Swan Analytical Instruments AMI Hydrazine User manual

AMI Hydrazine
Version 5.40 and higher
A-96.250.501 / 280915
Operator’s Manual

© 2015, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved
subject to change without notice
.
Customer Support
SWAN and its representatives maintain a fully trained staff of technical specialists
around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG
Studbachstrasse 13
8340 Hinwil
Switzerland
Internet: www.swan.ch
E-mail: [email protected]
Document Status
Title: Monitor AMI Hydrazine Operator’s Manual
ID: A-96.250.501
Revision Issue
00 June 2007 First Edition
01 April 2015 Update to Rev. 5.40, main board V2.4

AMI Hydrazine
A-96.250.501 / 280915 1
Table of Content
1. Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1. Warning Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2. General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.3. Restrictions for use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2. Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1. Description of the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2. Instrument Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.3. Instrument Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3. Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1. Installation Checklist Monitors . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2. Mounting of Instrument Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.3. Connecting Sample and Waste . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.4. Install the Reference Electrode . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.5. Install the Temperature Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.6. Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.6.1 Connection Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.6.2 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.7. Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.7.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.7.2 Alarm Relay. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.7.3 Relay Contacts 1 and 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.8. Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.8.1 Signal Output 1 and 2 (current outputs) . . . . . . . . . . . . . . . . . . 25
3.9. Interface Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.9.1 Signal Output 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.9.2 Hyperterminal Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.9.3 Profibus, Modbus Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.9.4 USB Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4. Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.1. Install the Diisopropylamine Bottle . . . . . . . . . . . . . . . . . . . . . . . . 28
4.2. Establish Sample Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.3. Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.4. Run in period . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.5. Correction of Hydrazine Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . 30

2A-96.250.501 / 280915
AMI Hydrazine
5. Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.1. Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.2. Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.3. Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.4. Changing Parameters and values. . . . . . . . . . . . . . . . . . . . . . . . . 34
6. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.1. Maintenance Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.2. Stop of Operation for Maintenance. . . . . . . . . . . . . . . . . . . . . . . . 36
6.3. Cleaning of protective filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
6.4. Cleaning of Hydrazine sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.5. Cleaning of Reference Electrode . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.6. Cleaning of Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.7. Changing Diisopropylamine Bottle . . . . . . . . . . . . . . . . . . . . . . . . 44
6.8. Changing the Diffusion Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.9. Calibration of Hydrazine Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6.9.1 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6.9.2 Zero Hydrazine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6.9.3 Process Hydrazine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.10. Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.11. Longer Stop of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7. Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.1. Error List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8. Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
8.1. Messages (Main Menu 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
8.2. Diagnostics (Main Menu 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
8.3. Maintenance (Main Menu 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
8.4. Operation (Main Menu 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
8.5. Installation (Main Menu 5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
9. Program List and Explanations. . . . . . . . . . . . . . . . . . . . . . . . . 61
1 Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
10. Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
11. Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
12. Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82

AMI Hydrazine
Safety Instructions
A-96.250.501 / 280915 3
AMI Hydrazine - Operator’s Manual
This document describes the main steps for instrument setup, oper-
ation and maintenance.
1. Safety Instructions
General The instructions included in this section explain the potential risks
associated with instrument operation and provide important safety
practices designed to minimize these risks.
If you carefully follow the information contained in this section, you
can protect yourself from hazards and create a safer work environ-
ment.
More safety instructions are given throughout this manual, at the
respective locations where observation is most important.
Strictly follow all safety instructions in this publication.
Target
audience
Operator: Qualified person who uses the equipment for its intended
purpose.
Instrument operation requires thorough knowledge of applications,
instrument functions and software program as well as all applicable
safety rules and regulations.
OM Location The AMI Operator’s Manual shall be kept in proximity of the instru-
ment.
Qualification,
Training
To be qualified for instrument installation and operation, you must:
read and understand the instructions in this manual as well as
the Material Safety Data Sheets.
know the relevant safety rules and regulations.

4A-96.250.501 / 280915
AMI Hydrazine
Safety Instructions
1.1. Warning Notices
The symbols used for safety-related notices have the following sig-
nificance:
DANGER
Your life or physical wellbeing are in serious danger if such
warnings are ignored.
Follow the prevention instructions carefully.
WARNING
Severe injuries or damage to the equipment can occur if such
warnings are ignored.
Follow the prevention instructions carefully.
CAUTION
Damage to the equipment, minor injury, malfunctions or incor-
rect process can be the consequence if such warnings are ig-
nored.
Follow the prevention instructions carefully.
Mandatory
Signs
The importance of the mandatory signs in this manual.
Safety goggles
Safety gloves

AMI Hydrazine
Safety Instructions
A-96.250.501 / 280915 5
Warning Signs The importance of the warning signs in this manual.
Electrical shock hazard
Corrosive
Harmful to health
Flammable
Warning general
Attention general

6A-96.250.501 / 280915
AMI Hydrazine
Safety Instructions
1.2. General Safety Regulations
Legal
Requirements
The user is responsible for proper system operation.
All precautions must be followed to ensure safe operation of the in-
strument.
Spare Parts
and
Disposables
Use only official SWAN spare parts and disposables. If other parts
are used during the normal warranty period, the manufacturer’s
warranty is voided.
Modifications Modifications and instrument upgrades shall only be carried out by
an authorized Service Technician. SWAN will not accept responsi-
bility for any claim resulting from unauthorized modification or alter-
ation.
WARNING
Electrical Shock Hazard
If proper operation is no longer possible, the instrument must be
disconnected from all power lines, and measures must be taken
to prevent inadvertent operation.
To prevent from electrical shock, always make sure that the
ground wire is connected.
Service shall be performed by authorized personnel only.
Whenever electronic service is required, disconnect instru-
ment power and power of devices connected to.
– relay 1,
– relay 2,
– alarm relay
WARNING
For safe instrument installation and operation you must read
and understand the instructions in this manual.
WARNING
Only SWAN trained and authorized personnel shall perform the
tasks described in this document.

AMI Hydrazine
Safety Instructions
A-96.250.501 / 280915 7
1.3. Restrictions for use
Alcalizing
Reagent
Correct the pH of the sample with Diisopropylamine only, however,
Diisopropylamine is not included in delivery. Please buy it from your
local supplier: E.g. VWR 6x 1l: 8.03646.1000, or SIGMA 38290-1L-F.
Sample
Require-
ments
No sand (or other polishing material) or oil are allowed in the sam-
ple.
The instrument is specified for a max. pressure of 2 bar (28 psi). If
the sample pressure exceeds 2 bar, please mount a pressure re-
duction in front of the instrument inlet.
Phosphates may interfere with the hydrazine measurement.
The pH value of the sample must be equal or higher than pH 7.
Flow
Interruption
High concentrations of Diisopropylamine reagent may attack the
acrylic material of the flow cell. This can cause cracks and render
the flow cell completely unusable. This will not happen during nor-
mal operation. However, if the sample flow is stopped, undiluted di-
isoproylamine will diffuse through the diffusion tube and can cause
damage. Therefore, the following rules should be observed if the
sample flow is stopped for more than one day:
Remove diisopropylamine bottle and close it. Keep it in a safe
place for further use.
WARNING
Diisopropylamine is corrosive.
Read the Safety Data Sheets (SDS) first.
Wear suitable protective clothing, gloves and eye/face protec-
tion.
Avoid inhalation of DIPA vapor.
In case of contact with eyes, rinse immediately with plenty of
water eyelid wide open for at least 10 min, summon medical
advice. In case of accident or if you feel unwell, summon
medical advice immediately (show the label where possible).
Screw an empty reagent bottle on the holder.
Let the sample flow for another 30 minutes to rinse out the Di-
isopropylamine dissolved in the diffusion tubing material.
Stop sample flow.

8A-96.250.501 / 280915
AMI Hydrazine
Product Description
2. Product Description
2.1. Description of the System
Microprocessor controlled system for the determination and control
of hydrazine or carbohydrazide used as boiler feedwater oxygen
scavengers.
Application Hydrazine is used in power plants as an oxygen scavenger. It helps
to remove oxygen, which may cause corrosion in the power plant.
The reaction of hydrazine with oxygen produces nitrogen and wa-
ter.
Signal
Outputs
Two signal outputs programmable for measured values (freely scal-
able, linear, bilinear, log) or as continuous control output (control
parameters programmable).
Current loop: 0/4–20 mA
Maximal burden: 510
Third signal output with the same specifications as option.
(Only possible if no communication interface is used.)
Relays Two potential-free contacts programmable as limit switches for
measuring values, controllers or timer for system cleaning with au-
tomatic hold function. Both contacts can be used as normally open
or normally closed.
Maximum load: 1 A/250 VAC
Alarm Relay One potential free contact.
Alternatively:
Open during normal operation, closed on error and loss of
power.
Closed during normal operation, open on error and loss of
power.
Summary alarm indication for programmable alarm values and in-
strument faults.
Input One input for potential-free contact to freeze the measuring value
or to interrupt control in automated installations. Programmable as
HOLD or OFF function.
Communica-
tion interface
(optional)
HyperTerminal Interface (RS232) for logger download.
USB Interface for logger download.
Profibus/Modbus interface (RS485) with Fieldbus protocol.

AMI Hydrazine
Product Description
A-96.250.501 / 280915 9
Safety
Features
No data loss after power failure. All data is saved in non-volatile
memory.
Over voltage protection of in- and outputs.
Galvanic separation of measuring inputs and signal outputs. The
analyzer is factory tested and ready for installation and operation.
Reagent
consumption
Reagent consumption (at 25°C):
< 1L diisopropylamine per month
Measuring
principle
3-electrode amperometry:
The sensor consists of two platinum electrodes and a reference
electrode. A voltage is applied to the platinum electrodes and kept
on a optimum level by the reference electrode. The hydrazine in the
sample generates a small current on the platine pin electrode,
which is proportional to the hydrazine concentration.
For an optimal sensitivity and stability, a rotor continuously cleans
the surfaces of the platinum electrodes (hydrodynamic cleaning).
To ensure a sufficient sample flow, a Hall-Sensor detects the rota-
tions of the rotor. If the rotation speed falls below a certain value, a
flow Alarm is triggered.
The signal of amperometric systems depends on flow rate. The
constant head guarantees a constant flow if the sample always
overflows into the longer constant head tube.
Temperature compensation is done automatically.
Fluidics The measurement of hydrazine requires a high pH of the sample.
The AMI Hydrazine buffers the pH to pH>10.5 by adding diisopro-
pylamine. The diisopropylamine is dissolved in the water while
flowing through a diffusion tube.
The sample enters at the sample inlet [E], passes the filter [G] and
the flow regulating valve [D], where the sample flow can be adjust-
ed. From there it flows through the diffusion tube [L] in the diisopro-
pylamine bottle [K], picking up diisopropylamine and reaching
pH>10.5. Afterwards, the sample is led directly into the shorter
overflow tube [C]. From there it flows through the hydrazine sensor
[F] and leaves the flow cell via sample outlet [H]. Excess sample
overflows from the shorter overflow tube into the constant head [A],
fills it until the sample overflows into the longer overflow tube. Sam-
ple must always overflow into the longer overflow tube to ensure
constant flow and pressure at the hydrazine sensor. The sample
leaves the flow cell via sample outlet [H].

10 A-96.250.501 / 280915
AMI Hydrazine
Product Description
A
B
C
D
E
F
Constant head
Overflow tube to waste (long)
Overflow tube to sensor (short)
Flow regulating valve
Sample inlet
Hydrazine sensor
G
H
I
J
K
L
Filter vessel
Sample outlet
Garb sample outlet
Grab sample valve
Diisopropylamine
Diffusions tube
A
B
C
D
F
E
G
H
I
J
K
L

AMI Hydrazine
Product Description
A-96.250.501 / 280915 11
2.2. Instrument Specification
Power Supply Voltage:
Power consumption:
100–240 VAC (± 10%)
50/60 Hz (± 5%)
or 24 VDC (± 10%)
max. 30 VA
Electronics Aluminium with a protection degree of IP 66 / NEMA 4X
housing Ambient temperature:
Limit range of operation:
Storage and transport:
Humidity:
Display:
-10 to +50 °C
-25 to +65 °C
-30 to +85 °C
10–90% rel., non condensing
backlit LCD, 75 x 45 mm
Measuring
range
Hydrazine
Range:
Accuracy:
Stability:
Response time:
0.1–600 ppb
5% of reading up to 200 ppb
± 15% up to 600 ppb
or ± 2 ppb (whichever is greater).
±5% of reading per month, or ±2ppb per
month
90% of change in 60 sec after the sample
entered the flow cell.
Measuring
range
temperature
Measuring range:
Resolution:
up to 60°C
0.1 °C
Sample
requirements
Flow rate:
Temperature:
Sample inlet pressure:
Sample outlet pressure:
pH value:
approx. 15 l/h
15 to 45 °C
0.15–2 bar
pressure free
equal or higher than pH 7.0
On-site The analyzer site must permit connections to:
requirements Sample inlet:
Sample outlet:
Serto 4x6 mm
Tube adapter 15x20 mm
Max. Altitude: 2000 m above sea level

12 A-96.250.501 / 280915
AMI Hydrazine
Product Description
Dimensions Panel:
Mounting hole distance
Screws:
Weight:
280x850x180 mm, stainless steel
254x824
8 mm diameter
12.0 kg
AMI Hydrazine
850 mm / 33½”
13 mm / ½”
4 x dia. 8 mm / ¼”
254 mm/ 10”
280 mm/ 11”
824 mm / 32
7/16
”

AMI Hydrazine
Product Description
A-96.250.501 / 280915 13
2.3. Instrument Overview
A
B
C
D
E
F
G
H
I
J
Panel
Transmitter
Bottle holder
Diisoproylamine bottle
Diffusion tube
Reference electrode
Grab sample valve
Grab sample outlet
Flow regulating valve
Hydrazine sensor
K
L
M
N
O
P
Q
R
S
T
Filter vessel
Sample outlet
Sample inlet
Hall sensor (not visible)
Flow cell block
Constant head
Overflow tube to sensor (short)
Overflow tube to waste (long)
Constant head cover
Temperature sensor
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
P
Q
R
S
T

14 A-96.250.501 / 280915
AMI Hydrazine
Installation
3. Installation
3.1. Installation Checklist Monitors
Check Instrument’s specification must conform to the National Electri-
cal Code, all state and local codes, and all plant codes and stan-
dards for electrical equipment.
On site require-
ments
100–240 VAC (10%), 50/60 Hz (5%) or 24 VDC (±10%),
isolated power outlet with ground connection and 30 VA.
Sample line with sufficient sample flow and pressure (see Instru-
ment Specification, p. 11).
Installation Mounting of Instrument Panel, p. 15.
Connecting Sample and Waste, p. 15.
Electrodes Install the Reference Electrode, p. 16.
Install the Temperature Sensor, p. 17.
Diisopropyl-
amine
Install the Diisopropylamine Bottle, p. 28.
Electrical wir-
ing
Note: Do not switch on the Instrument until all electrical
connections are made.
Connect all external devices like limit switches, current loops
and pumps, see Connection Diagram, p. 20.
Connect the power cord, see Power Supply, p. 21
Power-up Establish Sample Flow, p. 29.
Switch on power.
Adjust sample flow according to flow cell specifications.
Instrument
set-up
Program all necessary parameters see Programming, p. 29.
Program all parameters for external devices (interface,
recorders, etc.).
Program all parameters for instrument operation (limits,
alarms).
Run-in period Let the instrument run-in for approximately 30 min continuous
operation. See Run in period, p. 29.

AMI Hydrazine
Installation
A-96.250.501 / 280915 15
3.2. Mounting of Instrument Panel
The first part of this chapter describes the preparing and placing of
the system for use.
The instrument must only be installed by trained personnel.
Mount the instrument in vertical position.
For ease of operation mount it so that the display is at eye
level.
For the installation a kit containing the following installation
material is available:
– 4 Screws 8x60 mm
– 4 Dowels
– 4 Washers 8.4/24 mm
Mounting re-
quirements
The instrument is only intended for indoor installation.
Instrument Specification, p. 11
3.3. Connecting Sample and Waste
Sample inlet Slide the knurled nut [C] and the compression ferrule [B] over the
4x6 plastic tube [E]. Push it into the serto fitting [A] and tighten the
knurled nut.
Waste Connect the 1/2” tube [F] to the hose nozzle [D] and place it into a
pressure free drain.
A
B
C
D
E
F
Serto fitting at sample
inlet
Compression ferrule
Knurled nut
Hose nozzle at sample
outlet
Plastic tube 4x6
1/2” tube
A
D
E
B
C
F

16 A-96.250.501 / 280915
AMI Hydrazine
Installation
3.4. Install the Reference Electrode
The reference electrode is delivered separately and protected with
a waterfilled protective cap. The connector is fixed to the panel with
an adhesive tape and already connected to the front end PCB in
the AMI transmitter.
To install the reference electrode proceed as follows:
1Loosen the union nut [D].
2Remove the protective cap [B] from the reference electrode [C].
3Push the reference electrode through the union nut [D] into the
bore of the flow cell block [E] as far as it will go.
4Tighten the union nut.
5Remove the connector [A] from the panel and screw it onto the
reference electrode.
A
B
C
Connector
Protective cap
Reference sensor
D
E
Union nut
Flow cell block
C
B
D
E
A

AMI Hydrazine
Installation
A-96.250.501 / 280915 17
3.5. Install the Temperature Sensor
The temperature sensor is fixed to the panel with an adhesive tape
and already connected to the front end PCB in the AMI transmitter.
To install the temperature sensor proceed as follows:
1Remove the temperature sensor [A] from the panel.
2Put the temperature sensor in the designated hole of the con-
stant head cover [B].
3Push it into the hole as far as it will go.
A
B
C
Temperature sensor
Constant head cover
Constant head
A
B
C

18 A-96.250.501 / 280915
AMI Hydrazine
Installation
3.6. Electrical Connections
WARNING
Risk of electrical shock.
Do not perform any work on electrical components if the trans-
mitter is switched on. Failure to follow safety instructions could
result in serious injury or death.
Always turn off AC power before manipulating electric parts.
Grounding requirements: Only operate the instrument from
an power outlet which has a ground connection.
Make sure the power specification of the instrument corre-
sponds to the power on site.
Cable
thicknesses
In order to comply with IP66, use the following cable thicknesses
Note: Protect unused cable glands
Wire For Power and Relays: Use max. 1.5 mm2/ AWG 14
stranded wire with end sleeves.
For Signal Outputs and Input: Use 0.25 mm2/ AWG 23
stranded wire with end sleeves.
A
B
C
PG 11 cable gland: cable Øouter 5–10 mm
PG 7 cable gland: cable Øouter 3–6.5 mm
PG 9 cable gland: cable Øouter 4–8 mm
ABC
Table of contents