Suresense A202 User manual

User Manual
INFRARED THERMOMETER
Product: Infrared Thermometer
Product Code: A202
ReOrder Code: BHC-DIGTHER
Please note: This medical device must be used according to instructions to ensure accurate readings
Version: V1.1 (19/6/2020)

Suresense Infrared Thermometer User Manual V1.1 (19/6/2020)
2
Section 1 Safety
1.1 Instructions for the Safe Operation and Use
The following basic safety precautions should always be taken.
1. Close supervision is necessary when the thermometer is used by, on, or near children, handicapped
persons or invalids.
2. Use the thermometer only for the intended use described in this manual.
3. Do not use the thermometer if it is not working properly, or if it has suffered any damage
1.2 Warnings
• As with any thermometer, proper technique is crucial to obtaining accurate temperature readings.
Please read this manual thoroughly and carefully before using.
• Always operate the thermometer in an operating temperature range 5°C to 40°C (41°F to 104°F), and
relative humidity 15 to 93%.
• Always store the thermometer in a cool and dry place -25°C to 70°C (-13°F to 158°F) and relative
humidity 15% to 93%.
• The device requires no calibration.
• The device contains no user serviceable parts.
• The user must check that the equipment functions safely and see that it is in proper working
condition before being use
• The manufacturer does not require such preventive inspections by other persons.
• No modification of this equipment is allowed.
• The device is not suitable for use in the presence of flammable anaesthetic mixtures with air or with
oxygen or nitrous oxide.
• Manufacturer will provide circuit diagrams, component part lists, descriptions, calibration
instructions to assist to SERVICE PERSONNEL in parts repair.
• Do not clean or maintenance the device is in use.
• Avoid direct sunlight.
• Avoid dropping the thermometer. If dropped and believed to be damaged, please contact customer
services immediately.
• Do not touch the lens.
• Do not disassemble the thermometer.
• Basic safety precautions should always be observed, especially when the thermometer is used on or
near children and disabled persons.
• This thermometer is not intended to substitute for a consultation with your physician.
• This thermometer and the subject must remain in a stable environment for at least 30 minutes
before measuring the temperature.
• When the measured temperature falls within the fever temperature range of >37.9°C (100.2°F) and
<42.9°C (109.3°F), as indicated by the red LCD on display, please consult with your physician
immediately.
1.3 Definitions and symbols
Symbol
Description
Symbol
Description
Type BF Equipment
Batch code*
Information of manufacture, including
name and address
Date of manufacture*
Temperature limitation
Serial NO*
When the end-user wishes to discard
this product, it must be sent to separate
collection facilities for recovery and
recycling
Information of EU authorised
representative
Follow user manual
Caution:
The information you should know to
protect the equipment from possible
damage

Suresense Infrared Thermometer User Manual V1.1 (19/6/2020)
3
IP22
Anti-dust & Anti-water class
Note
The important information you should
know
Warning:
The information you should know to protect patients and medical staff from possible injury
*Batch code, Date of manufacturer and Serial No are printed on the label on the battery cover.
Section 2 Introduction
2.1 Restrictions of Use
This thermometer is clinically proven to produce accurate temperature measurements. However, please be
advised that the accuracy cannot be ensured when the thermometer is not clean. Check that the probe is
clean before taking a measurement.
2.2 Intended Use
A202 Infrared Thermometer is intended for the intermittent measurement and monitoring of human body
temperature from forehead measurement at home, clinics and hospital.
A control measurement using a conventional thermometer is recommended in the following cases:
1. If the reading is surprisingly low.
2. For new-born infants up to 100 days old.
3. For children under three years of age who have a weakened immune system or who react unusually
in the presence or absence of fever.
2.3 How does it work
The thermometer measures the infrared heat generated by the surface of the skin over the vessel and its
surrounding tissue. The thermometer then converts it into a temperature value.
NOTE:
The thermometer does not emit any infrared energy.
2.4 Highlighted Features
• Measurement that does not require probe cover, thereby saving cost of replacement.
• Automatically power off if left idle for 10 seconds.
• Memory function allows you to recall previous up to 25 previous results.
• Easy to read LCD with green backlight in a dark environment.
• Colour indication for body readings (green, yellow and red)
2.5 Meter Overview
1. Infrared Sensor
2. LED Light
3. Display Screen
4. ON Button/Measurement /
Memory/Set-up Button
5.Mode Switch
6. Battery Cover

Suresense Infrared Thermometer User Manual V1.1 (19/6/2020)
4
2.6 Display Screen
1. Body mode icon
2. Object mode icon
3. Memory icon
4. Temperature reading
5. Temperature unit
6. Battery icon
2.7 Display Mode
Two modes can be selected.
1.
Body Mode
This mode is used to measure the forehead temperature
2.
Object Mode
This mode is used to measure the object temperature.
2.8 Select the Temperature Unit
This thermometer offers both measurement units, °C or °F
Be sure the thermometer is OFF before selecting the temperature unit.
Long-press the Measurement Button for 8 seconds until the signal °C
or °F displayed on the LCD panel as figure shown. Re-press the
Measurement Button to select °C or °F unit. Then long-press the
Measurement Button for 3 seconds, it will automatically save your
choice and turn off.
2.9 Turn ON the LED Light
This meter provides LED Light to help users placing the thermometer at the correct position.
Be sure the thermometer is OFF before selecting the temperature unit.
1. Long-press the Measurement Button for 10 seconds until Signal --
LED displayed on the LCD panel.
2. Then short press the measurement to turn on or off LED light --
when LED is turned on, the blue light will indicate, while it is
turned off there is no light. Then long-press the Measurement
Button for 3 seconds, it will automatically save your choice and
turn off.

Suresense Infrared Thermometer User Manual V1.1 (19/6/2020)
5
2.10 Replacing the Battery
The thermometer comes with two 1.5 V AAA alkaline batteries. The thermometer will alert you when the
power is getting low, please follow the steps below to replace new batteries immediately.
1. Remove the battery cover as the arrow direction accordingly.
2. Remove the old batteries and replace with two 1.5V AAA size
alkaline batteries. Taking care to match the Positive (+) and
Negative (-) indications
3. Close the battery cover as the arrow direction accordingly.
NOTE:
1 Although the thermometer works when the battery is low, we still recommend that you change the
batteries to obtain an accurate result.
2 Remove the batteries if stored for a long period of time.
3 The batteries should be kept away out of children’s reach.
If they are swallowed, promptly see a doctor for help.
Section 3 Detailed Information
3.1 About Normal Body Temperature & Fever
The temperature in the forehead and temple area differs from the internal temperature, which is taken orally
or rectally. Vasoconstriction, an effect which constricts the blood vessels and cools the skin, can occur during
the early stages of a fever. In this case, the temperature measured by the A202 Infrared thermometer may be
unusually low. If the measurement therefore does not match the patient’s own perception or is unusually
low, repeat the measurement every 15 minutes. As a reference, you can also measure the internal body
temperature using a conventional oral or rectal thermometer. Body temperature can vary from one
individual/person to next. It also varies by location on the body and time of day. Below show the statistical
normal ranges from different sites. Please keep in mind that temperatures measured from different sites, even
at the same time, should not be directly compared. Fever indicates that the body temperature is higher than
normal. This symptom may be caused by infection, overdressing or immunization. Some people may not
experience fever even when they are ill. These include, but are not limited to, infants younger than 3 months
old, persons with compromised immune systems, persons taking antibiotics, steroids or antipyretics (aspirin,
ibuprofen, acetaminophen), or persons with certain chronic illnesses. Please consult your physician when
you feel ill even if you do not have fever.
Table*1 Normal Temperature Range of various body sites
Oral
0.6°C (1°F) or more above or below 37°C (98.6°F)
Rectal/ear
0.3°C to 0.6°C (0.5°F to 1°F) higher than oral temperature
Axillary (Armpit)
0.3°C to 0.6°C (0.5°F to 1°F) lower than oral temperature
3.2 Perform the Test
Using the device
(As a body reading)
Press “ON” Button to turn on the thermometer first.
Push the Mode switch to select Body mode.
The device will make a short beep when turned on and the
temperature unit (°C or °F) flashes to confirms the readiness to
perform the measurement.

Suresense Infrared Thermometer User Manual V1.1 (19/6/2020)
6
Position the thermometer 3cm from the forehead straight on.
The reading is displayed after a second and a long beep. A short beep
follows confirming it’s saved to the memory and it’s ready to take
another reading.
Press the “ON” button to turn off the unit or leave it idle for 10 seconds,
the unit will switch off automatically.
NOTE:
• As the forehead measurement temperature is likely to be affected by sweat, oil and the
surroundings, the reading shall be taken as a reference only.
• If the probe is placed at an angle close to the forehead measurement, the reading will be affected by
surrounding temperature. Babies’ skin reacts very quickly to the ambient temperature. Therefore, do
not take their temperature with the A202 Infrared thermometer during or after breastfeeding,
because the skin temperature may then be lower than the internal body temperature.
• If the measured temperature is < 32°C (89.6°F), the display will show with LO symbol.
• If the reading is >37.9°C(100.2°F) and < 42.9°C (109.3°F), the display will show together with red LCD
and six short beeps.
• The thermometer will automatically turn off if left idle for 10 seconds.
Using the device
(As an object reading)
Press “ON” Button to turn on the thermometer first.
Push the Mode switch to select Object mode.
The machine will make two short beeps indicating that the object
mode is turned on and the temperature unit flashes.
Move the probe close to the object and take measurements.
Make sure the probe is flat and close to the object, not at an angle.
Perform a measurement with a distance within 3 cm. When ready,
push measurement button to take measurements.
Read the result.
The measurement result will be done in 1 second. The reading is
shown together with LCD lighting and one long beep informs about
the temperature measurement, and after a while another, short,
confirming the saving of the result to the memory and readiness for
the next measurement.
Press the “ON” button to turn off the unit, Or leave it idle for 10 seconds,the unit will switch off
automatically.
3.3 Memory
Recalling the memory
This thermometer stores 25 most recent readings
Be sure the thermometer is OFF before recalling this memory.
Press for 2 seconds to enter the memory mode.

Suresense Infrared Thermometer User Manual V1.1 (19/6/2020)
7
Each time you press the Memory button, a result will be displayed in
the order of dates (latest result shown first), together with “MEM” and
number (from 1 to 25).
LCD in green or red will appear according to the memory reading.
When the memory is full, the oldest result is deleted as the new one
added. When the last record displayed in the display, press Memory
Button again to return the first record.
Exit the memory.
Long press ON Button to exit the memory or keep the meter in idle for
10 seconds to switch off automatically
Press the “ON” button to turn off the unit or leave it idle for 10 seconds,the unit will switch off
automatically.
Clear the memory
This thermometer stores 25 most recent readings
Removing one of the batteries, long press the measurement button,
then reload the battery until “---” on the LCD screen.
Section 4 Troubleshooting and Maintenance
4.1 Maintenance
• The probe is not waterproof. Please wipe with a clean and dry cotton swab to clean the probe on
the inside.
• The body of the thermometer is not water-resistant. Never put the thermometer under a running
tap or submerge it into water. Use a soft and dry cloth to clean it. Do not use abrasive cleaners.
• Store the thermometer in a cool and dry location. Free from dust and away from direct sunlight.
4.2 Troubleshooting
Error or Error Symbol
Error Description
Resolutions
No display On the LCD pane.
The battery has run out. Incorrect
battery polarity.
Replace the battery.
Please note:
The (+) side of the battery must
face
Measurement not possible (or an
abnormal value is displayed)
The thermometer is not ready.
Wait until the °C symbol is
displayed.
An abnormal temperature value is
displayed.
The probe tip is dirty or damaged.
Did you hear the beep after
pressing the ON button?
Clean the probe tip or get it
repaired. Wait until you hear the
beep before removing the
thermometer from the ear or
forehead
LO or HI symbol is displayed
The temperature measured is
outside the measuring Range.
LO-temperature<32°C (89.6°F).
HI-temperature ≥42.9°C (109.3°F).
Check that the probe tip is clean
and that the thermometer is
properly placed on the forehead.
Symbol is displayed
The battery has run out.
Replace the batteries
Symbol is displayed
The ambient temperature is
outside the operating
temperature range or is changing
too fast.
To ensure accurate
measurement, let the
thermometer rest at operating
temperature for 30 minutes prior
to use.

Suresense Infrared Thermometer User Manual V1.1 (19/6/2020)
8
Section 5 Specifications
Name:
Suresense Infrared Thermometer
Model:
A202
ReOrder:
BHC-DIGTHER
Battery
2pcs 1.5V AAA Alkaline Battery
Temperature unit
Default °C or Default °F
Display resolution
0.1°C or 0.1°F
Measurement range
Body Mode (Forehead): 32.0°C ~42.9°C (89.6°F
~109.3°F)
Object Mode: 0.0°C ~99.9°C (32°F ~211.8°F)
Accuracy
Body Mode (Forehead):
±0.2°C (±0.4°F): 36.0°C~39.0°C (96.8°F~102.2°F)
±0.3°C (±0.5°F): out of the range
Object Mode:
±1°C (±2°F): 0°C to 60.0°C (32.0°F to 140.0°F)
±4°C (±7.2°F): out of the range
Indicator light range (Body Mode only)
Green: <37.5C (<99.5F)
Yellow: ≥37.5°C - ≤37.9°C (≥99.5°F – ≤100.2°F)
Red: >37.9°C (>100.2°F)
Memory
25 Groups
Auto shut-off
10 seconds after use
Operating Environment
Temperature: +5°C to +40°C (41°F to 104°F)
Humidity: 15% to 93% R.H. (non-condensing)
Air Pressure: 70Kpa-106Kpa
Storage & Transport Environment
Temperature: –25°C to +70°C (-13°F to 158°F)
Humidity: 15% to 93% R.H. (non-condensing)
Dimensions
142.5mm ×38mm ×40mm
Weight:
76g(Net: 54g)
Section 6 Disposal
This product contains batteries and recyclable electronic waste. To protect the environment, do not
dispose of it in the household waste, but take it to appropriate
Section 7 Reference of Standards
7.1 Device Standards
Device Corresponds to the requirements of the standard for infrared thermometers
IEC 60601-1:2012
IEC 60601-1-2:2014
ISO 80601-2-56:2009
7.2 Classification
Anti-electric Shock Type: Internally powered equipment
Applied part: Type BF
Mode of operation: Continuous Operation
EMC: type B class I
7.3 Electromagnetic Compatibility
Device fulfils the stipulations of the standard IEC 60601-1-2
The stipulations of EU-Directive 93/42/EEC for Medical Devices Class Ila have been fulfilled.
* Technical alterations reserved!
Software identify no.: A202 V1.0.0
7.4 Manufacturer’s Declaration of the EMC
EN 60601-1-2:2015 ME EQUIPMENT and ME SYSTEMS identification,
marking and documents for Infrared Thermometer

Suresense Infrared Thermometer User Manual V1.1 (19/6/2020)
9
7.5 Instructions for use
The me equipment or me system is suitable for home healthcare environments and so on.
Warning:
Don’t near active HF surgical equipment and the RF shield room of an ME system for magnetic
resonance imaging, where the intensity of EM disturbances is high.
Warning:
Use of this equipment adjacent to or stacked with other equipment should be avoided because it
could result in improper operation. If such use is necessary, this equipment and the other equipment should
be observed to verify that they are operating normally.
Warning:
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.”
Warning:
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the equipment (A202 Infrared
Thermometer), including cables specified by the manufacturer. Otherwise, degradation of the performance
of this equipment could result.
If any:
a list of all cables and maximum lengths of cables (if applicable), transducers and other ACCESSORIES
that are replaceable by the RESPONSIBLE ORGANIZATION and that are likely to affect compliance of the ME
EQUIPMENT or ME SYSTEM with the requirements of Clause 7 (EMISSIONS) and Clause 8 (IMMUNITY).
ACCESSORIES may be specified either generically (e.g. shielded cable, load impedance) or specifically (e.g. by
MANUFACTURER and EQUIPMENT OR TYPE REFERENCE).
If any:
the performance of the ME EQUIPMENT or ME SYSTEM that was determined to be ESSENTIAL
PERFORMANCE and a description of what the OPERATOR can expect if the ESSENTIAL PERFORMANCE is lost
or degraded due to EM DISTURBANCES (the defined term “ESSENTIAL PERFORMANCE” need not be used).
Technical description
1. All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE with regard
to electromagnetic disturbances for the excepted service life.
2. Guidance and manufacturer’s declaration -electromagnetic emissions and Immunity
Table 1:
Guidance and manufacturer´s declaration – electromagnetic emission
Emissions test
Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations
Flicker emissions
IEC 61000-3-3
Not applicable
Table 2:
Guidance and manufacturer’s declaration - electromagnetic Immunity
Immunity Test
IEC 60601-1-2 Test level
Compliance level
Electrostatic discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst
IEC 61000-4-4
Power supply lines:±2 kV
input/output lines:±1 kV
Not applicable
Surge
IEC 61000-4-5
line(s) to line(s):±1 kV.
line(s) to earth:±2 kV.
100 kHz repetition frequency
Not applicable
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
0% 0.5 cycle
At 0º, 45 º, 90 º, 135 º, 180 º, 225
º, 270 º and 315 º
0% 1 cycle
And
70% 25/30 cycles
Single phase: at 0
0% 300 cycle
Not applicable

Suresense Infrared Thermometer User Manual V1.1 (19/6/2020)
10
Power frequency magnetic field
IEC 61000-4-8
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Conduced RF
IEC61000-4-6
150KHz to 80MHz: 3Vrms
6Vrms (in ISM and amateur radio
bands)
80% Am at 1kHz
Not applicable
Radiated RF
IEC61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE UTis the a.c. mains voltage prior to application of the test level.
Remark In the presence of strong electromagnetic field (Horizontal/Vertical, 80MHz~180MHz and
1.4GHz~2GHz, 10V/m applied on the product), it will have a change from 0.1°C to 0.3°C of the measured
value. In case of malfunction due to these strong electromagnetic fields, user need to wait till resume
normal operation without intervene.
Table 3:
Guidance and manufacturer’s declaration - electromagnetic Immunity
Radiated RF
IEC61000-4-3
(Test
specifications for
ENCLOSURE
PORT
IMMUNITY to
RF wireless
communications
equipment)
Test
Frequency
(MHz)
Band
(MHz)
Service
Modulation
Modulation
(W)
Distance
(m)
IMMUNITY
TEST
LEVEL
(V/m)
385
380 –
390
TETRA
400
Pulse
modulation
18 Hz
1,8
0.3
27
450
380 –
390
GMRS
460,
FRS 460
FM
± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
704 –
787
LTE
Band 13,
17
Pulse
modulation
217 Hz
0,2
0.3
9
745
780
810
800 –
960
GSM
800/900,
TETRA
800,
iDEN
820,
CDMA
850,
LTE
Band 5
Pulse
modulation
18 Hz
2
0.3
28
870
930
1720
1700 -
1990
GSM
1800;
CDMA
1900;
GSM
1900;
DECT;
LTE
Band 1,
3,
4, 25;
UMTS
Pulse
modulation
217 Hz
2
0.3
28
1845
1970

Suresense Infrared Thermometer User Manual V1.1 (19/6/2020)
11
Section 8 Warranty
• This thermometer is guaranteed for 18 months from the date of purchase against any
manufacturing defect, conditional upon normal household use.
• The intended service life of the product is 24 months.
• This product to be free of defects in workmanship and materials for a period on 18 months from the
date of purchase.
• During the warranty period, if this product is found to be defective, you may bring it together with
the purchase receipt and Warranty Certificate on a carry-in basis to manufacturer’s office during
normal business hours for warranty service.
• We will then repair or replace defective parts or exchange the whole product as we select, at no
charge to the original owner. After such repair, replacement or exchange, the product will be
warranted from the date of purchase.
• This warranty is valid only if the Warranty Certificate and Warranty Registration Card are duly
complete with date of purchase, serial number and dealer’s stamp, and if the Warranty Registration
Card is sent to local distributor office not later than 14 days from the date of purchase.
• This warranty is void if this product has been repaired or serviced by unauthorized person. This
warranty does not cover defects caused by misuse, abuse, accident, tampering, poor maintenance,
fire or any other acts beyond human control.
• Except as stated in the above paragraphs, A202 Infrared Thermometer disclaims all other warranties,
implied or expressed, including the warranties of merchantability of fitness for a particular purpose
with respect to the use of this product. Manufacturer shall not be liable for any direct, consequential
or incidental damages arising out of the use or inability to use product.
Section 9 Certificate of Guarantee
Bird Healthcare guarantees the Suresense Infrared Thermometer against any manufacturing defect for two
years from the date of purchase if it is returned to the dealer from whom it was purchased.
During this period the unit will be repaired or replaced free of charge if the fault is due to defective design
or assembly.
This guarantee does not cover any damage or defect caused by improper handling resulting from use that is
not in compliance with these instructions or from unauthorised attempts to repair it. Your local dealer cannot
declare this manufacturer’s guarantee invalid but may expand it by additional guarantees at their discretion
and at their own expense.
Shenzhen Aeon Technology Co., Ltd.
RM6H02, Block 27-29 Tianxia IC Industrial Park,
Majialong No.133 of Yiyuan road, Nantou Street
Nanshan District, Shenzhen, China.
Shanghai International Holding
Corp. GmbH (Europe)
Eiffestrasse 80, 20537 Hamburg, Germany
Distributed by
Bird Healthcare Pty Ltd.
18 Corporate Blvd,
Bayswater VIC Australia 3153
P: 1300 365 561 (+61 03 9720 2493)
F: 1300 949 815 (+61 03 9729 0168)