Sper scientific 850062 User manual

Test Tube pH Pen
850062
Test Tube pH Pen with
Automatic Temperature
Compensation (ATC)
850063

2
CONTENTS
INTRODUCTION ........................2
FEATURES.............................3
MATERIALS SUPPLIED ..................3
FRONT PANEL DESCRIPTION ............4
LCD DISPLAY. . . . . . . . . . . . . . . . . . . . . . . . . . .4
CALIBRATION ..........................5
MEASUREMENT PROCEDURES ..........7
TROUBLESHOOTING...................11
ERROR CODES. . . . . . . . . . . . . . . . . . . . . . . .12
BATTERY INSTALLATION
AND REPLACEMENT ...................12
CARE AND MAINTENANCE..............14
SPECIFICATIONS ......................15
WARRANTY ...........................16
INTRODUCTION
This Sper Scientic Test Tube pH Pen
(Model 850062/ 850063) is designed for easy
calibration and simple, accurate measurement.
The 12 cm long probe is capable of testing
samples even in narrow-mouthed containers.
The waterproof meter provides one-button,
three-point calibration with customizable
calibration values. Other features include
a data-hold function, automatic shuto and
low-battery indicator. Model 850063 also
measures temperature and provides automatic
temperature compensation.

3
FEATURES
• Waterproof IP65 housing
• Hold function
• One-button, three-point calibration
• Automatic buer recognition for
calibration accuracy
• Customizable calibration values
• Measuring stability indicator
• Protective storage bottle for sensor
• Low-battery indicator
• Automatic shuto
• Dual pH and temperature display
(Model 850063)
• Monitors temperature in °F or °C
(Model 850063)
MATERIALS SUPPLIED
• Meter with protective storage bottle
for sensor
• Two CR2032 button batteries
• Instruction Manual

4
FRONT PANEL DESCRIPTION
LCD DISPLAY
LCD DISPLAY
-The 1st display shows the measured pH
reading.
-The 2nd display shows the temp. (8692)
-
Auto Temp. Compensation
- =Data hold
Cal=Calibration mode
-ATC=
8692
8691
8691/92
8691/8692
KEYPAD
OPERATION
1. Remove the probe storage bottle from
meter to expose the electrode out. It is
n ormal if you find white crystals are
present on the c ap or electrode
assembly.
3
CAL
H
CAL
H
ATC
F
C
O
O
H
1. LCD
2. HOLD
3. CAL (calibration)
4. Power
850062
1. PH reading
2. Data-hold indicator
3. Calibration
4. Low-battery icon
850063
1. PH reading
2. Data-hold indicator
3. Calibration
4. Automatic Temperature
Compensation (ATC)
5. Low-battery icon
6. Temperature reading
7. Temperature units
1
2 3
4
LCD DISPLAY
-The 1st display shows the measured pH
reading.
-The 2nd display shows the temp. (8692)
-
Auto Temp. Compensation
- =Data hold
Cal=Calibration mode
-ATC=
8692
8691
8691/92
8691/8692
KEYPAD
OPERATION
1. Remove the probe storage bottle from
meter to expose the electrode out. It is
n ormal if you find white crystals are
present on the c ap or electrode
assembly.
3
CAL
H
CAL
H
ATC
F
C
O
O
H
1
2 3 4
1
2 3 54
6 7

5
CALIBRATION
The meter should be calibrated regularly with
standard USA, NIST or DIN buers to ensure
accuracy. If the meter is used frequently, daily
calibration is recommended.
Voltage Range for Each Calibration Point
pH 4.00 97 mV ~ 250 mV
pH 7.00 -60 mV ~ 60 mV
pH 10.00 -250 mV ~ -97 mV
Calibration Procedure
1. Press POWER button to turn the meter on.
2. Pour the pH buer into a clean container
and place the electrode into the buer. We
recommend starting with a middle range
buer such as pH 7. Make sure the electrode
is completely immersed in the buer.
3. Press CAL to enter Calibration Mode.
“CAL” displays for one second, followed by
the pH value. Model 850063 also displays
the temperature.
Calibration is necessary and should be
done regularly, recommended everyday
if the meter is used often. The unique
calibration design of the meter features
automatic buffer recognition to avoid errors.
CALIBRATION MODE (CAL)
7
1. Power on the meter.
2. Place the electrode into a buffer solution
(4,7or10), pH7 should be calibrated first
and then 4 or 10 pH for better accuracy
3. Press " " to enter pH calibration
mode. The icon "CAL” will appear on the
LCD for one second and then pH value
(4,7 or 10) will be displayed on the LCD.
8691/8692 Calibration
4. If probe or buffer is in error:
If the buffer is incorrectly inserted or
the probe is damaged or the probe
can not detect buffer in below voltage
range, the meter will escape calibration
mode automatically after 10 seconds.
Text "End" will appear on
the LCD in one second
and then will go back to
normal status.
CALATC
F
O

6
4. If the displayed value is dierent from the
standard buer that you are using, press
and hold down HOLD to change the value.
Each buer value may be adjusted within a
specied range:
a. 4.00 pH may be adjusted from
3.50 to 4.50
b. 7.00 pH may be adjusted from
6.50 to 7.50.
c. 10.00 pH may be adjusted from
9.50 to 10.50.
5. When the value stabilizes, the meter will
save the value. “SA” displays.
6. “End” displays and the meter returns to
Normal Mode.
7. Rinse the probe with deionized water,
a rinse solution, or tap water.
8. Repeat steps 2-7 to calibrate the meter with
the remaining two buer solutions.
8
PH 4.00 : 97mV..... 250mV
PH 7.00 : - 60mV..... 60mV
PH10.00 : - 250mV..... -97mV
Acceptable voltage range of each point:
5. If the probe recognizes the buffer:
If the probe successfully recognizes the
buffer, the buffer pH value (4 or 7 or 10)
will appear on the display in 2 seconds.
If the calibration buffer is not 4,7,10 but
another value, such as 7.01, just
press " " to change the value.
7. To save the calibration value:
When the electrode reads a stable value
and user stops pressing any keys, the
meter will automatically save the value
and then escape the calibration mode.
6. The adjustable cal. point range for 4.0
pH is from 3.50 to 4.50. For 7.0 pH is
from 6.50 to 7.50. For 10.0 pH is from
9.50 to 10.50
CALATC
F
O

7
Note…
If there is an error due to a damaged
or incorrectly inserted probe, or if the
buer voltage out of range, the meter will
automatically exit calibration mode in 10
seconds. “End” displays for one second and
the meter returns to Normal Mode.
MEASUREMENT PROCEDURES
Note…
The electrode should be soaked for at least
10 - 30 minutes in KCL solution before rst
use, or any time it has dried out. It is normal
for white crystals to be present on the cap or
electrode assembly.
4
Rotate the
storage bottle in
counter-clockwise
direction and then pull out
the bottle completely to expose the electrode.
Electrode
2. Dip the electrode into the test solution.
Press " " to power on the meter
and tir it to get a stable reading.
3. A small dot " " is flashing while the
meter is in measurement mode. The
LCD not only shows the measured pH
value but also displays the temperature
in model 8692.
8692
WARNING: For pH meters, please
always make the bottle is filled with
storage buffer to keep the electrode
wet and in a good storage condition.
ATC
C
O

8
1. Unscrew the storage bottle and remove it
from the probe, exposing the electrode.
2. Place the electrode into the test solution.
3. Press POWER button to turn the meter on.
4. Stir the probe gently to create a
homogeneous sample and shorten the
stabilizing time.
5. The pH decimal point will ash to indicate
that the meter is in Measurement Mode.
6. The pH value displays on the LCD. Model
850063 also displays the temperature.
7. Press POWER button to turn the meter o.
8. Rinse the electrode with deionized water, a
rinse solution, or tap water.
4
Rotate the
storage bottle in
counter-clockwise
direction and then pull out
the bottle completely to expose the electrode.
Electrode
2. Dip the electrode into the test solution.
Press " " to power on the meter
and tir it to get a stable reading.
3. A small dot " " is flashing while the
meter is in measurement mode. The
LCD not only shows the measured pH
value but also displays the temperature
in model 8692.
8692
WARNING: For pH meters, please
always make the bottle is filled with
storage buffer to keep the electrode
wet and in a good storage condition.
ATC
C
O

9
Data Hold Function
1. During measurement, press HOLD to
freeze the reading on the display. “H”
appears at the top of the LCD and the pH
decimal point stops ashing.
2. Press HOLD to return to Normal Mode.
Automatic Shuto
The meter automatically turns o after 20
minutes of inactivity. To disable this function:
1. If the meter is on, press POWER button to
turn the meter o.
2. Press both POWER button and HOLD for
two seconds.
3. The meter will turn on and “n” will display.
" "
4. Press to freeze current readings.
The " " icon will appear on the LCD
and the small dot will not flash . Press
" " again to release the hold mode.
5. Turn off the meter by pressing " "
button.
8692
5
7.Please see page 7 for the calibration
procedures.
6.Once finishing the measurement, clean
the electrode, put back the bottle, store
o
the meter under the 0~50 C.
This meter will shut off automatically 20
minutes of inactivity. For operating longer
time, you can disable the sleep mode.
To disable the auto power off:
AUTO POWER OFF
(SLEEP FUNCTION)
Before power on, " "
pressing " " +
keys simultaneously until a
"n" appeared on the screen
and then release keys to
return to normal mode.
H
H
ATC
F
O
" "
4. Press to freeze current readings.
The " " icon will appear on the LCD
and the small dot will not flash . Press
" " again to release the hold mode.
5. Turn off the meter by pressing " "
button.
8692
5
7.Please see page 7 for the calibration
procedures.
6.Once finishing the measurement, clean
the electrode, put back the bottle, store
o
the meter under the 0~50 C.
This meter will shut off automatically 20
minutes of inactivity. For operating longer
time, you can disable the sleep mode.
To disable the auto power off:
AUTO POWER OFF
(SLEEP FUNCTION)
Before power on, " "
pressing " " +
keys simultaneously until a
"n" appeared on the screen
and then release keys to
return to normal mode.
H
H
ATC
F
O

10
4. The meter returns to Normal Mode.
Automatic shuto is now disabled.
Note…
The meter will default to automatic shuto
each time it is turned o.
Automatic Temperature Compensation
(model 850063)
“ATC” displays at the top of the LCD. The
temperature is displayed in either °C or °F.
To select the temperature unit:
1. If the meter is on, press POWER button to
turn the meter o.
2. Press and hold both POWER button and
CAL until the temperature unit appears on
the LCD.
3. Press HOLD to toggle between °C and °F.
4. Press CAL to save the selection.
8
PH 4.00 : 97mV..... 250mV
PH 7.00 : - 60mV..... 60mV
PH10.00 : - 250mV..... -97mV
Acceptable voltage range of each point:
5. If the probe recognizes the buffer:
If the probe successfully recognizes the
buffer, the buffer pH value (4 or 7 or 10)
will appear on the display in 2 seconds.
If the calibration buffer is not 4,7,10 but
another value, such as 7.01, just
press " " to change the value.
7. To save the calibration value:
When the electrode reads a stable value
and user stops pressing any keys, the
meter will automatically save the value
and then escape the calibration mode.
6. The adjustable cal. point range for 4.0
pH is from 3.50 to 4.50. For 7.0 pH is
from 6.50 to 7.50. For 10.0 pH is from
9.50 to 10.50
CALATC
F
O

11
5. “SA” displays for one second and the meter
returns to Normal Mode.
TROUBLESHOOTING
Problem Action
The meter
will not turn
on / the
LCD display
is blank.
Press POWER button for more
than 0.1 seconds.
Check whether the batteries are
correctly inserted.
Remove the batteries for one
minute and then reinstall them.
Replace the old batteries with
new ones.
Response
time is slow
Clean the probe by immersing
the electrode in tap water for 10-
15 minutes. Rinse thoroughly with
distilled water or use a general-
purpose electrode cleaner.
Rapidly
uctuating
pH value
This is normal when the electrode
is exposed to air. The value will
stabilize when the electrode is
immersed in a solution.
8692 is capable of measuring with Automatic
Temperature Compensation. In 8692, "ATC"
will show in the middle top of the LCD.
AUTOMATIC TEMPERATURE
COMPENSATION (ATC) &
TEMP. UNIT SETTING
o o
To select the temp. unit ( C or F), turn off
the meter first. When the meter is off, press
" " and " " at the same time until
the C or F appears on the LCD. Press
" " to s elect the preferred unit and then
press " " to save. " " will appear on
the LCD for one sec and then go back to
normal.
Temp. Unit Setting 8685/86
6
Note: The disable sleep mode will be
invalid after every power off.
o
C

12
ERROR CODES
Error Code Description
The pH reading
displays “- - -”
The pH value is out of range:
the solution is too acidic or
too alkaline
The temperature
reading displays
“H” or “L”
(Model 850063)
The temperature value is out
of range: the temperature is
too high or too low
BATTERY INSTALLATION AND
REPLACEMENT
Note…
Remove the batteries when the meter will not
be used for one month or longer.
This meter uses two CR2032 button
batteries, which should provide over 100
hours of continuous use. When the low
battery indicator appears on the LCD, battery
replacement is needed.
Power on but no display
TROUBLESHOOTING
1) Make sure the time of pressing
power key is more than 100 mS.
2) Check the battery are in place and
make good contact and correct
polarity.
3) Replace with new batteries and try
again.
4) Move away the batteries for one
minute and then put back again.
10
Out of pH range, too
acidic/or too alkaline.
"H." Or "L."
Out of temp. range,
too cold/or too hot.
Slow response
Clean probe by immersing the electrode
in tap water for 10-15 minutes, then
rinse thoroughly with distilled water
or use a general purpose electrode
cleaner.
It is normal when the electrode is
not immersed in the water but
exposed in air.
pH value fluctuate quickly
LCD display “ - - - “
ATC
C
O
ATC
C
O
Power on but no display
TROUBLESHOOTING
1) Make sure the time of pressing
power key is more than 100 mS.
2) Check the battery are in place and
make good contact and correct
polarity.
3) Replace with new batteries and try
again.
4) Move away the batteries for one
minute and then put back again.
10
Out of pH range, too
acidic/or too alkaline.
"H." Or "L."
Out of temp. range,
too cold/or too hot.
Slow response
Clean probe by immersing the electrode
in tap water for 10-15 minutes, then
rinse thoroughly with distilled water
or use a general purpose electrode
cleaner.
It is normal when the electrode is
not immersed in the water but
exposed in air.
pH value fluctuate quickly
LCD display “ - - - “ ATC
C
O
ATC
C
O

13
1. If the meter is on, press POWER button to
turn the meter o.
2. Remove the battery cover on the back
of the meter by using the slot to turn it
clockwise. The cover will now come o
easily when you turn the timer over.
(Take care not to discard the washer in
the battery compartment.)
3. Remove the old batteries and replace
with 2 new CR2032 button cell batteries,
ensuring correct polarity.
To install the batteries:
1. Turn off the meter.
2.
3. Replace the old batteries with two new
button cells CR2032.
Loose the battery cover in clockwise
direction by coin (DON'T discard the
black washer!)
4. Make sure the batteries are in place
and the polarity is correct.
5. Put back the battery cover and
turn it
in counter-clockwise direction to lock
1. Please re-calibrate the meters after
changing batteries.
2. Remove battery from instruments
that you do not plan to use for a
month or more. Do not leave battery
in instrument.
NOTE:
2

14
4. Replace the battery cover, positioning
the tabs so that the cover is ush with
the meter.
5. Secure the cover by turning it
counter-clockwise.
Note…
Recalibrate the meter after replacing
the batteries.
CARE AND MAINTENANCE
The electrode should be soaked for at least
30 minutes before rst use, or any time it has
dried out. It is normal for white crystals to be
present on the cap or electrode assembly.
Always handle and store the electrode
with care.
• Do NOT touch, wipe or rub the glass bulb.
• The pH glass bulb must always be kept
moist. Place it in the protective storage bottle
whenever it is not in use.
• Rinse the electrode in deionized water, a
rinse solution, or tap water before each use.
• Rinse the electrode in deionized water
between measurements.

15
If the electrode has been exposed to a solvent
immiscible with water:
• Clean it with a solvent miscible with water,
such as ethanol or acetone.
• Rinse the electrode carefully with water.
SPECIFICATIONS
ATC
Range Res Accuracy
850062 N/A
0 ~ 14.00 0.01
±0.2
850063 0 ~ 50°C /
32 ~ 122°F ±0.1
Temperature 0 ~ 50°C / 32 ~ 122°F 0.1°C /°F ±1°C / 2°F
Operating
Temperature 0 ~ 50°C / 32 ~ 122°F
Operating
RH% < 80%
Storage
Temperaure -10 ~ 50°C / 14 ~ 122°F
Storage
RH% < 90%
LCD Display 29 x 22.5 mm
Dimensions
235 x 35 x 25 mm (9¼" x 1¼” x 1")
Pen - 115 x 35 x 25 mm (4½" x 1¼" x 1")
Probe - 120 x 12 mm (4¾" x ½")
Weight 71 g (2.5 oz)

16
WARRANTY
Sper Scientic warrants this product against
defects in materials and workmanship for
one (1) year from the date of purchase, and
agrees to repair or replace any defective unit
without charge. If your model has since been
discontinued, an equivalent Sper Scientic
product will be substituted if available. This
warranty does not cover damage resulting from
accident, misuse, or abuse of the product. To
obtain warranty service, ship the unit postage
prepaid to:
SPER SCIENTIFIC LTD.
8281 E. Evans Rd., Suite #103
Scottsdale, AZ 85260
(480) 948-4448
Be sure to include a description of the problem
and your return address. Register your product
online at www.sperwarranty.com within 10 days.
10/2/2015
This manual suits for next models
1
Table of contents
Other Sper scientific Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Helena Laboratories
Helena Laboratories QuickGel 1284 Operator's manual

BD
BD Phoenix M50 System Quick reference guide

MELAG
MELAG Cliniclave 45 D Technical manual

PermkinElmer
PermkinElmer NexION 300 ICP-MS manual

Gamry Instruments
Gamry Instruments Reference 3000 Operator's manual

Baker
Baker SterilGARD SG303 Operator's manual