Swan Analytical Instruments AMI User manual

AMI Sample Sequencer
Version 5.00 and higher
A-96.250.661 / 120418
Operator’s Manual

© 2017, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved
subject to change without notice.
Customer Support
SWAN and its representatives maintain a fully trained staff of technical specialists
around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG
Studbachstrasse 13
8340 Hinwil
Switzerland
Internet: www.swan.ch
E-mail: [email protected]
Document Status
Title: Monitor AMI Sample Sequencer Operator’s Manual
ID: A-96.250.661
Revision Issue
00 Sept. 2011 First release
01 July 2012 Extended with connection to AMI Silica
02 Feb. 2013 Extended with connection to AMI Phosphate II
03 March 2017 Extended with connection to AMI Silitrace

AMI Sample Sequencer
A-96.250.661 / 120418 1
Table of Contents
1. Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1. Warning Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2. General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2. Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1. Measuring Cycle at Batch Measurement . . . . . . . . . . . . . . . . . . . 9
2.1.1 AMI Silica and AMI Phosphate HL . . . . . . . . . . . . . . . . . . . . . . 9
2.2. Measuring Cycle at Continuous Measurement. . . . . . . . . . . . . . . 12
2.2.1 AMI Sodium A and AMI Sodium P . . . . . . . . . . . . . . . . . . . . . . 12
2.2.2 AMI Silitrace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.3. Instrument Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
2.4. Instrument Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3. Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.1. Installation Check List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.2. Install the AMI Sample Sequencer . . . . . . . . . . . . . . . . . . . . . . . . 24
3.3. Connect Sample and Waste . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.4. Connect the Instrument (Mode <AMI>) . . . . . . . . . . . . . . . . . . . . 31
3.4.1 Connect the I2C Bus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.4.2 Sample Sequencer Inputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.4.3 Sample Sequencer Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.4.4 Firmware Settings (Mode <AMI>) . . . . . . . . . . . . . . . . . . . . . . . 35
3.5. Connect the Instrument (Mode <Internal>). . . . . . . . . . . . . . . . . . 36
3.5.1 Connect the Two-Wire Signal Cable . . . . . . . . . . . . . . . . . . . . . 36
3.5.2 Sample Sequencer Inputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
3.5.3 Sample Sequencer Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
3.5.4 Firmware Settings (Mode <Internal>) . . . . . . . . . . . . . . . . . . . . 40
3.6. Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
3.7. Connection Diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
3.8. Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
3.9. Communication, Error Handling . . . . . . . . . . . . . . . . . . . . . . . . . . 44
3.9.1 Mode <AMI> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
3.9.2 Mode <Internal> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4. Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
4.1. Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

2A-96.250.661 / 120418
AMI Sample Sequencer
5. Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.1. Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.2. Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.3. Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
5.4. Changing Parameters or Values. . . . . . . . . . . . . . . . . . . . . . . . . . 49
6. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
6.1. Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
6.2. Stop of Operation for Maintenance. . . . . . . . . . . . . . . . . . . . . . . . 50
6.3. Tube connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.4. Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
6.5. Longer Stop of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
7. Error List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8. Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.1. Messages (Main Menu 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.2. Diagnostics (Main Menu 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
8.3. Maintenance (Main Menu 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
8.4. Operation (Main Menu 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
8.5. Installation (Main Menu 5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
9. Program List and Explanations. . . . . . . . . . . . . . . . . . . . . . . . . 58
1 Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
10. Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
11. Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
12. Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

AMI Sample Sequencer
Safety Instructions
A-96.250.661 / 120418 3
AMI Sample Sequencer - Operator’s
Manual
This document describes the main steps for instrument setup, oper-
ation and maintenance.
1. Safety Instructions
General The instructions included in this section explain the potential risks
associated with instrument operation and provide important safety
practices designed to minimize these risks.
If you carefully follow the information contained in this section, you
can protect yourself from hazards and create a safer work environ-
ment.
More safety instructions are given throughout this manual, at the
respective locations where observation is most important.
Strictly follow all safety instructions in this publication.
Target
audience
Operator: Qualified person who uses the equipment
for its intended purpose.
Instrument operation requires thorough knowledge of applications,
instrument functions and software program as well as all applicable
safety rules and regulations.
OM Location The AMI Operator’s Manual shall be kept in proximity of the instru-
ment.
Qualification,
Training
To be qualified for instrument installation and operation, you must:
read and understand the instructions in this manual as well as
the Material Safety Data Sheets.
know the relevant safety rules and regulations.

4A-96.250.661 / 120418
AMI Sample Sequencer
Safety Instructions
1.1. Warning Notices
The symbols used for safety-related notices have the following sig-
nificance:
DANGER
Severe injuries or death will result if such warnings are ignored.
Follow the prevention instructions carefully.
WARNING
Severe injuries or damage to the equipment can occur if such
warnings are ignored.
Follow the prevention instructions carefully.
CAUTION
Damage to the equipment, minor injury, malfunctions or incor-
rect process can be the consequence if such warnings are ig-
nored.
Follow the prevention instructions carefully.
Mandatory
Signs
The importance of the mandatory signs in this manual.
Safety goggles
Safety gloves

AMI Sample Sequencer
Safety Instructions
A-96.250.661 / 120418 5
Warning Signs The importance of the warning signs in this manual.
Electrical shock hazard
Corrosive
Harmful to health
Flammable
Warning general
Attention general

6A-96.250.661 / 120418
AMI Sample Sequencer
Safety Instructions
1.2. General Safety Regulations
Legal
Requirements
The user is responsible for proper system operation.
All precautions must be followed to ensure safe operation
of the instrument.
Spare Parts
and
Disposables
Use only official SWAN spare parts and disposables. If other parts
are used during the normal warranty period, the manufacturer’s
warranty is voided.
Modifications Modifications and instrument upgrades shall only be carried out by
an authorized Service Technician. SWAN will not accept responsi-
bility for any claim resulting from unauthorized modification or alter-
ation.
WARNING
Electrical Shock Hazard
If proper operation is no longer possible, the instrument must be
disconnected from all power lines, and measures must be taken
to prevent inadvertent operation.
To prevent from electrical shock, always make sure that the
ground wire is connected.
Service shall be performed by authorized personnel only.
Whenever electronic service is required, disconnect instru-
ment power and power of devices connected to.
–relay 1,
–relay 2,
– alarm relay
WARNING
For safe instrument installation and operation you must read
and understand the instructions in this manual.
WARNING
Only SWAN trained and authorized personnel shall perform the
tasks described in this document.

AMI Sample Sequencer
Product Description
A-96.250.661 / 120418 7
2. Product Description
Application
Range
The AMI Sample Sequencer is a complete system for the automat-
ic, continuous multiplexing of up to six sample streams to one pro-
cess analyzer. The AMI Sample Sequencer can be connected to
an:
AMI Phosphate HL
AMI Silica
AMI Silitrace
AMI Sodium A
AMI Sodium P
Operating
Modes
The AMI Sample Sequencer provides the following two operating
modes
AMI
Set the AMI Sample Sequencer to the mode <AMI> if it is con-
nected to an AMI Phosphate HL, AMI Silica or AMI Silitrace. In
this operating mode the AMI Sample Sequencer works as
slave, either of the analyzer or of an external PLC. Communi-
cation between the instruments takes place via I
2
C interface.
For further information see
Connect the I2C Bus, p. 31
.
Internal
Set the AMI Sample Sequencer to the mode <Internal> if it is
connected to an AMI Sodium A or AMI Sodium P. In this oper-
ating mode the AMI Sample Sequencer works as master. It
switches the sample streams according to the programmed
cycle time. There is no communication between the instru-
ments, only measuring values are transferred via signal out-
puts 1 and 2 of the analyzer to the signal inputs 1 and 2 of the
AMI Sample Sequencer. For further information see
Sample Sequencer Inputs, p. 37
.
Signal
Outputs
6 current outputs to transfer the measuring values of the AMI trans-
mitter to a PLC device.
Relay 6 signal outputs for indication of the 6-way valve position.
1 signal output for flow alarm.
Inputs 6 signal inputs to override programmed measuring sequences, by
activating or deactivating an input. The functionality depends on the
settings of the instruments.

8A-96.250.661 / 120418
AMI Sample Sequencer
Product Description
Safety
Features
No data loss after power failure. All data is saved in non-volatile
memory. Over voltage protection of in- and outputs. Galvanic sepa-
ration of measuring inputs and signal outputs.
Fluidics The sample enters the flow cell blocks [C, D, E] via the sample in-
lets [I]. Each sample inlet is equipped with a flow regulating valve
[G]. The sample flows through the active sample line via 6-way
valve [A] and flow meter [F] to the instrument. Excessive sample
(e.g. inactive sample lines) flows via overflow [H] to the atmospher-
ic drain [J].
A
B
C
D
6-way valve
Sample outlet
Flow cell block with sample
inlet 1 and 2
Flow cell block with sample
inlet 3 and 4
E
F
G
H
I
J
Flow cell block with sample
inlet 5 and 6
Flow meter
Flow regulating valve
Overflow
Sample inlets 1–6
Atmospheric drain
1
IIIIII
34
5
7
6
2
A
B
C
G
HHH
JJJ
GG GG G
DEF

AMI Sample Sequencer
Product Description
A-96.250.661 / 120418 9
2.1. Measuring Cycle at Batch Measurement
2.1.1 AMI Silica and AMI Phosphate HL
Example with two sample streams:
A measuring cycle always starts when the solenoid valve [C] clos-
es.
NOTICE:
• The different squares, black with/without underline and white
with/without underline are displayed on the display of the AMI
Sample Sequencer, for better understanding they are
integrated in the illustrations below.
• The meaning of the different squares is explained in
chapter 5, Operation.
Step 1 The solenoid valve [C] is open.
The 6-way valve [A] is on position 1 and the sample stream 1
flows via constant head [B] through the photometer [D]. No
measurement is performed at this time.
A
B6-way valve
Constant head
C
DSolenoid valve
Photometer
1
123456
34
5
7
6
2
A
B
C
D

10 A-96.250.661 / 120418
AMI Sample Sequencer
Product Description
Step 2 The solenoid valve [C] closes:
at the same time the 6-way valve [A] switches to sample stream
2, a new measuring cycle starts.
the sample of sample stream 1 remains in the photometer where
reagents are added and a measurement is performed.
During the measuring of sample stream 1, the tube [E] and the
constant head [B] are flushed with sample stream 2.
A
B
C
6-way valve
Constant head
Solenoid valve
D
EPhotometer
Sample inlet constant head
1
123456
34
5
7
6
2
A
B
C
D
E

AMI Sample Sequencer
Product Description
A-96.250.661 / 120418 11
Step 3 The measurement of sample 1 is finished.
The solenoid valve [C] is open and the photometer is flushed with
sample stream 2.
A valid measuring value of sample stream 1 is available.
A
B
C
6-way valve
Constant head
Solenoid valve
D
EPhotometer
Sample inlet constant head
1
123456
34
5
7
6
2
A
B
C
D
E

12 A-96.250.661 / 120418
AMI Sample Sequencer
Product Description
2.2. Measuring Cycle at Continuous Measurement
2.2.1 AMI Sodium A and AMI Sodium P
A new measuring cycle always starts when the 6-way valve [A] is
switched to the next programmed sample stream.
NOTICE: The following parameters are only available if the AMI
Sample Sequencer is set to the mode <Internal>.
Cycle time The time of switch over between the sample streams is defined as
-cycle time-. The cycle time includes the “Flush Time” and the mea-
suring time of the active sample line.
The measuring time can not be set, but can be calculated as fol-
lows:
Measuring time = “Cycle Time” - “Flush Time” + “Valid After”
The -cycle time- can be set individually for each sample line in 5
min steps from 5 min to 600 min. A sample line is deactivated if the
cycle time is set to 0 min.
Flush Time The parameter -Flush Time- defines how long the measuring cell of
the AMI Sodium A or an AMI Sodium P is flushed with the active
sample before the measurement starts. (see 5.1 Sequence, p. 60)
Valid After The parameter -Valid After- defines the remaining measuring time
of the sample stream x after the 6-way valve has switched to the
next sample stream (see 5.1 Sequence, p. 60).
A
B
C
D
E
Sample stream 1 on hold
Sample stream 2 on hold
Sample stream 1, high conc.
Sample stream 2, low conc.
Measuring signal
F
G
H
I
Cycle time
Flush time
Measuring time
Valid after
ppb
Time
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
G
I
G
IH
C
DE
F
HI GHI
A
B

AMI Sample Sequencer
Product Description
A-96.250.661 / 120418 13
Example with two sample streams:
NOTICE:
• The different squares, black with/without underline and white
with/without underline are displayed on the display of the AMI
Sample Sequencer, for better understanding they are
integrated in the illustrations below.
• The meaning of the different squares is explained in
chapter 5, Operation.
Step 1 The 6-way valve [A] is on position 1 and sample stream 1 [D]
flows via constant head [B] through the measuring cell [C] where
the sodium concentration is continuously measured.
A
B6-way valve
Constant head
C
DMeasuring cell
Sample 1
1
123456
34
5
7
6
2
A
B
C
D

14 A-96.250.661 / 120418
AMI Sample Sequencer
Product Description
Step 2 The 6-way valve [A] has switched to sample stream 2
– The -Valid after- time [I] is active. The instrument still
measures channel 1.
– The -Flush time- [G] is active. No valid measuring value is
available for channel 2 yet.
A
B
C
6-way valve
Constant head
Measuring cell
D
ESample 1
Mixing zone Sample 1 and 2
1
123456
34
5
7
6
2
B
A
E
D
C

AMI Sample Sequencer
Product Description
A-96.250.661 / 120418 15
Step 3 The -Valid after- time is finished:
– Sample stream 2 is selected and measured.
A
B6-way valve
Constant head
C
DMeasuring cell
Sample 2
1
123456
34
5
7
6
2
B
A
D
C

16 A-96.250.661 / 120418
AMI Sample Sequencer
Product Description
2.2.2 AMI Silitrace
The switch to the next channel is initiated by the AMI analyzer ac-
cording to the programmed switching time.
Example with two sample streams, switching time set to 30 min:
NOTICE:
• The different squares, black with/without underline and white
with/without underline are displayed on the display of the AMI
Sample Sequencer, for better understanding they are
integrated in the illustrations below.
• The meaning of the different squares is explained in
chapter 5, Operation.
Step 1 The 6-way valve [B] is on position 1 and sample stream 1 flows
through the flow cell block [D]. From there, it is pumped to the
photometer by the peristaltic pump [C].
A
BPhotometer
6-way valve
C
DPeristaltic pump
Flow cell block
1
12345
34
5
7
6
2
B
C
A
D

AMI Sample Sequencer
Product Description
A-96.250.661 / 120418 17
Step 2a The 6-way valve [B] has switched to sample stream 2:
– The instrument still measures sample 1
– Channel 2 is selected but not yet measured
A
B
C
Photometer
6-way valve
Peristaltic pump
D
ESample 1
Mixing zone Sample 1 and 2
1
123456
34
5
7
6
2
B
C
A
E
D

18 A-96.250.661 / 120418
AMI Sample Sequencer
Product Description
Step 2b The mixing zone [E] flows through the photometer [A]
– No valid measuring value is available.
A
B
C
Photometer
6-way valve
Peristaltic pump
D
ESample 1
Mixing zone Sample 1 and 2
1
123456
34
5
7
6
2
B
C
A
E
D
This manual suits for next models
1
Table of contents
Other Swan Analytical Instruments Recording Equipment manuals