pressure,
―an external defibrillator with sterile external paddles,
―sterile cover for the telemetry head.