
ANALOX O2 PORTABLE OXYGEN ANALYSER
Operators Manual MO2-807-02 Januar 2009 Page 5
3.0 Air Calibration
3.1 Air calibration is essential before ever use and is performed as
follows.
3.2 Ensure that an flow adaptors are removed and the reading on the
displa has stabilised.
3.3 Expose the anal ser to clean air for two minutes and adjust the
calibration knob until the displa reads the correct value using the
humidit chart ( ou can find the humidit chart on the inside of the
back cover ). If this is not possible refer to paragraph 3.4 or to
section 8.0 Maintenance.
3.4 It is possible that at ver high altitude normal calibration is not
achievable. In this event ou must ascertain the actual pressure in
BAR and multipl the atmospheric ox gen percent (20.9%) b this
pressure and set the reading during calibration to the calculated
level (this is the surface equivalent ox gen percentage). When
ou measure the level of ox gen in the sample ou must divide
the reading b the same atmospheric pressure value to obtain the
true percentage of Ox gen in our sample.
For Example: At an atmospheric pressure of 0.8 BAR the surface
equivalent ox gen percentage is 20.9% x 0.8 = 16.7%O2 Surface
Equivalent. If the reading ou then obtain from our sample is
32.0% ou must divide this b 0.8 to obtain the true Ox gen
percentage, 32.0/0.8 = 40.0%O2 True Percentage.
3.5
The anal ser is now read for ox gen measurement.
WARNING
The analyser is sensitive to Oxygen partial pressure.
Calibration ust always be carried out at the sa e
at ospheric pressure as Oxygen easure ent.