Sper scientific 850050 User manual

INTRODUCTION
This simple and easy to use waterproof
pH Pen measures from 2~12 pH.
DISPLAY / FACEPLATE
CALIBRATION
 Calibrate routinely before use, an-
ytime the readings seem erratic, and
whenever the batteries are replaced.
 Remove the protective cap.
 Clean the electrode with de-ionized/
distilled water. Note: Crystals may
form in the cap and these can be
soaked or rinsed away. Do not touch
or rub the glass bulb.
 Before first use or after extended
storage, soak the electrode in
distilled water for 10~30 minutes.
 Press >1 sec to turn on the unit.
 For 1-point calibration, dip/stir the
electrode into pH 7 buffer solution
and press the button. If the
probe is damaged or cannot detect
the buffer, the unit automatically exits
calibration mode after 10 seconds.
 During calibration, "CA" flashes on
the display.
 Then, "CAL" and the pH value are
shown.
 "SA" (save) is displayed as the
calibration point is saved.
 "En" (end) displays when the
calibration ends.
 For 2 or 3 point calibration, start with
pH 7, then repeat the above steps for
pH 4 and/or pH 10.
 For highest accuracy, calibrate at the
sample temperature.
 Repeat the above procedures until
readings stabilize.
 NOTE: If the buffer value is not
standard (for example 4.2), press ▲
to adjust the displayed value.
"SA" (save) is displayed, followed by
"En" (end).
MEASUREMENT
 Calibrate the pen as needed.
 Remove the protective cap and clean
the electrode with de-ionized/distilled
water. Do not touch the glass bulb.
 Press >1 sec to turn on the unit.
 Dip/stir the electrode into the sample.
 The pH value is displayed.
 "- - -" is displayed when the pH
sample is out of range.
 During measurement, press ▲to
freeze the reading on the LCD.
"Hold" is displayed. Press ▲again to
return to normal measurement.
 The pH Pen automatically shuts off
after 20 minutes of inactivity. To
disable this feature, start with the unit
off, then simultaneously press the
and ▲buttons. Release both buttons
when "n" is displayed. The unit
returns to measurement mode.
 After use, turn off the unit and rinse
the electrode.
 Protect the electrode's lifespan by
adding a few drops of KCL (potassium
chloride) solution into the protective
cap before storage.
 Always keep the protective cap on
when the unit is not in use.
Rev. 3/29/2011
Pull to
Remove the
Protective Cap
Electrode
FIRST USE: If the
unit fails to turn on,
check that the
batteries are working
and properly
installed.
HoldCal
7.0
ATC
87.5C
F
pH Value
Symbol will
flash when
battery is low
Hold
button
Calibration
button
On/Off
button
▲
HLD
CAL
LARGE DISPLAY PH PEN
850050
WARRANTY: Sper Scientific warrants
this product against defects in materials
and workmanship for a period of 1 year
from the date of purchase, and agrees
to repair or replace any defective unit
without charge. If your model has since
been discontinued, an equivalent Sper
Scientific product will be substituted if
available. This warranty does not cover
probes, batteries, or damage resulting
from accident, misuse, or abuse. To
obtain warranty service, ship the unit
postage prepaid to:
SPER SCIENTIFIC
7720 E Redfield Rd, Ste 7
Scottsdale, AZ 85260
Email: [email protected]
Include a description of the problem
and your return address along with the
returned item. Return the warranty card
within 10 days or register at:
www.sperscientific.com
BATTERY REPLACEMENT
 Replace all 4 LR44 batteries when the
low battery icon is flashing, or when
troubleshooting (below) fails.
 Unscrew the cover and install the
batteries with the flat positive (+) side
of the batteries facing the spring.
 Do not remove the O-ring.
 Remove the batteries when the unit is
not in use for an extended period.
TROUBLESHOOTING
 If there is no display, or you have
erratic readings after having soaked
the electrode in distilled/de-ionized
water for at least 1 hour prior to
calibration, check the continuity of the
batteries by removing them for 1
minute, then reinsert.
 Slow Response: Soak the electrode in
tap water for 10-15 minutes, then rinse
with distilled/de-ionized water.
 "- - -" displays when the pH sample is
too acidic or too alkaline.
 No Display: Be sure to press for at
least one second. If the problem
persists, see Battery Replacement
(above).
 If your readings are still erratic after
calibration, soak the electrode for 10
to 30 minutes in distilled water.
 Note: It is normal for the pH value to
fluctuate quickly when the electrode is
exposed to air and not immersed.
SPECIFICATIONS
Range 2~12 pH
Resolution 0.1 pH
Accuracy
Adjustable
Calibration
Points
pH 4
pH 7
pH 10
3.5~4.5
6.5~7.5
9.5~10.5
Operating
Environ.
31~122°F
0~50°C
<80 RH
Weight 2.4 oz, 68g with battery
Dimension 6½" x 1½" x 1"
165 x 35 x 32 mm
±0.3 pH

Other manuals for 850050

1

Other Sper scientific Accessories manuals

Sper scientific 330001 User manual

Sper scientific

Sper scientific 330001 User manual

Sper scientific 330006 User manual

Sper scientific

Sper scientific 330006 User manual

Popular Accessories manuals by other brands

HAMPTON BAY HB-7311-03 Use and care guide

HAMPTON BAY

HAMPTON BAY HB-7311-03 Use and care guide

HANYOUNG NUX PR Series instruction manual

HANYOUNG NUX

HANYOUNG NUX PR Series instruction manual

Delta VIS100-30G12D operating manual

Delta

Delta VIS100-30G12D operating manual

Tex Energy Tex-Flex 10 quick start guide

Tex Energy

Tex Energy Tex-Flex 10 quick start guide

T'nB SHPLUG01 user manual

T'nB

T'nB SHPLUG01 user manual

PRO-DESIGN B-SAFE 105 owner's manual

PRO-DESIGN

PRO-DESIGN B-SAFE 105 owner's manual

Astral Pool EF-4 Series TECHNICAL MANUAL. START-UP AND OPERATION

Astral Pool

Astral Pool EF-4 Series TECHNICAL MANUAL. START-UP AND OPERATION

Silvercrest 91210 Operation and safety notes

Silvercrest

Silvercrest 91210 Operation and safety notes

Sferalabs Exo Sense Pi quick reference

Sferalabs

Sferalabs Exo Sense Pi quick reference

Silvercrest SSSE 300 A1 operating instructions

Silvercrest

Silvercrest SSSE 300 A1 operating instructions

Air Weigh AW5800 installation guide

Air Weigh

Air Weigh AW5800 installation guide

superbrightleds BRI819P-B-D user manual

superbrightleds

superbrightleds BRI819P-B-D user manual

Silvercrest SSHK 100 B2 operating instructions

Silvercrest

Silvercrest SSHK 100 B2 operating instructions

Minusforty 22-UDGH owner's manual

Minusforty

Minusforty 22-UDGH owner's manual

HYDACELECTRONIC HPT 1000S instruction manual

HYDACELECTRONIC

HYDACELECTRONIC HPT 1000S instruction manual

RecorderGear PC100 user manual

RecorderGear

RecorderGear PC100 user manual

Avtech TMP-SDT-SEN Installation note

Avtech

Avtech TMP-SDT-SEN Installation note

Rcom USB 50 user manual

Rcom

Rcom USB 50 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.