Synthes Colibri II User manual

Universal and powerful battery system for a variety of applications
Colibri II
Instructions for Use


Colibri II Instructions for Use DePuy Synthes 1
Table of Contents
Introduction General Information 3
Colibri II Handpiece 7
Use 9
Attachments General Information 14
Drill Attachments 16
Screw Attachments 17
Ream Attachments 18
Other Rotating Attachments 20
Saw Attachments 25
Other Attachments 29
Care and Maintenance General Information 30
Cleaning and Disinfection 31
• Preparation Prior to Reprocessing 31
• Manual Cleaning Instructions 32
• Automated Cleaning Instructions
with Manual Pre-cleaning 36
Maintenance and Lubrication 39
Function Control 43
Packaging, Sterilization and Storage 44
Repairs and Technical Service 46
Disposal 47
Troubleshooting 48

2 DePuy Synthes Colibri II Instructions for Use
Table of Contents
System Specifications 50
Electromagnetic Compatibility 56
Additional Information 60
Ordering Information 61

Colibri II Instructions for Use DePuy Synthes 3
Efciently working cutting tools are the basis for success-
ful surgery. Therefore, it is mandatory to check used cut-
ting tools after every use for wear and/ or damage and to
replace them if necessary. We recommend using new Syn-
thes cutting tools for every surgery. Cutting tools must be
cooled with irrigation liquid to prevent heat necrosis.
The user of the product is responsible for proper use of
the equipment during surgery.
If the Colibri II is used in conjunction with an implant
system, make sure to consult the corresponding “Surgical
Techniques”.
For important information regarding electromagnetic
compatibility (EMC), please refer to the chapter “Electro-
magnetic Compatibility” in this manual.
The tool is classied as type BF against electrical shock
and leakage current. The tool is suitable for use on pa-
tients in accordance with IEC 60601-1.
This system requires regular maintenance service, at least
once a year, in order to maintain its functionality. This
servicehas to be performed by the original manufacturer
or an authorized site.
The manufacturer assumes no responsibility for damage
resulting from neglected or unauthorized maintenance.
Unusual Transmissible Pathogens: Surgical patients iden-
tied as at risk for Creutzfeldt-Jakob disease (CJD) and
related infections should be treated with single-use in-
struments. Dispose of the instruments used or suspected
of use on a patient with CJD after surgery and/or follow
the current national recommendations.
Intended use
The Colibri II is designed for the use in traumatology and
orthopedic surgery of the skeleton, i.e. drilling, reaming
or cutting bone.
Safety instructions
The surgeon has to evaluate if the machine is suitable
for an application, based on power limitation of the
machine, attachment and cutting tool regarding bone
strength/anatomical situation as well as handling of the
machine, attachment and cutting tool regarding bone
size. In addition, the contraindications of the implant
have to be respected. Please refer to the corresponding
“Surgical Techniques” of the implant system used.
The Colibri II is only to be used for surgery after careful
consultation of the instructions for use. It is recom-
mended that an alternative system is available to use
during application, as technical problems can never be
completely ruled out.
The Colibri II is designed for use by physicians and
trained medical personnel.
DO NOT use any component if damage is apparent.
DO NOT use any component if the packaging is damaged.
DO NOT use this equipment in the presence of oxygen,
nitrous oxide or a mixture consisting of ammable anes-
thetics and air.
To ensure the proper operation of the tool, only use Syn-
thes original accessories.
Before the rst and every subsequent use, power tools
and their accessories / attachments have to run through
the complete reprocessing procedure. Protective covers
and foils must be fully removed before sterilization.
For the tool to function properly, Synthes recommends
cleaning, disinfecting and servicing it after each use in ac-
cordance with the process recommended in the chapter
“Care and Maintenance”. Compliance with these speci-
cations can considerably extend the service life of the
tool. Only use Synthes oil (519.970) to lubricate the tool.
Introduction
General Information

4 DePuy Synthes Colibri II Instructions for Use
Precautions:
• To avoid injuries, the locking mechanism of the
tool has to be activated before every manipulation
and before placing the tool back down, i.e. the
mode switch has to be in OFF position.
• The tool must only be operated with a fully
charged battery. To do this, ensure that the battery
is charged in good time. We recommend that the
battery is replaced into the charger immediately
after surgery.
• The aseptic transfer is detailed on page 9ff. Alter-
natively for the Li-Ion battery 532.103 follow the
guidelines provided in the STERRAD®/ V-PRO®
sterilization guide (DSEM/PWT/0591/0081).
No other sterilization methods are allowed.
• Additionally, the batteries must never be washed,
rinsed or dropped. This will destroy the battery
with possible secondary damage (explosion haz-
ard!). Only use original Synthes batteries. Further
information can be found on page 12ff.
• Should the machine drop on the oor and have
visible defects, do not use it anymore and send it to
the Synthes service center.
• If a product drops on the oor, fragments may split
off. This represents a danger for the patient and
user as:
– these fragments may be sharp.
–unsterile fragments may enter the sterile eld or
hit the patient.
• Should the system have corroded parts, do not use
it anymore and send it to the Synthes service center.
Accessories/Scope of delivery
The Colibri II consists of a handpiece, one or several bat-
tery casings and batteries and a range of attachments
and accessories designed for the system.
For the system to operate properly, only Synthes cutting
tools should be used.
Special auxiliaries such as cleaning brushes and Synthes
oil are available for cleaning and servicing the system.
No oils from other manufacturers must be used. Only
Synthes oil (519.970) must be used.
Lubricants with other compositions can cause jamming,
can have a toxic effect or can have a negative impact on
the sterilization results. Only lubricate the power tool
and the attachments when clean.
Synthes recommends the use of the specically designed
Synthes Vario Case (68.001.255) and of the specically
designed Washing Basket (68.001.610) to sterilize and
store the system.
The following components are essential to ensure proper
operation:
• Handpiece (532.101)
• Battery Casing (532.132)
• Battery (532.103)
• Sterile Cover (532.104)
• Universal Battery Charger II (05.001.204)
• At least one attachment of the system
Please refer to the end of these Instructions for Use for
an overview of the components of the system.
Locating of the instrument or fragments of
instruments
Synthes instruments are designed and manufactured
to perform within the scope of their intended use. How-
ever, if a Power Tool or accessory/attachment breaks
during use, a visual inspection or a medical imaging de-
vice (e. g. CT, Radiation Devices, etc.) can aid in locating
the fragments and/or components of the instrument.
Storage and transport
Please use the original packaging for dispatch and trans-
port. If this is no longer available, please contact the
Synthes ofce.
Warranty/Liability
The warranty for the tools and accessories does not
cover damage of any kind resulting from improper use,
damaged seals or improper storage and transport. The
manufacturer excludes liability for damage resulting from
repairs or maintenance carried out by unauthorized sites.
The manufacturer assumes no responsibility for damage
resulting from neglected or unauthorized maintenance.
Introduction
General Information

Colibri II Instructions for Use DePuy Synthes 5
2
Do not reuse
Products intended for single use must not be
reused.
Reuse or reprocessing (e.g. cleaning and resteril-
ization) may compromise the structural integrity
of the device and/ or lead to device failure,
which may result in patient injury, illness or
death.
Furthermore, reuse or reprocessing of single use
devices may create a risk of contamination e.g.
due to the transmission of infectious material
from one patient to another. This could result in
the injury or death of the patient or user.
Synthes does not recommend reprocessing con-
taminated products. Any Synthes product that
has been contaminated by blood, tissue and/ or
bodily uids / matter should never be used again
and should be handled according to hospital
protocol.
Even though they may appear undamaged, the
products may have small defects and internal
stress patterns that may cause material fatigue.
Temperature
Relative humidity
Atmospheric pressure
S9 Duty cycle type according to IEC60034-1
IPX4 Ingress protection rating according to
IEC 60529
Explanation of the general symbols used
Caution
Read the provided Instructions for Use before
operating the device.
Consult the Instructions for Use before operat-
ing the device.
The device is classied as type BF against elec-
trical shock and leakage current. The device is
suitable for use on patients according to the
standards dened by IEC 60601-1
SM_109761
Do not immerse device in liquids.
10PB
Product is UL classied to the requirements
of both the United States and Canada
The device meets the requirements of directive
93/42/EEC for medical devices. It is authorized
by an independent notied body for which it
bears the CE symbol.
Li-Ion
This device contains Lithium-Ion batteries that
should be disposed of in an environmentally
friendly manner. The European Battery Directive
2006/66/EC applies to this device. See section
“Disposal” on page 47.
Cd MH
The European directive 2012/19/EC on waste
electrical and electronic equipment (WEEE)
applies to this device. This device contains ma-
terials that should be disposed of in accordance
with environment protection requirements.
Please observe national and local regulations.
See section “Disposal” on page 47.
Indicates Environment Friendly Use Period of
5years in China.
Indicates Environment Friendly Use Period of
10years in China.

6 DePuy Synthes Colibri II Instructions for Use
Manufacturer
Date of manufacture
Non sterile
Non sterile
Do not use if package is damaged.
Introduction
General Information

1
6
6 6
4
3
2
7
3
4
2
5
8
Colibri II Instructions for Use DePuy Synthes 7
Colibri II
Handpiece
1Attachment coupling
2Trigger for speed regulation
3Trigger for switching to reverse / oscillating drilling
4Mode selector switch OFF (lock), Oscillating Mode
(forward/oscillating), ON (forward, reverse)
5 Battery pack (battery casing with inserted battery)
6Release buttons for attachment
7Release buttons for battery casing
8Knob for the battery casing cover
Safety system
The Colibri II is equipped with a safety system that pre-
vents the machine from being unintentionally started. To
lock and unlock the tool, turn the mode selector switch
4to the appropriate setting on the front plate of the
handpiece: OFF, or ON position.
Protective systems
The Colibri II is equipped with three protective systems:
• A thermal overload safety system that shuts off the
tool if it becomes too hot during use. After cooling,
the tool can be used again.
• An exhaustive discharge protection ensures that the
battery does not completely discharge. This protects
the battery and extends its life.
• An internal fuse in the battery that blows in case of
unintended short cut. This prevents excessive heat, re
or explosion. If this happens, the battery can not be
used anymore.
Speed and rotational direction control
Mode selector switch in the ON position
The bottom trigger 2gradually increases and decreases
the forward/reverse speed. When the bottom and top
triggers 2and 3are pressed at the same time, the tool
immediately switches to reverse. When the bottom trig-
ger 2is released, the tool immediately stops.

8 DePuy Synthes Colibri II Instructions for Use
Colibri II
Handpiece
Precautions:
• The information contained in this Information for
Use concerns the Colibri II system. For more infor-
mation on the Colibri articles, please refer to the
Colibri Instructions for Use (036.000.173).
• To prevent injury, the machine must be locked
(OFF position) with the mode selector switch 4
when coupling and removing attachments and
tools, and before laying it down (see page 7).
• Always check correct functioning before use on
patient.
• Always have a back-up system to prevent problems
in case of decient system.
• Always wear personal protective equipment (PPE)
including safety goggles when working with the
Colibri II system.
• When the tool is not in use during the surgery, set
the handpiece on its side to ensure that it does not
fall over due to instability. Only place the power
tool in an upright position on the sterile table to
insert / remove attachments and cutting tools.
• After inserting a cutting tool, always check that it
is properly engaged by pulling it.
Warning: Do not place the Colibri II on a magnetic
surface since the machine might start unintention-
ally.
Mode selector switch in the oscillating drilling
position ( )
When the bottom and top triggers 2and 3are pressed
at the same time, the tool immediately switches to oscil-
lating rotation. When the top trigger 3is released, the
tool returns to normal forward rotation.
Compatibility between Colibri and Colibri II
Existing Colibri battery packs are compatible with
Colibri II handpiece
The small 12 VDC battery pack of the Colibri (532.003
with battery casing 532.002) as well as the large 14.4
VDC battery pack (532.033 with battery casing 532.032)
can both be used with the new Colibri II handpiece
(532.101).
Existing Colibri handpiece is compatible with
Colibri II battery pack
The existing Colibri handpiece (532.001) can be used
with the new battery pack of the Colibri II (532.103 with
battery casing 532.132).
For additional information on the 12 VDC battery pack
(532.002, 532.003 or 532.004), please refer to
section “Additional Information” on page 60 of these
Instructions for Use.

Colibri II Instructions for Use DePuy Synthes 9
Before initial use, brand-new tools and accessories must
undergo the entire reprocessing process and the batter-
ies should be charged. Completely remove protective
caps and lms.
Inserting the unsterile battery in the battery casing
The aseptic transfer is detailed below. Alternatively for
the Li-Ion battery 532.103 follow the guidelines provided
in the STERRAD/V-PRO sterilization guide
(DSEM/PWT/0591/0081).
To ensure sterility of the battery casing, the battery is
inserted into the battery casing by two people, one of
whom is wearing sterile garments:
1. The person with the sterile garments holds the ster-
ile battery casing. If the casing is not opened, the
same person presses the central button to unlock
(Fig. 1), turns the lid sideways (90°) as indicated by
the arrow (Fig. 2) and pulls to open (Fig. 3). Leave
the locking mechanism swung outward.
2. The person wearing the sterile garments places the
sterile cover on the battery casing (Fig. 4) and checks
if it is seated correctly. The sterile cover ensures that
the unsterile battery does not contact the outside of
the sterile casing.
Colibri II
Use
Figure 4Figure 3
Figure 1 Figure 2

10 DePuy Synthes Colibri II Instructions for Use
3. The person not wearing sterile garments carefully
guides the unsterile battery through the sterile cover
(Fig. 5). As an orientation guide, the two symbols of
the battery and the sterile cover should face each
other (Fig. 6). The same person presses it completely
into the battery casing to ensure a correct seat (Fig.
7). This person may not contact the outside of the
battery casing.
4. The person not wearing sterile garments grasps the
anges on the sterile cover and removes it from the
battery casing (Fig. 8).
5. The person wearing the sterile garments closes the
casing cover from the outside without contacting
the battery or the inside of the casing. After having
closed the casing cover, turn the lid sideways (90°)
until it clicks.
Figure 7 Figure 8
Figure 5 Figure 6
Colibri II
Use

Colibri II Instructions for Use DePuy Synthes 11
Precautions:
• Normally, one battery is sufcient for one surgery.
For safety, two battery packs (battery casing with
the battery) should be kept ready to ensure fast in-
traoperative change of batteries under sterile con-
ditions.
• Do not open a battery casing intraoperatively to in-
sert a new battery. Always replace the whole bat-
tery pack by another battery pack which should
have been prepared before the start of the surgery.
• Sterile battery casings that were in contact with
unsterile batteries during insertion of the batteries
must be resterilized before being used in the OR.
• To close the casing cover, press it rmly to ensure
that it is completely closed (Fig. 9 and 10) so that
the locking mechanism properly engages. Always
check that the cover is totally closed before using
the system.
• Sterilize the sterile cover after each use to ensure
aseptic conditions when inserting the unsterile
battery into the sterile battery casing.
Inserting the battery pack into the power tool
Guide the battery pack (battery casing with inserted
battery) from below into the shaft of the handpiece
(Fig.11). The shape of the battery casing prevents the
battery from being inserted incorrectly. Check if the
battery pack is seated correctly by gently pulling on it.
Removing the battery pack from the power tool
Simultaneously press the release buttons for the battery
casing with one hand (Fig. 12) and use the other hand to
remove the battery pack from the handpiece.
Figure 9
Figure 10
Figure 11 Figure 12

12 DePuy Synthes Colibri II Instructions for Use
Use
• Only use the battery for its intended use. Do not use
any battery which is not designed for the equipment.
• Only insert the battery pack (battery and battery cas-
ing) in the handpiece directly before using the Colibri II
system. This saves battery energy and prevents having
the need to change it during surgery.
• Do not apply force to the battery and do not let it fall.
This will destroy it with possible secondary damage.
• Never use a damaged or faulty battery; it can damage
the power tool.
• Do not use a faulty or damaged battery, as this may
damage the power tool. Test the status of the battery
using the Universal Battery Charger II.
• If the drive unit is defect (e.g. short-circuited) do not
insert a battery, as this will blow the internal fuse and
cause damage to the battery. Send the drive unit and
battery to the Synthes Service Center.
• Do not use a faulty or damaged battery, as this may
damage the power tool. Test the status of the battery
using the Universal Battery Charger II.
• If the drive unit is defect (e.g. short-circuited) do not
insert a battery, as this will blow the internal fuse and
cause damage to the battery. Send the drive unit and
battery to the Synthes Service Center.
Care and Maintenance
• The batteries must never be washed, rinsed or
dropped. This will destroy the batteries with pos-
sible secondary damage. Cleaning and Disinfec-
tion instructions of the batteries can be found in
the chapter ”Care and Maintenance”.
• The aseptic transfer is detailed on page 9ff. Alter-
natively for the Li-Ion battery 532.103 follow the
guidelines provided in the STERRAD/V-PRO
sterilization guide (DSEM/PWT/0591/0081).
No other sterilization methods are allowed.
Precautions:
• Do not expose batteries to heat or re. Avoid stor-
age in direct sunlight.
• Keep the batteries and the Universal Battery Char-
ger II clean and in a cool and dry place.
• Risk of re, explosion and burns. Do not disassem-
ble, crush, heat above 60°C /140°F or incinerate the
battery cells.
Precautions and warnings concerning
testing, measuring, charging, storing and usage of
Colibri II batteries (532.103)
Testing and Measuring
• Do not short-circuit the battery. Do not try to measure
the short-circuit current. This will blow the internal fuse
of the battery with irreversible damage of the battery.
• Never open or disassemble the battery.
Charging
• Only use the Synthes Universal Battery Charger II
(05.001.204) to charge the battery. The charger should
have a software version of 11.0 or higher. A label
placed on the bottom of the charger enables to
identify the latest software version of the charger.
Never charge the battery in another Synthes charger
or a charger from another manufacturer. This will
damage the battery.
• The batteries should always be charged before use.
• Place the battery into the charger immediately after
surgery.
Storage
• Always recharge the battery after each use. Do not
store an empty battery as this will shorten the life span
and will not be covered by warranty.
• When the battery is not being used, always store it in
the Synthes Universal Battery Charger II and turn the
charging station on. This will avoid the battery from dis-
charging, and it will be fully charged and ready to use.
Never store the battery in another Synthes charger nor
a charger from another manufacturer. This will damage
the battery.
• Never store the battery in the battery casing (532.132)
when it is connected to the Colibri II handpiece
(532.101) as this will discharge the battery.
• When storing batteries, ensure that they are separately
packed and do not store them with materials that
conduct electricity in order to prevent short-circuit.
This could damage the battery and generate heat
which can cause burns.
Colibri II
Use

Colibri II Instructions for Use DePuy Synthes 13
Oscillating drilling ( ) mode
To protect soft tissue when drilling and inserting
Kirschner wires, the Colibri II has an electronically con-
trolled oscillating mode.
To preset the oscillating mode, switch the mode selector
switch to position.
Pressing the bottom trigger causes the tool to rotate
clockwise as usual. Simultaneously pressing the top and
bottom triggers causes the tool to immediately switch
to oscillating mode. The clamped tool oscillates clock-
wise / anticlockwise. The speed can be changed by
means of the bottom trigger. After the top trigger is re-
leased, the tool returns to normal clockwise rotation.
Precautions:
• Oscillating mode may only be used with
the following attachments:
– AO/ASIF Quick Coupling (05.001.250)
–Chuck (05.001.252, 05.001.253)
– Quick Coupling for Kirschner Wires (532.022)
• Do not use the oscillating mode with the oscillat-
ing saw attachments!

14 DePuy Synthes Colibri II Instructions for Use
The Colibri II system offers a broad range of attach-
ments.
A wide range of rotating attachments has color-coded
rings, so that they can easily be identied. The table on
the next page lists the different types of attachments
available, the color coding as well as the speed of each
attachment.
Mounting the attachments
Insert the attachment into the attachment coupling
(Fig.1). If the positioning pins do not lock into place
right away, twist the attachment a bit to the right or left
until it locks into the correct position. Check if the at-
tachment is seated correctly by gently pulling on it.
Removing the attachments
Press the attachment release buttons 6 (see gure on
page 7) simultaneously and remove the attachment from
the coupling.
Precautions:
• To prevent injury, the power tool must be switched
off with the safety system (see page 7) during each
manipulation.
• Only use original attachments and tools from Syn-
thes. Damage that arises from using attachments
and tools made by other manufacturers is not cov-
ered by the warranty.
• Never use an attachment in reverse mode with an
old exible shaft as this could cause serious injury
to the patient.
• When the tool is not in use during the surgery, set
the handpiece on its side to ensure that it does not
fall over due to instability. Only place the power
tool in an upright position on the sterile table to
insert / remove attachments and cutting tools.
Attachments
General Information
Figure 1

Colibri II Instructions for Use DePuy Synthes 15
Article
Number
Product Speed Color coding
for speed
Drill
Attach-
ments
05.001.250 AO/ASIF Quick Coupling 1290 rpm Blue
05.001.252 Chuck (Drilling Speed), with Key, clamping range up
to B 4.0 mm
1290 rpm Blue
05.001.253 Chuck (Drilling Speed), with Key, clamping range up to
B7.3 mm
1290 rpm Blue
Screw
Attach-
ments
05.001.251 Screw Attachment with AO/ASIF Quick Coupling 350 rpm Red
Ream
Attach-
ments
532.017 AO/ASIF Quick Coupling for Medullary Reaming 350 rpm Red
532.018 Hudson Quick Coupling for Medullary Reaming 350 rpm Red
532.019 Trinkle Quick Coupling for Medullary Reaming 350 rpm Red
532.020 Trinkle Quick Coupling, modied, for Medullary
Reaming
350 rpm Red
532.015 Quick Coupling for DHS/DCS triple reamers 350 rpm Red
05.001.254 Chuck (Reaming Speed), with Key, clamping range up
to B 7.3 mm, with reverse motion
350 rpm Red
Other Ro-
tating At-
tachments
532.011 Mini Quick Coupling 3500 rpm None
532.012 J-Latch Coupling 3500 rpm None
532.022 Quick Coupling for Kirschner Wires 875 rpm None
05.001.187 Burr Attachment 17500 rpm None
511.3 0 0 Radiolucent Drive with Attachment 05.001.250 1250 rpm None
Saw
Attach-
ments
532.021 Oscillating Saw Attachment 17500 osc./min None
532.023 Oscillating Saw Attachment II (Crescentic Technique) 17500 osc./min None
532.026 Large Oscillating Saw Attachment 17500 osc./min None
Other
Attach-
ments
511.773 Torque Limiter, 1.5 Nm, for AO/ASIF Quick Coupling – N /A*
511.776 Torque Limiter, 0.8 Nm, with AO/ASIF Quick Coupling – N /A*
511.777 Torque Limiter, 0.4 Nm, with AO/ASIF Quick Coupling – N /A*
* Color coding on Torque Limiters does not refer to speed.
Technical data is subject to tolerances. Specifications are approximate and may
vary from one device to another or as a result of power supply fluctuations.

16 DePuy Synthes Colibri II Instructions for Use
AO/ASIF Quick Coupling (05.001.250)
For tools with an AO/ASIF coupling shaft.
Mounting and removing the tools
Insert the tool into the attachment from the front apply-
ing slight pressure and turning slightly. It is not necessary
to operate the coupling sleeve of the attachment.
To disconnect, push the coupling sleeve of the attach-
ment back and remove the tool.
Chucks
There are two Drill Attachments Chucks available as Drill Attachments for the Colibri II system.
Article number Clamping range Spare key Comments
05.001.252 0.5–4.0 mm 310.932 For drilling
05.001.253 0.5 –7. 3 mm 510.191 For drilling
Attachments
Drill Attachments
Inserting cutting tools
Open the jaws of the Chuck using the appropriate key or
by hand. Insert the shaft of the tool into the open drill
chuck and close it by twisting the chuck. Make sure that
the shaft lies central to the three jaws. Tighten the drill
chuck with the key. Make sure that the teeth of the key
engage correctly in the toothed rim of the chuck.
Removing cutting tools
Open the Chuck with the key and remove the tool.
Precautions:
• Check the cutting tool for wear and / or damage
after each use and replace if necessary.
• To ensure good xation of the tools, make sure the
toothed rims on the drill chuck and key are not
worn.
Warning: Do not use the Colibri II for acetabular
reaming.

Colibri II Instructions for Use DePuy Synthes 17
Screw Attachment, with AO/ASIF Quick Coupling
(05.001.251)
Mounting and removing the tools
Insert the tool into the attachment from the front by ap-
plying slight pressure and turning slightly. It is not neces-
sary to operate the coupling sleeve of the attachment.
To disconnect, push the coupling sleeve of the attach-
ment back and remove the tool.
Note: Theoretically, it is also possible to use the
AO/ASIF Quick Coupling (05.001.250) to insert
screws. However, the Screw Attachment (05.001.251)
has a lower speed and a higher torque and is there-
fore more suitable. Screws with a large diameter
may not be able to be inserted with the AO/ASIF
Quick Coupling as the torque may not sufce.
Precautions:
• Care should be taken when inserting screws with
the drive unit.
• Never fully insert screws with the drive unit. The
last turns or locking should always be done manu-
ally.
• Always use an appropriate torque limiting attach-
ment when putting locking screws into a locking
plate.
• The attachment is also suitable for application at a
lower rpm and /or higher torque.
Attachments
Screw Attachments

18 DePuy Synthes Colibri II Instructions for Use
All Colibri II Ream Attachments provide an
approximative maximal torque of 7.5 Nm (with
battery 532.103).
Quick Couplings for Medullary Reaming
AO/ASIF Quick Coupling (532.017)
Hudson Quick Coupling (532.018)
Trinkle Quick Coupling (532.019)
Trinkle Quick Coupling, modified (532.020)
The Quick Couplings for Medullary Reaming enable the
use of exible shafts with the appropriate coupling ge-
ometry. Reverse motion, which can damage the exible
shafts, is prevented by a special mechanical system.
Inserting cutting tools into the coupling
Maneuver the unlocking ring on the attachment back-
ward and insert the tool (such as a drill bit) while rotat-
ing it slightly until it locks into place. Release the ring.
Check if the tool is seated correctly in the coupling by
gently pulling on it.
Removing cutting tools
Push the unlocking ring on the attachment backward
and remove the tool.
Quick Coupling for DHS/DCS Triple Reamers
(532.015)
For DHS/DCS Triple Reamers; can also be used to open
the medullary cavity with most of the Synthes nailing
systems.
Mounting and removing the tools
To connect the tool, push the coupling sleeve forward
and then introduce the tool while turning slightly. To dis-
connect, push the coupling sleeve of the attachment for-
ward and remove the tool.
Attachments
Ream Attachments
Other manuals for Colibri II
1
Table of contents
Popular Drill manuals by other brands

Chicago Electric
Chicago Electric VARIABLE SPEED 18 VOLT CORD- LESS DRILL/HAMMER DRILL... operating instructions

Silverline
Silverline DIY Series manual

Rubi
Rubi TP-10 instruction manual

Bosch
Bosch GDX 18V-200 C Professional Original instructions

Parkside Performance
Parkside Performance PSBSAP 20-Li A1 instructions

Ryobi
Ryobi CDC1802 user manual