Super Systems DP 2000 User manual

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 1 of 30
Model DP 2000
Portable Digital Dew Point Analyzer
OPERATIONS MANUAL
Super Systems Inc.
7205 Edington Drive
Cincinnati, OH 45249
513-772-0060
Fax: 513-772-9466
www.supersystems.com

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 2 of 30
Super Systems Inc.
USA Office
Corporate Headquarters:
7205 Edington Drive
Shipping Address:
7245 Edington Drive
Cincinnati, OH 45249
Phone: (513) 772-0060
http://www.supersystems.com
Super Systems Europe
Unit E, Tyburn Trading Estate,
Ashold Farm Road, Birmingham
B24 9QG
UNITED KINGDOM
Phone: +44 (0) 121 306 5180
http://www.supersystemseurope.com
Super Systems México
Sistemas Superiores Integrales S de RL de CV
Acceso IV No. 31 Int. H Parque Industrial
Benito Juarez
C.P. 76120 Queretaro, Qro.
Phone: +52 442 210 2459
http://www.supersystems.com.mx
Super Systems China
No. 369 XianXia Road
Room 703
Shanghai, CHINA
200336
Phone: +86 21 5206 5701/2
http://www.supersystems.cn
Super Systems India Pvt. Ltd.
A-26 Mezzanine Floor, FIEE Complex,
Okhla Indl. Area, Phase –2
New Delhi, India 110 020
Phone: +91 11 41050097
http://www.supersystemsindia.com

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 3 of 30
Contents
INTRODUCTION –.....................................................................................................................................4
SPECIFICATIONS ......................................................................................................................................4
WARNINGS –.............................................................................................................................................5
STARTUP –.................................................................................................................................................5
OPERATION -.............................................................................................................................................6
INSTRUMENT DAMAGE ...........................................................................................................................8
WHAT IS DEW POINT? ............................................................................................................................10
HOW IT WORKS –....................................................................................................................................10
MAINTENANCE /SAFETY ISSUES –.....................................................................................................10
FACTORY CALIBRATION –.....................................................................................................................11
FIELD CALIBRATION –...........................................................................................................................11
RECHARGING THE MODEL DP2000 –...................................................................................................17
TROUBLESHOOTING -...........................................................................................................................17
UNIT DOESN’T POWER UP:...........................................................................................................................17
DEW POINT READING SHOWS +80°F (OR A HIGH READING)AND NEVER DROPS:..............................................18
NO VISIBLE FLOW IS SHOWN IN THE FLOW METER:........................................................................................23
DISPLAY APPEARS TO BE LOCKED UP AND NEVER CHANGES (NOT AT 80°F):...................................................24
DISPLAY READS -50°F (INDICATING THAT THE SENSOR FAILED): ...................................................................24
DISPLAY READS HIGHER THAN NORMAL AND DOES NOT MATCH OTHER DEW POINT EQUIPMENT:......................24
RETURNING THE UNIT TO SSI –...........................................................................................................24
WARRANTY..............................................................................................................................................26
SPARE PARTS –......................................................................................................................................27
APPENDIX “A” –......................................................................................................................................28
APPENDIX “B” –......................................................................................................................................29

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 4 of 30
INTRODUCTION –
Thank you for selecting Super Systems Inc. (SSi) and the DP2000 as your source
for accurate dew point measurements of:
●Endothermic Atmosphere
●Endothermic Generators
●Nitrogen / Methanol Atmosphere
●Nitrogen / Hydrogen Atmosphere
●Plant Air Systems
We have taken every precaution to protect this unit during shipment. It has been
packed in a customized foam carrier to protect it against mishandling during
shipment. Please retain this packing material to use when returning the
instrument to SSI for calibration or service. Carefully unpack the Model DP2000
Dew Point Analyzer, and if there are any signs of shipping damage, notify SSI and
the shipper immediately.
Keep this instruction book in a secure place and refer to it when there is a
question about the analyzer. An electronic version of this manual can be
downloaded for free from our website: www.supersystems.com.
SPECIFICATIONS
Dew Point Range: 0 to 80°F (-18 to 27°C)
Temperature Range:
0 to120°F (-18 to 49°C)
Power Supply:
Factory set to 115 or 240 VAC
Display Type:
LED Digital
Display Resolution:
+/- 1°F (+/- 0.1°C)
Retransmission Output:
4-20 mA (range is -50 to 80°F)
Battery Specifications:
12V Rechargeable
Battery Life:
Approx. 8 hours
Size (Closed Case):
11” x 10” x 7”
Weight:
8.2lbs (3.7 kg)

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 5 of 30
WARNINGS –
Although it is intended for use in an industrial environment, the DP2000 is a
sensitive piece of analysis equipment. Care should be taken not to drop the
analyzer or to operate it in a manner inconsistent with its intended use.
Open all sample ports and remove all soot and/or moisture from the lines
prior to attaching the sample tubing.
The analyzer must be stored at ambient temperature (65-80°F) for at least
four hours prior to operation.
When the unit is to be returned to SSI for service or any other reason, it
should be shipped in its original protective packaging. If this packaging is not
available, protect the instrument with at least four inches of foam or other
impact-absorbing material.
For maximum battery life, do not charge the unit until the “battery low”
indicator on the display is illuminated.
This unit is not designed to measure the dew points in corrosive gases, such
as ammonia (NH3), sulphur trioxide (SO3), chlorine (Cl), and hydrochloric acid
(HCl).
Please read and understand this Operations Manual before operating the
unit.
Failure to comply with these conditions may cause damage to the unit that will
not be covered under the warranty. Super Systems, Inc. is not responsible for
damage to this unit caused by disregard of these warnings, neglect, or misuse.
STARTUP –
The DP2000 Dew Point Analyzer has been calibrated and fully charged before it
was shipped from Super Systems, Inc. You can begin typical operation as soon as
the unit has been allowed to stabilize in a temperature similar to the
temperature in the heat-treating department. This is particularly important for
units that may have been sitting overnight in a delivery van in sub-zero weather,
since the rapid temperature change can cause condensation on the sensor which
will cause the unit to temporarily display inaccurate readings.

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 6 of 30
OPERATION -
The use of the DP2000 is somewhat dependent upon the application. Although
they are fundamentally alike, the operation procedures are different for sampling
heat treating furnaces and endothermic generators.
Flip the POWER switch to the on position. A green POWER ON light should
illuminate indicating that the unit is turned on. The LED display should also
illuminate showing “SSi” (Figure 1) followed by a dew point reading given in
degrees Fahrenheit (Figure 2).
Figure 1 Figure 2
Connect a sampling tube/hose to the barb fitting located on the side of the
DP2000 (Figure 3). Make sure that the hose or tubing fits over the fitting properly
and does not allow air to leak into the sampling line. A tubing and filter assembly
is supplied with the DP2000 (SSI P/N A20315 –Figure 5), which is designed to fit
the hose barb properly. Also, make sure that the sample gas is flowing through
the filter in the right direction. There is an arrow on the filter to indicate which
direction the gas should flow (Figure 4).
Figure 3 Figure 4

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 7 of 30
If the system from which gas is being sampled is under pressure, there should
be visible flow in the flow meter. Adjust the flow through the meter using the
adjusting knob on the meter.
If there is no visible flow after the sampling line is connected, the pump will need
to be turned on to “pull” a sample through the DP2000. Flip the PUMP switch to
turn on the sampling pump. The green PUMP ON light should illuminate when
the pump is on (Figure 6). You should also be able to hear the pump running
(depending on the noise level at your location).
Figure 5
Figure 6
Regardless of the application, allow the DP2000 to sample gas for two to three
minutes before recording any data. This assures that the sensor has achieved
equilibrium in the sample gas. Also, make sure that the in-line filter (P/N 31033
or equivalent) is clean and functional. Not only will this ensure that the sample
reading is not abnormally high (since soot tends to trap moisture), but it will also
prevent soot and other contaminants from entering the unit and damaging the
sensor.

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 8 of 30
The optimum flow rate of the sample gas
should be between 1.5 and 2.0 cubic feet per
hour (CFH) (Figure 7), although a flow rate as
low as 1.0 CFH is acceptable. If the unit is
reading less than 1.0 CFH, verify that there are
no obstructions to the flow such as a clogged
sample line or a poorly adjusted knob on the
DP2000’s flow meter. If these conditions do
not exist, the unit may need to be returned to
SSI for replacement of the pump. The pump’s life is dependent upon how
frequently it is used as well as the cleanliness of the sample gas. Insufficient
filtration will significantly reduce the operating life of the pump.
Heat Treat Furnace Sampling:
A gas sample may be extracted from a process
using the built-in pump. The sample tube from which the sample is taken out of
the furnace should extend into the furnace past the HOT face of the refractory.
For accurate results, a designated sample port should be used to extract the
sample. SSI offers a sample port assembly (part number 20263) which is ideal
for this purpose. If a designated sample port is not available, then a clean “burn-
off” port on an SSi Gold Probe, an industry leading oxygen sensor for
atmosphere control, can be used. Readings taken from the burn-off port on a
probe may be artificially high due to the presence of soot in the probe sheath.
Endothermic Generator Sampling:
For applications under pressure, the pump
should be switched off and the flow controlled by the small restriction valve on
the flow meter. A flow rate between 1.5 and 2.0 CFH is ideal. The sample should
be taken from the endothermic gas manifold after the gas has been cooled.
NOTE: Allow the sample port to blow out any soot and / or water before
connecting the sample tube. Failure to do so will result in inaccurate readings
and expose the sensor to potential damage.
INSTRUMENT DAMAGE
The two main causes of damage to the DP2000 are the ingestion of soot and
water. Both of these contaminants will cause erroneous readings in the short
term, and cause long-term damage to the sensor and internal components.
Soot / Particulate Contaminants
When taking a sample from a furnace or a generator, care should be taken to
reduce the amount of soot that enters the instrument. The in-line filter will trap
these particles, but cleaning the sample line before attaching the DP2000 will
Figure 7

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 9 of 30
increase the life of the filter. Furnace ports can be burned off by pumping air
through them while hot, or by removing them from the heat and mechanically
cleaning them. Generator ports should be opened before the instrument is
attached to allow any particulate buildup to be blown out. It is also helpful to tap
on the port while it is being blown out to eject any loose particles before the
instrument is attached.
If soot is allowed to collect on the dew point sensor in the instrument, it could
result in higher readings. This soot will also retain moisture than can corrode
the sensor over time. The sensor tip can be cleaned by carefully removing it
from the sample block (see Section 2.3 of the Field Calibration instructions) and
rinsing it in isopropyl alcohol. The power should be off while this is done, and the
power should remain off for at least 30 minutes after this procedure to allow all
of the alcohol to completely evaporate.
Water / Moisture Contamination
When a furnace or generator is being started up or cooled down, the resulting
gas will contain unusually high amounts of carbon dioxide and water. When the
gas cool, moisture will precipitate out and become condensation inside the
sample tubing assembly. Even if the furnace or generator is operating normally,
residual moisture may still be present in the sample tube or plumbing system.
In the same way that the ports are checked for soot (see above) they should be
checked for moisture before attaching the instrument. This is especially
important when taking a sample from a generator, since the sample port is
usually preceded by a significant amount of plumbing. All traces of moisture
should be eliminated before attaching the instrument. Failure to do so will result
in erroneous measurements and could result in damage to the analyzer.
The first signs of moisture in the instrument will be visible condensation in the
sample tubing and an unusually high dew point. The upper range of the sensor is
+80°F (27°C), so if that value is displayed on the instrument it is probably due to
the presence of moisture. If this moisture is not removed, it will cause the
sensor tip to corrode and will eventually require the sensor to be replaced.
To remove moisture from the instrument, the sample tubing and filter should be
removed from the instrument since they will probably be wet. A dry and inert gas
such as nitrogen or argon should then be flowed through the instrument (with
the pump off) for as much time as it takes to dry out. This dry-out time will
depend on the amount of moisture present in the instrument. The condition of
the sensor can be monitored by periodically reading the dew point from the
display and watching the value decrease over time. To test if it is operating
properly, verify the ambient dew point against a web-based weather station that
will report the ambient dew point for your area. If the displayed reading is within

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 10 of 30
three degrees (3°) of the reported dew point when the instrument is taken
outside, then all of the moisture has probably been successfully removed. The
wet filter and sample tubing can be re-attached after they have been completely
dried out.
To prevent the possibility of moisture damaging the instrument, be sure that the
measured dew point is below ambient levels before it is stored. If necessary,
nitrogen or argon can be used to dry out the instrument after use.
WHAT IS DEW POINT?
Dew point can be defined as the temperature at which the water vapor pressure
of the gas equals the saturated water vapor pressure. In other words, it is the
temperature at which condensation will just begin to occur as the gas is cooled.
Dew point and relative humidity are not the same measurement. Relative
humidity is the amount of water vapor in the air compared to the amount the air
could hold if it was totally saturated, and it is expressed as a percentage, not a
temperature. To determine dew point, two main variables are required: relative
humidity and temperature. The DP2000 measures both variables to compute the
displayed dew point.
HOW IT WORKS –
The dew point sensor is a “dielectric ceramic” that varies its electrical
capacitance with changes in relative humidity. The sensor is mounted in a short
probe, which is installed in a T-fitting that allows the sample gas to flow past the
sensor. The tip of this probe contains the dielectric ceramic relative humidity
(RH) sensor, as well as a built in temperature sensor to determine its dry bulb
temperature. Information from both of these sensors is used to compute the
resultant dew point, which is displayed on the digital LED display.
MAINTENANCE / SAFETY ISSUES –
One of the added features of the DP2000 is the ability to monitor the sensor’s
operating temperature through the built-in thermistor in the probe tip. The
temperature of the sample gas can be determined by pressing the switch labeled
“Sensor Temp”. It is spring loaded, so it will automatically return to displaying
dew point.
Maintaining proper sensor temperature will prevent the premature failure of the
sensor. The operating temperature of the sensor should remain below 140F
(60C) at all times. Periodic checks of the sensor temperature will verify that the

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 11 of 30
sensor is not being exposed to excessive heat. If these periodic checks show a
high sensor temperature, then the length of sample tubing should be increased
to allow for adequate cooling of the sample before it passes the sensor tip.
Continuous operation of the DP2000 Dew Point Analyzer will lead to premature
failure, since many the internal components are not designed for uninterrupted
use. If continuous monitoring of Dew Points is required, please contact Super
Systems, Inc. at 513-772-0060 to inquire about products intended specifically for
this application.
FACTORY CALIBRATION –
Factory calibration is recommended every six months if the unit is used
regularly. SSI’s calibration is NIST traceable and includes a numbered
“Certificate of Calibration”. This certificate also indicates the accuracy of the
analyzer before and after calibration. Please contact Super Systems, Inc. at 513-
772-0060 for more information regarding this service.
FIELD CALIBRATION –
It is also possible to calibrate the DP2000 in the field, which will require the
optional calibration kit (Part Number 31030). The instructions for a field
calibration are shown here, however please feel free to contact Super Systems at
513-772-0060 if you would like to review the process with us before you begin.
The calibration kit consists of two bottles of saturated salt solution in which each
bottle generates a precise relative humidity percentage (R.H. %) value. One
bottle is 11.3% R.H., and the other is 75.3% R.H. These two specific calibration
points are already pre-programmed into the microprocessor board.
1.0 Open the unit
1.1 Remove the aluminum faceplate of the DP2000 by removing the
single allen-head cap screw located at the bottom front of the
faceplate. After the screw has been removed, carefully lift the front of
the faceplate and slide it towards you about an inch. After the
faceplate has been removed, it can temporarily be rested in the lid of
the open case, to allow access to the components inside. This plate
will still be connected to the interior circuit boards, so care should be
taken to maintain all existing connections.
2.0 Locate the key components within the unit

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 12 of 30
2.1 The microprocessor board is located in the front left side of the unit.
This board contains three very small buttons that are used for
calibration. Two are next to one another, and they are marked “75.3%”
and “11.3%”, while the other has no label. The unmarked button is the
“Calibrate” button. The approximate locations of each button are
shown on this diagram:
2.2 The sensor-sampling chamber is located in the back left of the unit.
It is the gray rectangular box with brass barb fittings on either side
with a black plastic gland protruding from the center.
2.3 The sensor probe is positioned in the sensor-sampling chamber. It
is held in place by the nut on the black plastic gland.
3.0 Remove the sensor probe from the sensor sampling chamber.
3.1 Loosen the black plastic gland nut and slowly slide the sensor probe
out through the airtight seal. Care must be taken when removing this
sensor probe, since the tip is very delicate and can be easily damaged
if it is mishandled. Note that the probe has white mark at the wire
entry point, which must be aligned with corresponding white mark in
plastic gland when it is re-inserted in the sampling chamber.
4.0 Install the sensor probe into the 75.3% salt solution.
4.1 Slip the black sensor gland (supplied in the calibration kit) over the
sensor probe with the sensor tip protruding from the threaded end of
the gland and the sensor wires being flush with the top of the rubber
SPAN
BUTTON
(75.3)
ZERO
BUTTON
(11.3)
CAL
BUTTON
Dew Point Microprocessor Board

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 13 of 30
o-ring in the gland. Tighten the gland around the sensor. This does
not need to be done with a wrench or other tools, but it does need to be
tight enough to prevent ambient air from contaminating the humidity
level of the sampling chamber.
4.2 Remove the cap of the 75.3% salt solution and install the sensor
gland (with the sensor) into the salt solution. To increase the life of the
calibration salts, an effort should be made to minimize the amount of
time that the salt solution is exposed to the ambient air.
5.0 Allow the sensor to reach equilibrium with the calibration salt.
5.1 With the power to the unit still turned off, leave the sensor in the
calibration salt for a minimum of eighteen (18) hours. It is acceptable
to leave the sensor in the salt solution for a longer period of time, even
a few days, if desired.
6.0 Begin the 75.3% (Span) calibration process.
6.1 After leaving the sensor in the salt for at least eighteen (18) hours,
turn the unit on. The reading on the display is not important at this
point.
6.2 Simultaneously press the “75.3%” and “Calibration” buttons on the
microprocessor board.
7.0 Verify the 75.3% (Span) calibration.
7.1 Do not be concerned if the unit does not display 75.3, since it is not
supposed to match the value of the calibration salt.
7.2 Use the “Sensor Temp” switch on the faceplate of the unit to
determine and record the sensor temperature.
7.3 Look up this temperature on the “Theoretical Dew Point Values for
Calibration Verification” chart located in the back of this manual.
Appendix “A” will show the temperature values in Fahrenheit, and
Appendix B will show the temperature values in Celsius.
7.4 Next to the appropriate temperature, note the number in the
corresponding column titled “75.3%”. This should match with the dew
point that is shown on the display of the DP2000.

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 14 of 30
8.0 Determine the acceptability of the reading.
8.1 The value printed on the chart in Appendix A is a theoretical value,
and some variation can be expected. When a calibration is performed
at SSI, we certify (in writing) that the unit displays within +/- 1 degree
of the theoretical value after it has been calibrated. We would not
consider a calibration to be successful unless it is within +/- 1 degree,
however in the case of a field calibration, this degree of accuracy may
or may not be required. The degree of accuracy that is acceptable is
determined by the policy of the person performing the calibration.
NOTE: Keep in mind that the DP2000 only displays even numbers, and
not tenths of a degree. Therefore, a reading of 65°F could be as low as
64.50 or as high as 65.49.
9.0 Allow the sensor to achieve equilibrium at ambient atmosphere.
9.1 After the 75.3% (Span) calibration has been completed, remove the
sensor from the calibration salt and replace the cap on the salt.
9.2 Leave the sensor probe in the gland and while the unit is still on,
allow it to achieve equilibrium at the ambient atmosphere in the room.
This is accomplished by simply leaving the sensor exposed to ambient
air for between two and three minutes. You will know when this has
been accomplished when the numbers on the display begin to stabilize.
10.0 Install the sensor probe into the 11.3% salt solution.
10.1 Remove the cap of the 11.3% salt solution and install the sensor
gland (with the sensor) into the salt solution. To increase the life of the
calibration salts, an effort should be made to minimize the amount of
time that the salt solution is exposed to the ambient air.
10.2 Turn the unit off.
11.0 Allow the sensor to reach equilibrium with the calibration salt.
11.1 With the power to the unit still turned off, leave the sensor in the
calibration salt for a minimum of 24 hours. It is acceptable to leave the
sensor in the salt solution for a longer period of time, even a few days,
if desired.

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 15 of 30
12.0 Begin the 11.3% (Zero) calibration process
12.1 After leaving the sensor in the salt for at least twenty-four (24)
hours, turn the unit on. The reading on the display is not important at
this point.
12.2 Simultaneously press the “11.3%” and “Calibration” buttons on the
microprocessor board.
13.0 Verify the 11.3% (Zero) calibration
13.1 Do not be concerned if the unit does not display 11.3, since it is not
supposed to match the value of the calibration salt.
13.2 Use the “Sensor Temp” switch on the faceplate of the unit to
determine and record the sensor temperature.
13.3 Look up this temperature on the “Theoretical Dew Point Values for
Calibration Verification” chart located in the back of this manual.
Appendix “A” will show the temperature values in Fahrenheit, and
Appendix B will show the temperature values in Celsius.
13.4 Next to the appropriate temperature, note the number in the
corresponding column titled “11.3%”. This should match with the dew
point that is shown on the display of the DP2000.
14.0 Determine the acceptability of the reading
14.1 The value printed on the chart in Appendix A is a theoretical value,
and some variation can be expected. When a calibration is performed
at SSI, we certify (in writing) that the unit displays within +/- 1 degree
of the theoretical value after it has been calibrated. We would not
consider a calibration to be successful unless it is within +/- 1 degree,
however in the case of a field calibration, this degree of accuracy may
or may not be required. The degree of accuracy that is acceptable is
determined by the policy of the person performing the calibration.
NOTE: Keep in mind that the DP2000 only displays even numbers, and
not tenths of a degree. Therefore, a reading of 18°F (-7.8°C) could be
as low as 17.50°F (-8.06°C) or as high as 18.49°F (-7.51°C).

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 16 of 30
15.0 Allow the sensor to achieve equilibrium at ambient atmosphere
15.1 After the 11.3% (Zero) calibration has been completed, remove the
sensor from the calibration salt and replace the cap.
15.2 Leave the sensor probe in the gland and while the unit is still on,
allow it to achieve equilibrium at the ambient atmosphere in the room.
This should take between two and three minutes. You will know when
this has been accomplished when the numbers on the display begin to
stabilize.
16.0 Re-assemble the unit
16.1 After the calibration process has been completed, remove the
sensor probe from the gland and return it to the sensor-sampling
chamber, taking care to position it properly. The white mark on the
sensor probe should face towards the right of the sensor-sampling
chamber (at 3:00 if it were the face of a clock.). If the white mark is not
visible, then it should be placed so the sample flow directly strikes the
face of the mirror on the sensor tip (the sample flows from right-to-
left). In other words, the mirror should face the incoming gas stream.
16.2 Hand-tighten the black sensor gland to prevent air from leaking out
of the sampling chamber.
16.3 Slide the faceplate into position with the back posts going into the
corresponding holes as the faceplate slides back.
16.4 Verify that the system is leak proof by turning on the pump and
placing a finger over the sample inlet port. The flow meter on the side
of the unit will drop to zero if there are no leaks. If a leak is detected,
make sure that all tubing connections are tight, especially the black
sensor gland.
16.5 After the unit has passed the leak test, re-fasten the screw into the
faceplate and tighten it.
17.0 Make sure that all caps are replaced on the calibration salts, and return
the DP2000 to service.

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 17 of 30
RECHARGING THE MODEL DP2000 –
The internal 12V gel cell battery has enough power to run the analyzer for
approximately 8 hours, although after many charge/discharge cycles some loss
of battery life can be expected. When it is time to recharge the unit, a red
“Battery Low” light will illuminate. When this light comes on, the unit will
operate for an additional one to two hours before the battery is unable to power
the unit and it shuts off.
A full recharge will take 16 hours, however the analyzer can be used while it is
recharging. It can also be left on charge for as long as you want, with no harm to
the battery. A green “Recharging” light will come on to verify that the unit is
charging, and this light will go out automatically when it is completely charged.
For maximum battery life, it is recommended that the battery be discharged
before it is recharged. The battery does not need to be completely discharged
however, there is a correlation between the number of times that a battery is
charged and the life span of the battery. By keeping the number of times the
battery is recharged to a minimum, the battery life can be increased.
To charge the analyzer, plug the power cord into any 110 VAC outlet.
The instrument is not intended for use with 220 VAC power unless it has been
specifically set up accordingly at the factory. If this has been done, the serial
number plate will indicate 220 VAC operation.
TROUBLESHOOTING -
Unit doesn’t power up:
Verify that the battery has been charged. See RECHARGING THE MODEL DP2000
section to make sure that the battery has been charged according to instruction.
If the unit has been charged, but still won’t power up, attempt to power up the
unit with the power cord connected to a 110 VAC outlet. If the unit will power up
only when the power cord is plugged in, the battery may be bad. Please contact
Super Systems, Inc. at 513-772-0060 for more information regarding the
replacement of the battery.
Verify that the unit has power. Verify that the unit has been charged properly.
Connect the power cord and plug it into a 110 VAC outlet. If the unit still will not
power up with the power cord is plugged in, there may be significant damage to

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 18 of 30
the unit. Please contact Super Systems, Inc. at 513-772-0060 for more
information regarding this issue.
Dew point reading shows +80°F (or a high reading) and never drops:
The upper range of the sensor is +80°F (27°C). If that value is displayed, it likely
indicates the presence of moisture in the sample tubing or on the dew point
sensor tip. If this moisture is not removed, it will cause the sensor tip to corrode
and will eventually require the sensor to be replaced.
First, check the dew point sensor tip for obvious signs of moisture, corrosion,
and/or damage causing high readings. Remove power from the DP2000. Make
sure that it is not plugged into a power source and that the POWER switch is off.
Open the case by loosening the Allen head screw located in the bottom of the
face plate (Figure 8). Loosening this screw requires a 1/8” Allen wrench. Lift up
the face plate, but be careful of the wires connecting the face plate to the case
(Figure 9).
Figure 8 Figure 9
Loosen the black plastic compression fitting that holds the dew point sensor
(Figure 10).
Figure 10
Pull out the sensor and visually inspect it for visible moisture. Older units use a
sensor shown in Figures 11 and 12. New units use a sensor shown in Figure 13.

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 19 of 30
The older sensor has a plastic protective end that can be unthreaded and
removed for a more close examination. Removing the protective end exposes a
mirror and wire leads around it. The newer sensor has an aluminum protective
cover that cannot be removed. Look for signs of visible moisture on the sensor
or any other signs of contamination due to prolonged exposure to moisture.
Figure 11 Figure 12 Figure 13
Some examples of sensor damage due to prolonged exposure to excessive
moisture on an older model sensor are shown below (Figures 14 –16).
Figure 14 Figure 15 Figure 16
Water droplets present on the mirror indicates that the sensor has been exposed
to excessive moisture at some point. That moisture needs to be removed. DO
NOT ATTEMPT TO REMOVE THE MOISTURE FROM THE SENSOR WITH A CLOTH
OR COMPRESSED AIR AS THIS CAN PERMANENTLY DAMAGE THE SENSOR.
There are a few ways to remove the moisture from the sampling lines and
DP2000. The method used will depend on how much time the user can afford to
be without measurements from the DP2000 analyzer.
Drying out the Model DP2000:
The easiest way to remove moisture is to flow dry compressed air*, nitrogen, or
argon through the analyzer, until the dew point reading drops and stabilizes.
This can take up to two days depending on the amount of moisture in the
sampling lines and analyzer.
*
Dry compressed air is compressed air produced by a facility air compressor that
has flowed through a desiccant or refrigerated air dryer with a dew point of 40°F
(4°C) or less.

DP2000 Portable Dew Point Analyzer Operations Manual
Super Systems Inc. Page # 20 of 30
The in-line filter is the first place where moisture tends to collect. Upon
examination, there may not be any visible moisture in the filter element, but
there may still be a significant amount of moisture in the element. Remove the
in-line filter and replace it.This will help reduce the time required to dry out the
DP2000.
With the filter element replaced, test the DP2000 to see if it is operating properly.
Verify the ambient dew point against a web-based weather station that will report
the ambient dew point for your area. If the displayed reading is within three
degrees (3°) of the reported dew point when the instrument is taken outside,
then all of the moisture has probably been successfully removed. The wet filter
and sample tubing can be re-attached after they have been completely dried out.
The filter element will regain all of its original filtering properties after it has
dried out.
Option 1:
Connect a dry gas source to the sampling line of the DP2000. Verify that the
pressure is low (less than 2 psi) to prevent damage to the dew point sensor.
Adjust flow through the DP2000 to 1.5 to 2 CFH of flow indicated by the flow
meter on the side of the DP2000. Allow gas to flow through the unit and monitor
the dew point reading from the sensor. As the sample line and sensor dry out,
the dew point reading should drop and stabilize.
Option 2:
Moisture can be removed from the sampling line much more quickly if the lines
are disconnected and blown out with dry compressed air, nitrogen, or argon.
Below are steps to remove moisture with that method.
Isolate the sampling line from the atmosphere gas source to the DP2000 sample
inlet by disconnecting it on both ends. Higher pressures can be used for this
process as long as the lines have been disconnected at both ends and no sensing
equipment or filters are exposed to the high pressure gas. Blow gas through the
sampling line for as long as it takes to remove any visible moisture from the line.
Reconnect the line at both ends.
Lift the face plate of the DP2000 (reference Figure 8 & 9) and disconnect the
tubing between where the sample gas enters the unit and the Sensor Sampling
Block. Blow gas through the sampling line for as long as it takes to remove any
visible moisture from the line. Reconnect the tubing at both ends.
Table of contents
Other Super Systems Measuring Instrument manuals